Literature DB >> 35149877

No difference in patient-reported satisfaction after 12 months between customised individually made and off-the-shelf total knee arthroplasty.

Séverin Wendelspiess1,2, Raphael Kaelin1, Nicole Vogel3,4, Thomas Rychen1, Markus P Arnold1,2.   

Abstract

PURPOSE: A subset of patients is usually not satisfied after a total knee arthroplasty (TKA). Customised individually made (CIM) TKA are deemed to overcome drawbacks of classical off-the-shelf (OTS) TKA, but evidence is still sparse. The aim of this study was to compare satisfaction of patients with CIM and OTS TKA.
METHODS: This prospective cohort study compared clinical and patient-reported outcome measures (PROM) between patients with CIM and OTS TKA. The primary outcome was patient satisfaction after 12 months. Secondary outcomes were the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), the Forgotten Joint Score (FJS-12) and the EQ-5D-3L after 4 and 12 months.
RESULTS: Data were analysed from 74 CIM TKA and 169 OTS TKA between January 2017 and September 2020. Patients with CIM TKA were slightly younger, more often male, had a lower body mass index, a lower KSS and partially higher preoperative PROMs. Patient satisfaction after 12 months was high and comparable (CIM 87%, OTS 89%). All PROMs improved for both groups (p < 0.001) and did not differ after 12 months (p > 0.063). The majority of patients improved above the minimal important difference (range 65 to 89%) and reported a clear overall improvement (CIM 86%, OTS 87%). The postoperative KSS, notably regarding knee stability, was higher for CIM TKA (p < 0.001).
CONCLUSION: No difference was found in patient satisfaction between CIM and OTS TKA after 12 months. In both groups, patient satisfaction was high and PROMs improved considerably. LEVEL OF EVIDENCE: II, prospective cohort study.
© 2022. The Author(s).

Entities:  

Keywords:  Arthroplasty; Custom; Knee; Patient satisfaction; Patient-reported outcome measure; Patient-specific; Replacement

Mesh:

Year:  2022        PMID: 35149877      PMCID: PMC9418302          DOI: 10.1007/s00167-022-06900-z

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.114


Introduction

About 20% of patients are not satisfied after a total knee arthroplasty (TKA) [6, 12, 15]. A variety of predictors for dissatisfaction have been identified: female sex, lower grade of osteoarthritis, implant-related reasons, mental health problems, unfulfilled expectations, postoperative pain and limited function are only some of these [12, 15, 21, 29]. Moreover, the number of TKA is rising among younger and more active patients with high functional expectations [31]. Although TKA is a common, safe and cost-effective treatment for end-stage knee osteoarthritis [3, 12], the procedure does not entirely restore normal biomechanics and functional limitations may occur [26]. Classical off-the-shelf (OTS) TKA can cause implant overhang, malalignment and abnormal kinematics [20, 41]. One reason may be the high variability of knee phenotypes between individuals, ethnicities and sex [4, 7, 13, 19, 22]. Customised individually made (CIM) TKA have been developed to overcome these problems and improve outcome after TKA [5, 17, 44]. CIM TKA incorporate a bone-preserving approach using custom planning images, implants and instrumentation [44]. The personalised approach respects the patient’s individual knee anatomy and eliminates implant sizing compromises [18, 45]. Disadvantages of CIM TKA are a prolonged waiting time for the manufacturing, a higher radiation exposure [44] and limited intraoperative implant options [45]. To get a better understanding of the patients’ perspective, their satisfaction after TKA and potential problems in daily life, an evaluation of patient-reported outcome measures (PROM) is necessary. CIM TKA are relatively new, available in the US and Europe since 2011 [9] and at our hospital since 2015. First results are promising in terms of bone preservation [45], ligament balancing [45], alignment [2, 54] and patient satisfaction [32, 37]. However, sufficient data are still lacking: comparative studies are sparse and randomised controlled trials on this topic are absent [25, 51]. A prospective study comparing PROMs of patients with CIM and OTS TKA has not yet been published. The aim of this study was to analyse patient-reported satisfaction; further PROMs and clinical outcome of patients with CIM TKA were compared to OTS TKA. Our hypothesis was that CIM TKA are superior to OTS TKA.

