Dibenzo-18-crown-6-based carbon paste sensors for the nanomolar potentiometric determination of daclatasvir dihydrochloride: An anti-HCV drug and a potential candidate for treatment of SARS-CoV-2.

Yomna M Ahmed1, Sayed S Badawy1, Fatehy M Abdel-Haleem1,2.   

Abstract

Daclatasvir dihydrochloride (DAC) is an anti-hepatitis C virus (HCV) drug that has recently proven to be a promising candidate for the treatment of SARS-CoV-2. Still, there is a lack of sensitive potentiometric methods for its determination. In this work, carbon paste sensors based on dibenzo-18-crown-6 (DB18C6) were fabricated and optimized for the sensitive and selective potentiometric determination of DAC in Daclavirocyrl® tablets, serum, and urine samples. The best performance was obtained by two sensors referred to as sensor I and sensor II. Both sensors exhibited a wide linear response range of 5×10-9 - 1×10-3 mol/L, and Nernstian slopes of 29.8 ± 1.18 and 29.5 ± 1.00 mV/decade, with limits of detection, 4.8×10-9 and 3.2×10-9 mol/L, for the sensors I and II, respectively. Sensors I and II displayed fast response times of 5-8 and 5-6 s, respectively, with great reversibility and no memory effect. Moreover, the sensors exhibited a lifetime of 16 days. For the study of sensors morphology and elucidation of the interaction mechanism, the scanning electron microscope (SEM), Fourier-transform infrared spectroscopy (FTIR), and nuclear magnetic resonance (1H NMR) techniques were performed. A selectivity study was performed, and the proposed sensors exhibited good discrimination between DAC and potentially coexisting interferents with sensor II displaying better selectivity. Finally, sensor II was successfully applied for the determination of DAC in the above-mentioned samples, with recovery values ranging from 99.25 to 101.42%, and relative standard deviation (RSD) values ranging from 0.79 to 1.53% which reflected the high accuracy and precision.
© 2022 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Anti-HCV; Carbon paste electrodes; Daclatasvir; Dibenzo-18-crown-6; Potentiometry; SARS-CoV-2

Year:  2022        PMID: 35169329      PMCID: PMC8830182          DOI: 10.1016/j.microc.2022.107276

Source DB:  PubMed          Journal:  Microchem J        ISSN: 0026-265X            Impact factor:   4.821


Introduction

Hepatitis C virus infection is a worldwide health problem where its seroprevalence was estimated to be 1% of the world’s population[1]. HCV is a major cause of acute and chronic liver disease such as chronic hepatitis, cirrhosis, fibrosis, and hepatocellular carcinoma [2]. HCV infection is endemic in Egypt with a seroprevalence of 40% in some areas making it the highest prevalence rate in the world [3]. For this reason, Egypt developed a national strategy to control and treat HCV infection which aligns with WHO’s “global health sector strategy on viral hepatitis”, using highly effective regimens of direct-acting antiviral (DAA) therapy such as daclatasvir dihydrochloride, Scheme 1 [4].
Scheme 1

Structure of daclatasvir dihydrochloride. Chemical name: methyl N-[(1S)-1-[(2S)-2-[5-[4-[4-[2-[(2S)-1-[(2S)-2-(methoxycarbonylamino)-3-methyl-butanoyl]pyrrolidin-2-yl]-1H-imidazol-5-yl]phenyl]phenyl]-1H-imidazol-2-yl]pyrrolidine-1-carbonyl]-2-methyl-propyl]carbamate dihydrochloride. The numbers don’t indicate IUPAC nomenclature, but just for assigning the 1H NMR peaks. Ph: phenyl, Im: imidazolyl and Pyr: pyrrolidine.

Structure of daclatasvir dihydrochloride. Chemical name: methyl N-[(1S)-1-[(2S)-2-[5-[4-[4-[2-[(2S)-1-[(2S)-2-(methoxycarbonylamino)-3-methyl-butanoyl]pyrrolidin-2-yl]-1H-imidazol-5-yl]phenyl]phenyl]-1H-imidazol-2-yl]pyrrolidine-1-carbonyl]-2-methyl-propyl]carbamate dihydrochloride. The numbers don’t indicate IUPAC nomenclature, but just for assigning the 1H NMR peaks. Ph: phenyl, Im: imidazolyl and Pyr: pyrrolidine. Daclatasvir dihydrochloride inhibits HCV replication by binding to the N-terminus of NS5A[5], [6], and HCV NS5B RNA polymerase which might resemble RdRp of SARS-CoV-2[7]. Surprisingly, recent docking and clinical studies showed that DAC, in addition to another DAA called sofosbuvir, has remarkable binding interactions with SARS-CoV-2 enzymes, and reduces the mortality rates, improves clinical symptoms, and increases the chance of patients’ recovery[6], [7], [8]. According to Carolina et. al., DAC inhibits the polymerase reaction catalyzed by the SARS-CoV-2 RdRp complex, and inhibits the production of infectious SARS-CoV-2 titers across different cell types, including pneumocytes[9]. The inhibition of the infectious SARS-CoV-2 production in different cells by DAC, is especially significant during the initial stages of the disease and before the invasion of the virus into the parenchymal cells of the lung[6]. That’s why there is an urgent need for simple, fast, sensitive, and accurate method for determination of DAC in bulk, pharmaceutical formulations, serum, and urine. Several methods have been adopted for determination of DAC including, voltammetric [10], which has very narrow working concentration range (7.0 – 65.0 nM), spectral [11], and chromatographic methods [12], which are expensive, require sophisticated instruments, involve complex procedures, several sample manipulation steps, time-consuming, and in many cases are destructive to the sample. Potentiometry using ion-selective electrodes (ISEs) on the other hand offers more advantages such as low cost, portability of devices, fast response, wide dynamic range, and compatibility with online monitoring [13] [62]. To the best of our knowledge, only two potentiometric methods were reported for the determination of DAC [14], [15], but neither of them offered nanomolar determination of DAC, nor applied for its determination in serum samples. Thus, there is a need for development of a new potentiometric technique that offers more sensitivity, and applicable in body fluids. Carbon paste electrodes (CPEs) offer distinctive advantages over the conventional liquid polymeric membrane electrodes such as low ohmic resistance, the ability of surface renewal, and no need for internal reference solution[63]. In addition, carbon paste represents almost the most flexible electrode material for effective modification[17] [63] to improve the analytical performance of the sensor. Lipophilic ionophores represent one of the most common modifiers for CPEs, they include an increasing group of agents, both natural, such as valinomycin, and synthetic, such as crown ethers, that are capable of the selective ion binding[18]. In this study, CPEs based on DB18C6 as a neutral carrier, Scheme 2 , were synthesized and optimized with regard to their response characteristics according to the IUPAC guidelines[19] and tested for the determination of DAC in pharmaceutical formulation, serum and urine samples, achieving Nernstian response, reliable Nano-molar determination of the drug, and fast response.
Scheme 2

