Emile Talon1,2, Wilhelm Wimmer3,4, Arsany Hakim5, Claus Kiefer5, Manuela Pastore-Wapp5, Lukas Anschuetz1, Georgios Mantokoudis1, Marco D Caversaccio1,2, Franca Wagner5.
Abstract
PURPOSE: The use of magnetic resonance imaging (MRI) is often limited in patients with auditory implants because of the presence of metallic components and magnets. The aim of this study was to evaluate the clinical usefulness of a customized MRI sequence for metal artifact suppression in patients with BONEBRIDGETM BCI 602 implants (MED-EL, Innsbruck, Austria), the successor of the BCI 601 model.Entities:
Keywords: Artifact reduction; BONEBRIDGE; Magnetic resonance imaging; SEMAC-VAT WARP; WARP
Mesh:
Substances:
Year: 2022 PMID: 35072767 PMCID: PMC9474350 DOI: 10.1007/s00405-022-07272-3
Source DB: PubMed Journal: Eur Arch Otorhinolaryngol ISSN: 0937-4477 Impact factor: 3.236
Fig. 1BONEBRIDGE™ BCI 601 (top) and BCI 602 (bottom) (MED-EL, Innsbruck, Austria) implants with the required dimensions for the bone bed to accommodate the BC FMT transducers. Images courtesy of MED-EL
Fig. 2Retrosigmoid and mastoid implant positions used in the study
Fig. 5MR images from one of the two head samples. All MR images shown here are taken from an axial T1 sequence. a Pre-implantation scan used as a reference. b Post-implantation scan without metal artifact reduction sequences. c Post-implantation scan with SEMAC-VAT WARP artifact reduction sequence: implant placed on the mastoid. d Post-implantation scan with SEMAC-VAT WARP artifact reduction sequence: implant placed in retrosigmoid location
Percentages of artifact volumes with and without artifact reduction with respect to the total MRI volume for the two specimens and implant locations
| Without artifact reduction (%) | With artifact reduction: retrosigmoid location (%) | With artifact reduction: mastoid location (%) | |
|---|---|---|---|
| Head 1 | 56 | 27 | 30 |
| Head 2 | 55 | 25 | 26 |
Fig. 3Three-dimensional visualization of the artifact volumes reported in Table 1 for implants in Head 1 in a retrosigmoid (left figure) and mastoid (right figure) position. Green-shaded areas show the artifact extent without the artifact reduction sequence, while purple shading indicates the artifact extent when the SEMAC-VAT WARP sequence is applied
Fig. 4Three-dimensional visualization of simplified artifact dimensions on the registered MRI-CT images. The green sphere represents the artifact extent without the artifact reduction sequence (radius = 9 cm). The purple sphere shows the artifact extent when the SEMAC-VAT WARP sequence is applied (radius = 7 cm). The spheres are centered around the implant (shown in red) placed in the mastoid position
Differences in diagnostic usefulness between MRI scans with and without SEMAC-VAT WARP in the presence of the BB BCI 602 implant
| Implant position | Significance | |||
|---|---|---|---|---|
| Ipsilateral side | Retrosigmoid | − 3.21 | 0.001 | *** |
| Mastoid | − 3.02 | 0.002 | *** | |
| Contralateral side | Retrosigmoid | − 2.92 | 0.004 | *** |
| Mastoid | − 2.60 | 0.009 | *** |
Ipsi- and contralateral sides are compared, together with the two possible implant locations
***Statistically significant difference