Materials and methods

Study design, setting and recruitment

This is a single-side, observational, prospective cohort study. The study is in accordance with the World Medical Association Declaration of Helsinki [48]. All patients were recruited in our medical practice, based in a private hospital, and gave their written informed consent. Since 2017, all patients scheduled for any type of knee arthroplasty were asked to complete a set of PROMs. Patients with insufficient knowledge of German, English, French or Italian were excluded. Details regarding recruitment and procedures are published in the study protocol [52]. The current study included consecutive patients with a primary cruciate-retaining CIM TKA (iTotal CR G2, ConforMIS Inc., Bedford, MA, US) or cruciate-retaining OTS TKA (Attune® CR mobile-bearing, DePuy Synthes, Raynham, MA, US) who completed PROMs before the surgery and after 12 months. Patients with revision surgery or major re-operation on the affected knee were excluded. All TKA were performed between January 2017 and September 2020 by MPA (CIM and OTS) and by RK and TR (OTS), all of them well-experienced senior surgeons. CIM TKA patients chose their surgeon (MPA) accordingly because of their interest in the new technology. In rare cases, the patient was made aware of the possibility of a CIM TKA because of a marked joint line obliquity (tibial mechanical angle of ≤ 84° on long-leg radiographs) with an obvious anatomical difference in shape between the medial and lateral femoral condyles or hypoplasia of the lateral femoral condyle [13].

Surgical technique

All patients had the same peri- and postoperative anaesthesia and pain management protocol. Preoperative, all patients received 1 g of tranexamic acid and a single-shot antibiosis with a cephalosporin intravenous. All surgeries used a medial parapatellar approach without tourniquet. Soft tissue release was only performed to an extent that was needed for an appropriate ligament balancing. For both, CIM and OTS TKA, spacer blocks are applied to adjust the soft tissue balancing process. A lateral (in valgus knees) or medial release (in varus knees) was used accordingly to achieve the ideal ligament balancing status. No more than grade one release was needed. Patellar resurfacing was only done in cases with severe patellofemoral arthritis with lateralised patella (i.e. bone on bone, especially if the patellofemoral joint surface was uneven). Both implants apply mechanical alignment. The CIM implants are based on a preoperative computed tomography. Subsequently, the implant is manufactured and the surgeon provided with individualised instruments and a planning overview (iView®). The detailed surgical technique is described elsewhere [45]. The planning algorithm of the CIM TKA results in a hip–knee–ankle angle of 180° and a limited joint line obliquity provided by uneven medial and lateral inlay heights. For OTS TKA, a natural slope and rotation along the grinding marks on the arthritic tibial plateau is sought, followed by resection of the tibial plateau. After determining femoral rotation by the intramedullary balancer, the distal femur is resected first (extension gap). Subsequently, the posterior (flexion gap) and anterior femur condylar cut is made. All patients had the same postoperative rehabilitation protocol with immediate full weight-bearing on crutches until sufficient muscular stabilisation and a proper gait pattern was achieved.

Follow-up and outcome measures

Data were collected before the surgery and during routine control visits after 4 months and 12 months using Research Electronic Data Capture (REDCap®). Patients’ characteristics were extracted from the medical records. The surgeon graded the degree of osteoarthritis according to the Kellgren and Lawrence classification from 0 (no osteoarthritis) to 4 (severe osteoarthritis) [16] and completed the objective part of the Knee Society Score (KSS) ranging from 0 (worst) to 100 (best) points [40]. Comorbidities were classified according to the American Society of Anesthesiologists (ASA) from ASA I (normal healthy) to ASA V (moribund) [1]. The primary outcome was patient satisfaction after 12 months assessed on a five-point Likert scale (very satisfied, satisfied, neutral, unsatisfied or very unsatisfied). Secondary outcomes were all other PROMs: the Knee injury and Osteoarthritis Outcome Score (KOOS), the Forgotten Joint Score (FJS-12), the EQ-5D-3L, overall improvement and willingness to undergo the surgery again. The KOOS comprises five subscales on pain, symptoms, activities of daily living, sports and quality of life on scales from 0 (worst) to 100 (best) points [34]. The FJS-12 captures the patient’s ability to forget the artificial joint in everyday life on a scale from 0 (worst) to 100 (best) points [50]. The EQ-5D-3L measures health-related quality of life ranging from 0 (worst) to 1 (best) and includes a visual analogue scale (VAS) from 0 (worst imaginable health) to 100 (best imaginable health) [11]. Overall improvement was measured on a seven-point Likert scale (very much better, substantially better, a little better, no change, a little worse, substantially worse or very much worse) and patients were asked if they would undergo the surgery again (yes or no). Postoperative complications such as thromboembolic events, infections, re-operations, revisions or death were recorded as adverse events. Revision was defined as a second surgery to replace some or all parts of the original TKA.

