| Literature DB >> 35146030 |
Gerardo L Garcés1, Oscar Martel2, Alejandro Yánez2, Ignacio Manchado-Herrera3, Luci M Motta3.
Abstract
BACKGROUND: It is not clear whether the mechanical strength of adjustable-loop suspension devices (ALDs) in anterior cruciate ligament (ACL) reconstruction is device dependent and if these constructs are different from those of an interference screw.Entities:
Keywords: anterior cruciate ligament reconstruction; biomechanical testing; interference screw; suspensory fixation
Year: 2021 PMID: 35146030 PMCID: PMC8821986 DOI: 10.1177/23259671211031652
Source DB: PubMed Journal: Orthop J Sports Med ISSN: 2325-9671
Figure 1.UltraButton anterior cruciate ligament reconstruction in the testing machine.
Figure 2.UltraButton in the isolated device testing setup.
Figure 3.Mean cyclic displacement of the graft in the anterior cruciate ligament (ACL) in vitro reconstructions.
Specimen Testing Results
| TR | UB | IS |
| |||
|---|---|---|---|---|---|---|
| TR vs UB | TR vs IS | UB vs IS | ||||
| Displacement at cycle 1000, mm | 3.42 ± 1.34 | 3.39 ± 0.92 | 7.54 ± 3.18 | .999 |
|
|
| Ultimate failure load, N | 730 ± 115 | 781 ± 193 | 467 ± 94 | .704 |
|
|
| Displacement at ultimate failure, mm | 13.12 ± 8.79 | 14.49 ± 6.36 | 25.56 ± 6.68 | .910 |
|
|
| Yield load, N | 420 ± 72 | 547 ± 173 | 386 ± 51 |
| .877 |
|
| Stiffness, N/mm | 197 ± 58 | 248 ± 96 | 243 ± 83 | .496 | .586 | .992 |
| Failure mechanism | Cortical blowout (50%), suture tear (50%) | Cortical blowout (50%), tendon failure (50%) | Tendon slippage (90%), tendon failure (10%) | |||
Data are reported as mean ± SD unless otherwise indicated. Bold P values indicate statistically significant difference between groups (P < .05). IS, interference screw; TR, TightRope; UB, UltraButton.
Figure 4.TightRope device (A) before failure and (B) after cortical blowout of the button with subsequent tendon slippage.
Isolated Device Testing Results
| TR | UB |
| |
|---|---|---|---|
| Displacement at cycle 10, mm | 0.06 ± 0.08 | 0.12 ± 0.05 | .074 |
| Displacement at cycle 50, mm | 0.15 ± 0.09 | 0.26 ± 0.07 |
|
| Displacement at cycle 100, mm | 0.18 ± 0.08 | 0.34 ± 0.10 |
|
| Displacement at cycle 500, mm | 0.58 ± 0.51 | 0.50 ± 0.12 | .606 |
| Displacement at cycle 1000, mm | 0.71 ± 0.59 | 0.57 ± 0.15 | .486 |
| Ultimate failure load, N | 862 ± 64 | 1879 ± 126 |
|
| Displacement at ultimate failure, mm | 4.03 ± 1.40 | 4.37 ± 0.36 | .469 |
| Stiffness, N/mm | 542 ± 164 | 449 ± 38 | .112 |
| Failure mechanism | Suture tear (100%) | Suture tear (70%), button fracture (30%) |
Data are reported as mean ± SD unless otherwise indicated. Bold P values indicate statistically significant difference between groups (P < .05). TR, TightRope; UB, UltraButton.