Influences of assisted breathing and mechanical ventilator settings on tidal volume and alveolar pressures in acute respiratory distress syndrome: a bench study.

Renata Santos Vasconcelos1, Raquel Pinto Sales1, Juliana Arcanjo Lino1, Luíza Gabriela de Carvalho Gomes1, Nancy Delma Silva Vega Canjura Sousa1, Liégina Silveira Marinho1, Bruno do Valle Pinheiro2, Marcelo Alcantara Holanda1.   

Abstract

OBJECTIVE: To evaluate the influences of respiratory muscle efforts and respiratory rate setting in the ventilator on tidal volume and alveolar distending pressures at end inspiration and expiration in volume-controlled ventilation and pressure-controlled ventilation modes in acute respiratory distress syndrome.
METHODS: An active test lung (ASL 5000™) connected to five intensive care unit ventilators was used in a model of acute respiratory distress syndrome. Respiratory muscle efforts (muscle pressure) were configured in three different ways: no effort (muscle pressure: 0cmH2O); inspiratory efforts only (muscle pressure:-5cmH2O, neural inspiratory time of 0.6s); and both inspiratory and expiratory muscle efforts (muscle pressure:-5/+5cmH2O). Volume-controlled and pressure-controlled ventilation modes were set to deliver a target tidal volume of 420mL and positive end-expiratory pressure of 10cmH2O. The tidal volume delivered to the lungs, alveolar pressures at the end of inspiration, and alveolar pressures at end expiration were evaluated.
RESULTS: When triggered by the simulated patient, the median tidal volume was 27mL lower than the set tidal volume (range-63 to +79mL), and there was variation in alveolar pressures with a median of 25.4cmH2O (range 20.5 to 30cmH2O). In the simulated scenarios with both spontaneous inspiratory and expiratory muscle efforts and with a mandatory respiratory rate lower than the simulated patient's efforts, the median tidal volume was higher than controlled breathing.
CONCLUSION: Adjusting respiratory muscle effort and pulmonary ventilator respiratory rate to a value above the patient's respiratory rate in assisted/controlled modes generated large variations in tidal volume and pulmonary pressures, while the controlled mode showed no variations in these outcomes.

Entities:  

Mesh:

Year:  2022        PMID: 35081242      PMCID: PMC8889593          DOI: 10.5935/0103-507X.20210084

Source DB:  PubMed          Journal:  Rev Bras Ter Intensiva        ISSN: 0103-507X


INTRODUCTION

Ventilator-induced lung injury (VILI) is an iatrogenic cause of pulmonary damage related to excessive mechanical stress and/or strain imposed on the lung tissue during mechanical ventilation (MV).([ It is of particular concern for patients with acute respiratory distress syndrome (ARDS), as they present with severe lung edema and inflammation. Furthermore, mechanical alterations are heterogeneously distributed inside the lung parenchyma in ARDS, thereby predisposing the alveoli and small airways to excessive distension or pressures during tidal breathing.([ Setting the tidal volume (VT) to 4 to 6mL/kg of predicted or ideal body weight and limiting the distending pressures - both plateau (< 28 - 30cmH2O) and, particularly, the driving pressure (< 15cmHO) - during MV were associated with improved survival in ARDS.([ In fact, so-called protective ventilatory strategies may prevent or attenuate VILI by reducing both the stress and the strain on the lungs caused by MV.([ They are now the standard of care for the initial controlled MV of patients with ARDS.([ However, mortality rates remain high, ranging from 34% to 60%.([ Currently, there are no guidelines on ventilating ARDS patients with preserved respiratory drive and spontaneous respiratory efforts, i.e., assisted MV. Maintaining spontaneous breathing during MV may have beneficial effects, such as preventing diaphragmatic atrophy and dysfunction, avoiding respiratory monotony regarding VT variation, and recruiting juxta diaphragmatic alveoli, which usually collapse in severe ARDS.([ All these factors may contribute to the early liberation of the patient from the ventilator.([ On the other hand, assisted breathing during MV may result in higher VT and transpulmonary pressures, especially in areas close to collapsed alveoli, and may result in tidal recruitment and pendelluft ventilation, thus amplifying heterogeneous distensions of the lung parenchyma.([ Even in patients with good patient-ventilator synchrony, the target VT and the desired airway pressure limits may be frequently exceeded,([ thus compromising the effectiveness of protective ventilatory strategies. Furthermore, patient-ventilator asynchronies such as double triggering, also referred to as breath-stacking or ineffective efforts, may result in huge VT and transpulmonary pressures, which increase the risk of VILI.([ Researchers have found an association between patient-ventilator asynchronies and mortality in mechanically ventilated patients.([ On the other hand, controlled MV has been associated with VILI prevention or attenuation in experimental studies and, more importantly, with improved outcomes, including survival, in patients with moderate or severe ARDS.([ Three randomized controlled trials demonstrated the positive impact of early neuromuscular blockade in ARDS on functional parameters and mortality.([ Little attention has been given to the influences of inspiratory and expiratory muscle efforts, ventilatory modes - either volume-controlled ventilation (VCV) or pressure-controlled ventilation (PCV) - or the number of mandatory respiratory cycles (set respiratory rate - RR), all of which are combined, on VT and distending alveolar pressures during assisted MV. The main differences between VCV and PCV during assisted MV are the amount and type of flow delivered to the lungs, which may be higher with greater patient effort and exponential deceleration in the latter. In both modes, the set RR may cause patient-ventilator asynchronies when it is higher than the spontaneous RR of the patient. Therefore, we hypothesized that: first, assisted breaths invariably result in a VT higher than that in controlled breathing cycles and higher alveolar pressures, even in synchronic breathings; second, triggering and cycling asynchronies caused by setting the RR higher than the patient’s spontaneous RR, regardless of the ventilatory mode (VCV or PCV) or the intensive care unit (ICU) ventilator type, causes huge variations in VT and alveolar distending pressures; third, early active expiratory effort during inspiration may limit VT augmentation and the correspondent increase in alveolar distending pressures. The main objectives were to test the above hypothesis by evaluating the influences of respiratory muscle efforts - both inspiratory and expiratory - and of the RR setting in the ventilator - above or below the patient’s RR - on VT and alveolar distending pressures at end inspiration and expiration in both VCV and PCV in a mechanical simulated model of ARDS.