Ethical approval

The study was approved by the local ethics committee (reference: 2016-01777) [46].

Statistics

The statistical evaluation was done with IBM SPSS statistics for Windows, Version 28, Armonk, NY: IBM Corp and R, Version 4.0.5 [30]. Descriptive statistics are presented with means and standard deviation (SD) for continuous variables, frequency counts and percentages for categorical variables. Differences between pre- and postoperative data were tested with paired t-test for continuous variables and Wilcoxon signed rank test for categorical variables. To improve the interpretability of PROM results, the proportion of patients whose PROMs improved more than the minimal important difference (MID) was calculated. The MID is defined as the smallest difference in a score that patients perceive as important [38]. The following MID cutoff values were applied: 10.7 for KOOS symptoms, 16.7 for KOOS pain, 18.4 for KOOS activities of daily living, 12.5 for KOOS sports, 15.6 for KOOS quality of life [23], 10.8 for the FJS-12 [10, 14] and 0.15 for EQ-5D-3L [10]. Differences in subgroups were measured with unpaired t-test for continuous variables and with Mann–Whitney U test or Chi-Square test for categorical variables. Bivariate linear correlations were analysed with the Pearson test for continuous variables and the Spearman test for categorical variables. The correlation effect sizes were classified as low (r = 0.1), medium (r = 0.3) or strong (r = 0.5) [8]. The a priori power calculation is based on the KSS after 4 months. When the study was planned, reliable data on patient satisfaction or any other PROM after 12 months were not available. The calculated effect size was 0.5 and resulted in a sample size of 64 patients per group to assure a power of 0.8 with two-sided alpha = 0.05.

Results

Overall, 77 CIM and 182 OTS TKA were recruited and data of 74 CIM (63 patients, 30 female) and 169 OTS TKA (155 patients, 93 female) was analysed. Details regarding recruitment are shown in Fig. 1. Response rate to complete PROMs was 79% and loss to follow-up after 12 months was 4%. Patients’ characteristics are described in Table 1. Participants and patients who did not want to complete PROMs did not differ regarding patients’ characteristics (p > 0.159). Patients with CIM TKA were younger, more frequent male, had a lower body mass index, a lower KSS and partly higher PROMs preoperative (Tables 1 and 2).
Fig. 1

Flow chart of recruitment. CIM customised individually made, OTS off-the-shelf, TKA total knee arthroplasty, PROM patient-reported outcome measure, n number

Table 1

Patients’ characteristics

CIMn = 74OTSn = 169p value
Age, mean years (± SD)67.1 (± 8.4)69.7 (± 8.9)0.019
Age < 65 years, n (%)26 (35%)51 (30%)0.457
Body mass index, mean kg/m2 (± SD)26.6 (± 3.4)28.7 (± 4.9) < 0.001
Sex, n (%)0.035
 Male41 (55%)68 (40%)
 Female33 (45%)101 (60%)
Insurance, n (%) < 0.001
 Basic insurance3 (4%)101 (60%)
 Supplementary insurance71 (96%)68 (40%)
Side, n (%)0.579
 Left32 (43%)80 (47%)
 Right42 (57%)89 (53%)
Surgery, n (%)0.110
 Unilateral52 (70%)141 (83%)
 Bilateral22 (30%)28 (17%)
Kellgren and Lawrence classification, n (%)0.849
 10 (0%)1 (1%)
 21 (1%)5 (3%)
 320 (27%)43 (25%)
 453 (72%)120 (71%)
ASA classification, n (%)0.153
 I/II63 (85%)129 (76%)
 III10 (14%)39 (23%)
 IV/V1 (1%)1 (1%)
Operation time, mean min (± SD)84 (± 14)84 (± 18)0.823
Length of stay, mean days (± SD)6.4 (± 1.1)6.5 (± 1.1)0.500

CIM customised individually made, OTS off-the-shelf, n number, SD standard deviation, ASA American Society of Anesthesiologists

Table 2

KSS and PROMs before and after the surgery

Before surgery4 months12 months
CIMn = 74OTSn = 169p valueCIMn = 74OTSn = 169p valueCIMn = 74OTSn = 169p value
Mean (± SD)(95% CI)Mean (± SD)(95% CI)Mean (± SD)(95% CI)
KSS

52.8

(± 11.9)

57.1

(± 13.4)

0.018

(− 7.9 to − 0.8)

90.4

(± 7.7)

85.3

(± 9.1)

 < 0.001

(2.7 to 7.6)