Structure of dibenzo-18-crown-6. The numbering doesn’t indicate IUPAC nomenclature, but just for assigning the 1H NMR peaks.

Structure of dibenzo-18-crown-6. The numbering doesn’t indicate IUPAC nomenclature, but just for assigning the 1H NMR peaks.

Experimental

Materials

Pure analytical grade DAC, sofosbuvir (SOF), and ledipasvir (LED) were kindly provided by European Egyptian Pharmaceutical Industries (EEPI), Alexandria, Egypt. The pharmaceutical formulation Daclavirocyrl® (60 mg/Tablet) was purchased from a local drug store. Graphite powder (ACROS organics, New Jersy, USA), dioctyl phthalate (DOP, 97.0% Sigma-Aldrich), tricresyl phosphate (TCP, 98.0% Sigma-Aldrich), DB18C6, 98.0% Sigma-Aldrich, sodium tetraphenylborate (Na-TImB, 97% Aldrich chemical company), sodium tertakis(1-imidazolyl)borate (Na-TPB, 99% Alfa Aesar) and tridodecylmethylammonium chloride (TDMAC, 98% Sigma-Aldrich) were used in the construction of the sensors. Orthophosphoric acid (85 wt% H3PO4, Sigma-Aldrich), glacial acetic acid (CH3CO2H, Sigma-Aldrich), and boric acid (H3BO3, ACS grade, MISR-Scientific Company) were used in the preparation of Britton-Robinson (B-R) buffer. Copper sulphate pentahydrate (98.5%), chloride salts of sodium (99.8%), potassium (98%), ammonium (99%), magnesium (98%), calcium (97%), manganese (95%) and cobalt (97%) were purchased from ADWIC Cairo, Egypt. Ferric chloride hexahydrate (99%) was purchased from Merck, Germany. D-glucose anhydrous (99%) and starch (ACS grade), were purchased from Fisher Scientific UK. Dopamine hydrochloride (ACS grade), glycine (99%), and DL-leucine (99%) were purchased from Sigma-Aldrich, alanine monohydrochloride (ACS grade) was provided by B.D.H Ltd. London, and DL-histidine monohydrochloride (ACS grade) was purchased from Prolabo, Rue Pelée, Paris.

Apparatus

FTIR (SHIMADZU IR spectrometer), and NMR (Varian Mercury VX-300 NMR spectrometer). Surface images of electrodes were recorded using Quanta FEG 250 scanning electron microscope (FEI Company, Hillsboro, Oregon-USA). Potentiometric and pH measurements were performed using Jenway pH-mVmeter model 3310 (UK), with Ag-AgCl as the reference electrode. High-performance liquid chromatography (HPLC, YL9100 High-performance liquid chromatograph) was used as a reference method for the determination of DAC in commercial tablet samples. Lipophilicity (clogP) was calculated using the ALOGPS 2.1 software[20].

Standard solutions and sample preparation

A stock solution of DAC (1.0×10−3 mol/L) was prepared in Britton-Robinson buffer [21] pH 2.3 and the concentrations ranging from 1.0×10−3 mol/L to 1.0×10−9 mol/L were prepared by simple serial dilutions using deionized water (18 MΩ). For analytical application, five Daclavirocyrl® tablets were weighed, ground, and an average weight of two tablets were dissolved in a suitable volume and diluted to prepare 1.0×10−6 − 1.0×10−5 mol/L solutions. For application to body fluids, serum and urine samples were taken from a healthy volunteer (corresponding author) and they were diluted 100 times (taking 0.5 mL serum or urine and complete up to 50 mL) and spiked with a suitable volume of DAC solution to prepare 1.0×10−6 − 1.0×10−5 mol/L samples. All measurements were performed in a buffered medium. Stock solutions 1.0×10−3 M of SOF and LED were prepared; in addition, 1.0×10−1 mol/L stock solutions of each of NaCl, KCl, NH4Cl, CaCl2. MgCl2, CuSO4, CoCl2, MnCl2, FeCl3, dopamine, glycine, alanine, histidine, lysine, glucose, and soluble starch were prepared for the selectivity studies.

