Literature DB >> 35096688

Clinical and Radiological Safety of Retained Implantable Doppler Devices Used for Free Flap Monitoring.

Lucas Kreutz-Rodrigues1, Waleed Gibreel1, Brian T Carlsen1, Matthew A Frick1, Samir Mardini1, Karim Bakri1.   

Abstract

INTRODUCTION: Implantable Doppler devices are reliable adjuncts used for free flap monitoring. Occasionally, the probe/wire is not removed and remains in the soft tissues. The clinical safety of the retained probes and safety and compatibility with magnetic resonance imaging (MRI) have not been studied. We present a series of retained implantable Doppler probes examining clinic outcomes, safety and compatibility with MRI, and effect on MRI image quality.
METHODS: A retrospective review was conducted of patients who had an implantable Doppler device for free flap monitoring between July 2007 and August 2018. Routine post-operative imaging was reviewed for all patients to identify incidental findings of a retained probe. A subset of patients with retained implantable Doppler probes who underwent MRI was identified. Magnetic resonance images were reviewed to detect any degradation of image quality.
RESULTS: A total of 323 patients who had an implantable Doppler device placed were reviewed 18 (5.6%) patients were identified with a retained probe and were included in this study. Mean age was 49 years with mean follow-up of 34.4 months. One potential device-related complication occurred in 1 (5.6%) patient. A total of 32 MRI scans were performed in 8 patients with retained devices, including 6 patients who underwent a total of 21 MRIs of the surgical site. There were no complications related to the MRI scans, and we found no significant degradation of image quality.
CONCLUSION: Retained implantable Doppler probes were not associated with substantial adverse clinical outcomes nor affected MRI image quality of the surgical site.
© 2021 The Author(s).

Entities:  

Keywords:  Cook-Swartz; flap monitoring; free flap; implantable Doppler; microsurgery

Year:  2021        PMID: 35096688      PMCID: PMC8793759          DOI: 10.1177/22925503211006537

Source DB:  PubMed          Journal:  Plast Surg (Oakv)        ISSN: 2292-5503            Impact factor:   0.947


  14 in total

1.  On the heating of linear conductive structures as guide wires and catheters in interventional MRI.

Authors:  W R Nitz; A Oppelt; W Renz; C Manke; M Lenhart; J Link
Journal:  J Magn Reson Imaging       Date:  2001-01       Impact factor: 4.813

Review 2.  Magnetic resonance safety update 2002: implants and devices.

Authors:  Frank G Shellock
Journal:  J Magn Reson Imaging       Date:  2002-11       Impact factor: 4.813

Review 3.  Safety considerations in MR imaging.

Authors:  E Kanal; F G Shellock; L Talagala
Journal:  Radiology       Date:  1990-09       Impact factor: 11.105

4.  Eight-year experience of the Cook-Swartz Doppler in free-flap operations: microsurgical and reexploration results with regard to a wide spectrum of surgeries.

Authors:  Ariel Schmulder; Eyal Gur; Arik Zaretski
Journal:  Microsurgery       Date:  2011-01       Impact factor: 2.425

5.  Fixation of implantable Doppler probe with fibrin sealant after detachment of the silicone cuff.

Authors:  Jeong Tae Kim; Youn Hwan Kim; Sang Wha Kim
Journal:  J Plast Reconstr Aesthet Surg       Date:  2015-08-28       Impact factor: 2.740

Review 6.  An illustrative review to understand and manage metal-induced artifacts in musculoskeletal MRI: a primer and updates.

Authors:  J P Dillenseger; S Molière; P Choquet; C Goetz; M Ehlinger; G Bierry
Journal:  Skeletal Radiol       Date:  2016-02-02       Impact factor: 2.199

7.  The Cook-Swartz venous Doppler probe for the post-operative monitoring of free tissue transfers in the United Kingdom: a preliminary report.

Authors:  David W Oliver; Iain S Whitaker; Henk Giele; Paul Critchley; Oliver Cassell
Journal:  Br J Plast Surg       Date:  2005-04

Review 8.  Advances in MRI around metal.

Authors:  Pia M Jungmann; Christoph A Agten; Christian W Pfirrmann; Reto Sutter
Journal:  J Magn Reson Imaging       Date:  2017-03-25       Impact factor: 4.813

9.  Direct comparison of postoperative monitoring of free flaps with microdialysis, implantable cook-swartz Doppler probe, and clinical monitoring in 20 consecutive patients.

Authors:  Markus Winther Frost; Vachara Niumsawatt; Warren Matthew Rozen; Gete Ester Toft Eschen; Tine Engberg Damsgaard; Birgitte Jul Kiil
Journal:  Microsurgery       Date:  2014-10-06       Impact factor: 2.425

Review 10.  Biological effects and safety in magnetic resonance imaging: a review.

Authors:  Valentina Hartwig; Giulio Giovannetti; Nicola Vanello; Massimo Lombardi; Luigi Landini; Silvana Simi
Journal:  Int J Environ Res Public Health       Date:  2009-06-10       Impact factor: 3.390

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.