94.2

(± 7.5)

88.9

(± 8.6)

 < 0.001

(2.8 to 7.7)

KOOS symptoms

51.2

(± 16.8)

46.0

(± 18.0)

0.018

(0.4 to 10.1)

68.6

(± 15.3)

69.9

(± 16.4)

0.543

(− 5.8 to 3.1)

77.8

(± 16.0)

79.2

(± 14.9)

0.507

(− 5.6 to 2.8)

KOOS pain

49.6

(± 16.4)

41.9

(± 13.7)

 < 0.001

(3.7 to 11.7)

71.7

(± 16.9)

71.7

(± 16.8)

1.000

(− 4.7 to 4.7)

84.5

(± 16.1)

82.2

(± 15.8)

0.321

(− 2.2 to 6.6)

KOOS daily living

57.4

(± 15.8)

48.2

(± 15.7)

 < 0.001

(4.8 to 13.5)

78.0

(± 14.8)

77.1

(± 14.0)

0.671

(− 3.1 to 4.8)

87.2

(± 13.8)

84.3

(± 14.5)

0.151

(− 3.1 to 6.8)

KOOS sports

23.4

(± 18.8)

19.9

(± 18.3)

0.198

(− 1.8 to 8.7)

49.9

(± 24.5)

51.1

(± 25.7)

0.764

(−8.9 to 6.6)

68.1

(± 21.8)

62.4

(± 27.3)

0.141

(− 1.9 to 13.1)

KOOS quality of life

25.1

(± 13.0)

24.6

(± 14.1)

0.399

(− 3.3 to 4.3)

56.8

(± 20.6)

58.4

(± 20.3)

0.577

(− 7.2 to 4.1)

71.9

(± 21.5)

69.8

(± 21.8)

0.495

(− 3.9 to 8.1)

FJS-12

17.6

(± 12.4)

13.5

(± 13.1)

0.026

(0.5 to 7.6)

48.2

(± 25.6)

41.7

(± 26.0)

0.078

(− 0.7 to 13.8)

67.3

(± 25.0)

60.0

(± 29.2)

0.063

(− 0.4 to 15.1)

EQ-5D-3L

0.615

(± 0.184)

0.614

(± 0.182)

0.983

(− 0.050 to 0.051)

0.826

(± 0.158)

0.780

(± 0.149)

0.066

(− 0.003 to 0.088)

0.876

(± 0.142)

0.858

(± 0.140)

0.351

(− 0.020 to 0.057)

EQ-VAS

63.8

(± 22.3)

62.2

(± 20.5)

0.592

(− 5.5 to 6.9)

79.4

(± 13.7)

74.4

(± 15.5)

0.055

(− 0.1 to 8.9)

81.2

(± 15.6)

79.4

(± 14.9)

0.405

(− 2.4 to 5.9)

CIM customised individually made, OTS off-the-shelf, n number, SD standard deviation, CI confidence interval, KSS Knee Society Score, KOOS Knee injury and Osteoarthritis Outcome Score, FJS-12 Forgotten Joint Score, VAS visual analogue scale