METHODS

This bench study was conducted at the Respiration Laboratory of the Department of Internal Medicine of the Medical School of the Universidade Federal do Ceará, Brazil.

Simulated model

An ASL 5000™ mechanical simulator (IngMar Medical, Pittsburgh, EUA) was used. The respiratory model was configured to reproduce, as realistically as possible, the mechanical characteristics of an adult patient with moderate to severe ARDS with spontaneous breathing efforts.([ The following parameter settings were used: static compliance 25mL/cmH2O, and inspiratory airway resistance 10cmH2O/L/sec.([ Respiratory muscle efforts (muscle pressure - Pmus) were configured in three different ways: no effort (Pmus: 0cmH2O); inspiratory efforts only (Pmus:-5cmH2O, neural inspiratory time of 0.6s); and both inspiratory and expiratory muscle efforts (Pmus:-5/+5cmH2O, with neural inspiratory and expiratory times of 0.6s each).([ The simulated patient RR was set at 20 bpm.

Intensive care unit ventilators

Five ICU ventilators were used: Esprit V-1000 (Respironics™, Murrysville, EUA), DX 3012 (Dixtal™, Buenos Aires, Argentina), Servo I (Maquet™; Solna, Sweden), Puritan-Bennet 840 (Covidien Mansfield, MA, USA), and Savina 300 (Drager™, Lübeck, Germany). All ventilators used dual limbs (inspiratory and expiratory circuits) connected to a Y-adapter and an orotracheal tube (I.D 8.0mm) with no humidification system.([

Experimental protocol

The ICU ventilators were tested and calibrated according to their manufacturer’s recommendations. Volume-controlled ventilation and PCV were used. Volume-controlled ventilation was set to deliver a target VT of 420mL (6mL/kg for an IBW of 70kg) and an inspiratory time of 0.8s with a constant flow (square wave format) of 31L/min. Pressure-controlled ventilation was set to deliver a VT of 420mL, as in the VCV, by carefully titrating airway pressure above the positive end-expiratory pressure (PEEP) with the same inspiratory time of 0.8s. In both modes a PEEP of 10cmH2O and a pressure triggering sensitivity threshold of 2cmH2O below PEEP were set.([ In the Savina 300™ ventilator, the tests were also run with the AutoFlow® (AF) system in VCV mode (VCV-AF). In short, this system calculates the respiratory compliance in each breath and automatically delivers an initial inspiratory flow that equals the ratio of the target VT to the respiratory compliance.([ Another characteristic of this mode is that it allows spontaneous breathing during the breathing cycle, as the inspiratory and expiratory valves are kept open during the two phases of the breathing cycle.([

