Joseph D Mozingo1, Lindsay L Schuring1,2, Alexander J Mortensen1, Andrew E Anderson1,2,3,4, Stephen K Aoki1. 1. Department of Orthopaedics, University of Utah, Salt Lake City, Utah, USA. 2. Department of Biomedical Engineering, University of Utah, Salt Lake City, Utah, USA. 3. Department of Physical Therapy and Athletic Training, University of Utah, Salt Lake City, Utah, USA. 4. Scientific Computing and Imaging Institute, University of Utah, Salt Lake City, Utah, USA.
Abstract
BACKGROUND: The anterior center-edge angle (ACEA) is used to quantify anterior coverage of the femoral head by the acetabulum. However, its measurement has not been evaluated in a manner consistent with routine use, and the precise 3-dimensional (3D) anatomic location where it measures coverage is not known. PURPOSE: To determine the effect of patient positioning on ACEA measurement reliability, magnitude, and 3D location. STUDY DESIGN: Descriptive laboratory study. METHODS: Included were 18 adults; 7 participants had cam morphology and femoroacetabular impingement syndrome, and 11 participants had no radiographic evidence of hip abnormalities and no history of hip pain or injuries. Ultimately, 3D femur and pelvis models were generated from computed tomography images. Radiographs were generated with the models in different degrees of pelvic rotation, tilt, and obliquity relative to the standard false-profile view. The ACEA was measured by 2 raters by selecting the location of the bone edge on each radiograph. Selections were projected onto the pelvis model and expressed as a clockface location on the acetabular rim. The clockface was mirrored on left hips to allow a direct comparison of locations between hips. Interrater and intrarater reliability were quantified via the intraclass correlation coefficient (ICC). The effect of position on ACEA measurements and clockface locations was determined via linear regression. RESULTS: Intrarater and interrater reliability were excellent (ICC ≥0.97 for all). For every degree increase in rotation, tilt, and obliquity, the ACEA changed by +0.53°, +0.93°, and -0.04°, respectively. The mean clockface location (hour:minute:second) in the false-profile view was 2:09:32 ± 0:12:00 and changed by +0:02:08, -0:00:35, and -0:00:05 for every degree increase in rotation, tilt, and obliquity, respectively. CONCLUSION: ACEA measurements were reliable even with differences in patient positioning. Rotation and tilt were associated with notable changes in ACEA measurements. ACEA bone edge measurements mapped to the anterosuperior acetabular rim, typically in proximity to the anterior inferior iliac spine. Mapped location was most sensitive to rotation. CLINICAL RELEVANCE: Pelvic rotation and tilt affected ACEA measurements, which could alter the clinical classification and treatment of borderline abnormalities. Rotation in particular must be well controlled during patient imaging to preserve measurement reliability and accuracy and to describe coverage from the intended 3D rim location.
