High-resolution quantitative MRI of multiple sclerosis spinal cord lesions.

Amy R McDowell1, Natalia Petrova2, Daniele Carassiti2, Marc E Miquel3, David L Thomas4, Gareth J Barker5, Klaus Schmierer2,6, Tobias C Wood5.   

Abstract

PURPOSE: Validation of quantitative MR measures for myelin imaging in the postmortem multiple sclerosis spinal cord.
METHODS: Four fixed spinal cord samples were imaged first with a 3T clinical MR scanner to identify areas of interest for scanning, and then with a 7T small bore scanner using a multicomponent-driven equilibrium single-pulse observation of T1 and T2 protocol to produce apparent proton density, T1 , T2 , myelin water, intracellular water, and free-water fraction maps. After imaging, the cords were sectioned and stained with histological markers (hematoxylin and eosin, myelin basic protein, and neurofilament protein), which were quantitatively compared with the MR maps.
RESULTS: Excellent correspondence was found between high-resolution MR parameter maps and histology, particularly for apparent proton density MRI and myelin basic protein staining.
CONCLUSION: High-resolution quantitative MRI of the spinal cord provides biologically meaningful measures, and could be beneficial to diagnose and track multiple sclerosis lesions in the spinal cord.
© 2022 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  multiple sclerosis; myelin water fraction; spinal cord

Mesh:

Substances:

Year:  2022        PMID: 35014736      PMCID: PMC9208576          DOI: 10.1002/mrm.29152

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   3.737


INTRODUCTION

The myelin sheath is an extended and modified plasma membrane wrapped around axons in a spiral fashion. Myelin is composed of about 40% water, whereas the dry mass consists of about 80% lipids, primarily the glycolipid galactocerebroside, and 20% protein. Due to its importance for normal function of the nervous system, and severe consequences of myelin damage in demyelinating diseases (such as multiple sclerosis [MS]), the noninvasive in vivo assessment of myelin remains a key focus of imaging research, not least in the era of regenerative medicine. The spinal cord is most commonly affected in people with MS at presentation. , Lesions in the spinal cord have been shown to predict both a diagnosis of MS in patients with radiologically isolated, as well as clinically isolated, syndromes of demyelination , ; they are also associated with more severe disability. , As MS evolves over time, much of the permanent and deteriorating disability in people with MS affects lower body functions, including lack of sphincter control, sexual dysfunction, and impaired leg movement and coordination, resembling the clinical syndrome of progressive myelopathy. , , , Metrics such as spinal cord cross‐sectional area have been shown to be associated with MS‐related disability in vivo. However, the cross‐sectional area is affected by numerous elements of MS pathology including inflammation, edema, myelin content, the number of axons and gliosis. Magnetic resonance imaging is used widely in the assessment of MS, but standard qualitative T1‐weighted and T2‐weighted images are largely nonspecific for the underlying pathology. Quantitative MRI has the potential to distinguish specific tissue features, such as separating demyelination from axonal loss. Multicomponent relaxometry is a family of methods that can quantify the fractional amount of water in different tissue environments by exploiting the differences in relaxation time values. In the nervous system, these environments include free water in CSF, intracellular and extracellular water (IEW), and, most importantly in the context of demyelinating diseases, water “trapped” between the myelin bilayers. The ratio of myelin water signal to total water signal is termed myelin water fraction (MWF). Along with other related quantitative measures, including the relaxation times themselves, MWF has shown promise to identify specific features of MS pathology and monitoring disease progression. , The gold‐standard method for evaluating the MWF is analysis of a multi‐echo T2 relaxation curve. The multi‐echo T2 method has been previously compared to histology in post mortem brain at 1.5 T, and Laule et al demonstrated a strong correlation of MWF in MS tissue samples at 7 T with luxol fast blue staining intensity for myelin on histology. These studies support the hypothesis that myelin can be mapped in the brain using MWF. However, clinical translation of multi‐echo T2 has been hampered by the long acquisition time required. In contrast, gradient echo–based methods, such as the driven equilibrium single‐pulse observation of T1 and T2, are inherently fast. A multi‐component version of driven equilibrium steady‐state observation of T1 and T2 (mcDESPOT) has been used in clinical studies of MS, primary lateral sclerosis, and monitoring myelin development in children. , , , , However, histological validation has been lacking, and recent work has shown that mcDESPOT suffers from unpredictable bias, but as there was sufficient data for analysis, we present it here. , Given that significant efforts in MS research are directed toward development of strategies for myelin repair (remyelination), imaging biomarkers of myelin are desirable. To establish whether mcDESPOT could provide such an index (or indices), we undertook correlative experiments using postmortem spinal cord, to quantitatively compare MRI and histology data.

