Adverse events associated with AXIOS stents: Insights from the manufacturer and user facility device experience database.

Daryl Ramai1, Antonio Facciorusso2, Matthew DeLuca3, Monique Barakat4, Douglas G Adler5.   

Abstract

Background and
Objectives: The AXIOS stent is indicated for transgastric or transduodenal drainage of symptomatic pancreatic pseudocysts and symptomatic walled off necrosis. The AXIOS stent functions as a conduit which allows solid and liquid pancreatic fluid collections (PFC) contents to pass into the luminal GI tract and also allows the passage of standard and therapeutic endoscopes into the PFC to perform endoscopic debridement. We aim to investigate the number and type of complications associated with AXIOS stents. Materials and
Methods: We analyzed postmarketing surveillance data from the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database January 2016 to February 2021.
Results: During the study period, approximately 588 reports with 579 device issues and 250 patient complications were identified. Most device complications were due to stent positioning problems or stent malpositioning (n = 206; 35.6%), followed by stent migration (n = 72; 12.4%), premature deployment (n = 61; 10.5%), material integrity (n = 56; 9.6%), deployment failure (n = 47; 8.1%), and difficulty removing the stent (n = 45; 7.7%). The most reported patient adverse events were hemorrhage/bleeding (n = 81; 32.4%), perforation (n = 26; 10.4%), pain (n = 22; 8.8%), unspecified infection (n = 20; 8.0%), and death (n = 17; 6.8%). Conclusions: Findings from the MAUDE database highlight patient and device complications which endoscopists should be aware of before AXIOS stent placement.

Entities:  

Keywords:  AXIOS stent; endoscopy; pancreatic fluid collection

Year:  2022        PMID: 35083979      PMCID: PMC9258023          DOI: 10.4103/EUS-D-21-00096

Source DB:  PubMed          Journal:  Endosc Ultrasound        ISSN: 2226-7190            Impact factor:   5.275


INTRODUCTION

A novel large-diameter self-expanding metal stent with bilateral flanges, the AXIOS stent (Boston Scientific, Natick, MA, USA), has been designed especially for transmural drainage of pancreatic fluid collections (PFCs) including pancreatic pseudocysts and walled-off pancreatic necrosis (WON). The stent has been widely adopted for this indication and is used in the United States and around the world.[1234] The AXIOS stent is silicone covered with a nitinol-braided design which is deployed under EUS guidance, with or without the assistance of fluoroscopy.[3] The AXIOS stent has a “dumbbell-shaped” configuration which allows for wide flanges on both ends. This helps to prevent stent migration by providing an anchoring point and even distribution of pressure on the luminal walls within the PFC. The AXIOS stent is indicated for transgastric (TG) or transduodenal (TD) drainage of symptomatic pancreatic pseudocysts (≥6 cm in size) and symptomatic WON (≥6 cm in size with ≥70% fluid content) that are adherent to the gastric or bowel wall.[56] The AXIOS stent functions as a conduit which allows solid and liquid PFC contents to pass into the luminal GI tract and also allows the passage of standard and therapeutic endoscopes into the PFC to perform endoscopic debridement. Multiple large-scale studies evaluating the AXIOS stent have demonstrated high clinical efficacy, but adverse events have been reported.[1234] A large multicenter prospective cohort study demonstrated that EUS-guided placement of the AXIOS stent had a technical success rate of 98 % (95 % CI: 95 % – 100 %), and a clinical success rate of 93% (95 % confidence interval [CI]: 77 % – 100 %) for pancreatic pseudocysts and 81% (95 % CI 69 % – 94 %) for pancreatic walled off necrosis (WON), respectively.[7] However, the adverse event rate in this study was 9% (95% CI: 2%–16%), which included perforation and PFC infection.[7] In addition, a multicenter, international, retrospective review from 15 centers on patients who underwent placement of learning activity management system (LAMS) for the management of PFCs reported that seventy-nine LAMS-related adverse events occurred in 74 of 304 patients (24.3%), after a mean time of 25.3 days (median, 18 days; interquartile range: 6–30).[3] Another multicenter, retrospective study involving 14 centers of patients who underwent EUS-guided drainage of WON found that adverse events were observed in 9.8 % of LAMS and these were rated as severe in 2.0 % of cases.[8] The aim of this study was to evaluate events associated with the use of the AXIOS stent through the Food and Drug Administration (FDA's) Manufacturer and User Facility Device Experience (MAUDE) database from 2016 to 2021.

