| Literature DB >> 35174572 |
Jeanne E Poole1, Marye J Gleva2, Ulrika Birgersdotter-Green3, Kelley R H Branch1, Rahul N Doshi4, Tariq Salam5, Thomas C Crawford6, Mark E Willcox7, Arun M Sridhar1, Ghiath Mikdadi8, Sean C Beinart9, Yong-Mei Cha10, Andrea M Russo11, Ron K Rowbotham12, Joseph Sullivan12, Laura M Gustavson12, Kaisa Kivilaid13.
Abstract
INTRODUCTION: A wearable cardioverter defibrillator (WCD) is indicated in appropriate patients to reduce the risk for sudden cardiac death. Challenges for patients wearing a WCD have been frequent false shock alarms primarily due to electrocardiogram noise and wear discomfort. The objective of this study was to test a contemporary WCD designed for reduced false shock alarms and improved comfort.Entities:
Keywords: defibrillator; sudden cardiac death; ventricular arrhythmia
Mesh:
Year: 2022 PMID: 35174572 PMCID: PMC9305432 DOI: 10.1111/jce.15417
Source DB: PubMed Journal: J Cardiovasc Electrophysiol ISSN: 1045-3873 Impact factor: 2.942
A‐WCD detection algorithm performance using an ECG database during preclinical evaluation
| Rhythm | Test sample size (minimum required | Performance goal | Observed performance | 90% One‐sided lower confidence limit (minimum) |
|---|---|---|---|---|
| Shockable | ||||
| Coarse VF |
204 (200) | >90% sensitivity | 99.0% |
97.4% (87%) |
| Rapid VT |
62 (50) | >75% sensitivity | 98.4% |
93.9% (67%) |
| Nonshockable | ||||
| Normal sinus rhythm |
132 (100) | >99% sensitivity | 100% |
98.3% (97%) |
| AF, sinus bradycardia, SVT, heart block, idioventricular, PVCs |
219 (30) | >95% specificity | 96.3% |
94.1% (88%) |
| Asystole |
169 (100) | >95% specificity | 97.6% |
95.3% (92%) |
Note: Preclinical determination of VT/VF sensitivity and specificity was performed in accordance with the AHA recommendations for AED performance (sensitivity/specificity) as required by IEC 60601‐2‐4. A rich ECG database included rhythm segments derived from a variety of sources including spontaneous cardiac arrest from Emergency Medical System recordings, previously recorded ECGs from an EP lab, and prospective EP lab recordings with electrodes in the ASSURE WCD locations. Rhythm samples with noise greater than 25% of the QRS amplitude were excluded from the database as per the recommendations outlined by the AHA to test rhythm segments free of substantial noise.
Abbreviations: AED, automatic external defibrillator; AHA, American Heart Association; A‐WCD, ASSURE® WCD System; ECG, electrocardiogram; EP, electrophysiology; PVCs, premature ventricular complexes; SVT, supraventricular tachycardia; VF, ventricular tachycardia; VT, ventricular fibrillation.
American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy.
ASSURE system nominal therapy zone settings (VT 170 bpm, VF 200 bpm).
Figure 1ASSURE® WCD System (A‐WCD) noise management. The A‐WCD employs three levels of protection to achieve a low false alarm rate due to noise. Level 1 (blue)—minimize noise; Level 2 (red)—detect and remove noise that does occur; and Level 3 (yellow)—allow time for remaining noise to subside before alarming
Figure 2Electrocardiogram (ECG) sensing. (A) Five ECG electrodes are positioned circumferentially around the torso at the level of the subxiphoid process, labelled left front (LF), right front (RF), left back (LB), right back (RB), and right leg drive (RLD). Red arrows represent the four differential ECG vectors derived using RLD as a ground reference. (B) Garment interior depicting five embedded, cushioned ECG electrodes and defibrillation pads (two posterior and one anterior)
Figure 3Matched filter training. The black line shows the filter kernel. The colored lines show detected complexes
Figure 4ASSURE® WCD System (A‐WCD) segment with successful noise filtering. Example electrocardiogram segment from a 43‐year‐old female with dual‐chamber implantable cardioverter defibrillator (ICD) implanted for secondary prevention. ICD interrogation at the final 30‐day follow‐up revealed 3% atrial pacing and 99% ventricular pacing. (A) Signal contains baseline wander that is successfully removed by the algorithm filters. (B) Two channels were disqualified from segment analysis due to poor electrode contact (lead off). (C) This segment was categorized by the algorithm as normal sinus rhythm with a heart rate of 88 bpm and wide R‐wave. The episode was closed without a shock alarm event marker recorded
