Alterations in Corneal Biomechanical and Topographic Features After Accelerated Crosslinking: 1-Year Results.

Ahmet Kirgiz1, Sevil Karaman Erdur2, Kubra Serefoglu Cabuk1, Kursat Atalay3, Senay Asik Nacaroglu3.   

Abstract

OBJECTIVES: To determine the biomechanical and topographic alterations within the first year after accelerated crosslinking (CXL) treatment in patients with keratoconus.
METHODS: In this prospective study, 52 eyes of 52 patients with progressive keratoconus underwent accelerated CXL were included. All patients had a detailed preoperative ophthalmologic examination, including slit-lamp evaluation, Goldmann tonometry, fundoscopy, topography by Scheimpflug imaging (Sirius), and corneal biomechanical evaluation with a biomechanical waveform analysis device (ORA). Alterations in visual acuity and topographic findings were evaluated before the treatment and at 12 months follow-up. Corneal biomechanical features were obtained before the treatment, and at 1st, 3rd, 6th and 12th months.
RESULTS: Uncorrected-visual acuity and best-corrected visual acuity both statistically significantly improved at 12th month (p=0.001). There were no statistically significant differences in keratometry values, whereas maximum K (AKfront) and symmetry index front (SIfront) decreased significantly (p=0.015 and p=0.009, respectively). Corneal thinnest point and volume also decreased significantly at 12th month (p=0.001 for both). Goldmann-correlated intraocular pressure (IOPg) and corneal compensated IOP (IOPcc) values transiently increased in the first three months, while corneal hysteresis (CH) and the corneal resistance factor (CRF) transiently decreased, with the difference not statistically significant (p>0.05). However, central corneal thickness significantly decreased at the end of the 12th month (p=0.001).
CONCLUSION: Accelerated CXL seems to be effective in stopping the progression of keratoconus. Our findings showed transient alterations in biomechanical features, which will end with the preoperative values at the end of the 12th month. Further studies are needed to demonstrate the changes in corneal biomechanics in vivo. Copyright:
© 2019 by Beyoglu Eye Training and Research Hospital.

Entities:  

Keywords:  Accelerated crosslinking; corneal biomechanics; corneal hysteresis; corneal topography; keratoconus

Year:  2019        PMID: 35187443      PMCID: PMC8842046          DOI: 10.14744/bej.2019.44154

Source DB:  PubMed          Journal:  Beyoglu Eye J        ISSN: 2459-1777


Introduction

Keratoconus is a progressive, usually bilateral corneal ectasia characterized by corneal thinning and protrusion, progressive myopia, and irregular astigmatism (1). The only way shown to stop progression is the corneal collagen crosslinking (CXL) (2, 3). CXL is a minimally invasive procedure that increases the biomechanical strength of the cornea through the induction of photochemically triggered cross-links within the collagen network using riboflavin and ultraviolet-A radiation (4, 5). Conventional CXL was determined as a safe and efficient way of treatment in keratoconus in many studies (5, 6). However, the procedure lasts approximately one hour. Theoretically, the same energy dose can be delivered in a shorter time with an accelerated modality (7, 8). In accelerated CXL, reported as the rapid, effective and safe alternative in keratoconus treatment, the same energy dose ensuring a proportional biological effect is delivered in a shorter exposure time by setting different UV-A powers (9, 10). The UV-A power could be set to 30, 18 or 9 mW/cm2 to radiate cumulative dose of 5.4 J/cm2 in a shorter time, such as 3, 5 or 10 minutes, respectively. With the increasing use of these accelerated CXL protocols, studies on the long-term effects have begun to be published in recent years (11, 12). Although it is shown in ex vivo that the rigidity of the cornea is increased 4.5 times in the human cornea by CXL (13), it is important to evaluate the changes in corneal biomechanics after the treatment in vivo to understand the efficacy of the procedure. It is theoretically possible to evaluate the corneal biomechanical properties that may occur in the cornea with the ocular response analyzer (ORA) designed to obtain in vivo measurements (14). While several studies report the biomechanical changes after conventional CXL with the ORA device (15–21), the data about the results of accelerated CXL regarding the biomechanical features of the cornea are limited in the literature (22). In this study, we aimed to investigate the biomechanical and topographic alterations in 12 month period after accelerated collagen CXL treatment among patients with keratoconus.