BACKGROUND: The anterior center-edge angle (ACEA) is used to quantify anterior coverage of the femoral head by the acetabulum. However, its measurement has not been evaluated in a manner consistent with routine use, and the precise 3-dimensional (3D) anatomic location where it measures coverage is not known. PURPOSE: To determine the effect of patient positioning on ACEA measurement reliability, magnitude, and 3D location. STUDY DESIGN: Descriptive laboratory study. METHODS: Included were 18 adults; 7 participants had cam morphology and femoroacetabular impingement syndrome, and 11 participants had no radiographic evidence of hip abnormalities and no history of hip pain or injuries. Ultimately, 3D femur and pelvis models were generated from computed tomography images. Radiographs were generated with the models in different degrees of pelvic rotation, tilt, and obliquity relative to the standard false-profile view. The ACEA was measured by 2 raters by selecting the location of the bone edge on each radiograph. Selections were projected onto the pelvis model and expressed as a clockface location on the acetabular rim. The clockface was mirrored on left hips to allow a direct comparison of locations between hips. Interrater and intrarater reliability were quantified via the intraclass correlation coefficient (ICC). The effect of position on ACEA measurements and clockface locations was determined via linear regression. RESULTS: Intrarater and interrater reliability were excellent (ICC ≥0.97 for all). For every degree increase in rotation, tilt, and obliquity, the ACEA changed by +0.53°, +0.93°, and -0.04°, respectively. The mean clockface location (hour:minute:second) in the false-profile view was 2:09:32 ± 0:12:00 and changed by +0:02:08, -0:00:35, and -0:00:05 for every degree increase in rotation, tilt, and obliquity, respectively. CONCLUSION: ACEA measurements were reliable even with differences in patient positioning. Rotation and tilt were associated with notable changes in ACEA measurements. ACEA bone edge measurements mapped to the anterosuperior acetabular rim, typically in proximity to the anterior inferior iliac spine. Mapped location was most sensitive to rotation. CLINICAL RELEVANCE: Pelvic rotation and tilt affected ACEA measurements, which could alter the clinical classification and treatment of borderline abnormalities. Rotation in particular must be well controlled during patient imaging to preserve measurement reliability and accuracy and to describe coverage from the intended 3D rim location.
The anterior center-edge angle (ACEA) characterizes coverage of the femoral head by the acetabulum,
which is an important consideration in the diagnosis and treatment for patients
with acetabular dysplasia, pincer morphology, and other abnormalities of the hip.
Historically, an ACEA less than 20° has been deemed abnormal, whereas 20° to 25° is
considered borderline.
An accurate interpretation of the ACEA is dependent on a comprehensive
understanding of the effects of radiographic techniques and measurement errors.
Additionally, understanding how the 2-dimensional (2D) ACEA relates to 3-dimensional
(3D) anatomic features is critical for correctly identifying hip abnormalities and
choosing between arthroscopic, reorienting, and nonoperative treatment.Previous studies have shown the ACEA to be a reliable measurement.
However, reliability measurements included only images acquired in the
false-profile view (eg, cadaveric and computational studies
and a patient study that used a positioning device
). Thus, the reported reliability measurements do not account for the imprecise
nature of patient positioning during routine imaging.Less is known regarding the effect that patient positional error has on the ACEA, which
occurs as a result of improper pelvic rotation relative to the image cassette and/or
pelvic tilt and obliquity caused by an unnatural posture adopted during imaging. On the
basis that a 20° rotational error resulted in a change of <4° in the ACEA relative to
the false-profile view, Putnam et al
concluded that rotational errors were unlikely to alter clinical interpretation.
However, for the same magnitude of rotation, Li et al
found nearly double the change in the ACEA, which would be a concern when
evaluating patients who present with borderline abnormalities. In comparison, Zingg et al
found tilt to have a much greater effect than rotation, corroborated by Putnam et al,
in which a 20° tilt error yielded a 13° change in the ACEA relative to the
false-profile view. Importantly, these previous investigations were conducted in vitro
and did not consider the effect of pelvic obliquity on ACEA measurements. Accordingly,
the performance of this measurement under realistic conditions is not well
established.The ACEA measurement provides valuable information regarding the extent of femoral head
coverage by the acetabulum. However, the 3D location that the ACEA approximates, and
thus the region of coverage to which the ACEA corresponds, has not been well described.
Previous efforts have utilized computed tomography (CT) and magnetic resonance imaging
(MRI) to measure the ACEA on sagittal-plane
and custom reformatted
images. Although this may circumvent challenges with patient positioning
associated with radiography, the functional relative relationship of the femur and hip
from the standing position is not preserved. Indeed, ACEA measurements on planar CT/MRI
have been shown to exhibit weak correlations with their radiography-based counterpart.