METHODS

This study was covered by the UK Multiple Sclerosis Tissue Bank approval (Research Ethics Committee reference number 08/MRE09/31). Formalin‐fixed spinal cords were placed on a Perspex frame and inserted into glass tubes. The formalin was drained and the tubes then filled with perfluoropolyether (Fomblin; Solvay Solutions, United Kingdom) to provide a susceptibility‐matched signal‐free background (Supporting Information Figure S1). Qualitative structural MRI was performed on the whole cords at 3 T (Achieva Tx, Philips Healthcare, Best, Netherlands) using a 15‐element SENSE spine coil and a turbo spin‐echo sequence with the following parameters: TE = 16 ms, TR = 5000 ms, voxel size = 0.4 × 0.4 × 2.0 mm3). Nine areas of interest were identified by a neurologist with > 20 years of experience in reading MRI of people with MS (KS) for scanning at 7 T. The thoracic level 1 was marked with a radiologically opaque marker tied to the nerve root; measurements of distance from the marker for each lesion were noted and used to inform positioning of the glass tube for subsequent 7T scanning. Next, the cords were scanned using a 7T preclinical scanner (Agilent Technologies, Santa Clara, CA) and 39‐mm diameter quadrature transmit‐receive RF coil (Rapid Biomedical, Rimpar, Germany) designed for whole‐body rodent imaging. As the human spine is longer than the achievable FOV with this setup, for each area of interest the glass tubes were moved in or out of the RF coil as required to place the desired area at both the center of the RF coil and the isocenter of the magnet. We observed that the active area of the RF coil was longer than the linear region of the image encoding gradients. As the samples were long enough to protrude into this region, protons that were excited by the RF but not encoded by the gradients resulted in a “zipper artifact” in some images. This was minimized by careful rotation of the samples so that the artifacts did not intrude into areas required for analysis. In most samples, a small quantity of formalin remained after draining, which produced bright signal at the top of the sample tube. To avoid this wrapping into the tissue, the FOV was increased in the vertical direction. Spoiled gradient‐echo and balanced SSFP (with two RF phase cycles) sequences were acquired with the parameters given in Table 1, along with an actual flip‐angle acquisition for B1 correction. Images were resampled to 150 × 150 × 300 μm3 for processing. The images were processed using the open‐source Quantitative Imaging Tools (QUIT) C++ toolbox (available from http://github.com/spinicist/QUIT). T1 maps were generated using nonlinear least‐squares fitting to images with all flip angles, , and T2 and apparent proton density (aPD) maps were generated using the driven equilibrium single pulse observation of T(2) with full modelling method.
TABLE 1

Magnetic resonance imaging sequence acquisition parameters

SequenceTE/TR (ms)Flip anglesVoxel size (μm3)Matrix size
SPGR11.3/256, 8, 10, 15, 20, 25, 35, 45100 × 125 × 200160 × 160 × 160
bSSFP2/414, 17, 20, 25, 35, 50, 65100 × 125 × 200160 × 160 × 160
AFI2.62/7.555250 × 250 × 50064 × 64 × 64

Abbreviations: AFI, actual flip angle; bSSFP, balanced SSFP; SPGR, spoiled gradient echo.

