Dried blood spot (DBS) cards perform many functions for sampling blood that is intended for subsequent laboratory analysis, which include: (i) obviating the need for a phlebotomist by using fingersticks, (ii) enhancing the stability of analytes at ambient or elevated environmental conditions, and (iii) simplifying the transportation of samples without a cold chain. However, a significant drawback of standard DBS cards is the potential for sampling bias due to unrestricted filling caused by the hematocrit of blood, which often limits quantitative or reproducible measurements. Alternative microsampling technologies have minimized or eliminated this bias by restricting blood distribution, but these approaches deviate from clinical protocols and present a barrier to broad adoption. Herein, we describe a patterned dried blood spot (pDBS) card that uses wax barriers to control the flow and restrict the distribution of blood to provide enhanced sampling. These patterned cards reproducibly fill four replicate extraction zones independent of the hematocrit effect. We demonstrate a 3-fold improvement in accuracy for the quantitation of hemoglobin using pDBS cards compared to unpatterned cards. Patterned cards also facilitate the near quantitative recovery (ca. 95%) of sodium with no evidence of a statistically significant difference between dried and liquid blood samples. Similarly, the recovery of select amino acids was conserved in comparison to a recent report with improved intercard precision. We anticipate that this approach presents a viable method for preparing and storing samples of blood in limited resource settings while maintaining current clinical protocols for processing and analyzing dried blood spots.
Dried blood spot (DBS) cards perform many functions for sampling blood that is intended for subsequent laboratory analysis, which include: (i) obviating the need for a phlebotomist by using fingersticks, (ii) enhancing the stability of analytes at ambient or elevated environmental conditions, and (iii) simplifying the transportation of samples without a cold chain. However, a significant drawback of standard DBS cards is the potential for sampling bias due to unrestricted filling caused by the hematocrit of blood, which often limits quantitative or reproducible measurements. Alternative microsampling technologies have minimized or eliminated this bias by restricting blood distribution, but these approaches deviate from clinical protocols and present a barrier to broad adoption. Herein, we describe a patterned dried blood spot (pDBS) card that uses wax barriers to control the flow and restrict the distribution of blood to provide enhanced sampling. These patterned cards reproducibly fill four replicate extraction zones independent of the hematocrit effect. We demonstrate a 3-fold improvement in accuracy for the quantitation of hemoglobin using pDBS cards compared to unpatterned cards. Patterned cards also facilitate the near quantitative recovery (ca. 95%) of sodium with no evidence of a statistically significant difference between dried and liquid blood samples. Similarly, the recovery of select amino acids was conserved in comparison to a recent report with improved intercard precision. We anticipate that this approach presents a viable method for preparing and storing samples of blood in limited resource settings while maintaining current clinical protocols for processing and analyzing dried blood spots.