More importantly, such an approach does not leverage the 3D information inherent
to volumetric images.In this study, CT and dual fluoroscopy were used to generate participant-specific
false-profile radiographs based on the known relative position of the femur and pelvis
obtained in a standing position. The volumetric CT dataset can then be used to create
radiographs at precise computer-controlled values of pelvic rotation, tilt, and
obliquity. The first goal of this study was to quantify the reliability of ACEA
measurements and determine the sensitivity of such measurements to patient positioning.
The second goal was to identify the precise 3D location on the pelvis measured via the
ACEA and determine the effect of patient positioning on this location.
Methods
Participants
The participant cohort was selected from previously published imaging studies
from which CT and dual fluoroscopy images of the hip were available.
Briefly, after receiving approval from the University of Utah
institutional review board, we enrolled 18 participants between March 2013 and
January 2016; all participants provided informed consent in this single-center
study. The mean participant age and body mass index were 26 ± 5 years and 22.0 ±
2.8, respectively. Participants had no history of lower limb surgery and no
radiographic evidence of osteoarthritis. A total of 7 participants (5 male and 2
female) had a diagnosis of femoroacetabular impingement (FAI) syndrome.
The diagnosis was based on the presence of symptoms, a positive anterior
impingement examination finding, and the presence of cam morphology on radiographs.
There were 11 participants (6 male and 5 female) who had no radiographic
evidence of hip abnormalities and no history of hip pain or injuries.
Imaging
CT of the hip was performed via a 128-slice SOMATOM Definition scanner (Siemens)
using an established protocol.
The pelvis and femur were imaged at 120 kVp and 200 to 400 mAs, and
images were reconstructed with a 1.0-mm slice thickness. Images were resampled
to 3 times the original voxel resolution (resampled range, 0.20-0.25 × 0.20-0.25
× 0.33 mm), and each bone was segmented using Amira software (Version 6.0;
Thermo Fisher Scientific).
Next, a custom dual fluoroscopy system (Radiological Imaging Services)
validated for hip imaging
was used to obtain images of participants’ hips in a neutral standing
position with feet hip-width apart. Images were collected at 100 frames per
second, with energy settings ranging from 78 to 100 kVp and from 1.9 to 3.2 mAs
for tube voltage and current time product, respectively.
Data Processing
Surface meshes of the pelvis and proximal femur were generated from the CT
segmentations and then decimated and smoothed. Model-based tracking registered
the CT bone models to the fluoroscopic image pairs to determine the 3D position
of the femur and pelvis using previously described software.Following a validated approach,
digitally reconstructed radiographs (DRRs) of the hip were generated via
projection of the CT image volumes to simulate plain film radiographs at various
positions. CT image volumes were preprocessed such that voxels outside of the
segmented bone regions were assigned an intensity value of 0, with separate
volumes created for the femur and pelvis anatomies. Next, the femur and pelvis
anatomies were rigidly transformed to the neutral standing position determined
using model-based tracking. The anatomies were then rotated 65° about the
superior-inferior axis through the pelvic center to be in the standard
false-profile view (Figure
1), first described by Lequesne and de Sèze.
Additional positions were created by jointly rotating the femur and
pelvis anatomies relative to the standard false-profile view (45° to 85° in 5°
increments) and in the standard false-profile view with varying degrees of
pelvic tilt and obliquity (–10° to 10° in 2° increments) (Figure 1). Finally, the 3D anatomies
were projected to a 2D plane, analogous to an image cassette, for each joint
configuration.
Figure 1.
Hip position definitions. Top row: Pelvic rotation was defined as
rotation of the femur and pelvis about the superior-inferior axis
through the pelvic center, with the standard false-profile view set at
65° of rotation from a standard anteroposterior view. The pelvic center
was set as the midpoint of the imaged (white) and contralateral (black)
hip joint centers. Bottom row: Pelvic tilt was defined as rotation of
the pelvis about the medial-lateral axis, and pelvic obliquity was
defined as rotation of the pelvis about the anteroposterior axis passing
through the joint center of the imaged hip.