MATERIALS AND METHODS

We analyzed postmarketing surveillance data on AXIOS stents from the FDA MAUDE database to report device-related deaths, injuries, and modes of failure. The MAUDE database collects major adverse reports involving medical devices after FDA approval. Reporting can be mandatory (manufacturers, importers, and device-user facilities) or voluntary (healthcare professionals, patients, and consumers) and is freely and publicly accessible, see website: https://www.accessdata.fda. gov/scripts/cdrh/cfdocs/cfmaude/search. Cfm. The database is updated monthly with medical reports containing information on the device, event date, whether the device was returned to the manufacturer, and users’ and manufacturers’ event narratives. Events are classified on the basis of severity into four categories: death, injury, malfunction, or other. If a device is deemed defective, the FDA can issue safety alerts or recalls. Although this surveillance system cannot be used to establish definitive event rates, it can provide important insights into the most encountered complications and into potential mechanisms of medical devices. We queried the MAUDE database from January 2016 to February 2021. Individual reports were analyzed for date, device issues, and patient adverse including intensive care unit (ICU) admission, treatment, and mortality. Statistical analysis was performed using SPSS statistics version 27 (IBM SPSS Statistics for Windows, Version 24.0. Armonk, NY: IBM Corp.). No IRB approval needed for this study as the MAUDE Database is publicly accessible and deidentified.

RESULTS

Five hundred and eighty-eight reports with 579 device issues and 250 patient complications were identified. During the study period, the number of adverse events reported to the FDA increased from 17.2% to 24.5% in 2020. Most device complications were due to a stent positioning problems or stent malpositioning (n = 206; 35.6%), followed by stent migration (n = 72; 12.4%), premature deployment (n = 61; 10.5%), material integrity (n = 56; 9.6%), deployment failure (n = 47; 8.1%), and difficulty removing the stent (n = 45; 7.7%). Other less frequent device issues are reported in Table 1.
Table 1

Device issues reported

Device problem n
Entrapment of device16
Electrical problem6
Difficult to advance11
Material integrity56
Position problem206
Partial blockage25
Migration or expulsion72
Premature activation61
Detachment of device component29
Difficult to remove45
Activation failure47
Operates different than expected4
Device contaminated1

NOS: Not otherwise specified

Device issues reported NOS: Not otherwise specified The most reported patient adverse events were hemorrhage/bleeding (n = 81; 32.4%), perforation (n = 26; 10.4%), pain (n = 22; 8.8%), unspecified infection (n = 20; 8.0%), and death (n = 17; 6.8%). Other less frequently reported patient adverse events are reported in Table 2. Eight patients (1.4%) required ICU level of care, though unclear if due to underlying systemic disease or related to the stent only.
Table 2

Adverse events reported

Adverse events n
Edema1
Hematoma3
Fistula2
Syncope1
Erosion12
Abscess2
Fever10
Pain22
Nausea2
Vomiting6
Perforation26
Unspecified infection20
Peritonitis8
Sepsis4
Obstruction6
Laceration3
Cardiac arrest1
Pneumonia3
Abdominal distension1
Dyspnea1
Death17
Aneurysm/pseudoaneurysm9
Hemorrhage/bleeding81
Shock2
Thrombus1
Jaundice3
Stenosis3

NOS: Not otherwise specified

Adverse events reported NOS: Not otherwise specified Device problems and patient complications were further characterized according to the location of endoscopic placement (e.g. TG position and TD position). Approximately 332 (56.4%) reported adverse event cases reported using TG positioning, while 64 (10.9%) reported adverse event cases reported TD positioning; 192 (32.7%) cases did not provide sufficient data to categorized according to endoscopic placement [Tables 3 and 4]. TG was associated with a higher stent positioning problem compared to TD (TG 116 vs. TD 36); as well as higher bleeding events (TG 48 vs. TD 0).
Table 3

Device issues reported according to position

Device problemTransgastricTransduodenalUnspecified
Entrapment of device1024
Electrical problem420
Difficult to advance515
Material integrity30620
Position problem1163654
Partial blockage13012
Migration or expulsion38529
Premature activation32821
Detachment16112
Difficult to remove28215
Activation failure27614
Unexpected operation121
Device contaminated100