Blood is a complex
matrix, comprising cellular and liquid fractions,
that contains a wealth of diagnostically relevant biomarkers, which
are inclusive of the cells themselves (e.g., neutrophil count), DNA/RNA
(e.g., endogenous or from pathogens), and myriad solutes in plasma
(e.g., proteins, metabolites, free amino acids). For these reasons,
blood is often thought of as the ideal specimen for evaluating the
health status of a patient. Obtaining liquid blood samples in centralized
facilities or even local clinics is routine practice. In these settings,
a trained phlebotomist will collect milliliter volumes of blood by
venipuncture, which can either be immediately processed and tested
in the laboratory or stored for future analysis within a defined period
of time dependent on storage temperature.[1,2] However,
these same practices face unique challenges at the point-of-care or
in resource-limited settings. Specifically, the storage and transportation
of liquid blood are complicated by unreliable modes of transportation
and inadequate access to cold chain storage. These limitations often
require liquid samples to be discarded due to substantial degradation
or significant changes to critical hematological indices.In
contrast to liquid samples, storing blood in a porous matrix,
such as chromatography paper, enhances analyte stability at ambient
or elevated temperatures.[3,4] Dried blood spot (DBS)
cards offer simplified sampling using fingersticks and reliable transportation
of dried blood by mail, thus circumventing the need for cold chain
storage.[5,6] Traditional DBS cards, such as the Whatman
903 Protein Saver card, are a simple construction of a single sheet
of thick cellulose cardstock affixed to an envelope for sample identification
and handling.[7] Circles are printed onto
the surface of the paper using a thin layer of toner to provide guidance
for sample application. Fingerstick volumes of blood (e.g., 50–100
μL per spot) are applied to the card and allowed to dry for
a minimum of 4 h at ambient conditions (and ideally overnight), rendering
the card nonbiohazardous, before they are sealed and shipped through
the mail for laboratory analysis.[8] Self-sampling
low volumes of blood without the need for cold chain storage could
broadly expand access to basic health information by providing direct-to-consumer
testing, facilitating critical population screening, and biobanking
efforts.[9]Although traditional DBS
cards offer simple operation, require
low volumes of blood, and can be collected outside of the clinic,
they are severely limited by usability associated with unrestricted
sample application zones. This user error can result in nonuniform
or smeared blood spots, which will ultimately impact the quality of
subsequent laboratory analysis and represents a considerable barrier
for the ubiquitous use of traditional DBS cards.[9,10] Beyond
usability, traditional DBS cards do not account for differences in
hematocrit values (Hct)—the ratio of packed red blood cells
(RBCs) to total blood sample volume. The normal range of hematocrit
spans 36–50% and is affected by variables such as race, sex,
age, hydration, and overall health status.[11] Currently, the hematocrit value must be known prior to analysis
for the accurate quantitation of analytes using DBS cards. Whether
caused by filling imprecision or hematocrit, the uniformity of how
cells and liquid plasma are distributed throughout the paper cardstock
has a substantial impact on the overall utility of a DBS card.The hematocrit effect has been extensively reviewed as the main
obstacle to overcome for quantitative analysis using traditional DBS
cards.[12−18] Since the hematocrit value represents the ratio of cellular matter
to liquid plasma, blood samples with a high hematocrit value (e.g.,
55%) will be more viscous than samples with a low hematocrit value
(e.g., 30%). Variation in viscosity results in variable sample flow
and distribution through the paper, which negatively impacts the reproducibility
of sample volumes obtained from a single, fixed punch extraction from
DBS (i.e., sub-punch). Uncontrolled saturation or spreading of blood
through the DBS paper can also result in a heterogeneous distribution
of analytes throughout the area of the resulting DBS (i.e., volcano
effect).[19] Because analytes are typically
eluted from DBS via a sub-punch, any variation in sample volume and
distribution will manifest in downstream clinical measurements, causing
a lack of precision (i.e., intraspot agreement) or accuracy (i.e.,
agreement with liquid sample).Many methods for minimizing the
hematocrit effect in traditional
DBS cards have previously been reported.[20−22] Two distinct
approaches stand out: (i) whole spot analysis[23,24] and (ii) assay specific calibrants stored within the paper.[25] Both present viable options for minimizing the
hematocrit effect by quantitative removal of the entire blood spot
(dependent on the application of accurate sample volume) or the inclusion
of an internal standard at a known concentration to estimate extraction
efficiency. However, both methods are limited by the number of tests
that can be conducted from a single DBS spot. In each format, samples
can only be used to perform a single test due to the complete destruction
of the entire dried spot or assay-specific internal standard. Alternatively,
three-dimensional blood spheroids eliminate chromatographic effects
observed in traditional DBS and reduce the volume of blood required
per spot by utilizing functionalized hydrophobic paper.[24] This approach has successfully demonstrated
increased stability of enzymes and labile organic compounds at ambient
conditions. Recently, DBS technologies that operate independently
of the hematocrit by constricting sample volume have also been described.