Methods

In this prospective study, 52 eyes of 52 patients with progressive keratoconus who underwent accelerated CXL between July 2014 and January 2015, in Bagcilar Education and Research Hospital, Istanbul, Turkey were included. Patients who were followed up for 12 months regularly were included in this study. This study was approved by the local Ethics Committee. Progressive keratoconus was defined as an increase of 1.00 diopter (D) or more in the steepest keratometry (K) measurement or the loss of at least two lines of best-corrected distance visual acuity in last 1 year (23). The exclusion criteria were as follows: patients younger than 18 or older than 40 years of age, patients with a corneal thickness less than 400 μ at the thinnest point, patients with corneal scarring or severe dry eye, patients having a history of corneal surgery, or pregnant or lactating women during the course of this study. All patients had a detailed preoperative ophthalmologic examination, including slit-lamp evaluation, Goldmann tonometry, fundoscopy, topography by Scheimpflug imaging (Sirius, Schwind Eye-Tech-Solutions GmbH, Germany), corneal biomechanical properties and intraocular pressure evaluation by Ocular Response Analyzer (Reichert Technologies, Buffalo, NY, USA). Visual acuity was obtained with Snellen charts and converted to LogMAR units for analysis. Uncorrected visual acuity (UCVA) and best-corrected visual acuity (BCVA) were both recorded. The Sirius, a topographic device consisting of a combination of two rotating Scheimpflug cameras and a Placido disk, was used for topographical and central corneal thickness (CCT) measurements. Flat keratometry value (K1), steep keratometry value (K2), average keratometry value (Kavg), topographic cylindrical value (Cyl), apical keratoscopy front (AKfront), apical keratoscopy back (AKback), symmetry index front (SIfront), symmetry index back (SIback), corneal thinnest point (CTP), and corneal volume (CV) were recorded before the treatment and at 12 months follow-up. Apical keratoscopy front (AKfront) was defined as the steepest point of the anterior corneal surface, while apical keratoscopy back (AKback) was defined as the steepest point of the posterior corneal surface. Symmetry index front (SIfront), the symmetry index of the anterior curvature, was defined as the difference of the mean anterior tangential curvature of two circular zones centered on the vertical axis in the inferior and superior hemispheres. Symmetry index back (SIback), the symmetry index of the posterior curvature, was defined as the difference of the mean posterior tangential curvature of two circular zones centered on the vertical axis in the inferior and superior hemispheres. All measurements were taken with the manufacturer’s guideline by the same trained examiner. The Ocular Response Analyzer (ORA) reports two intraocular pressure values: Goldmann-correlated IOP (IOPg) and corneal compensated IOP (IOPcc). Two different bidirectional (inward and outward) applanation pressure measurements were recorded by ORA. The difference between these two pressures is the corneal hysteresis (CH), which is an indicator of corneal viscosity. Corneal resistance factor (CRF) is considered as the indicator of the overall resistance of the cornea that is mainly associated with the elastic properties of the cornea. Alterations in visual acuity and topographic findings of study participants were evaluated before the treatment and at 12th month of follow-up. Corneal biomechanical properties were obtained before the treatment and at 1, 3, 6, and 12 months following accelerated CXL.