Flow chart of recruitment. CIM customised individually made, OTS off-the-shelf, TKA total knee arthroplasty, PROM patient-reported outcome measure, n number Patients’ characteristics CIM customised individually made, OTS off-the-shelf, n number, SD standard deviation, ASA American Society of Anesthesiologists KSS and PROMs before and after the surgery 52.8 (± 11.9) 57.1 (± 13.4) 0.018 (− 7.9 to − 0.8) 90.4 (± 7.7) 85.3 (± 9.1) < 0.001 (2.7 to 7.6) 94.2 (± 7.5) 88.9 (± 8.6) < 0.001 (2.8 to 7.7) 51.2 (± 16.8) 46.0 (± 18.0) 0.018 (0.4 to 10.1) 68.6 (± 15.3) 69.9 (± 16.4) 0.543 (− 5.8 to 3.1) 77.8 (± 16.0) 79.2 (± 14.9) 0.507 (− 5.6 to 2.8) 49.6 (± 16.4) 41.9 (± 13.7) < 0.001 (3.7 to 11.7) 71.7 (± 16.9) 71.7 (± 16.8) 1.000 (− 4.7 to 4.7) 84.5 (± 16.1) 82.2 (± 15.8) 0.321 (− 2.2 to 6.6) 57.4 (± 15.8) 48.2 (± 15.7) < 0.001 (4.8 to 13.5) 78.0 (± 14.8) 77.1 (± 14.0) 0.671 (− 3.1 to 4.8) 87.2 (± 13.8) 84.3 (± 14.5) 0.151 (− 3.1 to 6.8) 23.4 (± 18.8) 19.9 (± 18.3) 0.198 (− 1.8 to 8.7) 49.9 (± 24.5) 51.1 (± 25.7) 0.764 (−8.9 to 6.6) 68.1 (± 21.8) 62.4 (± 27.3) 0.141 (− 1.9 to 13.1) 25.1 (± 13.0) 24.6 (± 14.1) 0.399 (− 3.3 to 4.3) 56.8 (± 20.6) 58.4 (± 20.3) 0.577 (− 7.2 to 4.1) 71.9 (± 21.5) 69.8 (± 21.8) 0.495 (− 3.9 to 8.1) 17.6 (± 12.4) 13.5 (± 13.1) 0.026 (0.5 to 7.6) 48.2 (± 25.6) 41.7 (± 26.0) 0.078 (− 0.7 to 13.8) 67.3 (± 25.0) 60.0 (± 29.2) 0.063 (− 0.4 to 15.1) 0.615 (± 0.184) 0.614 (± 0.182) 0.983 (− 0.050 to 0.051) 0.826 (± 0.158) 0.780 (± 0.149) 0.066 (− 0.003 to 0.088) 0.876 (± 0.142) 0.858 (± 0.140) 0.351 (− 0.020 to 0.057) 63.8 (± 22.3) 62.2 (± 20.5) 0.592 (− 5.5 to 6.9) 79.4 (± 13.7) 74.4 (± 15.5) 0.055 (− 0.1 to 8.9) 81.2 (± 15.6) 79.4 (± 14.9) 0.405 (− 2.4 to 5.9) CIM customised individually made, OTS off-the-shelf, n number, SD standard deviation, CI confidence interval, KSS Knee Society Score, KOOS Knee injury and Osteoarthritis Outcome Score, FJS-12 Forgotten Joint Score, VAS visual analogue scale The KSS after 4 and 12 months was higher for CIM TKA (p < 0.001, Table 2). Anatomical alignment and stability was better for CIM TKA (Table 3 and Supplementary material Table 6). For some patients (14%), the KSS after 12 months was missing: they did not attend the control visit because of freedom of complaints or restrictions due to the pandemic situation in 2020.
Table 3

Surgeon completed part of the KSS in detail before and after the surgery

KSSBefore surgery12 months
CIMn = 74OTSn = 169p valueCIMn = 66OTSn = 144p value
Anatomic alignment, n (%)0.1500.072
Neutral: 2–10° valgus12(16%)42(25%)64(97%)130(90%)
Varus: < 2° valgus46(62%)83(49%)1(2%)13(9%)
Valgus: > 10° valgus16(22%)44(26%)1(2%)1(1%)
Medial/lateral instability, n (%) < 0.001 < 0.001
None0(0%)19(11%)55(83%)71(49%)
Little or < 5 mm20(27%)95(56%)10(15%)60(42%)
Moderate or 5 mm45(61%)48(28%)1(2%)13(9%)
Severe or > 5 mm9(12%)7(4%)0(0%)0(0%)
Anterior/posterior instability, n (%) < 0.001 < 0.001
None5(7%)56(33%)60(91%)76(53%)
Moderate < 5 mm64(86%)109(65%)6(9%)66(46%)
Severe > 5 mm5(7%)4(2%)0(0%)2(1%)
Range of motion, mean° (± SD)118(± 15)116(± 15)0.268126(± 10)124(± 10)0.378
Flexion contracture, n (%)0.0880.234
None18(24%)53(31%)64(97%)132(92%)
1–5°20(27%)55(33%)1(2%)12(8%)
6–10°24(34%)40(24%)0(0%)0(0%)
11–15°9(11%)19(11%)1(2%)0(0%)
 > 15°3(4%)2(1%)0(0%)0(0%)
Extensor lag, n (%)0.1930.236
None18(24%)50(30%)64(97%)133(92%)
 < 10°38(51%)93(55%)1(2%)11(8%)
10–20°17(23%)26(15%)1(2%)0(0%)
 > 20°1(1%)0(0%)0(0%)0(0%)

CIM customised individually made, OTS off-the-shelf, KSS Knee Society Score, n number, SD standard deviation

