Yumi Kataoka1, Hitoshi Nishio2, Ryo Matsukiyo2, Ryoichi Kato2, Midori Hasegawa3, Takashi Kenmochi4, Ryoichi Shiroki5, Hiroshi Toyama6, Takashi Ichihara7, Shigeki Kobayashi8. 1. Faculty of Medical Quantum Science, Fujita Health University Graduate School of Health Sciences, Toyoake, Aichi, Japan. 2. Department of Radiology, Fujita Health University Hospital, Toyoake, Aichi, Japan. 3. Department of Nephrology, Fujita Health University, School of Medicine, Toyoake, Aichi, Japan. 4. Department of Organ Transplant Surgery, Fujita Health University, School of Medicine, Toyoake, Aichi, Japan. 5. Department of Urology, Fujita Health University, School of Medicine, Toyoake, Aichi, Japan. 6. Department of Radiology, Fujita Health University, School of Medicine, Toyoake, Aichi, Japan. 7. Department of Artifical Intelligence in Medical Imaging Development, Fujita Health University, School of Medicine, Toyoake, Aichi, Japan. 8. Faculty of Radiological Technology, Fujita Health University, School of Medical Sciences, Toyoake, Aichi, Japan.
Abstract
OBJECTIVES: Renal scintigraphy is widely used to evaluate residual function of a transplanted kidney from the donor. Dynamic computed tomography (CT) imaging can evaluate both kidney morphology and regional renal function. The aim of this study was to develop an imaging protocol and a calculation method using dynamic CT for assessing the effective renal plasma flow (ERPF) by model analysis, and to evaluate the validity of the obtained ERPF values. METHODS: Preoperative dynamic CT examination with a low radiation dose exposure system was performed for 25 renal transplant donors, and ERPF was calculated from the obtained images (CT-ERPF). To calculate CT-ERPF, we set the region of interest (ROI) in the renal cortex using automatic ROI-setting software developed in our laboratory. We compared the processing time with automatic and manual ROI settings. To evaluate the validity of CT-ERPF, we examined the correlation of age with CT-ERPF and compared with reported ERPF values. We also compared the uptake rates of technetium-99m-dimercaptosuccinic acid and CT-ERPF in terms of the right-to-left ratio. RESULTS: There was good agreement of CT-ERPF assessed using automatic and manual ROIs. CT-ERPF was negatively correlated with age and showed values below the reference ERPF range in 21 cases. The right-to-left ratio of CT-ERPF showed a significant correlation with that of technetium-99m-dimercaptosuccinic acid. CONCLUSIONS: Using our method, CT-ERPF was a useful indicator for preoperative evaluation of donor's renal function.
OBJECTIVES: Renal scintigraphy is widely used to evaluate residual function of a transplanted kidney from the donor. Dynamic computed tomography (CT) imaging can evaluate both kidney morphology and regional renal function. The aim of this study was to develop an imaging protocol and a calculation method using dynamic CT for assessing the effective renal plasma flow (ERPF) by model analysis, and to evaluate the validity of the obtained ERPF values. METHODS: Preoperative dynamic CT examination with a low radiation dose exposure system was performed for 25 renal transplant donors, and ERPF was calculated from the obtained images (CT-ERPF). To calculate CT-ERPF, we set the region of interest (ROI) in the renal cortex using automatic ROI-setting software developed in our laboratory. We compared the processing time with automatic and manual ROI settings. To evaluate the validity of CT-ERPF, we examined the correlation of age with CT-ERPF and compared with reported ERPF values. We also compared the uptake rates of technetium-99m-dimercaptosuccinic acid and CT-ERPF in terms of the right-to-left ratio. RESULTS: There was good agreement of CT-ERPF assessed using automatic and manual ROIs. CT-ERPF was negatively correlated with age and showed values below the reference ERPF range in 21 cases. The right-to-left ratio of CT-ERPF showed a significant correlation with that of technetium-99m-dimercaptosuccinic acid. CONCLUSIONS: Using our method, CT-ERPF was a useful indicator for preoperative evaluation of donor's renal function.