Surgical Technique

Accelerated crosslinking has been performed to all patients. Initially, a topical anesthetic agent (Alcaine; Alcon Laboratories, Inc.) was administered and the central 8.0 mm epithelium was removed with a blunt spatula. Riboflavin with dextran (0.1% in 20% dextran solution, Medicross, Germany) was administered topically every three minutes for 30 minutes. The cornea was aligned and exposed to UV-A 365 nm light for five minutes at an irradiance of 18 mW/cm2 (Peschke Meditrade, GmbH) (11). During UV-A exposure, riboflavin administration was continued every minute. After the treatment, the eye surface was washed with 20.0 mL of a balanced salt solution. Postoperatively, antibiotic (Vigamox; Alcon Co., Inc.), corticosteroid drops (Lotemax; Bausch & Lomb) and artificial tears (Refresh; Allergan) were administered; a soft contact lens bandage was placed. The contact lens was removed after the closure of the epithelial defect. Antibiotics and corticosteroid drops were continued four times daily for one week and two weeks, respectively.

Statistical Analysis

All analyses were performed with the Statistical Package for Social Sciences software (SPSS, Windows version 21.0; SPSS Inc.; Chicago, IL, USA). A P value of less than 0.05 was considered statistically significant. Topographic findings (K1, K2, Kavg, AKfront, AKback, SIfront, SIback, Cyl, CTP, and CV), BCVA and UCVA were compared using the paired Student-t test. Corneal biomechanical properties were compared using repeated-measures ANOVA with post hoc Tukey test. The correlations between CCT at 12 months, and each perioperative variable were evaluated using the Pearson correlation coefficient.

Results

Totally 52 eyes (28 right and 24 left) of 52 patients (18 female, 34 male) with a mean age of 24.15±6.75 (range: 18-39 years) years were included. Visual acuities and topographic findings of patients before the treatment and at 12 months follow-up are summarized in Table 1.
Table 1

Comparison of visual acuity and topographic findings between the preoperative and the 12 months of follow-up

Baseline Mean±SD12th month Mean±SDP*
UCVA (LogMAR)0.71±0.320.55±0.32<0.001
BCVA (LogMAR)0.42±0.220.26±0.14<0.001
K1 (D)45.4±1.845.4±20.805
K2 (D)48.6±2.248.6±2.40.837
Kavg (D)46.9±1.946.9±2.10.944
Cyl (D)3.2±1.43.1±1.40.198
AKfront (D)55.1±4.154.6±4.20.015
AKback (D)79.6±8.883.9±11.20.001
SIfront (D)6.1±2.85.8±2.80.031
SIback (D)1.5±0.71.6±0.70.013
CTP (μm)456±32424±480.001
CV (mm³)55.8±3.354.4±3.50.001

SD: standard deviation; UCVA: uncorrected visual acuity; BCVA: best corrected visual acuity; K1: flat keratometry value; K2: steep keratometry value; Kavg: average keratometry value; Cyl: topographic cylindrical value; AKfront: apical keratoscopy front; AKback: apical keratoscopy back; SIfront: symmetry index front; SIback: symmetry index back; CTP: corneal thinnest point; CV: corneal volume.

Paired samples t test.

Comparison of visual acuity and topographic findings between the preoperative and the 12 months of follow-up SD: standard deviation; UCVA: uncorrected visual acuity; BCVA: best corrected visual acuity; K1: flat keratometry value; K2: steep keratometry value; Kavg: average keratometry value; Cyl: topographic cylindrical value; AKfront: apical keratoscopy front; AKback: apical keratoscopy back; SIfront: symmetry index front; SIback: symmetry index back; CTP: corneal thinnest point; CV: corneal volume. Paired samples t test. There was a statistically significant improvement in the UCVA and BCVA after the 12-month visit (both p=0.001). The AKfront and SIfront values showed a significant decrease after 12 months compared with the preoperative values (p=0.015, and p=0.031, respectively). The CTP and CV also revealed a significant decrease at the end of 12-month visit (both p=0.001). Alterations in corneal biomechanical properties during the 12-month follow-up are given in Table 2. Regarding these alterations, IOPcc and IOPg values transiently increased in the first three months; whereas CH and CRF values transiently decreased in first three months, although the changes were statistically insignificant. However, CCT significantly decreased at the end of 12-month visit (p=0.001) (Fig. 1).
Table 2

Alterations in corneal biomechanical properties during follow-up after accelerated crosslinking

