| Literature DB >> 35057298 |
Paweł Kościelniak1, Marek Dębosz1, Marcin Wieczorek1, Jan Migdalski2, Monika Szufla3, Dariusz Matoga3, Jolanta Kochana1.
Abstract
A solid-contact ion-selective electrode was developed for detecting potassium in environmental water. Two versions of a stable cadmium acylhydrazone-based metal organic framework, i.e., JUK-13 and JUK-13_H2O, were used for the construction of the mediation layer. The potentiometric and electrochemical characterizations of the proposed electrodes were carried out. The implementation of the JUK-13_H2O interlayer is shown to improve the potentiometric response and stability of measured potential. The electrode exhibits a good Nernstian slope (56.30 mV/decade) in the concentration range from 10-5 to 10-1 mol L-1 with a detection limit of 2.1 µmol L-1. The long-term potential stability shows a small drift of 0.32 mV h-1 over 67 h. The electrode displays a good selectivity comparable to ion-selective electrodes with the same membrane. The K-JUK-13_H2O-ISE was successfully applied for the determination of potassium in three certified reference materials of environmental water with great precision (RSD < 3.00%) and accuracy (RE < 3.00%).Entities:
Keywords: ion-selective electrode; metal organic frameworks; potassium; potentiometry; water analysis
Year: 2022 PMID: 35057298 PMCID: PMC8781722 DOI: 10.3390/ma15020579
Source DB: PubMed Journal: Materials (Basel) ISSN: 1996-1944 Impact factor: 3.623
Figure 1Calibration graphs for the constructed electrodes with MOFs as solid contact and CW electrode in the main ion concentration range 10−5–10−1 mol L−1.
Figure 2The change in potential of the tested electrodes with MOFs as ion-to-electron transducers and a coated wire electrode over time. The measurements were carried out in 10−2 mol L−1 KCl.
Long-term potential stability of the studied solid contact and coated wire electrodes.
| Time (h) | Potential (mV) | ||
|---|---|---|---|
| K-JUK-13_H2O-ISE | K-JUK-13-ISE | CWE | |
| 0.00 | 361.58 | 348.27 | 308.28 |
| 67.00 | 339.90 | 324.70 | 274.21 |
| Drift (mV h−1) | 0.32 | 0.35 | 0.51 |
Figure 3Determination of the limit of detection for K-MOF-ISE and CWE.
Comparison of the potentiometric selectivity coefficients for proposed K-JUK-13_H2O-ISE, coated wire electrode (CWE) and K-OD-MWCNTs-ISE (with the same ion-selective membrane) (n = 3, α = 0.05).
| Interferent | KK.Jpot | ||
|---|---|---|---|
| K-JUK-13_H2O-ISE | CWE | K-OD-MWCNTs-ISE | |
| Li+ | −3.90 ± 0.09 | −4.19 ± 0.43 | −3.69 ± 0.66 |
| Ca2+ | −3.84 ± 0.46 | −3.71 ± 0.41 | −4.18 ± 0.30 |
| Mg2+ | −4.99 ± 0.20 | −4.97 ± 0.22 | −3.99 ± 0.76 |
| NH4+ | −2.21 ± 0.42 | −2.61 ± 1.48 | −1.64 ± 0.71 |
| Na+ | −4.23 ± 0.13 | −4.43 ± 0.01 | −3.89 ± 0.27 |
| H+ | −4.00 ± 0.13 | −4.02 ± 0.01 | −4.73 ± 1.91 |
Figure 4Water layer test of K-JUK-13_H2O-ISE and CWE.
Figure 5Impedance spectra of the studied solid-contact potassium-selective electrode with JUK-13_H2O as an ion-to-electron transducer.
Results of potassium determination in certified reference materials using the calibration curve method (Co and Cx—certified for CRM and found concentrations respectively; RE—relative error; RSD—relative standard deviation) (n = 6, α = 0.05).
| Sample | C0 | Cx | RE (%) | RSD (%) |
|---|---|---|---|---|
| Waste Water EU-H-3 | 1.02 | 1.05 ± 0.02 | 2.9 | 2.4 |
| Drinking Water EP-H-1 | 0.33 | 0.33 ± 0.01 | 1.9 | 1.7 |
| Ground Water ES-H | 0.074 | 0.073 ± 0.002 | −1.4 | 2.8 |