Measurements and outcomes

Each simulated scenario was recorded after stabilization of the respiratory pattern, which usually occurred rapidly, in less than 3 to 5 minutes, as was expected for a mechanical simulation. Thereafter, five consecutive minutes of the simulation were continuously recorded. Then, 20 representative breaths were selected for off-line analysis using ASL 5000™ software (LabVIEW; National Instruments; Austin, TX, USA). In total, 1.100 breaths were analyzed (3 scenarios, 2 modes, 2 settings of the mandatory RR, 5 ventilators, AF, 20 breaths per each) for the following variables: 1) VT delivered to the lungs, 2) alveolar pressures at the end of inspiration, 3) alveolar pressures at end expiration (effective or total PEEP) and 4) the difference between the alveolar pressure and the Pmus (which was considered a surrogate for the transpulmonary pressure as there is no pleural pressure in the mechanical model). Figure 1 shows the simulated scenarios. A total of 10 scenarios were studied for each ventilator. For the Savina 300™ ventilator, one additional scenario was tested in the AF.
Figure 1

Simulated scenarios according to the ventilatory modes and settings of the mandatory respiratory rate and the presence and types of spontaneous respiratory efforts.

Simulated scenarios according to the ventilatory modes and settings of the mandatory respiratory rate and the presence and types of spontaneous respiratory efforts. Figure 2 depicts representative curves that show how the respiratory variables were measured.
Figure 2

Two representative breathing cycles showing the tidal volume, alveolar pressure, and muscle pressure in the same plot.

The variables are described as medians and minimum and maximum values. Given the stability of the mechanical model and its almost negligible variability, we chose to make nominal comparisons between the obtained values without conducting comparative statistical tests.([ Differences that were considered potentially clinically relevant were highlighted and discussed. We predefined VT values > 560mL (8mL/kg) and end inspiration alveolar pressure (Palv) > 28cmH2O as clinically relevant.

RESULTS

Tables 1 to 4 show the results of the VT, Palv, total PEEP, and transpulmonary pressure end of inspiration for the five ventilators in all simulated scenarios, and figure 3 shows the difference between programmed (420mL) and observed VT in VCV and PCV mode. As expected, the VT remained constant with no variation during controlled MV (no effort, Pmus = 0). In general, during assisted MV, the VT, and alveolar pressures increased in both the VCV and PCV modes in all scenarios.
Table 1

Tidal volume values in the volume-controlled ventilation and pressure-controlled ventilation modes for all the ventilators and simulated scenarios

VCV mode Ventilators ICUVT (mL)PCV modeVT (mL)
Patient effortNo effort*Pmus-5cmH2OPmus-5cmH2O/+5cmH2ONo effort*Pmus-5cmH2OPmus-5cmH2O/+5cmH2O
Ventilator RRzero15251525zero15251525
SERVO IMaximum413 (-7)440 (+20)443 (+23)417 (-3)526 (+106)SERVO I418 (-2)439 (+19)483 (-63)421 (+1)546 (+126)
Median413 (-7)440 (+20)416 (-4)417 (-3)499 (+79)418 (-2)439 (+19)422 (+2)421 (+1)481 (+61)
Minimum413 (-7)440 (+20)409 (-11)417 (-3)440 (+20)418 (-2)439 (+19)413 (-7)421 (+1)476 (+56)
ESPRITMaximum353 (-67)358 (-62)392 (-28)362 (-58)461 (+41)ESPRIT353 (-67)419 (-1)422 (+2)424 (+4)508 (+88)
Median353 (-67)358 (-62)363 (-57)362 (-58)413 (-7)353 (-67)419 (-1)345 (-75)424 (+4)444 (+24)
Minimum353 (-67)358 (-62)356 (-64)362 (-58)367 (-53)353 (-67)419 (-1)341 (79)424 (+4)430 (+10)
DX 3012Maximum379 (-41)387 (-33)422 (+2)370 (-50)406 (-14)DX 3012418 (-2)434 (+14)476 (+56)417 (-3)490 (+70)
Median379 (-41)387 (-33)407 (-13)370 (-50)392 (-28)418 (-2)434 (+14)429 (+9)417 (-3)445 (-25)
Minimum379 (-41)387 (-33)394 (-26)370 (-50)370 (-50)418 (-2)434 (+14)418 (-2)417 (-3)438 (+18)
PB 840Maximum382 (-38)386 (-34)416 (-4)392 (-28)474 (+54)PB 840382 (-38)385 (-35)424 (+4)393 (-27)536 (+116)
Median382 (-38)386 (-34)393 (-27)392 (-28)449 (+29)382 (-38)385 (-35)371 (-49)393 (-27)467 (+47)
Minimum382 (-38)386 (-34)378 (-42)392 (-28)398 (-22)382 (-38)385 (-35)363 (-57)393 (-27)448 (+28)
SAVINAMaximum381 (-39)421 (+1)421 (+1)389 (-31)463 (+43)SAVINA411 (-9)414 (-6)428 (+8)382 (-38)459 (+39)
Median381 (-39)421 (+1)388 (-32)389 (-31)402 (-18)411 (-9)414 (-6)389 (-31)382 (-38)404 (-16)
Minimum381 (-39)421 (+1)367 (-53)389 (-31)400 (-20)411 (-9)414 (-6)372 (-48)382 (-38)398 (-22)
SAVINA AFMaximum461 (+41)450 (+30)508 (+88)452 (+32)556 (+136)-----
Median461 (+41)450 (+30)479 (+59)452 (+32)479 (+59)
Minimum461 (+41)450 (+30)430 (+10)452 (+32)468 (+48)