Surgeon completed part of the KSS in detail before and after the surgery CIM customised individually made, OTS off-the-shelf, KSS Knee Society Score, n number, SD standard deviation Patient satisfaction was high in both groups and did not differ after 4 and 12 months (p = 0.670 and p = 0.663, Table 4). All PROMs improved for both groups from baseline to 4 and 12 months, and from 4 to 12 months (p < 0.001). No differences in PROMs were found between the groups after 12 months (Table 2, Fig. 2). In both groups, the majority of patients improved above the MID after 12 months (Fig. 3). For overall improvement, most patients reported a “very much better” or “substantially better” overall knee state (CIM 86%, OTS 87%). Almost all patients would undergo the surgery again (CIM 95%, OTS 95%).
Table 4

Patient satisfaction after 4 and 12 months

4 months12 months
CIMn = 70OTSn = 160CIMn = 74OTSn = 168
Satisfied60(86%)141(88%)64(87%)150(89%)
 Very satisfied30(43%)77(48%)34(46%)78(46%)
 Satisfied30(43%)64(40%)30(41%)72(43%)
Not satisfied10(14%)19(12%)10(13%)18(11%)
 Neutral7(10%)18(11%)5(7%)15(9%)
 Unsatisfied1(1%)1(1%)4(5%)3(2%)
 Very unsatisfied2(3%)0(0%)1(1%)0(0%)

CIM customised individually made, OTS off-the-shelf, n number

Fig. 2

Boxplots of KSS and PROMs before and after the surgery. CIM customised individually made, OTS off-the-shelf, KSS Knee Society Score, KOOS Knee injury and Osteoarthritis Outcome Score, FJS-12 Forgotten Joint Score, VAS visual analogue scale

Fig. 3

Proportion of patients with PROM results above the MID after 12 months. CIM customised individually made, OTS off-the-shelf, MID minimal important difference, KOOS Knee injury and Osteoarthritis Outcome Score, FJS-12 Forgotten Joint Score

Patient satisfaction after 4 and 12 months CIM customised individually made, OTS off-the-shelf, n number Boxplots of KSS and PROMs before and after the surgery. CIM customised individually made, OTS off-the-shelf, KSS Knee Society Score, KOOS Knee injury and Osteoarthritis Outcome Score, FJS-12 Forgotten Joint Score, VAS visual analogue scale Proportion of patients with PROM results above the MID after 12 months. CIM customised individually made, OTS off-the-shelf, MID minimal important difference, KOOS Knee injury and Osteoarthritis Outcome Score, FJS-12 Forgotten Joint Score Adverse events are described in Table 5. Revision rate was 0% for CIM TKA and 2% (4/182) for OTS TKA.
Table 5

Number of postoperative adverse events

Adverse eventsCIMOTS
Arthrolysis2
Diagnostic arthroscopy to exclude an infection1
Pulmonary embolism1
Deceased1
Secondary patellar resurfacing2*
Complete revision2*
Quadriceps tendon rupture1*

CIM customised individually made, OTS off-the-shelf, n number

*Excluded from the analyses

Number of postoperative adverse events CIM customised individually made, OTS off-the-shelf, n number *Excluded from the analyses

Subgroup differences

Male patients were on average 1.6 years younger (p = 0.148) and mostly presented with higher preoperative PROMs, in total and in both TKA groups. Postoperative, no differences were found in PROMs compared to female patients (Supplementary material Table 8 and Fig. 4). Satisfaction after 12 months was slightly higher in female patients (female 90%, male 85%; p = 0.164) but not correlated to sex. One third of the patients were younger than 65 years. Subgroup analyses between patients younger and older than 65 years revealed no preoperative differences. After 4 months, younger patients presented with lower PROMs (Supplementary material Table 9 and Fig. 5). After 12 months, the KOOS symptoms and KOOS quality of life were still lower for younger patients. Satisfaction after 12 months did not differ (younger 88%, older 89%; p = 1.000) and was not correlated to age.

Correlation

Satisfaction after 12 months was not correlated to any variables of patients’ characteristics (p = 0.121 to 0.963), type of implant (p = 0.740), preoperative KSS (p = 0.446) or preoperative PROMs (p = 0.168 to 0.892). A medium to strong correlation existed between satisfaction after 12 months and all outcomes after 4 and 12 months (Supplementary material Table 6).