Hip position definitions. Top row: Pelvic rotation was defined as
rotation of the femur and pelvis about the superior-inferior axis
through the pelvic center, with the standard false-profile view set at
65° of rotation from a standard anteroposterior view. The pelvic center
was set as the midpoint of the imaged (white) and contralateral (black)
hip joint centers. Bottom row: Pelvic tilt was defined as rotation of
the pelvis about the medial-lateral axis, and pelvic obliquity was
defined as rotation of the pelvis about the anteroposterior axis passing
through the joint center of the imaged hip.
ACEA Measurement
A custom MATLAB program (Version R2017b; MathWorks) was used to display the DRRs
in random order and record ACEA measurements. Overall, 2 investigators (a
fellowship-trained orthopaedic surgeon [S.K.A.] and a fourth-year medical
student [A.J.M.]) independently performed all measurements, blinded to
participant and hip positioning. The ACEA was defined as the angle between 2
vectors passing through the femoral head center: (1) a vertical vector and (2) a
vector that intersected the anterior aspect of the acetabular bone edge, as
described by Crockarell et al
and Lequesne and de Sèze.
Bone edge points were identified and selected by the raters (Figure 2A). The location
of the femoral head center was predetermined, calculated as the center of the
best-fit sphere to the 3D femoral head mesh projected onto the 2D image plane
(Figure 2A). Raters
repeated measurements 1 week later on a subset of images, which included 4
images per hip position from FAI syndrome and healthy hips as well as left and
right hips.
Figure 2.
Anterior center-edge angle (ACEA) measurement and 3-dimensional location
mapping. (A) Representative digitally reconstructed radiograph (DRR) of
the hip. Raters selected the bone edge (red circle) landmark on the
pelvis. The femoral head center was defined as the center of the
best-fit sphere to the femoral head mesh projected onto the
2-dimensional image plane. The vertical line (black arrow) passing
through the femoral head center (black circle) was defined using the
dual fluoroscopy system. The ACEA was calculated as the angle between
the vertical line and a line extending from the femoral head center
through the selected bone edge point (red circle). (B) Bone edge
selections (red circle) were projected from the DRR to a corresponding
point on the bone models (red diamond) along a line following the
radiographic projection.
Anterior center-edge angle (ACEA) measurement and 3-dimensional location
mapping. (A) Representative digitally reconstructed radiograph (DRR) of
the hip. Raters selected the bone edge (red circle) landmark on the
pelvis. The femoral head center was defined as the center of the
best-fit sphere to the femoral head mesh projected onto the
2-dimensional image plane. The vertical line (black arrow) passing
through the femoral head center (black circle) was defined using the
dual fluoroscopy system. The ACEA was calculated as the angle between
the vertical line and a line extending from the femoral head center
through the selected bone edge point (red circle). (B) Bone edge
selections (red circle) were projected from the DRR to a corresponding
point on the bone models (red diamond) along a line following the
radiographic projection.
3D Location Mapping of the ACEA
Each rater-selected location was extended from the 2D point on the DRR to a 3D
line, following the radiographic projection in the direction from the simulated
image cassette to the x-ray source (Figure 2B). The minimum distance between
the line and the nodes on the 3D pelvis surface mesh (aligned to the orientation
of the DRR) was calculated. The 3D location of the ACEA measurement was defined
as the nearest cassette-facing node on the pelvis within a tolerance threshold.
The threshold was defined as the minimum line-to-node distance across all pelvic
nodes, summed with the median mesh edge length (range, 1.4-1.8 mm). If no nodes
of the acetabular rim were found within this threshold, the nearest node of the
pelvis was selected.To match the clinical description, ACEA 3D locations were projected onto a
clockface defined by the acetabular rim, and Cartesian coordinates were
expressed as a time (hour:minute:second). The clockface was defined by computing
the second principal curvature of the pelvis mesh using PostView software
(Version 2.1; University of Utah)
to isolate the acetabular lunate surface and rim.