Absence of muscle effort: Pmus = zero. Inspiratory effort: Pmus =-5cmH2O, Expiratory effort: Pmus = +5cmH2O. The spontaneous respiratory rate of the patient was set at 20bpm, and the target tidal volume was set at 420mL. In parentheses is the difference between the programmed tidal volume and the observed tidal volume.

Table 4

Transpulmonary pressure end of inspiration in the volume-controlled ventilation and pressure-controlled ventilation modes for all the ventilators and simulated scenarios

VCV mode ventilators ICUPalv - PmusPCV modePalv - Pmus
Patient effortNo effort*Pmus-5cmH2OPmus-5cmH2O/+5cmH2ONo effort*Pmus - 5cmH2OPmus-5cmH2O/+5cmH2O
Ventilator RRzero152515zero251525
SERVO IMaximum26.12727.426.328SERVO I26.427.42926.329
Median26.12726.326.326.726.427.426.826.326.2
Minimum26.12726.326.324.626.427.426.526.325.8
ESPRITMaximum24.224.725.924.725.6ESPRIT24.125.727.527.128.2
Median24.224.724.924.724.524.125.724.327.126
Minimum24.224.724.624.72324.125.724.327.125.4
DX 3012Maximum25.225.526.82525.4DX 301226.827.429.826.428.4
Median25.225.526.22524.926.827.42726.426.7
Minimum25.225.525.32524.126.827.426.726.426.4
PB 840Maximum24.825262525.6PB 84024.824.8262527.8
Median24.82524.82524.624.824.824.32525.7
Minimum24.82524.62522.424.824.823.82525.4
SAVINAMaximum25.32727.525.628SAVINA26.526.527.625.627.4
Median25.3272625.625.426.526.526.125.625.2
Minimum25.32725.325.625.426.526.525.525.625
SAVINA AFMaximum28.628.430.828.530.7-----
Median28.628.429.728.527.8
Minimum28.628.427.828.527.4

Absence of muscle effort: Pmus = zero. Inspiratory effort: Pmus =-5cmH2O, Expiratory effort: Pmus = +5cmH2O. The spontaneous respiratory rate of the patient was set at 20bpm.

Figure 3

Difference between programmed and observed tidal volume in volume-controlled ventilation and pressure-controlled ventilation mode for the five ventilators in all simulated scenarios. The tidal volume programmed was 420mL.

Two representative breathing cycles showing the tidal volume, alveolar pressure, and muscle pressure in the same plot. Tidal volume values in the volume-controlled ventilation and pressure-controlled ventilation modes for all the ventilators and simulated scenarios Absence of muscle effort: Pmus = zero. Inspiratory effort: Pmus =-5cmH2O, Expiratory effort: Pmus = +5cmH2O. The spontaneous respiratory rate of the patient was set at 20bpm, and the target tidal volume was set at 420mL. In parentheses is the difference between the programmed tidal volume and the observed tidal volume. Alveolar pressure at end-inspiration in the volume-controlled ventilation and pressure-controlled ventilation modes for all the ventilators and simulated scenarios Absence of muscle effort: Pmus = zero. Inspiratory effort: Pmus =-5cmH2O, Expiratory effort: Pmus = +5cmH2O. The spontaneous respiratory rate of the patient was set at 20bpm. Alveolar pressure at end-expiration or effective positive end-expiratory pressure in the volume-controlled ventilation and pressure-controlled ventilation modes for all ventilators and simulated scenarios Absence of muscle effort: Pmus = zero. Inspiratory effort: Pmus =-5cmH2O, Expiratory effort: Pmus = +5cmH2O. The spontaneous respiratory rate of the patient was set at 20bpm. Transpulmonary pressure end of inspiration in the volume-controlled ventilation and pressure-controlled ventilation modes for all the ventilators and simulated scenarios Absence of muscle effort: Pmus = zero. Inspiratory effort: Pmus =-5cmH2O, Expiratory effort: Pmus = +5cmH2O. The spontaneous respiratory rate of the patient was set at 20bpm. Difference between programmed and observed tidal volume in volume-controlled ventilation and pressure-controlled ventilation mode for the five ventilators in all simulated scenarios. The tidal volume programmed was 420mL.