Magnetic resonance imaging sequence acquisition parameters Abbreviations: AFI, actual flip angle; bSSFP, balanced SSFP; SPGR, spoiled gradient echo. Calculation of MWF from mcDESPOT requires the a priori specification of valid initial ranges of T1 and T2 for each water component. , It has been recognized that in postmortem tissue these relaxation times change following fixation. , , The single‐component T1 and T2 of white matter (WM) in a healthy cord was used as a reference point for the myelin water T1 and T2; healthy gray‐matter (GM) relaxation times were used as reference points for the intracellular/extracellular water T1 and T2; and literature values were used for the free‐water fraction. Fitting parameters including the ranges used for each water component are given in Table 2. Non‐overlapping ranges that encompassed the reference points or literature values for these parameters were used. It has been shown that neglecting the effect of TE in the spoiled gradient‐echo and balanced SSFP signal models lead to large errors in derived MWF values, so the TE‐corrected signal model was used.
TABLE 2

Minimum and maximum values that were set for T1 and T2 of each component during mcDESPOT fitting

MWFIEWFFWF
T1 (ms)50–350400–8001000–5000
T2 (ms)8–1414–25200–1500
Fraction constraints (%)0.1–50NA0.1–99
Myelin residence time (ms)10–250NANA

The expected T1 and T2 of MWF and IEWF were set significantly lower than for an equivalent in vivo experiment because formalin fixation is known to shorten these values. Fraction constraints constrain the MWF and FWF to within expected values to prevent convergence to an erroneous local minima. Myelin residence time is the mean time that a water molecule stays in the myelin pool before exchanging to the intracellular/extracellular water pool.

Abbreviation: mcDESPOT, multi‐component version of driven equilibrium steady‐state observation of T1 and T2.

Minimum and maximum values that were set for T1 and T2 of each component during mcDESPOT fitting The expected T1 and T2 of MWF and IEWF were set significantly lower than for an equivalent in vivo experiment because formalin fixation is known to shorten these values. Fraction constraints constrain the MWF and FWF to within expected values to prevent convergence to an erroneous local minima. Myelin residence time is the mean time that a water molecule stays in the myelin pool before exchanging to the intracellular/extracellular water pool. Abbreviation: mcDESPOT, multi‐component version of driven equilibrium steady‐state observation of T1 and T2. The MR parameter maps were then matched with histology using positional information (cord level and distance of the lesion from the marker) from the 3T whole‐cord scans, and checked against anatomical features (such as lesion shape and gray‐matter anatomy) in the histology and 7T images. MIPAV (http://mipav.cit.nih.gov/) was used to manually define regions of interest (ROIs) on aPD maps and transfer them to all other maps for ROI analysis. For histology, cords were dissected axially into blocks of 5‐mm thickness. Each tissue block was then marked with tissue dye to retain information on its spatial orientation and processed for embedding in paraffin. Ten‐micromillimeter‐thick sections were cut from each tissue block using a Shandon Finesse ME1 microtome (Thermo Scientific, United Kingdom). Care was taken to cut the sections in a plane perpendicular to the anterior spinal artery. The sections were mounted on Superfrost1 slides (VWR, United Kingdom) and left in a 608°C oven overnight. Serial sections of each block were stained for hematoxylin and eosin, phosphorylated neurofilaments (SMI‐31, mouse monoclonal, 1:1000; Abcam, United Kingdom) and myelin basic protein (MBP; SMI‐94, mouse monoclonal, 1:100; Covance, Princeton, NJ) following a modified protocol. , , Focal areas of complete myelin loss were identified on MBP immuno‐stained sections as demyelinated lesions in the white matter (WML) and gray matter (GML). Cellularity and axonal counts were determined by counting the number of cells in four square ROIs (size: 120 × 120 μm2) cast randomly onto WML and nonlesional white matter (NLWM) on hematoxylin and eosin and SMI‐31 stained sections, respectively (Supporting Information Figure S2). Cellularity and axonal counts were expressed as cells per cubic millimeter and axons per cubic millimeter, respectively. Unpaired two‐tailed t‐tests were performed to compare MWF, intracellular/extracellular water fraction (IEWF), and free water fraction (FWF) values in the NLWM and WML ROIs.

