| Literature DB >> 35141254 |
Su Jin Park1, Jae Rock Do1, Jae Pil Shin1, Dong Ho Park1,2,3.
Abstract
PURPOSE: This study evaluated the color contrast ratio (CCR) of the internal limiting membrane (ILM) using different color settings of digitally assisted vitreoretinal surgery (DAVS) with different indocyanine green (ICG) concentrations.s.Entities:
Keywords: 25 gauge pars plana vitrectomy; ILM peeling; epiretinal membrane (ERM); macular surgery; vitrectomy
Year: 2022 PMID: 35141254 PMCID: PMC8818890 DOI: 10.3389/fmed.2021.810070
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Concentrations of indocyanine green dye and parameters for digitally assisted vitreoretinal surgery groups.
|
|
| |||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| ICG dye concentration, % | 0.075 | 0.075 | 0.075 | 0.075 | 0.025 | 0.025 |
| Red, Green, Blue, % | 100,100,100 | 20,100,100 | 80,80,100 | 85,100,90 | 86,100,100 | 90,100,100 |
| Hue, ° | +2 | +2 | +2 | +2 | +2 | +20 |
| Saturation, % | 90 | 90 | 90 | 90 | 90 | 100 |
| Gamma | 1.2 | 1.2 | 1.2 | 1.2 | 1.2 | 0.9 |
| Screen distance, meter | 1.2 | 1.2 | 1.2 | 1.2 | 1.2 | 1.2 |
| Aperture, % | 50 | 50 | 50 | 50 | 50 | 50 |
| Endoillumination power, % | 35 | 35 | 35 | 35 | 35 | 35 |
ICG, indocyanine green; DAVS, digitally assisted vitreoretinal surgery; Ver., version.
Figure 1Color contrast ratio (CCR) measurement. After internal limiting membrane (ILM) peeling, retinal images were captured. CCR was measured by comparing RGB values between the indocyanine green-stained area (arrowhead) and the ILM-peeled area (asterisk) using the Eyedropper tool of Photoshop.
Clinical characteristics of standard operating microscope and digitally assisted vitreoretinal surgery groups at baseline.
|
|
|
|
|
|
|---|---|---|---|---|
| Age, years | 68.3 ± 6.7 | 69.3 ± 6.5 | 67.1 ± 6.2 | 0.749 |
| Male/female, | 10/15 | 7/5 | 3/10 | 0.199 |
| Indication of surgery (ERM/MH) | 22/3 | 11/1 | 12/1 | 0.656 |
| BCVA, logMAR | 0.76 ± 0.44 | 0.77 ± 0.61 | 0.54 ± 0.50 | 0.090 |
| IOP, mmHg | 13.8 ± 2.7 | 15.0 ± 4.5 | 13.4 ± 2.7 | 0.708 |
| Axial length, mm | 23.51 ± 0.81 | 23.85 ± 0.70 | 23.78 ± 1.19 | 0.204 |
Data are shown as the mean ± standard deviation unless indicated otherwise.
BCVA, best-corrected visual acuity; ICG, indocyanine green; logMAR, logarithm of the minimal angle of resolution; IOP, intraocular pressure; SOM, standard operating microscope; DAVS, digitally assisted vitreoretinal surgery; Ver., version; ERM, epiretinal membrane; MH, macular hole.
P-values were compared between SOM, DAVS version 1.1 and DAVS version 1.3 groups.
Kruskal-Wallis test.
Chi-square test.
Figure 2Comparison of color contrast ratio (CCR) among channels in digitally assisted vitreoretinal surgery system (DAVS) Ver. 1.1. CCR was compared in macular and vitreous images from the DAVS Ver. 1.1 group. The following four channels were compared in each captured macular and vitreous image: Default [Red (R) 100%, Green (G) 100%, and Blue (B) 100%]; Preset 1 (R 20, G 100, B 100%); Preset 2 (R 80, G 80, B 100%); and Preset 3 (R 85, G 100, B 90%). (A,B) Macular CCR was different among the four channels (P < 0.01). CCR was highest in Preset 3 and lowest in Preset 1 (P < 0.01). (C,D) In vitreous images, CCR did not differ among the four settings (P = 0.294). **P < 0.01.
Figure 3Comparison of color contrast ratio (CCR) using different color settings of digitally assisted vitreoretinal surgery system (DAVS) in different indocyanine green (ICG) concentrations. (A) Macular image captured using Preset 3 of DAVS Ver. 1.1 with 0.075% ICG. (B,C) Macular image captured using Customized Settings 1 and 2 of DAVS Ver. 1.3 with 0.025% ICG. (D) CCR of Customized Setting 2 of DAVS Ver. 1.3 with 0.025% ICG did not differ from that of Preset 3 of DAVS Ver. 1.1 with 0.075% ICG (P = 0.887). **P < 0.01.