Measurement of residual renal function of the donor kidney is necessary for
transplantation surgery. Renal function can be evaluated preoperatively by several methods,
including creatinine clearance, estimated glomerular filtration rate (eGFR), and renal
scintigraphy. Morphological evaluation is also performed using dynamic contrast-enhanced
computed tomography (dynamic-CT) with iodine contrast medium. In our hospital, three-phase
dynamic CT is performed for patients with renal disease, and includes pre-contrast of the first
phase, the nephrographic phase, and the excretory phase. Frennby et al.[1] reported that iodinated contrast medium had excellent
properties as a marker of GFR.A focus of our laboratory is the development of techniques for calculating renal
plasma flow from dynamic CT with model analysis. In previous studies, renal blood flow (RBF) was
accurately calculated using the deconvolution method with dynamic CT imaging.[2] This method requires dynamic CT imaging using 20 scans
of the first-pass image with one bolus injection of contrast medium. As the area under the curve
of the input/output function must be accurately calculated, it is necessary to satisfy the
sampling conditions for correct capture of the shape of the time attenuation curve (TDC) at the
interval and the number of dynamic scans. Therefore, if this method is applied as used in
320-row multidetector row CT, the exposure radiation dose may be markedly increased. Appropriate
minimization of radiation exposure by decreasing the number of scans is important. The purpose
of this study was to develop an imaging protocol and calculation method using dynamic CT for
calculation of effective renal plasma flow (ERPF) by model analysis, and to evaluate the
validity of the obtained ERPF values.
Methods
This study was approved by the Health Science Ethics Committee at Fujita Health
University and was conducted in accordance with the Declaration of Helsinki and Good Clinical
Practice. Written informed consent was obtained from all study subjects. Dynamic CT examination
(Aquilion ONE; Canon Medical Systems, Tokyo, Japan) with addition of low radiation dose was
performed for 25 renal transplant donors (nine men, sixteen women, 37–73 years old,
average age of 58.1 years), and ERPF measurements were performed from the obtained images
(CT-ERPF). Contrast medium was injected using a dual-shot GX 7 (Nemotokyorindo, Tokyo, Japan).
Data analysis was performed using MATLAB R2016a (MathWorks, Natick, MA, USA) for numerical
analysis programming language and OsiriX version 7.0.4 (OsiriX Foundation, Geneva, Switzerland)
for medical image management software.
Kinetic model of iodine contrast medium in kidney
In this study, the following pharmacokinetic model of iodine contrast medium in the
body was used for analysis. The contrast medium administered intravenously flows from the renal
artery to the glomerulus. Contrast medium filtered by the glomerulus flows to the renal tubule,
while contrast medium that has not been filtered flows to the renal vein. The kidney comprises
the renal cortex and renal medulla, with renal tissue, glomeruli, and tubules in the kidney
cortex, and only renal tissue in the renal medulla. The concentration of contrast medium in the
renal cortex at time t (Cc(t)) can be described by Equation 1:
where Ck(t) (mg I/g) is the concentration of contrast medium in
kidney tissue, and Cg(t) (mg I/g) is the concentration of contrast medium in
glomeruli. As time is required for the contrast medium to pass through the glomerulus in the
initial state after contrast administration, the contrast medium appears to accumulate in the
glomeruli. Thus, a microsphere model (Figure 1) can be
assumed.
Figure 1
Microsphere model. Cc(t) is the concentration of contrast medium in the renal
cortex at time t. F (ml/g/min) is the renal blood flow rate. As time is required for the
contrast medium to pass through the glomerulus in the initial state after administration, the
contrast medium seems to accumulate in the glomerulus. Thus, a microsphere model can be
assumed.