NOS: Not otherwise specified

Table 4

Adverse events reported according to position

Adverse eventsTransgastricTransduodenalUnspecified
Edema100
Hematoma201
Fistula101
Syncope001
Erosion507
Abscess101
Fever613
Pain9310
Nausea200
Vomiting312
Perforation10115
Unspecified infection7310
Peritonitis215
Sepsis310
Obstruction204
Laceration201
Cardiac arrest100
Pneumonia003
Abdominal distension010
Dyspnea100
Death629
Aneurysm/pseudoaneurysm405
Hemorrhage/bleeding48033
Shock101
Thrombus001
Jaundice003
Stenosis003

NOS: Not otherwise specified

Device issues reported according to position NOS: Not otherwise specified Adverse events reported according to position NOS: Not otherwise specified

DISCUSSION

We performed an analysis of the FDA MAUDE database for adverse events and device problems related to AXIOS stent placement. Hemorrhage/bleeding, perforation, pain, unspecified infection, and death were the most frequently reported adverse events. Stent positioning problems, migration or expulsion, premature deployment, and problems with material integrity leading to malfunction were the most frequently reported device problems. Unsurprisingly, deployment of the AXIOS stent in the TG position accounted for most reported complications (as PFCs are typically retrogastric). It should be noted that TD positioning of the AXIOS stent was utilized far less frequently, which may account for its lower rate of reported complications. To date, there are no head-to-head randomized clinical trials comparing the safety and efficacy of TG versus TD endoscopic placement of the AXIOS stent. This may be because the site chosen for cystenterostomy is made clinically based on the characteristics of the PFC, and the preference of the operator and likely cannot be randomized.[9] A substantial amount of literature exists supporting the clinical utility of the AXIOS stent. Small retrospective studies have demonstrated that the AXIOS stent has a higher technical success rate (98.9%) when compared to fully covered self-expanding metal stents and double pigtail stents.[2101112] These studies also concluded that the AXIOS stent achieves a PFC clinical resolution rate in 92.5% of cases.[2101112] The reported complication rate in these studies ranged from 5% to 9.4%.[2101112] AXIOS stents have been widely used for a variety of pancreatic and off-label nonpancreatic indications. These included but not limited to biliary drainage (EUS-guided choledochoduodenostomy and EUS-guided cholecystostomy), luminal bypass (EUS-guided gastroenterostomy), EUS-guided transmural drainage of PFCs (including walled off necrosis and pseudocysts), and for treatment of postsurgical fluid collections.[131415161718] Table 5 shows a summary of these indications and their rates technical and clinical success as well as adverse events. Overall, AXIOS stent placement has been associated with high clinical and technical success with adverse events ranging from 4% to 22%, depending on the indication. These results indicate that adverse events are significant and should be carefully reviewed between endoscopist and patient before stent placement.
Table 5

Indications and outcomes of lumen apposing metal stent placement

Type of studyNumber of patientsIndication for LAMSTechnical success (%)Clinical success (%)Adverse events (%)Reference
Meta-analysis737Pancreatic walled off necrosis-88.5% (95% CI 82.5-92.611.2% (95% CI 6.8-17.9)Mohan et al.
Meta-analysis332Pancreatic pseudocysts95.76%94.06%-Lyu et al.
Meta-analysis1437Biliary drainage91.5% (95% CI 87.7-94.2)87% (95% CI 82.3-90.6)17.9% (95% CI 14.3-22.2)Dhindsa et al.
Meta-analysis278Palliation of gastric outlet obstruction95.2% (CI 87.2-98.3)93.3% (CI 84.4-97.3)10.7% (CI 4.3-24.5)Chandan et al.
Meta-analysis1538Symptomatic gallbladder disease94.65% (95% CI 91.54-96.67)92.06% (95% CI 88.65-94.51)11.71% (95% CI 8.92-15.23)McCarty et al.
Retrospective46Malignant distal biliary obstruction93.5%97.7%11.6%Anderloni et al.
Multicenter retrospective47Postsurgical fluid collections93.6%89.3%4.25%Mudireddy et al.