The joint center was defined as the center of the best-fit sphere to the
lunate surface. The 6 o’clock position was defined as the midpoint of the
transverse acetabular ligament attachment sites
and projected to a plane fit to the nodes of the acetabular rim (Figure 3). The remaining
clockface positions were then defined relative to 6 o’clock. The clockface was
mirrored on left hips to allow a direct comparison of clockface locations
between left and right hips.
Figure 3.
Definition of the acetabular clockface for expressing the mapped location
of the anterior center-edge angle (ACEA) bone edge selection. Acetabular
landmarks: The lunate surface (red) and acetabular rim (blue) were
isolated on each participant’s mesh using PostView’s second principal
curvature tool. A sphere fit to the lunate surface was used to define
the joint and clockface centers (red circle). Clockface: 6 o’clock was
defined as the midpoint between the anteroinferior and posteroinferior
rim edges (blue “x”). The remainder of the clockface was defined
relative to the 6 o’clock position. The clockface was mirrored on left
hips to allow a direct comparison of clockface locations between left
and right hips.
Definition of the acetabular clockface for expressing the mapped location
of the anterior center-edge angle (ACEA) bone edge selection. Acetabular
landmarks: The lunate surface (red) and acetabular rim (blue) were
isolated on each participant’s mesh using PostView’s second principal
curvature tool. A sphere fit to the lunate surface was used to define
the joint and clockface centers (red circle). Clockface: 6 o’clock was
defined as the midpoint between the anteroinferior and posteroinferior
rim edges (blue “x”). The remainder of the clockface was defined
relative to the 6 o’clock position. The clockface was mirrored on left
hips to allow a direct comparison of clockface locations between left
and right hips.
Statistical Analysis
Reliability was quantified using 2-way mixed-effects, absolute-agreement,
single-measurement intraclass correlation coefficients (ICCs). All measurements
from the first session were used to calculate interobserver reliability. All
measurements from the second session with corresponding measurements from the
first session were used to calculate intraobserver reliability. The relationship
between measurements (ACEA angles, ACEA clockface locations) and pelvic
rotation, tilt, and obliquity was determined using linear regression. ACEA
measurements were averaged between raters for a given participant and hip
position. For regression analysis, measurements were also averaged across
participants.
Results
The intrarater reliability of the ACEA measurements was considered excellent
(ICC, 0.98 [rater 1] and 0.98 [rater 2]), and the interrater reliability was
also excellent (ICC, 0.97). The mean difference in ACEA measurements between the
raters was 1.3° (Table
1).
Table 1
Interrater and Intrarater Reliability for Anterior Center-Edge Angle
Measurements
Intraclass Correlation Coefficient (95% CI)
Absolute Difference, Mean ± SD, deg
Rater 1
0.98 (0.97-0.98)
1.16 ± 1.62
Rater 2
0.98 (0.97-0.99)
1.02 ± 1.07
Interrater
0.97 (0.96-0.97)
1.34 ± 1.60
Interrater and Intrarater Reliability for Anterior Center-Edge Angle
MeasurementsThe mean ACEA in the false-profile view was 51.1° ± 5.2°. For each degree increase in
rotation (ie, toward a true lateral view), the ACEA increased by 0.53° (Figure 4A). For each degree
increase in tilt (ie, anterior tilt), the ACEA increased by 0.93° (Figure 4B). For each degree
increase in obliquity (ie, away from midline), the ACEA decreased by 0.04° (Figure 4C).
Figure 4.
Anterior center-edge angle (ACEA) measurement as a function of hip position
(A) Rotation, (B) Tilt, (C) Obliquity. ACEA measurements were averaged
between raters for a given participant and hip position. Each point
represents the average across participants, and error bars indicate the 95%
CI. Data were fit using linear regression. AB, abduction; AD, adduction;
ANT, anterior; AP, anteroposterior view; LAT, lateral view; POST,
posterior.