Figure 5Initial detection and therapy timeline. The ASSURE® WCD System has two independently programmable therapy zones. Initial detection (gray) is the same for each zone (5/6 overlapping 4.8‐s segments of ventricular tachycardia [VT] or ventricular fibrillation [VF]). Confirmation (yellow) is dependent on zone; 2/2 segments for VF and 15/19 segments for VT. VT that is determined to be disorganized will be treated as slow VF requiring 2/2 segments for confirmation. Once VT or VF is confirmed, a shock alarm event marker is recorded, and the shock alarm sequence (red) is initiated. If 4/6 segments during the shock alarm period are VT or VF, a shock will be delivered. Four out of five non‐VT or VF segments will close an episode. After shock delivery, if an additional 3/5 VT or VF segments are detected, another shock will be delivered (up to five consecutive shocks) and the episode closed if 6/12 non‐VT or VF segments are detected. Electrocardiogram episode storage includes 2 min of onset before episode open, through confirmation and shock delivery and 1‐min post conversion. The shock alarm sequence consists of a triple‐sensory indicator: (1) a flashing red light and shock icon on the monitor, (2) an intense vibration from the alert button, and (3) siren and voice prompts. The rate thresholds are programmable down to 130 bpm for VT and 180 bpm for VF. For this study, the heart rate thresholds were set at nominal (VT at 170 bpm and VF at 200 bpm), and the shock alarm sequence was programmed off
Baseline patient characteristics
| Characteristics | Value ( |
|---|---|
|
| |
| Age (years), mean ± SD (range) | 61.2 ± 11.4 (29–89) |
| Female sex, | 40 (30.8) |
| Race, | |
| White | 83 (63.8) |
| Black/African American | 35 (26.9) |
| Not reported | 11 (8.5) |
| Other | 1 (0.8) |
|
| |
| Indication for an ICD, | |
| Primary prevention | 105 (80.8) |
| Secondary prevention | 25 (19.2) |
| Cardiomyopathy, | |
| Ischemic | 75 (57.7) |
| Nonischemic | 44 (33.8) |
| Mixed ischemic/nonischemic | 1 (0.8) |
| Primary valvular | 2 (1.5) |
| Arrhythmogenic right ventricular cardiomyopathy | 0 (0.0) |
| Hypertrophic cardiomyopathy | 3 (2.3) |
| Congenital | 0 (0.0) |
| Sarcoidosis | 0 (0.0) |
| Other | 5 (3.8) |
| Primary electrical disorder, | 0 (0.0) |
| Left ventricular ejection fraction %, mean ± SD | 28.2 ± 7.1 |
| Comorbidities | |
| Hypertension | 96 (73.8) |
| History of heart failure | 125 (96.2) |
| History of coronary artery disease | 95 (73.1) |
| Diabetes | 41 (31.5) |
| Chronic obstructive pulmonary disease | 28 (21.5) |
| Chronic kidney disease | 31 (23.8) |
| Body mass index, mean ± SD kg/m2 (range) | 31.4 ± 5.9 (20.5–54.4) |
|
| |
| Prior ICD‐detected VT/VF, | 71 (54.6) |
| Atrial fibrillation/flutter, | 51 (39.2) |
| Bradycardia requiring pacing support, | 9 (6.9) |
| Previous LifeVest use, | 21 (16.2) |
| QRS duration, mean ± SD (ms) (range) | 113.8 ± 24.7 (71.0–198.0) |
| Left bundle branch block, | 7 (5.9) |
| Right bundle branch block, | 14 (11.8) |
| Intraventricular conduction delay, | 13 (10.9) |
| Paced, | 11 (8.5) |
Abbreviations: ICD, implantable cardioverter defibrillator; VF, ventricular fibrillation; VT, ventricular tachycardia.
Categories not mutually exclusive.
Summary of 163 WCD recorded episodes in 18 patients
| CEC adjudicated rhythm type |
WCD episodes, |
WCD patients | WCD shock alarm event marker |
|---|---|---|---|
|
|
|
|
|
|
|
|
| |
| Atrial fibrillation | 1 (0.6%) | 1 (0.7%) | 0 |
| Atrial flutter | 0 (0.0%) | 0 (0.0%) | 0 |
| Paced rhythm | 29 (17.8%) | 4 (3.1%) | 2—False |
| Sinus rhythm | 85 (52.1%) | 11 (8.5%) | 1—False |
| SVT | 6 (3.7%) | 2 (1.5%) | 0 |
|
|
|
|
|
|
|
|
|
|
Abbreviations: CEC, Clinical Events Committee; SVT, supraventricular tachycardia; VF, ventricular fibrillation; VT, ventricular tachycardia.
Patients may have had one or more episodes.
Although noise was present, the underlying rhythm could be discerned.
In all episodes, the heart rate was ≤120 bpm in the presence of substantial noise.
SVT was defined as a non‐VT/VF rhythm with a heart rate >100 bpm. In one patient, a single episode was initially detected as VT, but following resolution of noise, no longer met the criteria for VT/VF. The other patient had five episodes in which substantial noise was present throughout. The underlying rhythm was adjudicated as SVT between 166 and 172 bpm.
Detection was due to oversensing of large amplitude atrial flutter waves.