CPEs preparation

CPEs were prepared by mixing the dry ingredients in certain weight percent (graphite, DB18C6, and ionic additive) for 10 min in a glass mortar with a specially made glass pestle, then the plasticizer was added stepwise while mixing. Then the paste was packed well in a piston-driven Teflon holder (resistance was measured to ensure well paste packing, the resistance of optimized sensors in this study was between 9 and 18 Ω)[22].

Effect of pH

The influence of the test solution pH on the potential of the prepared electrodes was investigated in two different DAC concentrations, namely, 1.0×10−4 and 1.0×10−3 mol/L. The pH of the drug solutions was varied using B-R buffer in the range of 2–5, and no further increase in pH was attempted because of the solubility limitation of the drug [23], [24]. The potential obtained at each pH value was recorded and plotted against that pH value.

Chemical bonding between DAC and DB18C6 using IR and NMR

FTIR was used for recording the spectra of DB18C6, DAC before and after mixing in dimethyl sulfoxide (DMSO). 1H NMR was also used for recording the spectra of DB18C6, DAC before and after mixing the solids in DMSO‑d 6, and the chemical shifts (δ) were related to that of the solvent.

Surface characterization of CPEs

SEM and energy dispersive X-ray analysis (EDX) were used for morphological characterization of the optimized sensors. Samples were mounted onto SEM stubs and the applied conditions were a 10.1 mm working distance, with an in-lens detector and an excitation voltage of 20 kV.

Response time, electrode memory, and reversibility

The practical response time of the optimized sensors was measured by successively immersing the electrodes in a series of DAC solutions. In each solution the DAC concentration is increased by a value of half a decade from 1.0×10−9 to 1.0×10−3 mol/L. Time reading was recorded at the same instance the electrode was introduced into the sample solution. Electrode memory and reversibility were tested by measuring the potential response of the sensors successively from high-to-low concentrations and vice versa (1.0×10−4 and 1.0×10−3 mol/L).

Water layer test

To test the formation of an aqueous layer, the sensors’ potential was recorded after successively introducing the electrodes into 1.0×10−3 mol/L DAC solution for 30 min, 1.0×10−3 mol/L Na+ solution for 30 min, and 1.0×10−3 mol/L DAC solution for 30 min.

Potentiometric selectivity

Selectivity coefficients were determined using the separate solution method (SSM) and the matched potential method (MPM). In SSM, 10−3 mol/L solution was used for both DAC and interfering ions (B), and the selectivity coefficient () was calculated according to the equation[25]: log = + (1- ) log , where EB and EDAC are the electrode potentials of 10−3 mol/L solution of interfering cations, and DAC, respectively, S is the slope of the calibration graph, and and are the charges of DAC and the interfering ion, respectively, and is the activity of DAC solution. For the application of MPM, the potential of a reference solution containing a fixed level of primary ion () is measured, and then a known activity () of the primary ion is added to that reference solution and the potential difference is recorded. In a second experiment, interfering ion with activity () is added to an identical reference solution until the same potential change is reached. The MPM selectivity coefficient is calculated using the equation: For larger selectivity [26], the range of measurements by this MPM was extended by using a more dilute reference solution (10−6 mol/L) and a more concentrated foreign species solution in the range of 10−3 – 10−1 mol/L[27].

Lifetime

The lifetime of the optimized sensors was tested by measuring the performance characteristics of the sensors for 20 days.

Analytical application

The proposed method was used for the determination of DAC in Daclavirocyrl® tablets, spiked serum, and urine samples. The analysis was performed using the standard addition method due to its ability to compensate for the effect of the matrix. In the standard addition method, small increments of DAC standard were added to 25 mL aliquot samples of different concentrations, and the potential difference was recorded for each addition. The concentration of the sample can be calculated by the following equation[28]:where Cx is the concentration to be determined, Vx is the volume of the sample, Cs and Vs are the concentration and volume of the added standard, respectively, △E is the potential change after adding Vs, and S is the slope of the calibration graph. HPLC method was used as a reference method for the determination of DAC in tablet formulation. Reversed-phase C18 column was used as the stationary phase, phosphate buffer (10 mM, 1 mL triethylamine L−1): acetonitrile (60:40 v/v) was used as the mobile phase with a flow rate of 2 mL/min with UV detection at 312 nm[29].