Anterior center-edge angle (ACEA) measurement as a function of hip position
(A) Rotation, (B) Tilt, (C) Obliquity. ACEA measurements were averaged
between raters for a given participant and hip position. Each point
represents the average across participants, and error bars indicate the 95%
CI. Data were fit using linear regression. AB, abduction; AD, adduction;
ANT, anterior; AP, anteroposterior view; LAT, lateral view; POST,
posterior.The mean clockface location for the ACEA in the false-profile view was 2:09:32 ± 12
minutes. For each degree increase in pelvic rotation, tilt, and obliquity, the
clockface location changed by 2 minutes and 7.8 seconds, 34.6 seconds, and 5.0
seconds, respectively (Figure
5). The maximum, median, and minimum clockface locations across positions
and participants were 3:20:23, 2:06:03, and 1:26:48 for rotation, respectively;
2:54:37, 2:03:53, and 1:34:56 for tilt, respectively; and 2:54:37, 2:05:46, and
1:35:22 for obliquity, respectively (Figure 5).
Figure 5.
Top: Anterior center-edge angle (ACEA) clockface location as a function of
hip position. ACEA measurements were averaged between raters for a given
participant and hip position. Each point represents the average clockface
location across participants, and error bars indicate the 95% CI. Data were
fit using linear regression. Bottom: Range of clockface locations based on
regression equations calculated over a consistent range of pelvic (A.1, A.2)
rotation, (B.1, B.2) tilt, and (C.1, C.2) obliquity (±10°). AB, abduction;
AD, adduction; ANT, anterior; AP, anteroposterior view; LAT, lateral view;
POST, posterior.
Top: Anterior center-edge angle (ACEA) clockface location as a function of
hip position. ACEA measurements were averaged between raters for a given
participant and hip position. Each point represents the average clockface
location across participants, and error bars indicate the 95% CI. Data were
fit using linear regression. Bottom: Range of clockface locations based on
regression equations calculated over a consistent range of pelvic (A.1, A.2)
rotation, (B.1, B.2) tilt, and (C.1, C.2) obliquity (±10°). AB, abduction;
AD, adduction; ANT, anterior; AP, anteroposterior view; LAT, lateral view;
POST, posterior.For a representative participant with FAI syndrome and healthy participant, the
absolute difference in the median clockface location among raters ranged from
0:01:25 to 0:10:01 for rotation, 0:00:58 to 0:02:56 for tilt, and 0:00:52 to 0:05:18
for obliquity (Figure 6).
The range in clockface locations (defined as the maximum clockface location minus
the minimum clockface location over the examined range of positions) was greatest in
rotation (Figure 6, A.1 and
A.2) 5.4% of bone edge selections mapped in proximity to the iliopubic eminence.
Figure 6.
Anterior center-edge angle (ACEA) clockface locations over the range of
pelvic (A.1, A.2) rotation (±20°), (B.1, B.2) tilt (±10°), and (C.1, C.2)
obliquity (±10°) for a representative participant with femoroacetabular
impingement (FAI) syndrome and a healthy participant.
Anterior center-edge angle (ACEA) clockface locations over the range of
pelvic (A.1, A.2) rotation (±20°), (B.1, B.2) tilt (±10°), and (C.1, C.2)
obliquity (±10°) for a representative participant with femoroacetabular
impingement (FAI) syndrome and a healthy participant.
Discussion
ACEA measurements exhibited excellent intrarater (ICCs, 0.98) and interrater
reliability (ICC, 0.97), and the mean absolute difference between raters was 1.3°
(Table 1). Murphy
et al
and Hanson et al
also found ACEA measurements to be highly consistent across raters, reporting
similar ICCs and median differences of 1° for bone edge measurements. Interestingly,
even when incorporating images with positional errors, we found ICCs that were
comparable to those in studies that only examined reliability in the standard
false-profile view.