In Figure 1, F (ml/g/min) reflects the renal
blood flow rate. At this time, contrast medium concentration in the kidney cortex at time t
(Cc(t)) is expressed by the amount of contrast medium that continues to accumulate
in the renal cortex from the arterial blood, as shown in Equation 2:
where Ca(t) (mg I/ml) is the concentration of contrast medium in the
arterial blood at time t. From Equation 2, (X(t) and Y(t)) are plotted to obtain an approximate
straight line as , . F can be calculated as the slope obtained from the approximate straight line
(Patlak plot method).[3,4]
Hematocrit correction
To calculate ERPF, hematocrit correction must be performed for renal blood flow F
calculated from Equation 2. Solving Equation 2 for F yields Equation 3:
where Ca(t) (mg I/ml) is the concentration of contrast medium in the
arterial blood of the renal artery. Because the concentration of contrast medium and the CT
value have a proportional relationship, Ca(t) is represented by the CT attenuation
value (HU). Pixels in the arterial blood obtained from the CT image have units of HU/voxel,
with one voxel represented by one pixel. The contrast medium exists in the plasma, and to
obtain Ca(t) from this CT pixel, plasma volume V (ml) of the voxel constituting this
pixel must be converted to the contrast medium concentration. This plasma volume is calculated
using the hematocrit. The renal artery is a large blood vessel corrected for the hematocrit
value of HCTLV in a large blood vessel. The amount of plasma contained in the voxel
of one pixel of the renal artery is (ml), and the concentration of contrast medium contained in the plasma within this
voxel is (HU/voxel/ml).Cc(t) is the concentration of contrast medium in the kidney cortex, and
represents the amount of contrast medium per unit weight. However, the units of
Cc(t) obtained from the CT image are CT value per voxel (HU/voxel). First, the units
of Cc(t) must be converted from CT value per voxel to CT value per unit weight. When
the density of renal tissue is assumed to be ρ (g/ml), the weight can be obtained by
multiplying the voxel by the density of the kidney tissue. Therefore, the CT attenuation value
per 1 g of kidney tissue is expressed as (HU·ml/voxel/g).Next, because glomerular capillaries are contained in the kidney cortex, hematocrit
correction is required as for Ca(t). Because the glomerular capillary is a small
blood vessel, the value is corrected for the hematocrit in a small blood vessel,
HCTSV. The amount of plasma contained in the voxel represented by one pixel of
kidney cortex is therefore (ml). The concentration of contrast medium contained in the plasma within this
voxel is expressed as (HU/voxel/g).ERPF corrects for the hematocrit in large and small blood vessels, and the
correction for kidney tissue density is shown in Equation 4:Substituting Equation 3 into Equation 4 yields Equation 5, which converts renal
blood flow F to ERPF:For this study, we performed hematocrit corrections using HCTLV=0.45 and
HCTSV=0.25.[5]
CT imaging protocol
In the preoperative examination of donors, CT imaging from the liver to the pelvis
was performed, with three phases of pre-contrast, nephrogram, and excretion. For analysis of
ERPF using the kinetic model (see previous section), dynamic CT imaging was added before the
renal artery phase. The study imaging protocol used dynamic CT with a tube voltage of
100 kV, auto exposure control (tube current, 84–135 mA), an X-ray rotation
speed of 0.5 s/rotation, and a 2.0-s interval for 20 s. Imaging was started 8 s
after starting injection. Contrast medium was injected at 600 mg I per kg body weight
within 25 s. The infusion rate was approximately 3.3–5.0 ml/s, with a boost
injection of 30 ml saline. We used FC04 AIDR 3D Strong (Aquilion ONE; Canon Medical
Systems) for the reconstruction function. The average dose at this time was 29.7 mSv, and
the average dose of dynamic imaging added on this protocol was 4.5 mSv. The contrast
medium used was iopamidol 370 (Bayer Pharmaceutical, Osaka, Japan), omnipaque 300 (Daiichi
Sankyo Co., Tokyo, Japan), iomepurol 350 (Eisai Co., Tokyo, Japan), or Optirei 240 (Fuji
Pharmaceutical, Tokyo, Japan).