Preoperative1st month3rd month6th month12th monthP*
IOPcc (mmHg)
 Mean±SD13.8±2.414.5±2.914.2±2.813.7±2.913.5±2.50.21
 Range7.6-19.39.5-22.29.6-21.26.8-22.17.6-19
IOPg (mmHg)
 Mean±SD10.6±2.810.8±3.310.6±3.310.5±2.910.4±2.80.50
 Range4.3-16.75.6-19.43.5-17.35-183.2-16.3
CH (mmHg)
 Mean±SD8.5±1.48.1±1.38±1.68.4±1.58.6±1.20.39
 Range4.3-12.25.9-12.94.6-136-13.56-12
CRF (mmHg)
 Mean±SD7.3±1.77±1.77±1.77.2±1.57.4±1.60.58
 Range2.6-11.24-12.71.9-10.34-11.54.3-11
CCT¶ (mμ)
 Mean±SD476±39444±49437±46451±47459±470.001
 Range420-609371-634342-612366-621378-633

SD: standard deviation; IOPcc: corneal compensated intraocular pressure; IOPg: Goldmann-correlated intraocular pressure; CH: corneal hysteresis; CRF: corneal resistance factor; CCT: central corneal thickness ¶ measured with Sirius device.

Repeated measures ANOVA, post-hoc Tukey test.

Figure 1

Alterations in central corneal thickness (CCT) at postoperative 12 months after accelerated crosslinking.

Alterations in corneal biomechanical properties during follow-up after accelerated crosslinking SD: standard deviation; IOPcc: corneal compensated intraocular pressure; IOPg: Goldmann-correlated intraocular pressure; CH: corneal hysteresis; CRF: corneal resistance factor; CCT: central corneal thickness ¶ measured with Sirius device. Repeated measures ANOVA, post-hoc Tukey test. Alterations in central corneal thickness (CCT) at postoperative 12 months after accelerated crosslinking. As regards the correlations, there was significant correlation between the mean CCT at 12th month and baseline CCT, CTP, and CV (r=0.888, r=0.769, r=0.523, respectively, p=<0.001 for all); whereas no significant correlation was found between the mean CCT at 12th month and other parameters (p>0.05 for all).

