| Literature DB >> 35119608 |
Sufian Hommayda1, Timothy Hamann2, Marc Töteberg-Harms3,4,5.
Abstract
PURPOSE: Cataract surgery combined with excimer laser trabeculotomy (phaco-ELT) leads to a significant reduction in intraocular pressure by enhancing trabecular outflow. The aim of this study is to compare two laser systems for ELT (AIDA vs. ExTra).Entities:
Keywords: Cataract; Excimer laser trabeculostomy; Excimer laser trabeculotomy; Glaucoma; Laser; Minimally invasive glaucoma surgery; Phacoemulsification
Mesh:
Year: 2022 PMID: 35119608 PMCID: PMC9122876 DOI: 10.1007/s10792-021-02140-1
Source DB: PubMed Journal: Int Ophthalmol ISSN: 0165-5701 Impact factor: 2.029
Technical data of the ELT laser systems
| AIDA | ExTra | |
|---|---|---|
| Laser type | XeCL excimer laser, laser class 4 | XeCL excimer laser, laser class 4 |
| Wavelength | 308 nm | 308 nm |
| Cooling | Air | Air |
| Pulse energy | 1.2 mJ at fiber tip | 1.3 mJ at fiber tip |
| Pulse duration | 60–120 ns | 60–120 ns |
| Repetition rate | 20 Hz | 20 Hz |
| Laser spot size | 200 μm | 210 μm |
| Cannula diameter | 500 μm | 500 μm |
| Cannula material | Stainless steel | Stainless steel |
| Length of fiber | 200 cm | 200 cm |
(AIDA = AIDA laser system; ExTra = ExTra laser system; XeCl = Xenonchloride; nm = nanometer(s); mJ = millijoule(s), ns = nanosecond(s); Hz = hertz; μm = micrometer(s); cm = centimeter(s))
Demographical data of the ExTra laser system and the AIDA laser system group
| ExTra ( | AIDA ( | Total | Percentage | ||
|---|---|---|---|---|---|
| 0.195 | |||||
| Male | 27 (28.7%) | 81 (36.8%) | 108 | 34.4 | |
| Female | 67 (71.3%) | 139 (63.2%) | 206 | 65.6 | |
| 0.805 | |||||
| RE | 52 (55.3%) | 117 (53.2%) | 169 | 53.8 | |
| LE | 42 (44.7%) | 103 (46.8%) | 145 | 46.2 | |
| 0.095 | |||||
| ELT | 1 (1.1%) | 0 | 1 | 0.3 | |
| Phako-ELT | 93 (98.9%) | 220 (100%) | 313 | 99.7 | |
| 0.425 | |||||
| Primary open angle glaucoma | 43 (45.7%) | 127 (57.7%) | 170 | 54.1 | |
| Pseudoexfoliative glaucoma | 42 (44.7%) | 74 (33.6%) | 116 | 36.9 | |
| Pigmentdispersion glaucoma | 3 (3.2%) | 4 (1.8%) | 7 | 2.2 | |
| Angle closure glaucoma | 4 (4.3%) | 7 (3.2%) | 11 | 3.5 | |
| Low tension glaucoma | 1 (1.1%) | 4 (1.8%) | 5 | 1.6 | |
| Other glaucomas | 1 (1.1%) | 2 (0.9%) | 3 | 1 | |
| 14 (14.9%) | 21 (9.5%) | 35 | 11.2 | 0.177 | |
| 11 (11.7%) | 27 (12.3%) | 38 | 12.2 | 1.0 |
(*cataract excluded; AIDA laser system; ExTra = ExTra laser system; LE = left eye; RE = right eye; ELT = excimer laser trabeculotomy; phaco-ELT = phacoemulsification cataract extraction with intraocular lens implantation combined with ELT)
Results
| Baseline Mean ± SD | 1 day | 1 week | 1 month | 3 month | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| IOP [mmHg] | ||||||||||||
| ExTra | 20.3 ± 5.9 | 17.8 ± 8.4 | 0.02 | 15.4 ± 6.0 | < 0.001 | 14.9 ± 4.9 | < 0.001 | 13.7 ± 4.4 | ||||
| AIDA | 18.7 ± 6.3 | 18.5 ± 8.5 | 0.44 | 15.9 ± 7.2 | 0.02 | 16.3 ± 5.3 | < 0.001 | 15.8 ± 4.7 | ||||
| 0.43 | 0.53 | 0.7 | 0.15 | |||||||||
| AGD | ||||||||||||
| ExTra | 2.0 ± 1.3 | 1.5 ± 1.4 | < 0.01 | 1.6 ± 1.6 | 0.09 | 1.6 ± 1.3 | 0.19 | 1.5 ± 1.5 | ||||
| AIDA | 2.0 ± 1.3 | 1.3 ± 1.4 | < 0.001 | 1.3 ± 1.5 | 0.002 | 1.6 ± 1.5 | 0.07 | 2.1 ± 1.5 | ||||
| 0.98 | 0.28 | 0.43 | 0.92 | |||||||||
| BCVA [logMAR] | ||||||||||||
| ExTra | 0.5 ± 0.4 | 0.6 ± 0.3 | 0.01 | 0.4 ± 0.3 | 0.05 | 0.3 ± 0.3 | 0.003 | 0.2 ± 0.2 | ||||
| AIDA | 0.3 ± 0.3 | 0.4 ± 0.3 | < 0.001 | 0.4 ± 0.3 | 0.74 | 0.3 ± 0.3 | 0.03 | 0.3 ± 0.3 | ||||
| 0.004 | 0 | 0.65 | 0.83 | |||||||||
(AIDA laser system; ExTra = ExTra laser system; IOP = intraocular pressure; AGD = anti-glaucoma drugs; BCVA = best-corrected visual acuity; mmHg = millimeters of mercury)
IOP spikes
| ALL ( | EXTRA ( | AIDA ( | ||
|---|---|---|---|---|
| 1 day | 41 (13.4%) | 10 (10.6%) | 31 (14.1%) | 0.407 |
| 1 week | 3 (1.0%) | 1 (1.1%) | 2 (0.9%) | 0.898 |
| 1 month | 4 (1.6%) | 1 (1.1%) | 3 (1.4%) | 0.829 |
| 1 d to 1 month | 44 (14.0%) | 11 (11.7%) | 33 (15.0%) | 0.442 |
(AIDA = AIDA laser system; ExTra = ExTra laser system; IOP = intraocular pressure; IOP spike = increase of IOP ≥ 10 mmHg from baseline)