RESULTS

Formalin‐fixed spinal cords of 3 people with MS (1 male, 2 females), and 1 male control subject with no neurological disease, were used. The MS cords were from donors who died at the age of 67–87 years. Their disease duration had been 8–44 years. The MS cords had been fixed for 1127–1441 days. The control subject died at the age of 89 years. Duration of fixation of this cord was 1366 days. Figure 1 shows matched histology and quantitative MR parameter maps of spinal cords from the 3 MS patients and 1 control subject. Excellent visual correspondence was observed between the anatomic detail on histology images and aPD maps. Both T1 and T2 maps showed less contrast between lesion and healthy tissue than aPD, whereas the MWF produced high contrast, matching well with the myelin histology in the cords from people with MS. The IEWF map showed, as expected, the inverse of the MWF map (lower in myelinated areas and higher elsewhere). Surprisingly, the FWF map was nonzero in healthy tissue area, especially in the control cords, but decreased to zero in lesion areas. Cord MS454 had no gray‐matter areas that could definitely be determined as normal‐appearing gray matter. Myelin basic protein staining, aPD, MWF, IEWF, and FWF images only are shown together in Supporting Information Figure S3 to facilitate comparison of the maps.
FIGURE 1

Histology and MR maps of three multiple sclerosis (MS) spinal cords, from top to bottom; sections taken from spinal cord 1 at levels C8, T3, T5 and T11, spinal cord 2 at T11, and spinal cord 3 at two positions in C7. Control cord sections at C8 and T9 are shown on the bottom two rows. The histological slides stained with myelin basic protein (MBP), hematoxylin and eosin (H&E), and SMI at the level of each lesion found are shown next to apparent proton density (aPD), T1, T2, myelin water fraction (MWF), intracellular/extracellular water fraction (IEWF) and free water fraction (FWF) maps at the same level. From left to right: MBP, H&E, and SMI stained sections, aPD, T1, T2, MWF, IEWF, and FWF. All images were scaled between identical limits for each contrast so a comparison could be made (aPD 0–5.0 × 105 PU; T1 0–800 ms; T2 10–20 ms; MWF 0%–0.5%; IEWF 0%–1%; and FWF 0%–0.05%). Cntrl, control; PD, proton density

Histology and MR maps of three multiple sclerosis (MS) spinal cords, from top to bottom; sections taken from spinal cord 1 at levels C8, T3, T5 and T11, spinal cord 2 at T11, and spinal cord 3 at two positions in C7. Control cord sections at C8 and T9 are shown on the bottom two rows. The histological slides stained with myelin basic protein (MBP), hematoxylin and eosin (H&E), and SMI at the level of each lesion found are shown next to apparent proton density (aPD), T1, T2, myelin water fraction (MWF), intracellular/extracellular water fraction (IEWF) and free water fraction (FWF) maps at the same level. From left to right: MBP, H&E, and SMI stained sections, aPD, T1, T2, MWF, IEWF, and FWF. All images were scaled between identical limits for each contrast so a comparison could be made (aPD 0–5.0 × 105 PU; T1 0–800 ms; T2 10–20 ms; MWF 0%–0.5%; IEWF 0%–1%; and FWF 0%–0.05%). Cntrl, control; PD, proton density Quantitative histology results and mean values of the MR parameters from the ROIs are given in Table 3. Cellularity and axon count and MBP staining fraction were reduced in lesions compared with normal WM; MWF also showed a decrease in lesion areas, whereas the IEWF increased. We observed a noticeable decrease in the FWF fraction. It should be noted that the MBP staining fraction can only be compared within a single batch; hence, there is variable staining between the pathological cords and the control cord, which was stained at a later date. Therefore, the MBP values were not entered into the correlation analysis for the control cord though it is still included in the table for completeness.
TABLE 3