Results and Discussion

Effect of electrode composition and modifiers

Performance characteristics of ion-selective electrodes largely depend on the nature and amount of different components of the electrodes; so, several CPEs -containing 1% DB18C6 as a neutral carrier ionophore- were prepared and tested for the determination of DAC. From the data in Table S1(in supplementary materials), sensor 1 containing only DB18C6 with no ionic additives exhibited a Sub-Nernstian response towards DAC of 14.8 mV/decade. It is well-established that, neutral carrier-based ISEs are greatly influenced by the presence of ionic sites. These sites facilitate the ion-exchange process at the membrane/solution interface [18] [64]. For this reason, 1% TDMAC cationic site was added in sensor 2; but the slope decreased to 10.7 mV/decade indicating that DAC is cationic in the sample solution; and thus it is more suitable to use anionic additive instead. 1% Na-TPB anionic site was introduced to sensor 3, resulting in a Super-Nernstian slope of 48.7 mV/decade. The amount of Na-TPB was reduced to 0.5% in sensor 4 showing more promising results in terms of the Nernstian response (29.8 mV/decade), the wide linear dynamic range (5.0×10−9 – 1.0×10−3 mol/L), and the low detection limit (4.8×10−9 mol/L). The great enhancement of the electrode response by Na-TPB is due to the fact that anionic additives offer many advantages to neutral carrier-based ISEs such as improving the selectivity to the primary ion, reducing paste resistance [30], [31], reduction or elimination of interference by co-ions, and minimizing electrode response time[32]. The type of plasticizer is known to influence the electrochemical properties of CPEs, as it dissolves the ionophore and anionic additive and affects the overall mobility of paste ingredients[20] [64]. It also influences the dielectric constant (ε) of the paste and affects the extraction of analyte species from the sample solution[33] [64]. As mentioned earlier, good performance characteristics were obtained with sensor 4 when TCP (ε = 6.9)[34] [64] was used as a plasticizer. However, when the plasticizer was changed to DOP (ε = 5.1)[34] [65] in sensor 5 the slope and upper detection limit diminished to 24.9 mV/decade and 1.0×10−4 mol/L, respectively, and the lower detection limit was increased to 2.8×10−8 mol/L. It has been reported that the nature of aromatic groups linked to boron in the anionic site has a great influence on target ion recognition and selectivity[35] [66], and thus another borate anionic additive salt (Na-TImB) was used in sensors 6–10. Sensors 6–8 were fabricated using TCP plasticizer, but the performance characteristics of these electrodes were worse than that obtained by sensor 4 (containing Na-TPB), as they have narrower linear dynamic ranges. However, when the plasticizer was changed from TCP to DOP in sensor 9 with 0.25% Na-TImB, it showed Nernstian behavior (29.5 mV/decade) with a wide linear range (5.0×10−9 – 1.0×10−3 mol/L) and a low detection limit (3.2×10−9 mol/L). Furthermore, when the percent of Na-TImB was increased to 0.75% in sensor 10, the slope, linear range, and the detection limit diminished to 22.2 mV/decade, 5.0×10−8 – 1.0×10−4, and 2.8×10−8 mol/L, respectively. The two optimized sensors -sensors 4 and 9, referred to as sensors I and II, respectively- differ in the anionic site type, and the plasticizer type. Sensor I containing Na-TPB (clogP = 6.89) showed the best performance in the presence of TCP (clogP = 4.87) plasticizer, and on the other hand sensor II containing Na-TImB (clogP = 0.54) did the same but in presence of DOP (clogP = 6.86) plasticizer. In other words, the results imply a synergistic effect between a less lipophilic plasticizer and a more lipophilic anionic additive, and vice versa. This could be explained based on that there is some specific interaction[36] between each (Na-TPB and TCP) and (Na-TImB and DOP), or that both components control the overall membrane lipophilicity, which in turn affects the process of complexation between the ionophore and the target ion. In terms of the performance characteristics of the fabricated sensors, sensors 4 and 9 showed the best Nernstian slopes, wide linear ranges, and lowest detection limits; therefore, these sensors were chosen for further studies. These sensors will be referred to as sensor I and sensor II, respectively, and their corresponding calibration graphs are shown in Fig. 1 .
Fig. 1

Calibration graphs of sensors I (●) and II (○).

Calibration graphs of sensors I (●) and II (○). As shown in Fig. 2 , sensors I and II were not sensitive to pH change up to pH 4, after which at pH 4.5 white turbidity was observed in the sample solution, accompanied by a large decrease in electrodes’ potential. This turbidity could be due to the precipitation of DAC due to its solubility limitation [23], [24], and thus causes a decrease in the concentration of dissolved DAC cations which explains the decrease in the measured potential. The decrease in potential is also confirmed by the pKa values of DAC, 3.82 and 6.09 [37], as after pH 3.82 the ionization state of DAC changes which will, in turn, affect the potential value recorded. Similar behavior was reported by Derar et. al. [14].
Fig. 2

The effect of pH on the potential response of sensors I and II.

The effect of pH on the potential response of sensors I and II.

Chemical bonding between DAC and DB18C6

Supramolecular chemistry has drawn great attention in the field of ISEs based on molecular recognition. The mechanism of action of supramolecular-based potentiometric sensors depends on host-guest chemistry, where supramolecular assemblies are held together with reversible noncovalent interactions that encompass a wide range of binding energies ranging from electrostatic interactions, H-bonding, hydrophobic, and π-π interactions[38]. Recent studies show that DAC exhibits unusual complexation behavior with different forms of supramolecular cyclodextrin hosts, namely γ-cyclodextrin[39] and methylated β-cyclodextrins[40]. Pederson has discovered dibenzo-18-crown-6 in 1967 [41], and since then it has been widely studied for metal complex formation; it was only recently used for drug ion sensors[42]. DB18C6 has a great ability to form inclusion complexes with different species, such as Na+, K+ [43], Ca2+ [44]; these species will be discussed in the selectivity study. According to the literature, 18-crown-6 has proven to be a good receptor for the protonated N-terminus of an amino acid[45]; such protonated N sites are present in DAC structure, revealing that DB18C6 is a suitable host for the analyte under study. To study the nature of chemical bonding between the ionophore DB18C6 and DAC, FTIR and 1H NMR spectra were recorded. On evaluating Table 1 and Fig. 3 , broadening in peak 1 was observed, which indicates the H-bonding interaction between both compounds. Besides, the lowering in N—H stretching vibrational frequency from 3441.01 cm−1 in free DAC to 3425.58 cm−1 in the mixture can also indicate NH-π interaction[46]. Furthermore, a redshift was observed in peaks 2 and 3 (aromatic C—H peaks) suggesting the presence of CH-π and π-π interaction[47], although theory predicts a blueshift for such interactions, experimental results presented here reports otherwise, moreover, Lemmens et. al. [47] and Erlekam et. al[48] also documented redshifts.
Table 1

FTIR data for DAC and a mixture of DB18C6 and DAC.