Assisted breaths with inspiratory efforts only

There were no trigger asynchronies when only inspiratory efforts were present (Pmus =-5cmH2O) and the mandatory RR was set to a lower value (15 bpm) than the simulated spontaneous RR of the patient (20 bpm). When triggered by the simulated patient, the median VT was 27mL lower than the set VT (range - 63 to +79mL) considering all five ventilators, and there was a variation in alveolar pressure at the end of inspiration, with a median of 25.4cmH2O (range 20.5 and 30cmH2O). The transpulmonary pressure at the end of inspiration increased with the variation in VT (24.3 to 29). The tidal volume variation, in this case, was due to asynchronous events.

Assisted breaths with inspiratory and expiratory efforts

The results in the simulated scenarios with both spontaneous inspiratory and expiratory muscle efforts (Pmus =-5cmH2O followed by +5cmH2O) were as follows: first, with a mandatory RR lower than the simulated patient’s efforts, the median VT was higher than controlled breathing, but it was lower than the cases observed in the assisted MV when only inspiratory efforts were present in 6 of the 11 ventilator settings. On the other hand, Palv increased significantly above 28cmH2O, and the transpulmonary pressure end of inspiration values was similar to those obtained when only the inspiratory effort was present; second, VT increased significantly in 7 of the 11 settings (median value of 429, maximum 546 and minimum 367) with a mandatory RR higher than the spontaneous one, while the Palv, total PEEP and Palv-Pmus values showed the greatest variation among the simulated scenarios but with median values even lower or similar to those obtained in the situation of lower RR setting with synchronous assisted breathings. These alterations were related to two factors. In the case with lower mandatory RR, the presence of expiratory muscle effort reduced or attenuated the increment in VT in relation to the inspiratory effort. In the second case, the presence of expiratory effort combined with higher mandatory RR resulted in patient-ventilator asynchronies. In the scenario with the AF function active in the VCV mode of the Savina 300 ventilator, the VT was significantly higher than the VT measured when this function was inactive. Similar VT, Palv, PEEP, and Palv-Pmus values were observed when VCV and PCV modes were compared in the same ventilator. Interestingly, the effective PEEP remained at approximately 10cmH2O in most scenarios, except when the mandatory RR was set higher and when both inspiratory and expiratory efforts were present, which resulted in both increments and decrements (under pressurization). Only the Savina 300 ventilator with the AF function active in the VCV mode (VCV-AF) did not show decrements in PEEP below the value set.