Discussion

The most important finding of the present study was that patient satisfaction after 12 months was high and did not differ between CIM and OTS TKA. Our hypothesis could not be confirmed. Further, the percentage of satisfied patients in both groups was in the same range or slightly higher than reported in previous studies [6, 29, 47, 55]. Preoperative, patients in the CIM TKA group tend to be younger, male and presented with less subjective impairment. In our hospital, CIM TKA are currently only covered for patients with a supplementary insurance, which explains the difference in insurance status and might be an indicator for a higher socio-economic status. However, data about the socio-economic status of our patients are not available. The current study is the first to prospectively compare CIM and OTS TKA regarding patient satisfaction and PROMs. A high satisfaction rate and clear improvements over time in all PROMs were found; however, there were no differences between both groups. Another retrospective study found comparable KOOS results, also without any differences, but with a higher satisfaction in patients with CIM TKA [32]. The results of the KOOS sports were lower compared to the other KOOS subscales, which can possibly be explained by the advanced age at which patients no longer perform activities like running, jumping or kneeling. Advanced PROMs, like the FJS-12, are more discriminating and might be more appropriate for younger and active patients [49]. Patients with CIM TKA tended to have a higher FJS-12, indicating that these patients are less aware of their artificial joint in everyday life. However, the FJS-12 was also noticeably lower in younger patients. The postoperative KSS was high and, likewise in another study [32], higher for CIM TKA. CIM TKA were mostly neutrally aligned, which is due to the implant design, instruments and the planning of the surgery [45]. Stability was clearly better in CIM TKA, notably the issue of midflexion instability was eliminated in most of the cases [33, 43]. The shape of the CIM implant resembles the original condylar shapes, consequently, the interaction between the individually given ligaments and the implanted prosthesis results in better kinematical behaviour. It has been shown, that CIM TKAs are associated with an increased manipulation rate due to sporadic lower range of motion [53], which we could not confirm. Understandably, surgeons tend to fit CIM TKA too tight in the beginning, because of the initial fear to resect too much bone and due to the limited implant options. Experience helps to find the ideal laxity references intraoperative which results in a low manipulation rate. A higher rate of revisions and adverse events was observed in patients with OTS TKA. It might be due to the smaller sample of CIM TKA. A longer follow-up is needed to clearly show differences between the two groups. Regardless, our findings are in accordance with data from the Swiss Implant Registry: the 2-year revision rate is 3.4% (iTotal: data not yet available; Attune®: 3.6%), with patellar problems occurring in about every third patient scheduled for revision [42]. Other studies on CIM TKA found improved alignment [18, 36, 45], decreased blood loss and hospital stay [39] and reduced total costs [27]. A large retrospective study recently found a high satisfaction rate (89%, mean follow-up 2.8 years) and high implant survivorship (98.5%) [37]. However, based on the current evidence, CIM TKA are not clearly superior [24, 25, 51] and long-term results are not yet available. A recent systematic review revealed the lack of strong methodological studies [51]. Dissatisfaction after TKA remains challenging and influencing factors are not yet fully understood. In addition to implant design and clinical outcome, PROMs need to be further investigated. Strengths of this study are the prospective comparative design and the number of recruited patients, as the availability of CIM TKAs is relatively recent. The number of loss to follow-up was very low and a broad set of PROMs was used. Nevertheless, our study has some limitations and needs to be interpreted accordingly. A selection bias could not be prevented as we are located in a private hospital and CIM TKA patients require a supplementary insurance. Given that and the possible intolerance of patients, a randomised controlled trial has not yet been practical. However, the Attune® implant in the OTS TKA group is with over 3000 implants per year by far the most applied implant in Switzerland [42]. Because of the observational study design, the influence of other confounders cannot be excluded. The study focussed on patient satisfaction and PROMs, hence, knee alignment was not examined in detail, e.g. the hip–knee–ankle angle. Relevant differences between the two groups in postoperative alignment, classified according to the KSS, were not found. The follow-up is only short-term as long-term data are preferable but not yet available. However, for studies with PROMs as primary outcome, it was shown that a 12-month follow-up after TKA is adequate as results are consistent [28, 35]. This study showed that patient satisfaction is high for CIM and OTS TKA, and both achieve good clinical outcome and PROMs.

