Optimizing Visual Performance With Digitally Assisted Vitreoretinal Surgery.

Gerardo González-Saldivar, David R Chow.   

Abstract

BACKGROUND AND
OBJECTIVE: Evaluate depth of field (DOF) and lateral resolution (LR) on digitally assisted vitreoretinal surgery (DAVS) with different camera aperture values and TV viewing distances to determine maximal optical performance. PATIENTS AND METHODS: Five retinal surgeons objectively evaluated DOF and LR at low (10×), medium (15×), and high (20×) magnification on a conventional microscope (CM) and DAVS. DAVS was evaluated at different camera apertures (30%, 50%, and 75%) and TV viewing distances (1.2 meters [m], 1.5 m, and 2.0 m). A comparison between CM and optimized DAVS values was carried out.
RESULTS: On DAVS, DOF significantly increased as camera aperture size was reduced (75% to 50%, P = .022; 75% to 30%, P ≤ .001; and 50% to 30%, P = .001) and as microscope magnification was reduced (high to medium, P = .009; high to low, P ≤ .001; and medium to low, P = .002). LR significantly increased as microscope magnification was increased (low to medium, P = .010; low to high, P ≤ .001; and medium to high, P = .015) and as TV display distance decreased (2.0 m to 1.5 m, P = .019; 2.0 m to 1.2 m, P = .009; and 1.5 m to1.2 m, P = .185). When comparing optimized DAVS against CM, DOF was significantly larger on DAVS (P = .019). LR on DAVS outperformed the CM, but statistical significant was not reached (P = .185).
CONCLUSIONS: The authors' study is the first to evaluate different aperture, magnification, and TV distance settings, and when optimized, DAVS can significantly outperform a CM with respect to DOF and LR. Optimized visual performance for DAVS is obtained with a 30% camera aperture and a TV viewing distance no further than 1.5 m. Maximal microscope magnification to exploit LR during macular surgery and low-to-medium magnification to increase DOF for overall vitrectomy surgery is recommended. [Ophthalmic Surg Lasers Imaging Retina. 2020;51:S15-S21.]. Copyright 2020, SLACK Incorporated.

Entities:  

Mesh:

Year:  2020        PMID: 32348530     DOI: 10.3928/23258160-20200401-02

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging Retina        ISSN: 2325-8160            Impact factor:   1.300


  2 in total

1.  Optimal Display Positions for Heads-Up Surgery to Minimize Crosstalk.

Authors:  Kotaro Tsuboi; Yukihiko Shiraki; Yuichiro Ishida; Takashi Shibata; Motohiro Kamei
Journal:  Transl Vis Sci Technol       Date:  2020-12-17       Impact factor: 3.283

2.  Customized Color Settings of Digitally Assisted Vitreoretinal Surgery to Enable Use of Lower Dye Concentrations During Macular Surgery.

Authors:  Su Jin Park; Jae Rock Do; Jae Pil Shin; Dong Ho Park
Journal:  Front Med (Lausanne)       Date:  2022-01-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.