DISCUSSION

The main findings of the present study can be summarized as follows: assisted breaths resulted in a higher VT than those of controlled breathing cycles and in higher alveolar pressures, even in synchronic breathings; triggering asynchronies caused by setting the RR to a level higher than that of the patient’s spontaneous RR, regardless of the ventilatory mode - VCV or PCV - or the type of ICU ventilator, caused huge variations in VT and alveolar pressures at the end of inspiration; and early active expiratory effort during mechanical inspiration may limit VT augmentation and the correspondent increase in the alveolar distending pressures when there are no triggering asynchronies. On the other hand, when triggering asynchronies were present, the combination of inspiratory and expiratory efforts caused huge variations in VT, alveolar distending pressures, and PEEP, including over- and under-pressurization of the last parameter. The AF system of the VCV mode was associated with higher VTs and alveolar pressures. Our results confirm previous findings demonstrating the effects of assisted breathing on the amount of VT and the pressure delivered to the lungs. Morais et al.([ studied an experimental model of severe ARDS using mechanically ventilated rabbits and pigs and observed that muscle effort increased lung injury, especially in the dependent lung, where greater stress and local pulmonary stretch were generated. This phenomenon was minimized by using high PEEP, which may offset the need for muscle paralysis. Moraes et al.([ studied an experimental model of mild to moderate ARDS and found that high VT was associated with VILI and that VT control appeared to be more important than RR control to attenuate VILI. In the present study, in situations of inspiratory and expiratory muscle effort, VT was above 6mL/kg only in situations with the RR set at 25 breaths/min. However, when the RR was set at 15 breaths/min for the same effort pattern, there was an increase in alveolar pressure without an increase in VT. Our hypothesis for this result is that the presence of expiratory muscle effort had a limiting effect on VT, thereby preventing the value from exceeding 6mL/kg. Biehl et al.([ emphasize that patient-ventilator asynchrony often limits the use of low VT in ARDS patients requiring high minute ventilation, where adjustments of ventilator settings and sedative agents are modestly effective in limiting asynchrony, often requiring the use of neuromuscular blockade. The present study showed that in the mechanical model, simulated neuromuscular blockade, inspiratory muscle effort, and inspiratory/expiratory muscle effort had similar effects on VT variation when the model used an RR lower (RR at 15 breaths/min) than that of the ventilator (RR at 25 breaths/min). In addition, different patterns of VT variation and pulmonary pressures were found only in conditions where muscle effort was associated with RR higher (25 breaths/min) than that of the ventilator (RR at 20 breaths/min). Thus, it is reasonable to consider that the presence of muscle effort does not necessarily potentiate lung injuries due to excessive VT. Respiratory rate setting is a key parameter in the management of MV, especially in patients who develop ARDS. Studies have reported that most patients with respiratory failure require a rate between 20 and 30 cycles/min, according to their needs.([ However, experimental studies with animals have shown that a higher RR may intensify VILI and that ventilated lungs with a lower RR produced less edema and perivascular hemorrhage than those ventilated with a higher RR.([ The results of the present study corroborate these findings, as setting the RR in the ventilator to a value above the patient’s RR generated variations in VT and pulmonary pressures, including values above the limits considered safe for the protective ventilatory strategy. It should be noted, however, that the study used a mechanical model in which the patient’s RR had a fixed pattern and did not vary according to their metabolic needs. Richard et al.([ compared a bench study with an in vivo study and showed that in both the mechanical and patient models, VT and its variability seemed to be influenced by the relationship between the patient’s RR and the RR setting in the ventilator -, i.e., the higher the rate, the lower the possibility of synchronous breathing cycles. In addition to highlighting the importance of adjusting VT, these findings also demonstrate the influence of RR on the variability of this ventilatory parameter because adjusting the RR of the ventilator to a value above the RR of the patient generates variations in VT and pressures since there is a respiratory effort by the patient. Plateau pressure or alveolar pressure cannot and should not be considered a surrogate for pulmonary stress, as there is evidence of similar stress values for completely different VTs.([ In the present study, the values of alveolar pressure at the end of inspiration, until recently described in the literature as a predictor of lung injury when above 30cmH2O, only increased to this limit when there was Pmus-5/+5cmH2O with RR set at both 15 breaths/min and 25 breaths/min, especially in the PCV mode. Briel et al.([ found that the reduction in VILI-related atelectrauma is associated with the optimization of PEEP values. However, the appropriate level of PEEP remains a matter of controversy. Randomized clinical trials, multicenter studies, and meta-analyses have not confirmed that PEEP above 12cmH2O reduces the mortality of ARDS patients.([ However, it is known that a very low end-expiratory lung volume may be related to cyclic opening and the collapse of unstable alveolar units. In this context, the detrimental effects of ventilation can be alleviated by the application of PEEP to prevent cyclic derecruitment of alveoli. However, PEEP should not be high enough to lead to excessive inflation. In the present study, PEEP levels remained close to the values of 10cmH2O in both modes, except for conditions where there was asynchrony, in which PEEP reached values higher than those set when Pmus-5cmH2O, thus suggesting hyperinflation due to the presence of auto PEEP, and values lower than those set when Pmus-5/+5cmH2O, thus suggesting system depressurization. According to Lasocki et al.,([ the AF system is based on an attractive principle: it seeks to ensure adjusted VT while maintaining the advantages of PCV. Despite this potential advantage, clinical trials have not been conducted, and its clinical efficacy compared with conventional VCV has not been formally demonstrated. In the present study, the use of the AutoFlow® system showed no advantages for patient-ventilator asynchrony compared with conventional VCV and PCV modes. In asynchronous situations, the VCV-AF mode delivered higher tidal volumes, thereby generating higher pulmonary pressures, which could potentially aggravate the development of VILI. The higher VT supply may be explained by the fact that the inspiratory flow generated in all experimental conditions in the VCV-AF mode was relatively higher than that in the other ventilatory modes. In view of the complexity of studying critically ill patients with ARDS, the present bench study used an experimental model that aggregated variations in respiratory muscle effort patterns, ventilatory modes, and ventilator RR and allowed us to assess their impact on VT and pulmonary pressures during assisted MV in a mechanical ARDS model. Given the difficulties of conducting bedside studies, the ASL 5000™ lung simulator allows the development of studies using a very realistic simulation with good reproducibility and reliability and no risks for patients. The clinical implications of this study include reaffirming the impact of the influence of muscle effort on VT variations and pulmonary pressures in ARDS patients; highlighting the importance of adjusting the ventilator’s RR, which is often neglected in clinical practice, thereby leading to patient-ventilator asynchrony; and emphasizing the importance of a careful choice of ventilatory mode and its management. This study has some limitations. It used a mechanical model of the respiratory system. Bench conditions are not equivalent to patients whose efforts, pulmonary compliance, and respiratory system resistance can be highly variable, and the model had fixed inspiratory and expiratory Pmus and RR, i.e., the mechanical model did not react to a ventilatory demand, which prevented us from assessing patients’ physiological response to metabolic demands. Therefore, the results need to be confirmed in patients. Other limitations of the study are that patient-ventilator asynchronies were not evaluated, and compliance of the ventilator circuits was not measured, which may justify the VT difference between them.