Creation of input function (Ca(t))
An ROI was set for each of the right and left renal arteries of the obtained
dynamic contrast CT data to generate a TDC (attenuation). The range from the TDC until the CT
value started to rise was defined as baseline. An average within the baseline range was
calculated as the baseline value, which was subtracted from the TDC. This corrected TDC was
processed with five-point smoothing to average the CT values of five points before and after
the plot. This value was defined as the input function Ca(t) (Figure 2). The smoothing processing was important because the original
baseline corrected TDC curve can become unstable because of noise. Determining the appropriate
baseline value was important for calculating CT-ERPF.
Figure 2
Creation of input function for computed tomography-effective renal plasm flow (CT-ERPF)
calculation. A time density curve (TDC) was created from the CT value of the renal artery
region of interest (ROI) setting. We visually set the ROI of the renal artery before inflow of
contrast medium and used this as the baseline range. CT values of TDC within the pre-contrast
range were averaged to create the baseline value. The baseline value was subtracted from the
created TDC, and five-point smoothing was performed to obtain the input function of the
concentration of contrast medium in the arterial blood at time t
(C
Creation of output function (Cc(t))
In the phase in which the kidney cortex was being enhanced, three ROIs of the
kidney cortex were set in the cross-section of the central part of the kidney, and three TDCs
were generated from the obtained ROI. The range from the TDC until the CT value started to rise
was defined as the baseline range. CT values of the TDC for each pixel within the baseline
range of the kidney cortex were averaged to determine the baseline value (Figure 3). Determining this baseline value was also important for calculating
CT-ERPF.
Figure 3
Baseline range setting for the output function used for CT-ERPF calculation. Three pointed
ROIs were set in the renal cortex and the TDC was created. The range before the increase in
TDC was used as the baseline range, and the average of the three TDCs used as the output
function.
Automatic setting of the renal cortex
The ROI for the renal cortex was automatically set using the following procedure.
For removal of bone, the maximum CT value of the first dynamic image after contrast medium
administration was calculated, and 10% of the maximum CT value was taken as the threshold
value. Pixels with a CT value higher than the threshold value were considered bone regions and
removed. For removal of renal artery/renal vein/kidney tissue, a threshold value was obtained
using the maximum CT value of the kidney cortex TDC and the CT value in the equilibrium phase.
For the CT image in the equilibrium phase, pixels below the threshold value were removed as the
renal artery, renal vein, and extrarenal tissue, and both kidneys were extracted. The extracted
right and left kidneys were then distinguished based on the left and right renal centroid
coordinates. For removal of the renal medulla, in the renal cortex, the CT value in the
arterial phase is higher than that in the equilibrium phase, while the converse is true for the
renal medulla. Using this relationship, pixels with higher CT values in the equilibrium phase
than those in the arterial phase were removed as the renal medulla, allowing the boundary
between the kidney cortex and renal medulla to be determined, and the cortical area extracted.
The glomerulus may also exist at the boundary between the cortex and medulla. Thus, a process
to incorporate the boundary between the cortex and medulla where the glomerulus exists within
the cortex was performed. For ROI creation in the renal cortex, the boundary between the
extracted renal cortex and renal vein was visually confirmed, and any mislabeled pixels were
corrected manually, to create an ROI of the entire renal cortex. An ROI was also created for
the entirety of each left and right kidney, and a TDC generated for each pixel. The baseline
value was subtracted from this TDC, and as for creation of the input function, five-point
smoothing was performed to obtain the output function Cc(t) (Figure 3).