DAC (cm−1)DB18C6 + DAC (cm−1)PeakObservationSuggested interactionRef.
3441.013425.581Peak broadening and decrease in wavenumberH-bonding and NH-π interaction[46]
2993.523001.242Increase in wavenumberCH-π and/or π-π interaction[47]
2908.652916.373Increase in wavenumberCH-π and/or π-π interaction[47]
Fig. 3

FTIR spectra of DB18C6 (red), DAC (blue) and DB18C6 + DAC (black).

FTIR data for DAC and a mixture of DB18C6 and DAC. FTIR spectra of DB18C6 (red), DAC (blue) and DB18C6 + DAC (black). 1H NMR spectra, of DB18C6, DAC, and their mixture are shown in Fig. S1 and the corresponding chemical shifts of the peaks are represented in Table S2. π-π interaction mentioned earlier is further confirmed by the up-field shifts of NMR peaks assigned for the aromatic protons of the phenyl groups in both molecules and in the two imidazole H1 in DAC. In addition, the hydrogens of –CH2 moieties in the ring structure of DB18C6 also experienced up-field shifts, which can be explained by the possibility of CH-π interactions [49] between these hydrogens in the crown ether and the aromatic phenyl or imidazole in DAC. Similar up-field shifts were observed in pyrrolidine protons H-1,2,3, suggesting CH-π interactions between these protons in DAC and the aromatic phenyl groups in DB18C6. In addition, the secondary amide proton in DAC experienced a downfield shift, which can be attributed to H-bonding [50] between NH in DAC and the oxygen atoms in the DB18C6 cavity. DB18C6 can interact at both sides of DAC, or in the middle of the compound, as in Scheme 3 .
Scheme 3

Representation of the complexes formed between DAC and DB18C6. The inclusion complex between the cavity (represented by the oval shape for simplification) of DB18C6 and one of DAC terminals (a), and another possible complexation mode showing the host and guest in parallel positions (b). Dark grey and silver: carbon, white: hydrogen, red: oxygen: blue: nitrogen.

Representation of the complexes formed between DAC and DB18C6. The inclusion complex between the cavity (represented by the oval shape for simplification) of DB18C6 and one of DAC terminals (a), and another possible complexation mode showing the host and guest in parallel positions (b). Dark grey and silver: carbon, white: hydrogen, red: oxygen: blue: nitrogen.

Surface characterization

The signal transduction and thus the response of electrochemical sensors is related to their physicochemical morphology and surface architecture that connects the ionophore with the analyte in sample solution at the nanometer scale[51]. The SEM images of sensors I and II in Table 2, Table 3 show that the pastes exhibit a granular sub microstructure indicating the high surface area of the sensors. On comparing SEM images of the two sensors, it can be observed that sensor II has fewer white spots than sensor I, suggesting higher solubility of electrode components in the plasticizer[52], which may be the reason that sensor II offer lower detection limit than sensor I.
Table 2

SEM images of sensor I and sensor II.

SEM images
Sensor ISensor II
Table 3

SEM images with elemental mapping of sensor I and sensor II.

SEM images with elemental mapping
sensor Isensor II
SEM images of sensor I and sensor II. SEM images with elemental mapping of sensor I and sensor II. Mapping data in Table 3 show the surface distribution of different elements, C (representing graphite, plasticizer, and DB18C6), O (present in DB18C6), B (belong to the ionic site, Na-TPB or Na-TImB), N in Na-TImB, and P (specific for TCP plasticizer). As can be seen, the different elements are evenly distributed at the surface of the paste with high homogeneity, indicating the good solubility of the ionophore and exchanger in the plasticizer. EDX and corresponding w% of different elements of sensors I and II are shown in Fig. S2, the appearance of well-defined peaks of elements related to the ionophore and ionic site approves the effective modification of the carbon pastes. One of the critical factors in the use of ISEs in routine analysis is the so-called response time. Practical response time has been defined as the time that elapses between the instant at which the ISE and the reference electrode are brought in contact with the sample solution, during which potential (E) changes by 90% of the final value[53]. As presented in Fig. 4 sensor I showed a response time of <5 sec for low concentrations (5×10−9 – 1×10−6 mol/L), and about 8 sec for high concentrations (1×10−5 – 1×10−3 mol/L), while sensor II showed a response time of about 5 sec for low concentrations, and about 6 sec for high concentrations. The fast response by the sensors may be attributed to the assumption that the additive (Na-TPB or Na-TImB) lowers the activation barrier for the cation-exchange reaction at the membrane/solution interface and thus reduces the time response after an activity step[54]. In addition, the small electrode resistance plays an important role in shortening time responses[55].
Fig. 4

The dynamic response time of sensors I (●) and II (○).

The dynamic response time of sensors I (●) and II (○). The repeatability of potential reading was tested by immersing each sensor in successive high-to-low concentrations (1×10−4 and 1×10−3 mol/L). As shown in Fig. 5 , the sensors possess excellent reversibility i.e., have no memory effect, this may be explained by the high reversibility of ionophore-analyte complexation and fast exchange kinetics of association-dissociation interaction at the interface between the electrode surface and the sample solution[56] [67].
Fig. 5

Electrode memory and reversibility of sensor I (●) and II (○).

