Simin Mahinrad1, Can Ozan Tan2,3, Yue Ma4,5, Maria Aristova4,6, Andrew L Milstead1, Donald Lloyd-Jones7, Susanne Schnell4,8, Michael Markl4,6, Farzaneh A Sorond1.
Abstract
BACKGROUND: Dual-venc 4D flow MRI, recently introduced for the assessment of intracranial hemodynamics, may provide a promising complementary approach to well-established tools such as transcranial Doppler ultrasound (TCD) and overcome some of their disadvantages. However, data comparing intracranial flow measures from dual-venc 4D flow MRI and TCD are lacking.Entities:
Keywords: 4D flow MRI; cerebrovascular hemodynamics; healthy subjects; transcranial Doppler ultrasound
Mesh:
Year: 2022 PMID: 35146822 PMCID: PMC9363520 DOI: 10.1002/jmri.28115
Source DB: PubMed Journal: J Magn Reson Imaging ISSN: 1053-1807 Impact factor: 5.119
Average MRI Scan Parameters (range in parenthesis)
| Parameters | 3D TOF MRA | Dual‐venc 4D flow MRI |
|---|---|---|
| TR, msec | 22 | 5.6 (5.5–5.6) |
| TE | 3.75 | 3.29 (3.17–3.29) |
| Flip angle, ° | 18 | 15 |
| Temporal resolution, msec | N/A | 78.3 (77–78.4) |
| Spatial resolution, mm3 | [0.5 × 0.26 × 0.26] | [1 × 1.04 × 1.04] |
| Low Venc, m/sec | N/A | 0.5 (0.5–0.6) |
| High Venc, m/sec | N/A | 1.0 (1.0–1.2) |
TR = repetition time; TE = echo time.
FIGURE 1Dual‐venc 4D flow MRI and transcranial Doppler ultrasound workflow. 4D flow MRI workflow example shows: (a) the volume coverage of the dual‐venc 4D flow MRI data based on TOF MRI; (b) the dual‐venc flow raw data including magnitude images and velocity encoded phase images in x, y, and z directions; (c) preprocessed data after eddy current correction, noise filtering, and velocity anti‐aliasing; (d) visualization and quantification of data by (d1) visualization of streamlines; (d2) PC‐MRA vessel segmentation; (d3) extraction of centerlines; (d4) placement of 2D analysis planes and (d5) lumen segmentation and velocity profile on a 2D plane. Transcranial Doppler workflow example shows (e) insonation of the middle cerebral artery via the transtemporal window; (f) the Doppler signal of the middle cerebral artery with the color bar representing signal intensity; (G) the digitization of the Doppler signal by Fast Fourier transform displayed by Windaq software; and (h) the middle cerebral artery flow velocities over time. FOV = field of view; TOF = time‐of‐flight; TCD = transcranial Doppler; MCA = middle cerebral artery; RMCA = right middle cerebral artery; LMCA = left middle cerebral artery.
FIGURE 2Example of intracranial flow analysis from dual‐venc 4D flow MRI in a representative participant. (a) 3D visualization of time‐resolved arterial blood flow during early systole (t = 39.2 msec) in the circle of Willis (color coded by velocity magnitude from red to blue). (b) Peak velocity values over one cardiac cycle for the M1 segment of middle cerebral arteries. Peak velocity values represent the median over the M1 segment of the middle cerebral artery for several analysis planes with 1 mm distance for this representative participant. MCA = middle cerebral artery; ACA = anterior cerebral artery; PCA = posterior cerebral artery; ICA = internal carotid artery.
Interobserver Agreement of MCA Peak Velocity as Measured by Dual‐venc 4D Flow MRI
| LMCA | RMCA | |
|---|---|---|
| Observer 1 mean (±SD), m/sec | 0.65 (0.16) | 0.62 (0.12) |
| Observer 2 mean (±SD), m/sec | 0.62 (0.15) | 0.64 (0.13) |
| Mean difference (±SD), m/sec | 0.02 (0.05) | −0.02 (0.05) |
| ICC, | 0.973 | 0.962 |
| Coefficient of variance, % | 6% | 5% |
ICC = intraclass correlation coefficient; LMCA = left middle cerebral artery; RMCA = right middle cerebral artery.
FIGURE 3Bland–Altman plots for inter‐observer agreement of MCA‐M1 peak velocity by dual‐venc 4D flow MRI. LMCA = left middle cerebral artery; RMCA = right middle cerebral artery.
Comparison Between TCD and Dual‐venc 4D Flow MRI in Measurement of MCA Peak Velocity
| LMCA | RMCA | |
|---|---|---|
| TCD mean (±SD), m/sec | 0.76 (0.16) | 0.77 (0.13) |
| 4D flow MRI mean (±SD), m/sec | 0.64 (0.15) | 0.61 (0.12) |
| Mean difference TCD – MRI (±SD), m/sec | 0.13 (0.24) | 0.15 (0.16) |
| Coefficient of variance, % | 26% | 22% |
4D flow MRI results are from the observer 1.
LMCA = left middle cerebral artery; RMCA = right middle cerebral artery.
FIGURE 4Bland–Altman plots of dual‐venc 4D flow MRI compared with TCD for MCA‐M1 peak velocity measurement. TCD = transcranial Doppler ultrasound; LMCA = left middle cerebral artery; RMCA = right middle cerebral artery.