Correction of the time difference between input and output functions
The methodology assumes no time lag between input function Ca(t)
calculated from the renal artery and output function Cm(t) calculated from the renal
medulla. However, contrast medium in the renal artery flows into the glomerulus and reaches the
renal medulla in approximately 1–2 s. Therefore, to accurately perform model
analysis, the time difference between the input function Ca(t) and renal medullary
output function Cm(t) was calculated and corrected.[6]For creation of an ERPF image and value for one or both kidneys, renal blood flow F
was calculated with the Patlak method (Figures
1–3) using Equation 4. The linear
approximation range of the Patlak analysis was performed at three points after the rising edge
of the output function Cc(t). Hematocrit correction was performed on the calculated
values for each pixel to prepare an ERPF image. ERPF values of the left and right kidneys, and
of both kidneys, were calculated by integrating the ERPF value for each pixel within the ROI
set in the renal cortex. For evaluating the accuracy of the automatic ROI of the renal cortex,
we compared the processing time for manual and automatic ROIs and the match rate. An ROI of the
renal cortex prepared manually and confirmed by radiologists (manual ROI) was evaluated with
the following three items:(1) Match rate between the manual and automatic ROIs. The coincidence rate between
the manual and automatic ROIs was calculated using Equation 6:(2) Evaluation of CT-ERPF calculated using manual and automatic ROIs. The
correlation of CT-ERPFs calculated using manual ROI with automatic ROI was determined by
plotting (X, Y) the coordinates (X-axis: CT-ERPF calculated using a manual ROI; Y-axis: CT-ERPF
calculated using an automatic ROI) to determine a correlation formula, a correlation
coefficient, and a p value.(3) Comparison of processing time between the manual and automatic ROIs. The manual
ROIs were performed in nine of 25 cases under the guidance of a radiologist employed at Fujita
Health University Hospital. The required time to set the ROI in the renal cortex was compared
between manual and automatic settings in nine cases.
Evaluation of ERPF calculated by CT
The calculated ERPF was compared with reported reference values.[7-11] The
conventional paraaminohippuric acid clearance (CPAH) was defined as the reference
ERPF. We also evaluated the age-dependence of ERPF values in our cases. To confirm the validity
of CT-ERPF values, the right-to-left ratio of technetium-99m-dimercaptosuccinic acid
(99mTc-DMSA) kidney uptake and CT-ERPF values before surgery were evaluated.
99mTc-DMSA scintigraphy was performed in 19 cases, and was performed first on the
same day of dynamic CT.
Statistical analysis
All statistical analyses were performed using statistical software (Matlab R2016a;
MathWorks). Statistical comparisons were performed between the CT-ERPF value calculated by
automated ROI and the CT-ERPF value calculated by manual ROI, and between the right-to-left
ratio of CT-ERPF values and the right-to-left ratio of renal uptake of 99mTc-DMSA.
Correlation coefficients of those ratios were obtained, and data analyzed by Pearson’s
correlation coefficient. A p-value <0.05 was considered significant. Bland–Altman
analysis was performed to compare differences in the ratios. Bias was determined as the average
value of differences in CT-ERPF values calculated from both methods, and limits of agreement
were set from –1.96 to +1.96 standard deviations of the CT-ERPF value.
Results
Evaluation of automatic ROI in the renal cortex
The average agreement rate with the manual ROI before manual correction to the
automatic ROI was 81.2%, and the average agreement rate after manual correction was 82.9%. In
both kidneys, the linear equation was y=1.010×–12.83 (p<0.05), with a linear
relationship close to 0.999 and 1, indicating a significant positive correlation. The
correlation of CT-ERPF calculated using the manual and automatic ROIs, and the
Bland–Altman analysis, are shown in Figure 4. The
bias between ERPF calculated using automatic and manually set ROIs was 8.85 ml/min. The
limits of agreement also changed from 30.42 ml/g/min to –12.7 ml/g/min.
Figure 4
Correlation of CT-ERPFs calculated using manual and automatic ROIs. Left: CT-ERPF values
calculated from the manual ROI and the automatic ROI plotted as a correlation. The linear
slope is close to 1, with a p value <0.05, indicating a significant positive correlation.
Right: Bland–Altman plots showing a bias of 8.85 ml/min and limits of agreement of
–12.7–30.42 ml/g/min.
Comparison of processing times for manual and automatic ROI
Processing times for setting ROIs in the renal cortex using manual and automatic
methods are shown in Table 1. When all ROIs of the renal
cortex were manually set, the average was 16.7 h per case, while the processing time using
automatic setting was only 1.36 h on average. Automated ROI processing consisted of both
automatic processing and manual correction. The average processing time for automatic
processing was 0.0245 h, while the average manual correction processing time was
1.34 h.