Discussion

This study is one of the few studies in the literature evaluating biomechanical features along with topography changes following accelerated CXL reporting at least 1-year results (22). Both UCVA and BCVA statistically significantly improved in 12th months of follow up with significant alterations in AKfrontand SIfront values of the cornea, demonstrating the efficacy of this method in controlling keratoconus. On the other hand, CH and CRF remained unchanged after 12 months of the treatment suggesting these parameters may not represent the exact biomechanical effects of accelerated CXL. In CXL treatment, reactive oxygen species are thought to be produced, leading to the formation of covalent cross-links between collagen molecules, with a resulting stiffening of the stromal tissue. This strengthening of the corneal stroma decelerates the progression of keratoconus and ectasia and, in many cases, improves patients’ visual, refractive, and topographic outcomes (5, 24). In that aspect, some alterations especially in corneal biomechanical properties are expected. In the relevant literature, there are several studies about the alterations in topographic and corneal biomechanical properties after conventional CXL. Goldich et al. (15) analyzed 10 eyes in the first six months after conventional CXL treatment and consistent with our results, they also reported transient nonsignificant elevations in CH and CRF values after CXL. Sedaghat et al. (16) compared CH and CRF values in the first six months after conventional CXL for keratoconus in 56 eyes of 51 patients and did not determine any significant alterations in CH, CRF, CCT and IOPg, but IOPcc was reported to be decreased significantly. In another study of Goldich et al. (17), 14 patients were assessed after conventional CXL for 24 months and a significant improvement in BCVA with a decrease in the mean cylinder without any significant alterations in minimal corneal thickness, endothelial cell density, corneal hysteresis, or corneal resistance factor was reported. Greenstein et al. (18) investigated the corneal biomechanical changes in 69 eyes after conventional CXL for one year, and consistent with our results, they reported that there were not any alterations in CH or CRF values. In this study, we determined transient alterations in the first three months in IOPcc, IOPg, CRF and CH after accelerated CXL which were not statistically significant but a significant decrease in CCT in the 12th month of follow-up. In another recent study investigating corneal topographical and biomechanical properties in the 1st year after conventional CXL in 97 eyes and compatible with our results they also reported a significant improvement in UCVA and BCVA values with a significant decrease in CCT and corneal volume without any significant alterations in CH and CRF. However, diversely from our results, they reported a significant decrease in average keratometry value (19). Interestingly, Beckman et al. (20) investigated the biomechanical effects of conventional CXL in 29 eyes of 28 patients and reported increased CH values with applanation resonance tonometer but not with ORA. They also determined a significant decrease in CCT in 1st and 6th months of follow-up after treatment. Terai et al. (24) defined ORA as providing invaluable information for defining biomechanical conditions of the cornea. We used ORA to determine the corneal biomechanical properties in this study and did not determine any significant alterations in the corneal biomechanical properties during the first year after accelerated CXL. Consistent with our results, Çağıl et al. (25) did not find any significant alterations in CH and CRF but observed a significant decrease in CCT at postoperative six months after conventional CXL. In a recent and the only one study investigating the corneal biomechanical alterations after accelerated CXL, Uzel et al. (22) found out that second-generation parameters (p1 area, p2 area, h2, and dive 2) were more sensitive than CH and CRF with a new ORA software (version 2.04). CH is an indicator of corneal viscosity, and CRF is considered as the indicator of the overall resistance of the cornea that is mainly associated with the elasticity of the cornea. After CXL, accelerated or conventional, decrease in CH and/or CRF values is expected since CXL increases the biomechanical strength of the cornea. However, in this study, we determined that, though transient alterations in CH or CRF might be seen in first months after accelerated CXL, there was not a statistically significant increase at the end of 12th month compared with the preoperative values. There may be some reasons for these insignificant results. Firstly, the decrease in CCT may have prevented the accurate measurement of corneal biomechanics. Secondly, the ORA device may be inadequate in evaluating the corneal biomechanical properties due to the changing in corneal collagen configuration after CXL. The transient alterations seen in the first three months in CH and CRF may be due to the insufficient ORA signals which were probably affected by optical irregularity. The data about the effects of accelerated CXL are increasing in the literature. Elbaz et al. (26) assessed the topographical, visual, and refractive effects of accelerated CXL in keratoconus at 6th and12th months of follow-up and reported that there was not a statistically significant alteration in corrected distance visual acuity and mean refractive cylinder, but uncorrected distance visual acuity was improved. Moreover, all corneal parameters, including Ksteep, Kflat, average K, corneal astigmatism, were stable at 6 and 12 months in all patients. Mita et al. (27) evaluated the effectiveness of accelerated CXL for keratoconus in first six months after treatment in 39 eyes of 22 patients and reported a statistically significant improvement in UCVA and a significant decrease in the mean maximum keratometry readings without any changes in the endothelial cell density. Although alterations in 1st year after accelerated CXL were investigated in a few studies before, there are also some recent studies reporting the differences between conventional CXL and accelerated CXL. Chow et al. (28) compared one-year outcomes of 2 CXL methods in patients with progressive keratoconus and reported insignificantly flattened maximum and minimum keratometry values and significantly reduced corneal thickness in both groups. Moreover, similar to our results significant improvements in UCVA and BCVA were reported. Hashemi et al. (29) compared the six months results of 2 CXL methods in 62 eyes and reported similar clinical efficacies. Tomita et al. (30) reported that there were no statistically significant differences in the postoperative changes in the keratometric readings or the corneal biomechanical responses between the accelerated and conventional CXL. In correlation analysis, we found no statistically significant relation between CCT at 12 months versus preoperative corneal biomechanical properties. However, there was a significant correlation between CCT at 12 months versus preoperative CCT, CTP and CV. In that aspect, in all patients, preoperative CCT, CTP and CV should be evaluated carefully. Unlike with our results, Viswanathan et al. (31) determined a positive correlation of both CH and CRF with CCT in the normal, keratoconic and crosslinked eyes. Maximum keratometry value (AKfront) was defined as one of the topographic indicators of CXL success and reported to be stable or decreased after CXL (32). Likewise, significantly decreased AKfront values were also reported in this study in 12th-month control after the treatment showing that accelerated CXL is effective in stopping the progression of keratoconus. The main limitations of our study were the small sample size and lack of a control group consisting of patients who underwent conventional CXL. However, the investigation of both topographic and corneal biomechanical properties in the 12 month period is the strength of this study. In conclusion, in the 12th month after accelerated CXL, determination of improved UCVA and BCVA, together with the stable K values, decreased AKfront and SIfront suggests the long-term efficacy of this procedure. Our findings showed transient alterations in biomechanical features, which will end with the preoperative values at the end of 12th month. Further studies with larger groups and longer follow-up periods are required to determine the exact effects of accelerated CXL on corneal topography and advanced technologies are needed to assess the corneal biomechanical alterations more accurately.
  31 in total

