Sirisha Tadimalla1,2, Daniel J Wilson3, David Shelley3, Gavin Bainbridge3, Margaret Saysell3, Iosif A Mendichovszky4, Martin J Graves4, J Ashley Guthrie3, John C Waterton5,6, Geoffrey J M Parker5,7, Steven P Sourbron8.
Abstract
BACKGROUND: Three-dimensional variable flip angle (VFA) methods are commonly used for T1 mapping of the liver, but there is no data on the accuracy, repeatability, and reproducibility of this technique in this organ in a multivendor setting.Entities:
Keywords: T1 mapping; VFA; bias; liver; precision; repeatability; reproducibility
Mesh:
Year: 2022 PMID: 35224803 PMCID: PMC9545852 DOI: 10.1002/jmri.28127
Source DB: PubMed Journal: J Magn Reson Imaging ISSN: 1053-1807 Impact factor: 5.119
Details of Scanners Used in This Study
| Scanner | Vendor | Model | RF Coil |
|---|---|---|---|
| S1 | Siemens Healthineers | 3.0 T Prisma VE11C | 18 channel array |
| S2 | Siemens Healthineers | 1.5 T Aera VE11A | 18 channel array |
| P1 | Philips Medical Systems | 3.0 T Achieva | 32 channel array |
| P2 | Philips Medical Systems | 1.5 T Ingenia | dStream Torso |
| G1 | GE Healthcare | 3.0 T Discovery MR750 | 32 channel array |
| G2 | GE Healthcare | 1.5 T Discovery MR450 | 32 channel array |
The 95% CI on Overall Mean Bias, Repeatability, Reproducibility, and Spatial Heterogeneity of T1 Values in the Normal Liver
| MOLLI | VFA BH | VFA FB | |
|---|---|---|---|
| Bias (%) | 1.7 ± 3.7** | 31 ± 8.0 | 29 ± 8.1 |
| Relative repeatability Error (%) | 2.4 ± 0.7** | 11 ± 3.1 | 9.4 ± 2.0 |
| Relative reproducibility error (%) at 1.5 T | 34 ± 9.2 | 24 ± 8.1 | |
| Relative reproducibility error (%) at 3.0 T | 14 ± 5.8* | 22 ± 4.0 | 29 ± 7.1 |
| Spatial heterogeneity (%) | 11 ± 1.5** | 25 ± 1.9 | 18 ± 2.3 |
MOLLI, modified Look‐Locker imaging; rANOVA, repeated measures ANOVA; VFA, variable flip angle.
Paired t‐test (VFA BH vs. VFA FB) P‐value < 0.0001.
**rANOVA P‐value < 0.0001; *rANOVA P‐value < 0.01.
Repeatability and Reproducibility of VFA T1 Values in the Liver, Compared to Literature Values in Phantoms and Other Organs
| Application Area | Corrections | Scanners | Relative Repeatability Error | Relative Reproducibility Error |
|---|---|---|---|---|
| Liver (from this study) | None |
1.5 T: 1 × S, 1 × P, 1 × G 3.0 T: 1 × G, 1 × P, 1 × S | 10% | 28% |
| Breast (reference |
| 3.0 T: 3 × S | 11% | 14% |
| Brain (R1) (reference |
| 3.0 T: 4 × S, 2 × P | 10%–16% | 10%–20% |
| Prostate (transitional zone) (reference | With and without | 3.0 T: 2 × S | 12% (14% without correction) | 14% (18% without correction) |
VFA, variable flip angle.
Literature values were converted to the definition used in this study.
FIGURE 1Bias, repeatability, reproducibility, and spatial heterogeneity of variable flip angle (VFA) T1 measurements across field strengths. The box plots indicate the median and interquartile range (IQR), and the whiskers encompass all points within 1.5 times the IQR. Each color‐coded datapoint represents a single volunteer.
Results (P‐values) of Repeated Measures ANOVA (Levene's Tests in Parentheses) on the Effect of Field Strength on Bias, Repeatability, Reproducibility, and Spatial Heterogeneity
| Sequence | Bias | Repeatability | Reproducibility | Spatial Heterogeneity |
|---|---|---|---|---|
| MOLLI |
| 0.29 (0.36) | ‐ |
|
| VFA BH | 0.16 (0.30) | 0.99 (0.77) |
| 0.30 (0.26) |
| VFA FB |
| 0.69 (0.75) | 0.68 (0.17) |
|
Significant results are highlighted in bold font.
MOLLI, modified Look‐Locker imaging; VFA, variable flip angle.
FIGURE 2Bias, repeatability, reproducibility, and spatial heterogeneity of variable flip angle (VFA) T1 measurements across vendors. The box plots indicate the median and interquartile range (IQR), and the whiskers encompass all points within 1.5 times the IQR. Each color‐coded datapoint represents a single volunteer.
Results (P‐Values) of Repeated Measures ANOVA (Levene's Tests in Parentheses) on the Effect of Vendor on Bias, Repeatability, and Spatial Heterogeneity
| Sequence | Bias | Repeatability | Reproducibility | Spatial Heterogeneity |
|---|---|---|---|---|
| MOLLI |
| 0.96 (0.88) |
|
|
| VFA BH | 0.08 ( | 0.69 (0.80) |
| 0.31 (0.65) |
| VFA FB | 0.69 ( | 0.40 (0.13) |
|
|
Significant results are highlighted in bold font.
MOLLI, modified LL imaging; VFA, variable flip angle.
Vendors P and S only.
FIGURE 3T1 values in the normal liver at 1.5 T and 3.0 T. Each color‐coded datapoint represents a single volunteer. Between‐subject variation (BSV, expressed as a percentage) is given for each scanner and method on the right.
The 95% CI of the mean T1 values in the liver at 1.5 T and 3.0 T
| Literature reference | MOLLI | VFA BH | VFA FB | |
|---|---|---|---|---|
| T1 at 1.5 T (msec) | 602 | 586 ± 33 | 780 ± 91 | 705 ± 39 |
| T1 at 3.0 T (msec) | 752 | 782 ± 36 | 1036 ± 148 | 1053 ± 147 |
MOLLI, modified Look‐Locker imaging; VFA, variable flip angle.