The ADX Test Card by Accel Diagnostics utilizes a microfluidic network
and magnetic beads to collect, distribute, and analyze blood.[26] The HemaSpot HF comprises precut paper wedges
contained within a plastic housing, which hold a finite volume of
sample.[27] Similarly, the HemaPEN[28] and HemaXis DB10[29] integrate multiple fixed-volume capillary tubes to standardize the
volume of blood applied to a porous matrix. While these devices provide
enhanced control over the application of sample volume, they do not
conform with current clinical collection or automated punching and
elution protocols.In order to improve the utility of DBS cards
with an intent for
widespread use, current clinical protocols for sample collection and
subsequent analysis should be maintained. Therefore, innovation should
build upon the major benefits of traditional DBS technology (i.e.,
single layer of cardstock). An attractive approach for enhanced sampling
is controlling the flow of blood samples in the cardstock with hydrophobic
wax barriers.[30] Defining specific areas
for (i) sample addition, (ii) distribution, and (iii) storage by wax
patterning presents a method for addressing the limitations of current
DBS technologies without creating additional clinical barriers. Ideally,
blood sampling would be performed via a self-administered fingerstick,
simple collection onto a solid matrix, drying, and delivery to a laboratory
for testing without significant degradation of the sample at ambient
conditions.Herein, we describe the creation of patterned dried
blood spot
(pDBS) cards to address the limitations of traditional DBS cards directly
related to the hematocrit effect. Patterning traditional DBS cardstock
with hydrophobic wax barriers regulates sample application, distribution,
and volume control while operating independently of the hematocrit
over a broad range of clinical values (20–60%). A user simply
needs to apply a volume of blood to the center of the card and the
sample will automatically distribute to four replicate punch zones.
Providing more spots for analysis while also maintaining reproducible
spreading across physiological hematocrit values can (i) increase
the number of technical replicates or (ii) increase the number of
clinical assays performed from one sample of whole blood without concern
for significant punch-to-punch variation.We first investigated
the capacity of pDBS cards for quantitative
sampling by estimating the volume of blood contained in a standard
6 mm paper punch and reported minimal variation even when the sample
input deviates from the World Health Organization (WHO) recommended
volume of 75 μL.[5] Next, we demonstrated
enhanced usability and spot uniformity independent of the hematocrit
for samples collected with pDBS cards compared to traditional, unpatterned
cardstock. We highlighted a broad class of analytes to showcase this
approach including the quantitation of hemoglobin by UV–vis
spectrophotometry, sodium by inductively coupled plasma atomic emission
spectroscopy (ICP-AES), and specific amino acids by high-performance
liquid chromatography (HPLC). pDBS cards permit the enhanced sampling
of small volumes of blood that can be generated from a fingerstick
and represent a reproducible method capable of performing multiple
tests without requiring multiple sample collections or altering established
laboratory workflows. We anticipate the quantitative nature of this
self-sampling method of blood collection will empower patients by
providing critical, accurate diagnostic information at home or in
low-income economies without impacting existing clinical procedures.
Experimental Design: Card Design and Fabrication
pDBS cards comprise a single layer of cardstock impregnated with
wax to form three distinct features: (i) sample addition zone, (ii)
lateral distribution channels, and (iii) four replicate, collection
punch zones (Figure A). We designed our cards to accommodate a sample input volume of
75 μL and output punch diameter of 6 mm in accordance with the
WHO recommended specifications for DBS sampling. The design features
(e.g., lateral channels) and geometries were informed by our previous
experience with whole blood in paper for measuring the hematocrit.[31,32] Whole blood is transported from the sample addition zone along the
lateral channels via capillary action and fills four replicate collection
punch zones at the end of each channel. Extending the lateral channels
past the collection punch zones allowed for the complete saturation
of the punch zone for more accurate sampling compared to traditional
DBS cards. Wax printing is typically performed by the direct deposition
of wax onto relatively thin (≤250 μm), smooth papers
followed by the application of heat to allow the wax to coat the paper
fibers.[33] For papers >250 μm thick,
standard printing practices cannot deposit sufficient wax to form
complete hydrophobic barriers (Figure S1A). Incomplete barriers resulted in uncontrolled sample flow and represent
a challenge for patterning DBS papers. Alternative methods for patterning
thick materials with photoresist, paraffin,[34] and silanes[35] have been reported previously.