Electrode memory and reversibility of sensor I (●) and II (○). Early solid contact electrodes had potential drifts and did not exhibit good detection limits which was recently suggested to be due to the poor adhesion between membrane and underlying electron conductor which can lead to the formation of a water layer, acting as a reservoir for electrolyte [57]. The water layer test or the so-called Morf test was introduced by Morf et al. [58] to test the potential stability of solid contact electrodes such as CPEs. As shown in Fig. 6 , both sensors exhibited stable EMF reading in 10−3 M DAC solution for 30 min, after which the solution was changed to 10−3 M Na+ solution and the sensors also showed a fast response and potential stability, and the decrease in potential proves the selectivity of the sensors towards DAC. Upon changing the solution back to 10−3 M DAC the potential is restored with high stability proving the absence of water film reservoirs formed between the carbon paste and stainless-steel rod contact. It turns out that the high hydrophobicity of carbon paste and its tight packing -which is aided by the modern design of piston-driven Teflon holders that yield pastes with good mechanical properties[59]- prevent the formation of a water layer.
Fig. 6

Water layer test of sensors I (●) and II (○).

Water layer test of sensors I (●) and II (○).

Selectivity

The selectivity of sensors I and II was evaluated by calculating log , assessed by the SSM and MPM against several inorganic, organic species that may coexist with DAC and two of the formulated drugs SOF and LED, since this method will be applied in the determination of DAC in pharmaceutical tablet, serum, and urine samples. As shown in Table 4 and Fig. S3, it can be noted that all tested species don’t significantly interfere with the determination of DAC; this is confirmed as all log values were less than −1, except for K+ in the case of sensor I using SSM, where the value was −0.57. This behavior is not surprising as many studies proved that the ionophore, DB18C6, has a great selectivity towards K+ [43]. However, when the MPM was applied, it gave a better selectivity coefficient value of −2.73 against K+. This large difference may be due to the fact that SSM has many limitations in the case of ions with non-Nernstian responses and that of unequal charges[60].
Table 4

Logarithm the selectivity coefficient for the proposed electrodes using separate solution method and matched potential method.

Foreign species (B)Log KDAC,Bpot
Sensor I
Sensor I′
Sensor II
Sensor II′
SSMMPMSSMMPMSSMMPMSSMMPM
Na+−1.00−4.07−0.91−3.98−1.80−4.38−1.52−3.92
K+−0.57−2.73−1.72−3.60−2.37−2.98−2.38−3.29
NH4+−1.10−4.16−0.86−3.85−1.90−4.92−1.18−3.94
Ca2+−3.43−4.79−2.36−3.66−6.43−4.95−3.83−3.63
Mg2+−5.00−4.57−3.05−3.93−5.13−4.91−3.95−3.55
Cu2+−3.67−2.71−1.78−2.68−3.93−4.16−2.23−3.86
Co2+−4.53−4.38−2.82−3.75−5.40−4.40−3.09−3.97
Mn2+−3.93−4.74−2.30−3.86−4.43−4.88−2.98−4.11
Fe3+−1.46−2.59+1.19+0.18−1.40−3.59+1.06+0.13
SOF−1.63−1.13−1.49−1.12
LED−1.08−1.30−1.42−1.45
Dopamine−3.25−3.01−4.29−2.91
Glycine−4.07−3.13−4.37−3.20
Alanine−4.38−3.97−4.52−3.97
Histidine−4.17−3.49−4.42−3.67
Lysine−3.99−3.57−4.14−3.51
Glucose−4.29−3.06−5.09−3.22
Soluble starch−4.71−3.97−5.02−3.59
Logarithm the selectivity coefficient for the proposed electrodes using separate solution method and matched potential method. On the other hand, sensor II showed no interference from K+ as it offers a much lower log value of −2.37 using SSM and −2.98 using MPM. The presence of Na-TImB in sensor II plays an important role in enhancing the selectivity of the electrode. Alaviuhkola T. et. al. reported that the nature of aromatic unit bound to boron in the ionic sites strongly affects the selectivity of the ISE[36], the boron center in TPB is attached to four phenyl groups, while in TImB it is linked to four N atoms, this could result in high extraction of DAC from the solution phase due to the higher extent of binding to the imidazole groups in Na-TImB through H-bonding, CH-π and/or π-π interaction than in case of Na-TPB. In addition, the binding affinity of DB18C6 towards K+ is affected by the solvent[43], thus, the better selectivity results obtained by sensor II could be related also to the use of DOP plasticizer with low polarity which facilitates the interaction with DAC. In sensor I the more polar TCP has greater interaction with K+ than the less polar DOP[33]. Since the used ionophore DB18C6 was reported to respond to several cations, sensors I′ and II′ were made with the same composition as sensors I and II, respectively, only lacking the ionophore, to investigate its effect on the selectivity of the electrodes. The composition (% w/w) of sensor I′ is 69.75% graphite, 29.75% TCP, and 0.5% Na-TPB, and that of sensor II′ is 69.875% graphite, 29.875% DOP, and 0.25% Na-TImB. On comparing the selectivity coefficient values obtained using sensors with (I and II) DB18C6 and without (I′ and II′) DB18C6, it can be noticed from Table 4 that DB18C6 has improved the selectivity of the sensors against all tested species, except for K+, where the sensors without the ionophore exhibited better selectivity for DAC over K+, in other words, the incorporation of DB18C6 in the electrode cocktail increased its response to K+. This phenomenon, as mentioned earlier, is due that DB18C6 has the strongest binding affinity towards K+ among all other alkali metal cations[43]. From the data in Table 4, it is evident that sensor II has generally better selectivity towards DAC than sensor I, and for this reason, sensor II was chosen to be applied for the determination of DAC in pharmaceutical and biological samples. It is well-established that the primary reason for limited lifetimes of ISEs is caused by the loss of plasticizer, ionophore, and/or the ionic site from the sensor due to leaching of these components into the sample[33]. The lifetime of sensors II was tested by measuring the performance characteristics of the sensor for 20 days. It was found that after 16 days the sensor exhibited good response slope of 25.3 mV/decade, as observed in Table S3. Sensor II showed great sensitivity as it was used for 20 days, and the lower detection limit (LDL) was only increased from 3.2×10−9 to 5.5×10−9 mol/L, while the upper detection limit (UDL) was reduced from 1×10−3 to 1×10−6 mol/L, this deviation from linearity at the UDL may be due to Donnan failure[61]. The optimized sensor was successful in determining DAC in pharmaceutical formulation (tablet) and in spiked biological fluids (serum and urine) using standard additions method; the results are summarized in Table 5 . All the recovery values were between 99.25 and 101.42% and the RSD values were between 0.79 and 1.53% indicating high precision and accuracy and thus high reliability.
Table 5