1.  The Bunsen-Roscoe law for the human eye at very short durations.

Authors:  G S BRINDLEY
Journal:  J Physiol       Date:  1952-09       Impact factor: 5.182

2.  Determining in vivo biomechanical properties of the cornea with an ocular response analyzer.

Authors:  David A Luce
Journal:  J Cataract Refract Surg       Date:  2005-01       Impact factor: 3.351

3.  Long-term results of riboflavin ultraviolet a corneal collagen cross-linking for keratoconus in Italy: the Siena eye cross study.

Authors:  Aldo Caporossi; Cosimo Mazzotta; Stefano Baiocchi; Tomaso Caporossi
Journal:  Am J Ophthalmol       Date:  2010-02-06       Impact factor: 5.258

4.  Short-term comparison of accelerated and standard methods of corneal collagen crosslinking.

Authors:  Hassan Hashemi; Akbar Fotouhi; Mohammad Miraftab; Hooman Bahrmandy; Mohammad Amin Seyedian; Kazem Amanzadeh; Shahab Heidarian; Hamidreza Nikbin; Soheila Asgari
Journal:  J Cataract Refract Surg       Date:  2015-03       Impact factor: 3.351

5.  High-irradiance accelerated collagen crosslinking for the treatment of keratoconus: six-month results.

Authors:  Mariko Mita; George O Waring; Minoru Tomita
Journal:  J Cataract Refract Surg       Date:  2014-06       Impact factor: 3.351

6.  Equivalence of biomechanical changes induced by rapid and standard corneal cross-linking, using riboflavin and ultraviolet radiation.

Authors:  Silvia Schumacher; Lydia Oeftiger; Michael Mrochen
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-11-25       Impact factor: 4.799

7.  Increased corneal hysteresis after corneal collagen crosslinking: a study based on applanation resonance technology.

Authors:  Jeannette Beckman Rehnman; Anders Behndig; Per Hallberg; Christina Lindén
Journal:  JAMA Ophthalmol       Date:  2014-12       Impact factor: 7.389

8.  Can we measure corneal biomechanical changes after collagen cross-linking in eyes with keratoconus?--a pilot study.

Authors:  Yakov Goldich; Yaniv Barkana; Yair Morad; Morris Hartstein; Isaac Avni; David Zadok
Journal:  Cornea       Date:  2009-06       Impact factor: 2.651

9.  Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking.

Authors:  Gregor Wollensak; Eberhard Spoerl; Theo Seiler
Journal:  J Cataract Refract Surg       Date:  2003-09       Impact factor: 3.351

10.  Evaluation of Corneal Biomechanical Changes After Collagen Crosslinking in Patients with Progressive Keratoconus by Ocular Response Analyzer.

Authors:  Raciha Beril Küçümen; Berna Şahan; Canan Aslı Yıldırım; Ferda Çiftçi
Journal:  Turk J Ophthalmol       Date:  2018-09-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.