However, wax printing offers several advantages such as rapid prototyping,
excellent reproducibility, and is suitable for mass production.[33,36] We utilized a double-sided wax transfer method[37] to successfully pattern papers commonly used for traditional
DBS cards (e.g., Whatman CF-12, Ahlstrom 226, Munktell TFN) (Figure S1B). First, we printed the top and bottom
designs onto laminate sheets using a Xerox ColorQube 8580 wax printer.
Next, we aligned a sheet of chromatography paper with the top and
bottom designs using a custom acrylic alignment jig. Finally, we used
a Promo Heat CS-15 T-shirt press (45 s at 137 °C) to transfer
the wax from the laminate sheets to the paper to form hydrophobic
barriers through the full thickness of the paper.
Figure 1
Schematic of a patterned
dried blood spot (pDBS) card. (A) Top
and bottom of the card are uniquely patterned and include three distinct
features: (i) sample addition zone (outlined in green), (ii) lateral
distribution channels for sample splitting (outlined in cyan), and
(iii) four replicate collection punch zones (outlined in red). Collection
punch zones are removed via a standard 6 mm office hole punch prior
to analysis. (B) Scanned image of actual pDBS card.
Schematic of a patterned
dried blood spot (pDBS) card. (A) Top
and bottom of the card are uniquely patterned and include three distinct
features: (i) sample addition zone (outlined in green), (ii) lateral
distribution channels for sample splitting (outlined in cyan), and
(iii) four replicate collection punch zones (outlined in red). Collection
punch zones are removed via a standard 6 mm office hole punch prior
to analysis. (B) Scanned image of actual pDBS card.Patterning each side with a unique design allowed for the
partial
coating of the cellulose fibers through approximately half the thickness
of the paper to reduce the void volume of the sample addition zone
and lateral distribution channels in pDBS cards (Figure B). This process provided an
added benefit of minimizing sample input volume while maximizing sample
collection volume from the punch zones. After the addition of whole
blood, we dried pDBS cards under ambient conditions in a biosafety
cabinet (ca. 16 h), whereby they can be used immediately or sealed
in a foil pouch with silica desiccant packets and a humidity indicator
card for long-term storage. All data presented herein were collected
using pDBS cards fabricated from TFN grade cardstock. We chose to
demonstrate the utility of our cards for sampling a range of analytes
(e.g., hemoglobin, sodium, and select amino acids) and technique groups
(e.g., UV–vis spectrophotometry, ICP-AES, and HPLC).
Results
and Discussion
Effects of Evaporation on the Quantitation
of Hemoglobin
Evaporation at ambient conditions is the driving
force for drying
samples of blood in DBS cards. Sealing—or partially sealing—sections
of our pDBS cards influenced the location and extent of evaporation.
Additionally, altering the bottom design of the pDBS card can affect
evaporation by controlling the amount of unpatterned card area that
is exposed to the environment. We iteratively added or removed a layer
of laminate to the top and bottom sides of the pDBS card and evaluated
the effects of evaporation on the quantitation of hemoglobin using
a modification of the standard Drabkin’s assay (Figure S2). The bottom design either (i) excluded
(designs A and B) or (ii) included (designs C and D) the lateral distribution
channels. Evaluating these design features across a range of hematocrit
is critical for understanding the effects of evaporation since these
samples have varying volumes of liquid plasma (e.g., 52.5 μL
of plasma in 75 μL of 30% hematocrit blood vs 37.5 μL
of plasma in 75 μL of 50% hematocrit blood). Excluding the lateral
channels and sample addition zone in the bottom design reduced the
total void volume of the unpatterned area and eliminated evaporation
from the bottom side of the lateral channels and sample addition zone.