However, differences between raters were smallest when the pelvis was rotated
toward an anteroposterior (AP) view (mean absolute difference, 0.70° from –20° to
–15° of rotation) and increased the more the pelvis was rotated toward a lateral
view (mean absolute difference, 4.00° from 15° to 20° of rotation) likely because of
increased overlap in the projected anatomy, obscuring the bone edge.Accurate patient positioning is difficult to achieve using the false-profile
radiographic view in which the pelvis is rotated 65° from the AP view. As a result,
physicians must frequently interpret false-profile radiographs with variable amounts
of pelvic rotation. Additionally, postural perturbations due to constrained space
and positioning during image acquisition can alter pelvic tilt and obliquity. In
this study, changes in pelvic rotation and tilt were both associated with changes in
the ACEA, whereas changes in pelvic obliquity had a minimal effect.For each degree of anterior tilt, Putnam et al
and Zingg et al
found a 0.63° to 0.65° increase in the ACEA (measurements were made to the
sourcil edge). The effect on bone edge measurements was more pronounced, with a
0.93° increase in the ACEA for every degree of anterior tilt in the present study
(Figure 4B). Putnam et al
found that pelvic rotation had a smaller effect on the ACEA for every degree
than did pelvic tilt. This trend was also observed herein. Putnam et al
and Li et al
reported changes in the sourcil ACEA of 0.18° and 0.4° for every degree of
rotation, respectively, compared to 0.53° for every degree of rotation for the bone
edge ACEA in the present study (Figure 4A). Clinicians should take into account the increased
sensitivity to pelvic tilt when using the ACEA to measure anterior acetabular
coverage, especially when the bone edge is used as the anterior reference point.
Putnam et al
concluded that “aberrant pelvic rotation will likely not result in a
clinically meaningful difference in ACEA measurements,” whereas Li et al
asserted that rotation significantly affects ACEA measurements.Although rotation had less of an effect on the ACEA than did tilt overall (Figure 4, A and B), the findings of Li et al
and the present study perhaps call into question the conclusion of Putnam et al,
particularly in cases of borderline abnormalities. On examination of the end
ranges of rotation, rotation toward a true lateral view resulted in greater errors
in ACEA measurements than did rotation toward an AP view (Figure 4A), also observed by Li et al.
Specifically, 20° of rotation toward a true lateral view resulted in a mean
difference in the ACEA of 15.6° relative to the standard false-profile view compared
to only 6.4° when rotated toward an AP view. Based on these findings, ACEA
measurements taken using radiographs with suspected malrotation, particularly toward
a lateral view, should be interpreted cautiously.The ACEA in the false-profile view is generally accepted as a 2D surrogate measure of
anterior acetabular coverage; however, the precise 3D location to which it maps on
the acetabular rim has not been described. The results of the present study
demonstrate that in the standard false-profile view, ACEA measurements mapped to the
anterosuperior acetabular rim in proximity to the anterior inferior iliac spine
(Figure 5), with mapped
locations that were highly consistent between raters (Figure 6). Pelvic rotation had the largest
effect on the ACEA rim location for every degree, with a 20° error resulting in a
mean shift in the clockface location of approximately 42.5 minutes (Figure 5A.1). Rotation toward
an AP view resulted in smaller ACEA values, which were located more superiorly on
the acetabular rim, underestimating femoral head coverage relative to the standard
false-profile view. In comparison, rotation toward a true lateral view resulted in
larger ACEA values, which were located more anteriorly on the acetabular rim,
overestimating femoral head coverage relative to the standard false-profile view.
