| Literature DB >> 35079248 |
Tanvir Pasha1, Hanan Naniparuthayil Hassan2, Nikhila Radhakrishna1, Muhammed Shafeeque N3, Ajin Shaji Varghese4, Lokesh Viswanath1, Nithin Bhaskar Valuvil1, Sathiyan S1,2, Govardhan H B1, Ibrahim Khaleel1, Siddanna Rudrappa Palled1, Naveen Thimmaiah1.
Abstract
PURPOSE: To evaluate feasibility and safety of execution of optimized intra-cavitary brachytherapy (ICBT) plan of first fraction in subsequent fractions in high-volume, low-resource centers.Entities:
Keywords: anatomical variation; cervical cancer; intracavitary brachytherapy; organ at risk
Year: 2021 PMID: 35079248 PMCID: PMC8782065 DOI: 10.5114/jcb.2021.112113
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Comparative plans
Fig. 2Coronal section of CT scan of ICBT application with isodose distribution. Represented isodose levels are 100% dose (orange), 90% (blue), and 80% (green)
Fig. 4Sagittal section of CT scan of ICBT application showing dose in color wash, blue shows prescribed dose 650 cGy
Patients’ profile (n = 30)
| Age (years) | Range | 32-55 |
| Median | 46 | |
| Stage (number of patients) | I | 0 |
| IIA | 3 | |
| IIB | 23 | |
| IIIA | 0 | |
| IIIB | 4 | |
| Histologic type (number of patients) | Squamous cell carcinoma | 30 |
Intra-application variation between plan X and plan B
| Dosimetric parameters | Mean dose ±SD (cGy) | Mean dose ±SD (cGy) | Paired |
|---|---|---|---|
| Left point A | 538.15 ±104.96 | 513.55 ±79.60 | 1 |
| Right point A | 535.30 ±94.32 | 512.55 ±81.69 | 1 |
| D2cc bladder | 553.57 ±111.23 | 517.74 ±88.45 | 0.068 |
| D2cc rectum | 429.55 ±111.03 | 411.61 ±109.56 | 1 |
| D2cc sigmoid | 411.69 ±152.73 | 361.35 ±114.01 | 0.021* |
statistically significant
Fig. 5Box-and-Whisker plot representation of intra-application variation plan X and plan B in sigmoid dose (SI = sigmoid)
Intra-application variation between plan A and plan X
| Dosimetric parameters | Mean dose ±SD (cGy) | Mean dose ±SD (cGy) | Paired |
|---|---|---|---|
| Left point A | 538.15 ±104.96 | 541.80 ±87.13 | 1 |
| Right point A | 535.30 ±94.32 | 542.03 ±90.57 | 1 |
| D2cc bladder | 553.57 ±111.23 | 511.23 ±85.47 | 0.007* |
| D2cc rectum | 429.55 ±111.03 | 424.31 ±81.95 | 1 |
| D2cc sigmoid | 411.69 ±152.73 | 373.31 ±135.80 | 0.224 |
statistically significant
Fig. 6Box-and-Whisker representation of inter-application dose variation of rectum plan Y and plan C (R = rec- tum)
Fig. 7Box-and-Whisker representation of inter-application dose variation of the sigmoid in plan Y and plan C (SI = sigmoid)
Inter-application variation between plan C and plan Y
| Dosimetric parameters | Mean dose ±SD (cGy) | Mean dose ±SD (cGy) | Paired |
|---|---|---|---|
| Left point A | 536.61 ±82.48 | 559.17 ±114.25 | 0.273 |
| Right point A | 541.23 ±82.95 | 556.70 ±114.60 | 0.436 |
| D2cc bladder | 500.00 ±89.40 | 506.64 ±104.50 | 0.678 |
| D2cc rectum | 401.06 ±83.53 | 452.46 ±123.97 | 0.002* |
| D2cc sigmoid | 340.84 ±117.90 | 387.79 ±141.36 | 0.007* |
statistically significant