Reducing the void volume improved reproducibility for card filling.
Further covering the lateral channels on the top side of the pDBS
card minimized evaporation along the channel and effectively concentrated
the blood sample in the collection punch zones. The preconcentration
of blood in the collection punch zones resulted in a higher percent
deviation for the quantitation of hemoglobin (Table S1), which we expect is due to the volume dependency
of the Drabkin’s assay.[38] Both designs
B and C had comparable performance even though design B included no
laminate covering the unpatterned area and design C was completely
laminated (except at the sample addition zone). We chose to move forward
with design B for two reasons: (i) it yielded the lowest percent error
for both 30% and 50% hematocrit samples and (ii) reduced the number
of laminate layers necessary, which simplified the manufacturing and
operational processes.
Estimation of Sample Volume in 6 mm Paper
Punch
After
finalizing the form factor of our pDBS card and minimizing the effect
of evaporation through unique bottom patterning, we measured the volume
of a dried sample contained in an individual 6 mm paper punch in order
to correlate the concentration of an analyte to the total sample of
blood. The accurate comparison of liquid reference samples to our
pDBS card is dependent on the sample volume contained within a sub-punch.
This type of measurement has been accomplished using a variety of
methods including ion suppression by liquid chromatography-tandem
mass spectrometry[20] and the electrical
conductivity of DBS extract by a ring disk electrode.[10] We utilized the volume dependency of the Drabkin’s
assay to estimate the output sample volume in our pDBS card.[39] First, we constructed a series of calibration
curves (Figure S3A) using liquid hemoglobin
standards with varied sample input volumes (3–11 μL)
to establish a relationship between the linear slope of the calibration
curve and sample volume (Figure S3B). Then,
we calibrated our pDBS cards with hemoglobin standards and estimated
the sample volume contained in a 6 mm paper sub-punch using the resultant
linear relationship and slope of the calibration curve in our pDBS
card (Figure S3C). All hemoglobin samples
reproducibly filled the pDBS cards (Figure S3D). We estimated that each 6 mm paper sub-punch contained 10.3 ±
0.4 μL of whole blood, representing a total output sample volume
of approximately 41.2 μL from an input volume of 75 μL
of blood. These data are supported by the calculated theoretical volume
(ca. 10.6 μL) occupied in a 6 mm diameter disc of grade TFN
(ca. 470 μm thick with a nominal porosity of 80%). The low variation
(<5%) observed in the sample volume contained in a paper sub-punch
indicated consistent sample distribution in pDBS cards.Deviating
from the recommended sample input volume of 75 μL can negatively
impact the quantitation of analytes such as hemoglobin. To simulate
under- and overfilling, we applied a range of sample volumes 60–90
μL in 5 μL increments at a single hematocrit (Figure S4A). Our pDBS cards reproducibly filled
four replicate punch zones with a sample volume ≥65 μL
(Figure S4B). The average deviation for
replicate cards with sample input varying ±15 μL was only
12.0% compared to that of the liquid reference sample. This result
provided confidence that slight variations in the sample input volume
(e.g., from direct addition of a fingerstick rather than sample addition
by volumetric pipet) will not substantially impact quantitative results
if volumetric sample application is unavailable at the site of collection.
pDBS Cards Fill Independent of Hematocrit Value
We
aimed to further evaluate the effect of sample input on the quantitation
of hemoglobin by surveying the physiological range of hematocrit values
(20–60%). We anticipated that controlling the total area of
the pDBS card through patterning would minimize the negative effects
of variable sample spreading caused by the hematocrit. The direct
comparison of pDBS cards and unpatterned TFN clearly demonstrated
how the hematocrit influenced the results of standard assays such
as the quantitation of hemoglobin (Table ). Patterned cards yielded ≤7% error
across the full range of hematocrit values, while unpatterned cards
yielded 3-fold higher percent error at low hematocrit (21% error at
20% hematocrit, Table ). Inter- and intracard variations (i.e., spot-to-spot variation)
were consistent between both card types (Table S2), which suggested that the deviation in the quantitation
of hemoglobin can largely be attributed to uncontrolled sample spreading
of blood in unpatterned cards. Agreement between pDBS cards (Figure A) and unpatterned
TFN (Figure B) with
the reference liquid blood is represented by Bland–Altman plots.[40] The observed bias was reduced in pDBS cards
(−0.7 g/dL) compared to that of TFN (−1.0 g/dL). Similarly,
the limit of agreement was narrower for pDBS (2.2 g/dL) than for TFN
(3.0 g/dL) in comparison to the reference method.