CONCLUSION

Adjusting respiratory muscle effort and pulmonary ventilator respiratory rate to a value above the patient’s respiratory rate in assisted/controlled modes generated large variations in tidal volume and pulmonary pressures, while the controlled mode showed no variations in these outcomes. On the other hand, the presence of expiratory muscle effort had a limiting effect on tidal volume and prevented the value from exceeding 6mL/kg. The pulmonary ventilator model influences ventilation even when similarly adjusted, which reinforces the need to standardize the pulmonary ventilator model in multicenter studies. The volume-controlled ventilation, volume-controlled ventilation with the AutoFlow® system, and pressure-controlled ventilation modes showed similar ventilation behavior. However, tidal volume and pulmonary pressures were slightly higher in the pressure-controlled ventilation and volume-controlled ventilation with the AutoFlow® system modes, thus suggesting that these modes require greater careful management during the use of protective mechanical ventilation with low tidal volume regulation.
Table 2

Alveolar pressure at end-inspiration in the volume-controlled ventilation and pressure-controlled ventilation modes for all the ventilators and simulated scenarios

VCV mode Ventilators ICUPalvPCV modePalv
Patient effortNo effort*Pmus-5cmH2OPmus-5cmH2O/+5cmH2ONo effort*Pmus-5cmH2OPmus-5cmH2O/+5cmH2O
Ventilator RRzero15251525zero15251525
SERVO IMaximum26.227.327.431.332.7SERVO I26.427.327.231.331
Median26.227.326.431.325.526.427.326.431.326.3
Minimum26.227.32231.322.926.427.324.831.324.2
ESPRITMaximum24.224.725.929.730ESPRIT24.126.725.932.131.6
Median24.224.724.929.724.624.126.724.332.125.9
Minimum24.224.721.329.72324.126.722.532.124.3
DX 3012Maximum25.225.522.23030.6DX 301226.827.427.431.731.8
Median25.225.526.13024.526.827.426.731.726.6
Minimum25.225.526.93021.926.827.424.631.723.7
PB 840Maximum24.925.126.130.230.5PB 84024.924.924.93029.5
Median24.925.124.930.223.524.924.9243025.4
Minimum24.925.120.530.220.824.924.921.83023.3
SAVINAMaximum2527263030SAVINA2626263030
Median25272530252626253025
Minimum252723302326223022
SAVINA AFMaximum2828303332-----
Median2828283327
Minimum2828263326

Absence of muscle effort: Pmus = zero. Inspiratory effort: Pmus =-5cmH2O, Expiratory effort: Pmus = +5cmH2O. The spontaneous respiratory rate of the patient was set at 20bpm.

Table 3

Alveolar pressure at end-expiration or effective positive end-expiratory pressure in the volume-controlled ventilation and pressure-controlled ventilation modes for all ventilators and simulated scenarios

VCV mode ventilators ICUAlveolar pressure at end-expiration or PEEPePCV modeAlveolar pressure at end-expiration or PEEPe
Patient effortNo effort*Pmus-5cmH2OPmus-5cmH2O/+5cmH2ONo effort*Pmus - 5cmH2OPmus-5cmH2O/+5cmH2O
Ventilator RRzero15251525zero15251525
SERVO IMaximum10.110.210.410.310.2SERVO I10.110.310.610.210.4
Median10.110.210.410.38.310.110.310.410.28.4
Minimum10.110.27.410.37.210.110.37.710.27.2
ESPRITMaximum10.410.712.310.615.8ESPRIT10.410.512.610.715.8
Median10.410.71110.610.310.410.511.110.710.4
Minimum10.410.77.210.6810.410.57.410.77.4
DX 3012Maximum10.410.612.110.714.8DX 301210.410.712.110.712.8
Median10.410.610.810.710.110.410.711.210.710.2
Minimum10.410.67.610.77.610.410.78.210.78
PB 840Maximum109.910.19.99.9PB 840101010.19.910
Median109.99.99.98.810109.99.98.5
Minimum109.98.49.98.610108.19.97.8
SAVINAMaximum10.410.81210.712.7SAVINA10.710.61210.712.7
Median10.410.811.210.71110.710.611.110.711
Minimum10.410.810.910.79.710.710.610.910.79.5
SAVINA AFMaximum10.710.612.810.612.6-----
Median10.710.611.310.610.8
Minimum10.710.611.210.69.8

Absence of muscle effort: Pmus = zero. Inspiratory effort: Pmus =-5cmH2O, Expiratory effort: Pmus = +5cmH2O. The spontaneous respiratory rate of the patient was set at 20bpm.

