| Literature DB >> 35124622 |
Zhangmin Shen1, Ying-Hsiang Chou1,2,3, Cheng-Chang Lu1,2, Ren-Jing Huang1,2, Ching-Hsiu Ke4,5, Lung-Kwang Pan4, Chien-Yi Chen1,2.
Abstract
BACKGROUND: In the last 40 years, the number of deaths due to cancer has been the highest in TaiwanEntities:
Keywords: Full factorial design; environmental radiations; half value layer (HVL); thermoluminescence dosimeter (TLD); tomotherapy facility
Mesh:
Year: 2022 PMID: 35124622 PMCID: PMC9028642 DOI: 10.3233/THC-THC228044
Source DB: PubMed Journal: Technol Health Care ISSN: 0928-7329 Impact factor: 1.205
Figure 1.(a) TOMO treatment room (b) the floor plan of the TOMO facility.
Figure 2.TLDs were attached on the ceiling of the TOMO facility during the survey.
Experimental levels and factors for the optimization experiments of the settings of the TLD Harshaw 3500 reader
| Experimental factors | Level 1 | Level 2 |
|---|---|---|
| Initial temperature ( | 130 | 140 |
| Heating rate ( | 9 | 11 |
| Maximum temperature ( | 230 | 250 |
Variable in a 2 full factorial design
| Group | Factor/level | ||
|---|---|---|---|
| A initial temperature ( | B Heating rate ( | C Maximum temperature ( | |
| 1 | 130 | 11 | 230 |
| 2 | 130 | 11 | 250 |
| 3 | 130 | 9 | 230 |
| 4 | 130 | 9 | 250 |
| 5 | 140 | 11 | 230 |
| *6 | 140 | 11 | 250 |
| 7 | 140 | 9 | 230 |
| 8 | 140 | 9 | 250 |
*G6 is original setting.
The coefficient estimate among the eight groups
| G | Linear regression equation | Correlation coefficient | Stdev |
|
|
|---|---|---|---|---|---|
| 1 | Y | R | 0.468 | 0.92267 | 0.590 |
| 2 | Y | R | 2.464 | 0.72186 | |
| 3 | Y | R | 0.510 | 0.84044 | 0.434 |
| 4 | Y | R | 1.351 | 0.72564 | |
| 5 | Y | R | 1.839 | 0.67562 | |
| 6 | Y | R | 1.072 | 0.80806 | |
| 7 | Y | R | 1.592 | 0.77861 | |
| 8 | Y | R | 1.189 | 0.64446 |
Figure 3.The derived linear regression fit of the eight groups.
Figure 4.Three-dimensional distributions of the treatment room averaged environmental radiations, mapped via colored profiles.
The averaged environmental radiations (mSv/mo)
| Location | Radiation (mSv/mo) | |
|---|---|---|
| Control area | ||
| Treatment room | ||
| G (0, 0) | 1730 | |
| A ( | 171 | |
| B (3.16, | 211 | |
| Treatment room (inside slide door) | 104 | |
| Public area | ||
| Waiting room | ||
| Treatment room (outside ceiling) D | 0.16 | |
| Sofa (W5) | 0.15 | |
| Clock (W6) | 0.15 | |
| Achieves room | ||
| Above cabinet (A1) | 0.13 | |
| Above cabinet (A2) | 0.15 | |
| Control room | ||
| Computer (C3) | 0.16 | |
| Ceiling (C4) | 0.14 | |
| MRI | ||
| Control room (M7) | 0.11 | |
| Computer room (M8) | 0.16 | |
| Taichung City (TC1) | 0.15 | |
| Taichung City (TC2) | 0.14 | |
Figure 5.Half-value layer of the TOMO treatment room.