Table 1
Comparison of pDBS Card and Unpatterned
Cardstock (TFN) for the Quantitation of Hemoglobin at Various Hematocrit
Valuesa
pDBS
card
unpatterned
TFN
reference method
Hct (%)
[Hgb] ± SD (g/dL)
error (%)
[Hgb] ± SD (g/dL)
error (%)
[Hgb] ± SD (g/dL)
20
6.2 ± 0.2
–7
5.3 ± 0.1
–21
6.7 ± 0.2
30
9.4 ± 0.2
+1
9.1 ± 0.4
–3
9.4 ± 0.9
40
12.3 ± 0.3
–5
12.3 ± 0.3
–5
12.9 ± 0.5
50
14.7 ± 0.5
–7
15.2 ± 0.5
–4
15.9 ± 0.6
60
18.2 ± 0.5
–7
17.4 ± 0.4
–11
19.6 ± 0.3
Data represent the average of
20 replicates ± standard deviation. Reference values using reference
method (liquid blood) are provided.
Figure 2
Bland and Altman plots for (A) pDBS card and (B) unpatterned TFN
from Table . Limits
of agreement are represented by dotted lines from −1.96 to
+1.96 s.
Data represent the average of
20 replicates ± standard deviation. Reference values using reference
method (liquid blood) are provided.Bland and Altman plots for (A) pDBS card and (B) unpatterned TFN
from Table . Limits
of agreement are represented by dotted lines from −1.96 to
+1.96 s.Patterned cards reproducibly filled
four replicate collection punch
zones (6 mm diameter) across the full range of hematocrit values (Figure A). Since pDBS cards
filled independently of the hematocrit, four replicate punches can
always be collected for analysis and enable more tests to be performed
from a single card. In stark contrast, the diameter of the blood spot
in unpatterned TFN decreased with increasing hematocrit (20–60%
hematocrit) (Figure B). A direct consequence of the decreased blood spot diameter in
unpatterned DBS is one less technical replicate punch of dried blood
under idealized conditions (Figure C).
Figure 3
pDBS cards fill uniformly independent of hematocrit value.
Representative
images of (A) pDBS cards and (B) unpatterned TFN at various hematocrit
values (19.9–60.2%, N = 5). (C) pDBS cards
fill four replicate 6 mm collection punch zones compared to only three
replicate punches from unpatterned TFN independent of hematocrit value.
pDBS cards fill uniformly independent of hematocrit value.
Representative
images of (A) pDBS cards and (B) unpatterned TFN at various hematocrit
values (19.9–60.2%, N = 5). (C) pDBS cards
fill four replicate 6 mm collection punch zones compared to only three
replicate punches from unpatterned TFN independent of hematocrit value.
Quantitation of Sodium by ICP-AES
Blood sodium levels
are routinely measured as part of a basic metabolic panel that often
includes additional electrolytes such as calcium, chloride, and potassium.