Determination of DAC in tablets, and spiked serum and urine using standard addition method.

SampleSensor II
Taken (µmol/L)Recovery ± SDRSD
Daclavirocyrl® (60 mg/tablet)199.71 ± 0.890.90
599.25 ± 0.650.66
10100.97 ± 1.371.35
Spiked serum1100.35 ± 0.790.79
5101.10 ± 1.191.18
10101.42 ± 0.760.75
Spiked urine1100.13 ± 1.261.26
5100.98 ± 1.541.53
10100.47 ± 1.301.30

SD: standard deviation, RSD: relative standard deviation, N = 4 replicates.

Determination of DAC in tablets, and spiked serum and urine using standard addition method. SD: standard deviation, RSD: relative standard deviation, N = 4 replicates. The results obtained from the determination of 5 µmol/L DAC in Daclavirocyrl® by sensors II were compared to HPLC[29] as a reference method, and the HPLC chromatogram is represented in Fig. S4. Statistical F- and t-tests were applied, and the calculated values, as observed in Table 6 , are less than the critical theoretical values, indicating that there is no significant difference between the precisions and means of both methods at 95% confidence limit (P = 0.05), implying the success of the proposed method in the accurate and precise determination of DAC.
Table 6

Statistical comparison between data obtained by sensors II, and HPLC reference method [29] for 5 µmol/L Daclavirocyrl® tablet sample.

MethodRecoverya ± SDF-calculated (9.28)bt-calculated (2.45)c
HPLC [29]100.23 ± 0.53
Sensor II
Standard addition method99.25 ± 0.651.512.32

aaverage of four determinations, b tabulated F-value at P = 0.05, c tabulated t-value at P = 0.05 and 6 degrees of freedom.

Statistical comparison between data obtained by sensors II, and HPLC reference method [29] for 5 µmol/L Daclavirocyrl® tablet sample. aaverage of four determinations, b tabulated F-value at P = 0.05, c tabulated t-value at P = 0.05 and 6 degrees of freedom. By comparing to previously reported potentiometric methods for the determination of DAC[14], [15], this work offers improvement in performance characteristics of the optimized sensors in terms of, response time, dynamic linear range, and detection limit (Table 7 ).
Table 7

Comparison between performance characteristics of optimized sensors I and II in this work and the sensors developed in previously reported potentiometric methods[14], [15].

Comparison
This work
Dear et.al. [14]
Eldin et. al. [15]
SensorCPE (Sensor I)CPE (Sensor II)SC-SPISEPVC
Sensing element(s)DB18C6 and Na-TPBDB18C6 and Na-TImBDAC-TPB and MWCNTDAC-TPB
Slope, mV/decade29.829.531.028.7
Linear range, M5×10−9 − 10−35×10−9 − 10−310−5 − 10−210−6 − 10−3
LOD, nM4.93.2870110
Response time, s5–86–8< 1540

SC-SPISE − solid-contact screen printed ion-selective electrode, PVC − conventional polyvinyl chloride membrane electrode, DAC-TPB − daclatasvir-tetraphenylborate ion-exchanger, MWCNTs − multiwalled carbon nanotubes.

Comparison between performance characteristics of optimized sensors I and II in this work and the sensors developed in previously reported potentiometric methods[14], [15]. SC-SPISE − solid-contact screen printed ion-selective electrode, PVC − conventional polyvinyl chloride membrane electrode, DAC-TPB − daclatasvir-tetraphenylborate ion-exchanger, MWCNTs − multiwalled carbon nanotubes.