Although the majority of ACEA measurements mapped to the acetabular rim, 5.4% of
bone edge selections mapped in proximity to the iliopubic eminence. Mapped locations
that were off-rim typically occurred in the presence of excess rotation of the
pelvis toward a true lateral view. This likely contributed to the greater
discrepancy in clockface locations relative to the standard false-profile view when
at the end range of rotation toward a true lateral view compared to rotation toward
an AP view. Thus, radiographs with a moderate amount of pelvic rotation toward a
true lateral view may fail to provide meaningful estimates of coverage.Although ACEA measurements were most sensitive to changes in pelvic tilt, the results
of this study indicated that alterations in tilt did not significantly affect the 3D
anatomic correlate of the ACEA (Figure 5B.2). Indeed, the range of clockface locations measured across
participants was narrower for tilt (Figure 5B.2 and Figure 6, B.1 and B.2) than rotation (Figure 5A.2 and Figure 6, C.1 and C.2) and nearly
identical to the maximum, median, and minimum locations measured for obliquity. In
comparison, rotation had a sizable effect on both the ACEA and clockface location.
Thus, it is imperative that pelvic rotation is well standardized before imaging,
with less severe consequences erring on the side of rotation toward an AP view.
Limitations
The main limitation of this study was the use of a relatively small sample size
that included participants with normal acetabular coverage with variable femoral
anatomy. The cohort, which included FAI syndrome and healthy hips, was selected
as a convenience sample from previously published imaging studies.
In contrast to other studies that examined ACEA reliability,
which were conducted in vitro, the neutral position of the hip in our
study was assigned based on a participant’s true standing position obtained from
dual fluoroscopy images. Thus, despite having fewer participants, our study had
the advantage that the methods enabled pelvic rotation, tilt, and obliquity to
be defined relative to a functionally relevant participant-specific starting
position. In the future, the mapping approach described herein could be applied
to clarify the 3D clockface position for a wider spectrum of underlying hip
abnormalities, including those with abnormal coverage. Another limitation of
this study was that DRRs, not clinically acquired radiographs, were used to
collect ACEA measurements. Although images lacked soft tissue, this approach
enabled us to precisely control femur and pelvis positioning, which is a
challenge in a clinical setting and would require obtaining several images.
Conclusion
This study demonstrated that ACEA bone edge measurements exhibited excellent
reliability. Although intrarater reliability was mostly insensitive to errors in
pelvic positioning, the discrepancy between raters increased as the pelvis was
rotated toward a true lateral view. The ACEA measurement was most sensitive to
changes in pelvic tilt. However, changes in pelvic rotation also resulted in errors
that could be problematic when evaluating borderline abnormalities, with rotation
toward a lateral view having a particularly detrimental effect. Furthermore, changes
in pelvic rotation had the largest effect on the 3D anatomic correlate of the 2D
ACEA measurement, which was normally found to lie on the acetabular rim near the
anterior inferior iliac spine. In conjunction with 3D imaging, knowledge of the
limitations of the ACEA measurement and the quantitative ceiling on its performance,
in addition to the mapped 3D acetabular rim location, will help improve preoperative
and intraoperative understanding of these measurements moving forward.
Authors: Michael M Murphy; Penny R Atkins; Evangeline F Kobayashi; Andrew E Anderson; Travis G Maak; Anatoliy V Nechyporenko; Stephen K Aoki Journal: Arthroscopy Date: 2019-11 Impact factor: 4.772
Authors: Ashley L Kapron; Stephen K Aoki; Christopher L Peters; Steve A Maas; Michael J Bey; Roger Zauel; Andrew E Anderson Journal: J Appl Biomech Date: 2014-02-25 Impact factor: 1.833
Authors: Niccolo M Fiorentino; Michael J Kutschke; Penny R Atkins; K Bo Foreman; Ashley L Kapron; Andrew E Anderson Journal: Ann Biomed Eng Date: 2015-12-08 Impact factor: 3.934
Authors: Joey A Hanson; Ashley L Kapron; Kathryn M Swenson; Travis G Maak; Christopher L Peters; Stephen K Aoki Journal: J Hip Preserv Surg Date: 2015-06-13