The accurate quantitation of sodium is critical for controlling blood
pressure and evaluating proper nerve and muscle function.[41] Additionally, because sodium is found both intra-
and extracellularly, it represented an attractive analyte class to
further evaluate the quantitative capabilities of pDBS cards. The
concentration of sodium in blood samples obtained from pDBS cards
(1715 ± 21 ppm) was nearly identical with the concentration in
the reference liquid sample (1810 ± 24 ppm), suggesting that
there is no apparent loss or evaporative concentration of sodium to
the TFN paper (Figure ). A two-tailed Student’s t test yielded
a p-value of 0.26, providing no evidence of a statistically
significant difference in sodium concentration between the dried and
liquid blood samples. The clinical reference range for sodium in blood
is 135–145 mEq/L.[42] Both the dried
and liquid blood samples fell below the expected range with 74.6 and
78.7 mEq/L sodium, respectively. Both samples were prepared using
nitric acid digestion, which included multiple liquid handling and
quantitative transfer steps, which could account for the low observed
concentrations. While the range of concentrations of sodium extracted
from pDBS punches (683.9 ppm) was more dispersed than those from liquid
samples of blood (392.4 ppm), the standard deviation was slightly
less. The comparison of variances (F-test) yielded a p-value of 0.27, indicating no significant difference between the
variance of the data sets. Therefore, the precision of pDBS card microsampling
could be amenable to use of calibration standards for quantitative
results.
Figure 4
Quantitation of sodium by ICP-AES. Ten replicate samples were prepared
using (i) pDBS cards or (ii) reference method with liquid whole blood.
A two-tailed Student’s t test yielded a p-value of 0.26, providing no evidence of a difference in
recovered sodium concentration between pDBS card (dried blood) and
the reference method (liquid blood).
Quantitation of sodium by ICP-AES. Ten replicate samples were prepared
using (i) pDBS cards or (ii) reference method with liquid whole blood.
A two-tailed Student’s t test yielded a p-value of 0.26, providing no evidence of a difference in
recovered sodium concentration between pDBS card (dried blood) and
the reference method (liquid blood).
Quantitation of Amino Acids by HPLC
Amino acid analysis
via DBS sampling is commonly used for the detection of various inborn
errors of amino acid metabolism including phenylketonuria (PKU) in
newborns. Efforts to streamline and improve the quantitation of amino
acids from DBS have been extensively reported.[43−45] For demonstrative
purposes, we selected three representative hydrophobic amino acids
(e.g., tryptophan, leucine, and proline) and one basic—or positively
charged—amino acid (e.g., lysine) for analysis. The recovery
of each amino acid from pDBS cards was determined by the ratio of
extracted analyte concentration (μM) and liquid reference concentration
(μM) as analyzed by HPLC. Two distinct sample groups (e.g.,
20 and 40% hematocrit) were selected to represent (i) a high liquid-to-cell
ratio—which can be prone to underestimating analytes of interest—and
(ii) the average hematocrit obtained from our panel of healthy donors,
respectively. Each amino acid yielded excellent recovery for both
blood sample groups (Table ). While most samples fell in the range 82–93% recovery,
two samples yielded higher concentrations when extracted from pDBS
cards compared to liquid reference samples (proline 115%, lysine 102%).
Resultant loss and variability in analyte recovery may be attributed
to the number of liquid handing steps required to extract, process,
and derivatize samples prior to analysis by HPLC. However, all reported
values in Table are
in agreement with other reports where the recovery of amino acids
ranged from 84.2 ± 22.2 to 96.0 ± 12.0%.[46] Additionally, the evaluation of interassay precision (i.e.,
card-to-card comparison) demonstrated coefficients of variation (%
CV) for tryptophan of 0.8–5.2%, leucine of 2.6–6.7%,
proline of 1.0–5.5%, and lysine of 5.4–5.7% (Table S3). These % CV values are considerably
improved in comparison to recently reported % CV values for amino
acid analysis by traditional DBS sampling using similar methods (e.g.,
% CV for leucine 8.3–15.3%).[44] The
successful quantitation and improved interassay precision of select
amino acids by HPLC supported the enhanced sampling capabilities of
pDBS cards.