Table1
Processing time for setting ROIs in the renal cortex
Case number
Manual ROI (hours)
Automatic processing+manual correction (hours)
Automatic processing
Manual correction
Total time
11
16.0
0.0240
1.19
1.21
12
24.0
0.0238
1.65
1.68
17
24.0
0.0236
1.33
1.35
32
16.0
0.0250
0.75
0.77
33
16.0
0.0246
0.96
0.98
40
13.0
0.0230
1.24
1.26
42
12.0
0.0260
1.22
1.24
46
20.0
0.0250
1.76
1.79
47
14.0
0.0253
2.57
2.59
Average
16.7
0.0245
1.34
1.36
The required time to set ROIs in the renal cortex part manually and automatically
on nine cases. Manual ROIs were confirmed by a radiologist who was engaged in Fujita Health
University.
CT-ERPF images and CT-ERPF values
A representative CT-ERPF image is shown in Figure
5. The left panel shows a contrast-enhanced CT image, the right panel the CT-ERPF
image, the upper panel shows the right kidney, and the lower panel the left kidney. The
calculated CT-ERPFs for both kidneys are shown in Figure
6. Previous non-Japanese studies have reported normal adult conventional ERPFs
(CPAH) of 442–694 ml/min/1.73 m2,[7,8] while
Japanese studies have reported values of
439–817 ml/min/1.73 m2.[9-11] In the present study, all cases
were Japanese adult renal transplant donors (men: 37–73 years old; woman: 43–71
years old). Thus, we adopted a reference range of ERPF (CPAH) of
439–817 ml/min/1.73 m2.
Figure 5
CT image and CT-ERPF image. Left: Contrast-enhanced CT images. Right: CT-ERPF images. The
upper image shows the right kidney. The lower image shows the left kidney.
Figure 6
Correlation of age with CT-ERPF in the 25 renal transplant donors. The correlation
coefficient R2=0.40 (p<0.05) indicates a weak negative correlation. The
transverse line shows a lower value of the reference ERPF.
Twenty-one cases showed values below the reference range. The results of
comparisons of the right-to-left ratio of 99mTc-DMSA kidney uptake with those of
CT-ERPF for 19 of the 25 cases are shown in Figure 7. An
excellent correlation was observed between 99mTc-DMSA uptake and CT-ERPF.
Bland–Altman plots had a bias of 0.0087, and the limits of agreement were
–0.30–0.32. One case was outside of this range. In this case, the sizes of the left and
right kidneys were significantly different compared with other patients.
Figure 7
Relationship between CT-ERPF and technetium-99m-dimercaptosuccinic acid
(99mTc-DMSA) uptake. Left: Correlation of the right-to-left ratio of CT-ERPF with
the right-to-left ratio of 99mTc-DMSA uptake. The correlation coefficient
R2=0.92 (p<0.05) indicates an excellent positive correlation. Right:
Bland–Altman plots showing a bias of 0.0087 and limits of agreement of
–0.30–0.32.
Discussion
Automatic ROI setting of renal cortex contours
In our method, the total processing time for automatic ROIs was markedly reduced
compared with that for manual ROIs. Nevertheless, future studies are required to further reduce
the total processing time for routine clinical use.
Time difference correction between input and output functions
RBF can be accurately calculated by the deconvolution method using TDC.[2] In this method, dynamic CT imaging requires 20 scans
of the first-pass image with one bolus injection of contrast medium. In our proposed model
analysis method, dynamic imaging is performed in the initial phase of contrast medium flowing
into the cortex, and the input function from the renal artery image and the cortical area are
used to calculate the output function for each pixel TDC. A time difference appears in the
image because of differences in the flow path of the contrast medium. As the model uses
calculations assuming no time difference between the paths, difference correction is
necessary.[8] According to Krier
et al.,[12] output function of the
cortex appears as the initial vascular phase, then in the proximal tubule, followed by the
distal tubule. To accurately capture the shape of the vascular phase, the latter two phases
must be separated in the deconvolution method. Thus, at least two phase TDCs of the vascular
and proximal tubules are required. Compared with the deconvolution method, the Patlak method
calculates RBF using several points of the initial vascular phase of TDC. Thus, the number of
dynamic scans (sampling number) can be reduced.Kudo et al. performed hematocrit correction with HCTLV=0.45, HCTSV=0.25, and a
brain tissue density of 1.04 g/ml to measure cerebral blood flow.[5] Thus, in the present study the concentration of
contrast medium in the renal cortex was converted from the CT value per voxel to the amount of
contrast medium per unit weight,[6] and the
density of renal tissue was the same as that of Kudo et al.[5] To obtain the ERPF, we used hematocrit correction of large and small
blood vessels, and used Equation 7 to consider the density of kidney tissue.