Conclusion

In this work, Dibenzo-18-Crown-6-based carbon paste sensors were developed for the determination of DAC in its pure form, Daclavirocyrl® tablets, serum, and urine samples. The best analytical performance of two sensors referred to as sensor I and sensor II, which exhibited wide linear response range of 5.0×10−9 – 1.0×10−3 mol/L, with Nernstian slopes of 29.8 ± 1.18 and 29.5 ± 1.00 mV/decade, LOD of 4.8×10−9 and 3.2×10−9 mol/L, and instantaneous response times of 5–8 and 5–6 s, respectively, with great reversibility and no memory effect. The proposed sensors showed great selectivity towards DAC in presence of many metal cations, sugars, amino acids, dopamine, and other anti-HCV drugs (SOF and LED). In addition, the two sensors were applied for the determination of DAC in the above-mentioned samples, with high accuracy (recovery values ranging from 99.25 to 101.42%), and precision (relative standard deviation in the range of 0.79 to 1.53%). Finally, the proposed method was compared to a reference method (HPLC), and statistical analysis of data revealed compatible results of the methods.

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
  33 in total

1.  PVC membrane, coated-wire, and carbon-paste ion-selective electrodes for potentiometric determination of galantamine hydrobromide in physiological fluids.

Authors:  Fatehy M Abdel-Haleem; Mohamed Saad; Ahmed Barhoum; Mikhael Bechelany; Mahmoud S Rizk
Journal:  Mater Sci Eng C Mater Biol Appl       Date:  2018-04-04       Impact factor: 7.328

2.  t-Butyl calixarene/Fe2O3@MWCNTs composite-based potentiometric sensor for determination of ivabradine hydrochloride in pharmaceutical formulations.

Authors:  Fatehy M Abdel-Haleem; Eman Gamal; Mahmoud S Rizk; Rasha M El Nashar; Badawi Anis; Hussam M Elnabawy; Ahmed S G Khalil; Ahmed Barhoum
Journal:  Mater Sci Eng C Mater Biol Appl       Date:  2020-06-05       Impact factor: 7.328

3.  Strong N-H...pi hydrogen bonding in amide-benzene interactions.

Authors:  Philipp Ottiger; Chantal Pfaffen; Roman Leist; Samuel Leutwyler; Rafał A Bachorz; Wim Klopper
Journal:  J Phys Chem B       Date:  2009-03-05       Impact factor: 2.991

4.  A Stability-Indicating UPLC Method for the Determination of Potential Impurities and Its Mass by a New QDa Mass Detector in Daclatasvir Drug Used to Treat Hepatitis C Infection.

Authors:  Varaprasad Jagadabi; P V Nagendra Kumar; Kasthuri Mahesh; Srinivasu Pamidi; L A Ramaprasad; D Nagaraju
Journal:  J Chromatogr Sci       Date:  2019-01-01       Impact factor: 1.618

Review 5.  Daclatasvir: A NS5A Replication Complex Inhibitor for Hepatitis C Infection.

Authors:  Michael A Smith; Randolph E Regal; Rima A Mohammad
Journal:  Ann Pharmacother       Date:  2015-10-20       Impact factor: 3.154

6.  RP-HPLC method development and validation for quantification of daclatasvir dihydrochloride and its application to pharmaceutical dosage form.

Authors:  Syed Suffdar Hussain Shah; Muhammad Iqbal Nasiri; Humera Sarwar; Aatka Ali; Syed Baqir S Naqvi; Sohail Anwer; Muhammad Kashif
Journal:  Pak J Pharm Sci       Date:  2021-05       Impact factor: 0.684

7.  Binding selectivity of dibenzo-18-crown-6 for alkali metal cations in aqueous solution: A density functional theory study using a continuum solvation model.

Authors:  Chang Min Choi; Jiyoung Heo; Nam Joon Kim
Journal:  Chem Cent J       Date:  2012-08-08       Impact factor: 4.215

Review 8.  Electrochemical Biosensors - Sensor Principles and Architectures.

Authors:  Dorothee Grieshaber; Robert MacKenzie; Janos Vörös; Erik Reimhult
Journal:  Sensors (Basel)       Date:  2008-03-07       Impact factor: 3.576

9.  In vitro antiviral activity of the anti-HCV drugs daclatasvir and sofosbuvir against SARS-CoV-2, the aetiological agent of COVID-19.

Authors:  Carolina Q Sacramento; Natalia Fintelman-Rodrigues; Jairo R Temerozo; Aline de Paula Dias Da Silva; Suelen da Silva Gomes Dias; Carine Dos Santos da Silva; André C Ferreira; Mayara Mattos; Camila R R Pão; Caroline S de Freitas; Vinicius Cardoso Soares; Lucas Villas Bôas Hoelz; Tácio Vinício Amorim Fernandes; Frederico Silva Castelo Branco; Mônica Macedo Bastos; Núbia Boechat; Felipe B Saraiva; Marcelo Alves Ferreira; Steffen Jockusch; Xuanting Wang; Chuanjuan Tao; Minchen Chien; Wei Xie; Dinshaw Patel; Aitor Garzia; Thomas Tuschl; James J Russo; Rajith K R Rajoli; Carolina S G Pedrosa; Gabriela Vitória; Letícia R Q Souza; Livia Goto-Silva; Marilia Zaluar Guimarães; Stevens K Rehen; Andrew Owen; Fernando A Bozza; Dumith Chequer Bou-Habib; Jingyue Ju; Patrícia T Bozza; Thiago Moreno L Souza
Journal:  J Antimicrob Chemother       Date:  2021-06-18       Impact factor: 5.790

View more
  1 in total

1.  Molecularly-imprinted polymer-base bulk optode for the determination of ivabradine hydrochloride in Procoralan®.

Authors:  Fatehy M Abdel-Haleem; Mahmoud S Rizk; Menna M El-Beshlawy
Journal:  RSC Adv       Date:  2022-06-15       Impact factor: 4.036

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.