H Déchaud, R Bador, F Claustrat, C Desuzinges.
Abstract
We describe an immunofluorometric assay for prolactin based on lanthanide labeling of a monoclonal antibody and measuring time-resolved fluorescence. In this "sandwich"-type assay, the label (Eu3+) was bound to the second antibody by means of a simple, rapid method involving the anhydride of diethylenetriaminepentaacetic acid. To measure the photoluminescence of europium (or other lanthanides), we have developed a time-resolved fluorometer with a nitrogen laser as the pulsed excitation source. During the assay, the solid-phase antibody immobilized inside a polystyrene tube is incubated with the plasma sample and the second antibody in a one-step procedure. Results for 67 human plasmas correlated well (r = 0.98) with those by an immunoradiometric method.Entities:
Mesh:
Substances:
Year: 1986 PMID: 3521942
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327