CI: Confidence interval; LAMS: Lumen-apposing stent

Indications and outcomes of lumen apposing metal stent placement CI: Confidence interval; LAMS: Lumen-apposing stent A multicenter, international, retrospective review concluded that the adverse events associated with LAMS are not negligible and should be considered before utilized for the drainage of PFCs.[3] However, while the placement of AXIOS stents is associated with risks, AXIOS stents appear to represent an overall safer and more effective approach when compared to percutaneous and surgical methods, and endoscopic drainage of pancreatic pseudocysts and necrosis is, and will likely remain, first-line therapy.[19] Our analysis of the MAUDE database revealed 81 patients who experienced hemorrhage/bleeding associated with AXIOS stent placement or removal. The mechanism by which AXIOS-related hemorrhage occurs is the subject of significant discussion and controversy. A majority of the reports described significant hemorrhage of the splenic artery either during endoscopic placement or upon stent removal. 81 of the 81 [100%] patients who experienced hemorrhage/bleeding underwent AXIOS stent placement in the TG position. It is believed that the intimate anatomical relationship of the splenic artery and posterior gastric wall may contribute to the heightened adverse events reported in the TG position. Furthermore, many patients with severe pancreatitis develop splenic and/or portal vein thrombosis and collateral vasculature in the retroperitoneum that could also be sources of bleeding. Our study showed that there were perforations (n = 26), pain (n = 22), and unspecified infections (n = 20) following AXIOS stent placement. Importantly, we also identified 17 deaths during the study period. It must be noted that, at this time, there is limited information available regarding these deaths, and patients with pancreatic pseudocysts and/or pancreatic necrosis represent a high-risk patient cohort. There are several limitations to this study: (1) MAUDE database reporting is inconsistent, and complications may be underreported; (2) details regarding procedures are limited, making it difficult to establish the exact cause of the reported events, such as operator error, interaction with other devices, or device defect; (3) the total number of reported devices used by the reporting facility or from the manufacturer during the research time frame is unavailable, so the incidence of adverse events cannot be estimated; and (4) MAUDE database does not report the total number of AXIOS stent placements performed annually in the United States.

CONCLUSIONS

AXIOS stent placement is widely performed for the drainage of PFCs. Our analysis of the FDA MAUDE database revealed the type, number, and trends of reported device adverse events. Endoscopists should be aware of the risk of hemorrhage/bleeding, perforation, pain, and infection as potential major complications of Axios stent placement.

Financial support and sponsorship

Nil.

Conflicts of interest

Douglas G. Adler is an Co-Editor-in-Chief of the journal. The article was subject to the journal's standard procedures, with peer review handled independently of this editor and his research groups.
  16 in total

1.  A novel lumen-apposing metal stent for endoscopic ultrasound-guided drainage of pancreatic fluid collections: a prospective cohort study.

Authors:  Daisy Walter; Uwe Will; Andres Sanchez-Yague; Dirk Brenke; Jochen Hampe; Helge Wollny; Jose Miguel Esteban López-Jamar; Gertrud Jechart; Peter Vilmann; Joan B Gornals; Sebastian Ullrich; Martin Fähndrich; Alberto Herreros de Tejada; Félix Junquera; Ferran Gonzalez-Huix; Peter D Siersema; Frank P Vleggaar
Journal:  Endoscopy       Date:  2014-09-30       Impact factor: 10.093

2.  International multicenter comprehensive analysis of adverse events associated with lumen-apposing metal stent placement for pancreatic fluid collection drainage.

Authors:  Alessandro Fugazza; Amrita Sethi; Arvind J Trindade; Edoardo Troncone; John Devlin; Mouen A Khashab; Frank P Vleggaar; Auke Bogte; Ilaria Tarantino; Pierre H Deprez; Carlo Fabbri; José Ramón Aparicio; Paul Fockens; Rogier P Voermans; Will Uwe; Geoffroy Vanbiervliet; Antoine Charachon; Christopher D Packey; Petros C Benias; Yasser El-Sherif; Christopher Paiji; Dario Ligresti; Cecilia Binda; Belén Martínez; Loredana Correale; Douglas G Adler; Alessandro Repici; Andrea Anderloni
Journal:  Gastrointest Endosc       Date:  2019-11-20       Impact factor: 9.427

Review 3.  Review of management options for pancreatic pseudocysts.