Conclusions

This prospective cohort study found no difference in satisfaction between patients with CIM and OTS TKA after 12 months. In both groups, satisfaction was high and PROMs improved considerably. We believe that it is important to further investigate patient-reported outcome and factors that affect the outcome to gain a better understanding of satisfaction after TKA. Below is the link to the electronic supplementary material. Supplementary file1 (DOCX 39 KB) Supplementary file2 Boxplots of KSS and PROMs before and after the surgery: comparison female and male patients. KSS Knee Society Score, KOOS Knee injury and Osteoarthritis Outcome Score, FJS-12 Forgotten Joint Score, VAS visual analogue scale (PDF 15 KB) Supplementary file3 Boxplots of KSS and PROMs before and after the surgery: comparison younger and older patients. KSS Knee Society Score, KOOS Knee injury and Osteoarthritis Outcome Score, FJS-12 Forgotten Joint Score, VAS visual analogue scale (PDF 15 KB)
  42 in total

1.  Mismatched knee implants in Indonesian and Dutch patients: a need for increasing the size.

Authors:  Nicolaas C Budhiparama; Imelda Lumban-Gaol; Nadia Nastassia Ifran; Pieter C J de Groot; Dwikora Novembri Utomo; Rob G H H Nelissen
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2020-03-11       Impact factor: 4.342

2.  Functional knee phenotypes: a novel classification for phenotyping the coronal lower limb alignment based on the native alignment in young non-osteoarthritic patients.

Authors:  Michael T Hirschmann; Lukas B Moser; Felix Amsler; Henrik Behrend; Vincent Leclerq; Silvan Hess
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-04-11       Impact factor: 4.342

Review 3.  Patient Dissatisfaction Following Total Knee Arthroplasty: A Systematic Review of the Literature.

Authors:  Rajitha Gunaratne; Dylan N Pratt; Joseph Banda; Daniel P Fick; Riaz J K Khan; Brett W Robertson
Journal:  J Arthroplasty       Date:  2017-07-21       Impact factor: 4.757

4.  Patient satisfaction after total knee arthroplasty: who is satisfied and who is not?

Authors:  Robert B Bourne; Bert M Chesworth; Aileen M Davis; Nizar N Mahomed; Kory D J Charron
Journal:  Clin Orthop Relat Res       Date:  2010-01       Impact factor: 4.176

Review 5.  Considerable inter-individual variability of tibial geometric ratios renders bone-implant mismatch unavoidable using off-the-shelf total knee arthroplasty: a systematic review and meta-analysis.

Authors:  Lucas Beckers; Jacobus H Müller; Jeremy Daxhelet; Salvatore Ratano; Mo Saffarini; Tarik Aït-Si-Selmi; Michel P Bonnin
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2021-06-01       Impact factor: 4.342

6.  Alignment in total knee arthroplasty - A comparison of patient-specific implants with the conventional technique.

Authors:  Dariusch Arbab; Pia Reimann; Martin Brucker; Bertil Bouillon; Christian Lüring
Journal:  Knee       Date:  2018-06-20       Impact factor: 2.199

Review 7.  Application of minimal important differences in degenerative knee disease outcomes: a systematic review and case study to inform BMJ Rapid Recommendations.

Authors:  Tahira Devji; Gordon H Guyatt; Lyubov Lytvyn; Romina Brignardello-Petersen; Farid Foroutan; Behnam Sadeghirad; Rachelle Buchbinder; Rudolf W Poolman; Ian A Harris; Alonso Carrasco-Labra; Reed A C Siemieniuk; Per O Vandvik
Journal:  BMJ Open       Date:  2017-05-11       Impact factor: 2.692

Review 8.  Patient Satisfaction After Total Knee Replacement: A Systematic Review.

Authors:  Cynthia A Kahlenberg; Benedict U Nwachukwu; Alexander S McLawhorn; Michael B Cross; Charles N Cornell; Douglas E Padgett
Journal:  HSS J       Date:  2018-06-05

9.  Minimal important differences for the WOMAC osteoarthritis index and the Forgotten Joint Score-12 in total knee arthroplasty patients.

Authors:  N Holtz; D F Hamilton; J M Giesinger; B Jost; K Giesinger
Journal:  BMC Musculoskelet Disord       Date:  2020-06-23       Impact factor: 2.362

Review 10.  Contemporary knee arthroplasty: one fits all or time for diversity?

Authors:  Johannes Beckmann; Malin Kristin Meier; Christian Benignus; Andreas Hecker; Emmanuel Thienpont
Journal:  Arch Orthop Trauma Surg       Date:  2021-07-16       Impact factor: 3.067

View more
  1 in total

1.  Preoperative Factors Affecting Patient-reported Outcome Measures for Total Knee Arthroplasty.

Authors:  Yuya Mawarikado; Yusuke Inagaki; Tadashi Fujii; Hidetaka Imagita; Takahiko Fukumoto; Takanari Kubo; Mimo Shirahase; Akira Kido; Yasuhito Tanaka
Journal:  Prog Rehabil Med       Date:  2022-09-15
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.