N Spooner1, A Olatunji2, K Webbley2.
Abstract
When using dried blood spot (DBS) sampling for the quantitative bioanalysis of circulating concentrations of drugs, metabolites and endogenous analytes, it is important that a fixed volume of blood is deposited to overcome the issues associated with blood hematocrit (HCT) and homogeneity. The volumetric performance of the KTH DBS collection device was tested with radiolabelled [14C]-diclofenac. It was demonstrated that the device deposits a fixed volume of blood (13.5μL) regardless of the HCT (25-65%), or lipid content of the blood sample. Further, it was found that the precision and accuracy of the derived dried blood samples were at least as good as those of a positive displacement pipette. The device was found to be easy to use and gave acceptable dried samples for 92.9% of the tests performed (n=42).Entities:
Keywords: Bioanalysis; Dried blood spot; Hematocrit; Microsampling; Patient centric sampling
Mesh:
Substances:
Year: 2017 PMID: 29154197 DOI: 10.1016/j.jpba.2017.11.036
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935