  37 in total

1.  Epidemiology, Patterns of Care, and Mortality for Patients With Acute Respiratory Distress Syndrome in Intensive Care Units in 50 Countries.

Authors:  Giacomo Bellani; John G Laffey; Tài Pham; Eddy Fan; Laurent Brochard; Andres Esteban; Luciano Gattinoni; Frank van Haren; Anders Larsson; Daniel F McAuley; Marco Ranieri; Gordon Rubenfeld; B Taylor Thompson; Hermann Wrigge; Arthur S Slutsky; Antonio Pesenti
Journal:  JAMA       Date:  2016-02-23       Impact factor: 56.272

Review 2.  Acute Respiratory Distress Syndrome: Advances in Diagnosis and Treatment.

Authors:  Eddy Fan; Daniel Brodie; Arthur S Slutsky
Journal:  JAMA       Date:  2018-02-20       Impact factor: 56.272

3.  Effects of inspiratory rise time on triggering work load during pressure-support ventilation: a lung model study.

Authors:  Shinya Murata; Kazuki Yokoyama; Yushi Sakamoto; Ken Yamashita; Jun Oto; Hideaki Imanaka; Masaji Nishimura
Journal:  Respir Care       Date:  2010-07       Impact factor: 2.258

Review 4.  Higher vs lower positive end-expiratory pressure in patients with acute lung injury and acute respiratory distress syndrome: systematic review and meta-analysis.

Authors:  Matthias Briel; Maureen Meade; Alain Mercat; Roy G Brower; Daniel Talmor; Stephen D Walter; Arthur S Slutsky; Eleanor Pullenayegum; Qi Zhou; Deborah Cook; Laurent Brochard; Jean-Christophe M Richard; Francois Lamontagne; Neera Bhatnagar; Thomas E Stewart; Gordon Guyatt
Journal:  JAMA       Date:  2010-03-03       Impact factor: 56.272

5.  Evolution of Diaphragm Thickness during Mechanical Ventilation. Impact of Inspiratory Effort.

Authors:  Ewan C Goligher; Eddy Fan; Margaret S Herridge; Alistair Murray; Stefannie Vorona; Debbie Brace; Nuttapol Rittayamai; Ashley Lanys; George Tomlinson; Jeffrey M Singh; Steffen-Sebastian Bolz; Gordon D Rubenfeld; Brian P Kavanagh; Laurent J Brochard; Niall D Ferguson
Journal:  Am J Respir Crit Care Med       Date:  2015-11-01       Impact factor: 21.405

6.  Positive end-expiratory pressure setting in adults with acute lung injury and acute respiratory distress syndrome: a randomized controlled trial.

Authors:  Alain Mercat; Jean-Christophe M Richard; Bruno Vielle; Samir Jaber; David Osman; Jean-Luc Diehl; Jean-Yves Lefrant; Gwenaël Prat; Jack Richecoeur; Ania Nieszkowska; Claude Gervais; Jérôme Baudot; Lila Bouadma; Laurent Brochard
Journal:  JAMA       Date:  2008-02-13       Impact factor: 56.272

7.  Higher versus lower positive end-expiratory pressures in patients with the acute respiratory distress syndrome.

Authors:  Roy G Brower; Paul N Lanken; Neil MacIntyre; Michael A Matthay; Alan Morris; Marek Ancukiewicz; David Schoenfeld; B Taylor Thompson
Journal:  N Engl J Med       Date:  2004-07-22       Impact factor: 91.245

8.  Potentially harmful effects of inspiratory synchronization during pressure preset ventilation.

Authors:  J C M Richard; A Lyazidi; E Akoumianaki; S Mortaza; R L Cordioli; J C Lefebvre; N Rey; L Piquilloud; G F Sferrazza Papa; G F Sferrazza-Papa; A Mercat; L Brochard
Journal:  Intensive Care Med       Date:  2013-08-09       Impact factor: 17.440

9.  Variability of Tidal Volume in Patient-Triggered Mechanical Ventilation in ARDS.

Authors:  Sophie Perinel-Ragey; Loredana Baboi; Claude Guérin
Journal:  Respir Care       Date:  2017-08-01       Impact factor: 2.258