CT-ERPF compared with reference ERPF
CT-ERPF values using the Patlak method in our study tended to be lower (21 of 25
cases) than reported reference ERPFs. Russell et al. used these measurements
interchangeably with the following conversions: ERPF=CPAH; clearance of
iodine-131-ortho-iodohippurate (COIH)=0.90 CPAH; clearance of
technetium-99m-mercaptoacetyltriglycine (99mTc-MAG3) (CMAG3)=0.59; and
COIH=0.53 CPAH.[7,8] We considered that CT-ERPF was not equal to
CPAH, and that CT-ERPF could be converted by the additional factor. CT-ERPF values
decreased with age. Further, Russell et al. reported a decrease in ERPF with aging, which
was calculated using the following equation for patients >40 years: ERPF=568–5.83 (age
–40)±126.[7,8] Dujardin et al.[13] also
reported an inverse correlation of age with RBF calculated from contrast MRI data using the
deconvolution method. This was because of a moderate age dependence of these values in males.
Further, arterial blood flow from Doppler ultrasound measurements tended to decrease with
age.[14,15] We also found a trend towards a decrease in CT-ERPF with age, which may
reflect a progressive decrease in renal function.
CT-ERPF versus 99mTc-DMSA uptake
In the present study, there was an excellent correlation of the right-left ratio of
CT-ERPF with that of 99mTc-DMSA uptake. Taylor reported an excellent agreement of
99mTc-DMSA uptake and COIH with serum creatinine
≤2.0 mg/dl,[16] while Momin
et al. reported
that 99mTc-DMSA and technetium-99m-diethylenetriaminepentaacetic acid
(99mTc-DTPA) scanning methods provided similar relative renal functions
values.[17] Thus, CT-ERPF values may reflect
a relative renal function similar to the GFR in the donor’s kidney.
Study limitations
A limitation of our study is that measurement of RBF (e.g., CPAH) was
difficult in routine examination in our hospital. For this reason, we performed indirect
comparisons and evaluations of the distribution of normal values and nuclear medicine
examination findings. Historically, examination by 99mTc-MAG3 is common for renal
blood flow examination. However, in our university hospital, the 99mTc-DMSA kidney
uptake test was routinely performed for split renal function. Another limitation is that GFR
(e.g., eGFR, 24 h creatinine clearance, 99mTc-DTPA scan) should be used to
evaluate renal function. Our present data are preliminary findings, and simultaneous
measurements of CT-ERPF and CT-GFR are in progress in our laboratory using the same method.
These measurements may be useful for assessing kidney diseases. Finally, the dose of radiation
exposure on CT was higher compared with radio-isotope examinations, for example,
99mTc-DTPA and 99mTc-MAG3.[18] However, we suggest that CT provides detailed anatomical information before
transplantation surgery compared with other examinations. Thus, in limited cases, such as in
renal transplantation donors, CT is a useful technique and should be used.
Conclusions
Dynamic CT enables calculation of CT-ERPF using our proposed method and is a
feasible technique in renal transplantation donors.
Authors: L O Lerman; M R Bell; V Lahera; J A Rumberger; P F Sheedy; A Sanchez Fueyo; J C Romero Journal: Am J Hypertens Date: 1994-09 Impact factor: 2.689