| Literature DB >> 35097002 |
Vincenzo Russo1, Antonello D'Andrea2, Stefano De Vivo3, Anna Rago3, Gianluca Manzo2, Antonio Bocchetti1, Andrea Antonio Papa3, Valerio Giordano2, Ernesto Ammendola3, Berardo Sarubbi3, Paolo Golino1, Antonio D'Onofrio3, Gerardo Nigro1.
Abstract
INTRODUCTION: Little is known about the clinical performance of single-chamber leadless pacemaker (LLPM) in patients without atrial fibrillation (AF) as pacing indication. The aim of this study was to describe the clinical characteristics of patients who underwent single chamber LLPM implantation at three tertiary referral centers and to compare the safety and effectiveness of the single-chamber LLPM among patients with or without AF.Entities:
Keywords: atrial fibrillation; atrioventricular block; complications; effectiveness; leadless pacemaker; safety; sinus node dysfunction; syncope
Year: 2022 PMID: 35097002 PMCID: PMC8795374 DOI: 10.3389/fcvm.2021.781335
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
Baseline characteristics of study population.
|
|
|
|
| |
|---|---|---|---|---|
| Age, years | 76.7 ± 11.24 | 78.1 ± 10.8 | 75.5 ± 11.2 |
|
| Male, | 90 (64.3) | 41 (61.2) | 49 (67.1) |
|
| Hypertension, | 98 (70) | 54 (80.5) | 44 (60.3) |
|
| Diabetes, | 43 (30.7) | 19 (28.3) | 24 (32.9) |
|
| COPD, | 22 (15.7) | 7 (10.4) | 15 (20.5) |
|
| Dyslipidemia, | 70 (50) | 34 (50.7) | 36 (49.3) |
|
| CKD, | 28 (20) | 9 (13.4) | 16 (21.9) |
|
| Dialysis, | 11 (7.8) | 2 (3.0) | 9 (12.3) |
|
| Anemia, n, (%) | 16 (11.4) | 13 (19.4) | 3 (4.1) |
|
| Malignancy, | 18 (12.8) | 10 (14.9) | 8 (10.9) |
|
| DCM, | 30 (21.4) | 14 (20.9) | 16 (21.9) |
|
| CAD, | 41 (29.3) | 16 (23.9) | 25 (34.2) |
|
| Pre-syncope, | 96 (68.6) | 51 (76.1) | 45 (61.6) |
|
| Syncope, | 61 (43.6) | 19 (28.3) | 42 (57.5) |
|
| Infectious Leads extraction, | 16 (11.4) | 3 (4.5) | 13 (17.8) |
|
COPD, chronic obstructive pulmonary disease; CKD, chronic kidney disease; DCM, dilated cardiomyopathy; CAD, coronary artery disease.
Intraoperative data and electrical parameters.
|
|
|
|
| |
|---|---|---|---|---|
|
|
|
| ||
| Implant duration, minutes | 45.21 ± 18.59 | 43.10 ± 13.23 | 49.24 ± 21.56 |
|
| Fluoroscopy time, minutes | 9.05 ± 6.23 | 9.09 ± 5.16 | 9.24 ± 7.11 |
|
| R wave amplitude, mV | 12.08 ± 4.93 | 11.32 ± 4.75 | 12.19 ± 4.84 |
|
| Ventricular threshold, V | 1.25 ± 0.75 | 1.45 ± 0.63 | 1.12 ± 1.24 |
|
| Ventricular impedance, Ohm | 792.4 ± 214.4 | 788.22 ± 228.78 | 784.58 ± 201.65 |
|
Perioperative complications.
|
|
|
|
| |
|---|---|---|---|---|
| Pericardial effusion, | 2 (1.4) | 0 (0) | 2 (2.6) |
|
| Inguinal hematoma, | 1 (0.7) | 1 (1.5) | 0 (0) |
|
| Femoral Pseudoaneurysm, | 1 (0.7) | 1 (1.5) | 0 (0) |
|
| Device dislocation, | 1 (0.7) | 0 (0) | 1 (1.3) |
|
| High ventricular threshold, | 2 (1.4) | 1 (1.5) | 1 (1.3) |
|
Figure 1Study of the flowchart.
Electrical parameters and clinical events at follow-up.
|
|
|
|
| |
|---|---|---|---|---|
|
| ||||
| R wave amplitude, mV | 13.75 ± 5.04 | 11.8 ± 5.2 | 10.9 ± 4.8 |
|
| Ventricular threshold, V | 1.2 ± 0.4 | 0.53 ± 0.45 | 0.55 ± 0.37 |
|
| Ventricular impedance, Ohm | 716.9 ± 187.4 | 707.9 ± 168 | 711 ± 187 |
|
| Ventricular pacing (%) | 40 ± 29 | 31 ± 16 | 52 ± 36 |
|
|
| ||||
| Syncope, | 5 (3.8) | 2 (3.3) | 3 (4.3) |
|
| Cardiac hospitalization, | 7 (5.4) | 3 (4.9) | 4 (5.8) |
|
| All-cause death, | 10 (7.7) | 5 (8.2) | 5 (7.2) |
|
Figure 2Cumulative risk of combined endpoint (syncope, cardiac hospitalization, and mortality) between the atrial fibrillation (AF) and non-AF groups.