Table 2
Quantitation of Amino Acids by HPLCa
20%
hematocrit
40%
hematocrit
pDBS
reference
recovery (%)
pDBS
reference
recovery
(%)
[Trp] ± SEM (μM)
116 ± 6
124 ± 5
93
127 ± 1
137 ± 5
93
[Leu] ± SEM (μM)
30 ± 2
36 ± 2
85
39 ± 1
36 ± 1
90
[Pro] + SEM (μM)
73 ± 4
89 ± 6
82
98 ± 1
85 ± 3
115
[Lys] ± SEM (μM)
37 ± 2
36 ± 2
102
35 ± 2
40 ± 2
87
Data for liquid blood reference
represent the average of five replicates ± standard error of
the mean (SEM). Data for pDBS cards represent the average of four
replicates ± SEM. Recovery was calculated with respect to liquid
blood reference samples.
Data for liquid blood reference
represent the average of five replicates ± standard error of
the mean (SEM). Data for pDBS cards represent the average of four
replicates ± SEM. Recovery was calculated with respect to liquid
blood reference samples.
Conclusions
We aimed to develop a device that can improve the sampling of whole
blood at the point-of-care while maintaining current clinical protocols
for DBS analysis. Our approach comprised wax-patterned DBS cardstock
to restrict the flow and distribution of whole blood with four defined
extraction zones. Controlling the flow of blood in the pDBS card allowed
for reproducible filling across the full range of hematocrit values
and reduced the sampling bias for pDBS cards compared to unpatterned
TFN cardstock. Specifically, the accuracy for the quantitation of
hemoglobin with low hematocrit (20%) was improved by 3-fold using
pDBS cards. Sampling was further improved by spatially defining extraction
zones, which consistently produced four replicate 6 mm diameter sub-punches
from a single application of blood (75 μL), independent of the
hematocrit value. We designed these cards to accommodate the direct
application of fingerstick volumes of blood and modeled ideal conditions
by dispensing blood using a volumetric pipet. The highly controlled
nature of this method of sample dispensing may be reflective of the
conserved inter- and intracard variations reported for both pDBS and
traditional DBS cards. We anticipate that the patterned features of
pDBS cards will maintain uniform filling and address the reported
challenges associated with applying fingersticks to DBS at the point-of-care.Surveying common DBS analytical techniques such as ICP-AES and
HPLC indicated good agreement with liquid reference samples for the
quantitation of sodium and select amino acids, respectively. Additionally,
we were able to process and analyze samples of whole blood without
changing the recommended handling procedures for DBS cards (i.e.,
amenable to automated punching machines). Standardizing the sample
output from pDBS cards could expand the number of tests performed
from a single sample collection or permit increased numbers of technical
replicates compared to traditional unpatterned DBS cards. Beyond the
classes of analytes and techniques demonstrated in this manuscript,
quantitative DBS sampling has the potential for myriad applications
related to molecular amplification (e.g., screening for viral diseases),
nutritional evaluations, immunologic studies, pharmacokinetics, therapeutic
drug monitoring, and genetic testing.[47]Since pDBS cards are exposed to ambient conditions during
sample
application, spreading, and drying, we expect that performance may
vary under certain environmental conditions at the time of collection,
as similarly experienced with traditional DBS cards. For example,
sample spreading may be reduced due to extremely dry conditions (relative
humidity below 10%) or high temperatures, which could cause excessive
evaporation. However, this effect is commonplace for DBS technologies
and is not identified as a major obstacle for ubiquitous use.[6] Additionally, hydrophobic wax barriers used in
this manuscript are incompatible with organic solvents (e.g., methanol,
acetonitrile, ethanol) and harsh surfactants (e.g., Triton X-100 and
sodium dodecyl sulfate). Removing paper zones from the pDBS card via
punches prior to elution of analyte will avoid leeching wax, or other
material used to generate hydrophobic barriers, into the final tested
sample. While the pDBS card presented here was used for sampling whole
blood, we anticipate that we could expand on this approach to collect
and store additional sample types such as saliva, tears, or blood
plasma to provide enhanced sampling and quantitative analysis in a
workflow that connects the point-of-care to a clinical laboratory
infrastructure.