Authors:  Christos Agalianos; Ioannis Passas; Ioannis Sideris; Demetrios Davides; Christos Dervenis
Journal:  Transl Gastroenterol Hepatol       Date:  2018-03-21

4.  Fully covered self-expanding metal stents versus lumen-apposing fully covered self-expanding metal stent versus plastic stents for endoscopic drainage of pancreatic walled-off necrosis: clinical outcomes and success.

Authors:  Ali A Siddiqui; Thomas E Kowalski; David E Loren; Ammara Khalid; Ayesha Soomro; Syed M Mazhar; Laura Isby; Michel Kahaleh; Kunal Karia; Joseph Yoo; Andrew Ofosu; Beverly Ng; Reem Z Sharaiha
Journal:  Gastrointest Endosc       Date:  2016-08-24       Impact factor: 9.427

5.  Endoscopic Therapy With Lumen-apposing Metal Stents Is Safe and Effective for Patients With Pancreatic Walled-off Necrosis.

Authors:  Reem Z Sharaiha; Amy Tyberg; Mouen A Khashab; Nikhil A Kumta; Kunal Karia; Jose Nieto; Uzma D Siddiqui; Irving Waxman; Virendra Joshi; Petros C Benias; Peter Darwin; Christopher J DiMaio; Christopher J Mulder; Shai Friedland; David G Forcione; Divyesh V Sejpal; Tamas A Gonda; Frank G Gress; Monica Gaidhane; Ann Koons; Ersilia M DeFilippis; Sanjay Salgado; Kristen R Weaver; John M Poneros; Amrita Sethi; Sammy Ho; Vivek Kumbhari; Vikesh K Singh; Alan H Tieu; Viviana Parra; Alisa Likhitsup; Craig Womeldorph; Brenna Casey; Sreeni S Jonnalagadda; Amit P Desai; David L Carr-Locke; Michel Kahaleh; Ali A Siddiqui
Journal:  Clin Gastroenterol Hepatol       Date:  2016-05-14       Impact factor: 11.382

6.  Endoscopic gallbladder drainage for symptomatic gallbladder disease: a cumulative systematic review meta-analysis.

Authors:  Thomas R McCarty; Kelly E Hathorn; Ahmad Najdat Bazarbashi; Kunal Jajoo; Marvin Ryou; Christopher C Thompson
Journal:  Surg Endosc       Date:  2021-07-06       Impact factor: 4.584

7.  Comparison Between Lumen-Apposing Metal Stents and Plastic Stents in Endoscopic Ultrasound-Guided Drainage of Pancreatic Fluid Collection: A Meta-analysis and Systematic Review.

Authors:  Yunxiao Lyu; Ting Li; Bin Wang; Yunxiao Cheng; Liang Chen; Sicong Zhao
Journal:  Pancreas       Date:  2021-04-01       Impact factor: 3.327

8.  A pilot study of a 20-mm lumen-apposing metal stent to treat pancreatic fluid collections: First reported multicenter use of a new device.

Authors:  Douglas G Adler; Shawn Mallery; Stuart Amateau; Jose Nieto; Linda Jo Taylor; Ali Siddiqui
Journal:  Endosc Ultrasound       Date:  2019 Mar-Apr       Impact factor: 5.628

9.  EUS-guided gastroenterostomy versus enteral stenting for gastric outlet obstruction: Systematic review and meta-analysis.

Authors:  Saurabh Chandan; Shahab R Khan; Babu P Mohan; Aun R Shah; Mohammad Bilal; Daryl Ramai; Neil Bhogal; Banreet Dhindsa; Lena L Kassab; Shailendra Singh; Suresh Ponnada; Andrew K Nguyen; Stephanie McDonough; Douglas G Adler
Journal:  Endosc Int Open       Date:  2021-02-22

Review 10.  EUS-guided biliary drainage: A systematic review and meta-analysis.

Authors:  Banreet Singh Dhindsa; Harmeet Singh Mashiana; Amaninder Dhaliwal; Babu P Mohan; Mahendran Jayaraj; Harlan Sayles; Shailender Singh; Gordon Ohning; Ishfaq Bhat; Douglas G Adler
Journal:  Endosc Ultrasound       Date:  2020 Mar-Apr       Impact factor: 5.628

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.