Zhe-Wei Yang1, Thi-Thu-Hien Pham2, Chin-Chi Hsu1, Chi-Hsiang Lien1, Quoc-Hung Phan1.
Abstract
A graphene-based surface plasmon resonance (SPR) prism coupler sensor is proposed for the rapid detection of immunoglobulin G (IgG) antibodies. The feasibility of the proposed sensor is demonstrated by measuring the IgG concentration in phantom mouse and human serum solutions over the range of 0-250 ng/mL. The results show that the circular dichroism and principal fast axis angle of linear birefringence increase in line with increases in IgG concentration over the considered range. Moreover, the proposed device has a resolution of 5-10 ng/mL and a response time of less than three minutes. In general, the sensor provides a promising approach for IgG detection and has significant potential for rapid infectious viral disease testing applications.Entities:
Keywords: graphene; immunoglobulin G detection; surface plasmon resonance; virus infectious disease
Mesh:
Substances:
Year: 2022 PMID: 35214258 PMCID: PMC8962983 DOI: 10.3390/s22041362
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1(a) Schematic illustration of graphene-based SPR sensor and (b) Raman spectroscopy analysis of single-layer graphene.
Figure 2Schematic illustration of dual-retarder Mueller polarimetry system.
Output polarization states generated by dual-retarder Muller polarimetry system.
| Phase Retardation of LCVRs | State of Polarization | |
|---|---|---|
| δ1 | δ2 | |
| 90° | 270° | 0° |
| 0° | 0° | 45° |
| 90° | 90° | 90° |
| 90° | 180° | R- |
Figure 3Measurement results obtained for (a) circular dichroism (R) and (b) principal fast axis angle (α) of mouse IgG samples with concentrations of 0–250 ng/mL.
Figure 4Measurement results obtained for (a) circular dichroism (R) and (b) principal fast axis angle of LB (α) of human IgG samples with concentrations of 0–250 ng/mL.
IgG sensing performance of existing detection methods.
| Actual Glucose Concentration (mg/dL) | Sensing Medium | Range of Detection (ng/mL) | Resolution (ng/mL) | Ref. |
|---|---|---|---|---|
| Immunoassay and magnetic bio-separation | Pig IgG serum | 0–25 × 103 | 5 × 103 | [ |
| Enzyme-linked immunosorbent assay | Human IgG serum | 0–1000 | 100 | [ |
| Enzyme-linked immunosorbent assay | Mouse, goat IgG serum | 0–100 | 20 | [ |
| SPR-graphene oxide | Rabbit IgG serum | 0–100 × 103 | 62.5 | [ |
| SPR-graphene oxide | Human IgG serum | 0–100 × 103 | 5 × 103 | [ |
| SPR-single-layer graphene | Human, mouse IgG serum | 0–250 | 5–10 | Present study |