Histology and MR indices for each region of interest

SampleMBP stain %Nuclei/mm2 Axons/mm2 aPD (arb units)T1 (ms)T2 (ms)MWF %IEWF %FWF %
WMLNLWMWMLNLWMWMLNLWMWMLNLWMWMLNLWMWMLNLWMWMLNLWMWMLNLWMWMLNLWM
MS1 C81.7883.9217361858796920 1746.0+05 4.5+05 708.3445.321.4163.618.596.280.40.21.2
MS1 T35.1778.5614582546876714 6014.9+05 3.4+05 518.9390.617.414.714.128.185.369.70.62.3
MS1 T53.3587.8618921944232611 7826.1+05 4.1+05 604.3407.219.815.48.723.39174.40.32.3
MS1 T111.988.3127602778951415 1225.8+05 4.2+05 590.9394.919.515.2924.990.876.70.24.8
MS2 T112.8963.421702135793412 6045.6+05 4.2+05 578.2517.417.417.51314.8868312.2
MS3 C07A15.7161.4612151615828118 8435.4+05 4.2+05 663.6449.718.415.45.52393.875.70.71.3
MS3 C07B0.5943.3912331389763915 6255.1+05 5.0+05 634.7478.919.115.5718.292.679.80.42
Control C7NA73.91NA27 018NA15 634NA4.5+05 NA403.5NA13.9NA29.8NA67.1NA3.1
Control T8NA44.47NA27 456NA16 571NA4.4+05 NA410.4NA14.3NA29.1NA68.5NA2.5

There are no lesions in the control cords. Hence, these entries are marked not applicable (NA). All three histology measures were reduced in lesions compared with normal white matter.

Histology and MR indices for each region of interest There are no lesions in the control cords. Hence, these entries are marked not applicable (NA). All three histology measures were reduced in lesions compared with normal white matter. Graphs of the MWF, IEWF, and FWF values in the NLWM, NLGM, and WM and GM lesion are shown in Supporting Information Figure S4. MWF, IEWF, and FWF were all significantly different between NLWM and WML (p < 0.0001). Graphs of correlation (Pearson’s correlation) between histological and MR indices over all WML and NLWM in MS cords, and control WM in control cords (axon counts vs aPD, T1, MWF, and MBP stain vs T1, MWF) are shown in Figure 2. All are significant (Pearson’s R and p‐values found in Supporting Information Table S1). Correlations between MWF and aPD in individual WM types (WML, NLWM, and control WM) and histological indices are shown in Supporting Information Figures S5 and S6, respectively. Correlation between both cell nuclei and axon counts and MWF in WML were not significant. Correlation between nuclei counts and MWF in NLWM and control WM was just significant (p = .046), but not significant between axons and MWF in NLWM and control WM. This significance is likely due to the difference in nuclei between control data and NLWM rather than a correlation with MWF.
FIGURE 2

Correlation between histological and MR indices over all white‐matter lesion (WML) and nonlesional white matter (NLWM) in MS cords and control white matter in control cords. Correlation of axons against aPD, T1, T2, and MWF (A–D, respectively) and MBP against aPD, T1, T2, and MWF (E–H, respectively). Control cord white‐matter data are shown by the red data points, NLWM by black dots, and WML by green dots. The best fit line is shown with 95% confidence intervals. R, Person’s R value

Correlation between histological and MR indices over all white‐matter lesion (WML) and nonlesional white matter (NLWM) in MS cords and control white matter in control cords. Correlation of axons against aPD, T1, T2, and MWF (A–D, respectively) and MBP against aPD, T1, T2, and MWF (E–H, respectively). Control cord white‐matter data are shown by the red data points, NLWM by black dots, and WML by green dots. The best fit line is shown with 95% confidence intervals. R, Person’s R value

DISCUSSION

Our results show that with high‐resolution scanning at 7 T, good correspondence between quantitative MRI and histological metrics is observed. Apparent proton density maps show excellent contrast between healthy and pathological tissue, and T1 and T2 maps provide quantitative measures that correlate with quantitative histology measures. The MWF maps show clear differentiation between lesion and nonlesional tissue. The nonzero FWF in the healthy control cord is an unexpected finding, as fixed tissue has not been believed to contain water protons with a long T2. This may indicate infiltration by the fixative or may simply be due to bias in the fitting procedure. The FWF component in the control cords displays better contrast between WM and GM than the MWF, for instance, which could be caused by either of these mechanisms. The measured FWF is lower in lesions in pathological cords, which may also be due to said bias or could indicate the assumption that the FWF is nonexchanging (i.e., that water cannot move between the FWF and intracellular/extracellular water pools) is invalid. In lesions it is possible that biological barriers between these pools break down such that they behave as one, with a common T1 and T2. Our samples were not rinsed of fixative before acquisition, as this could cause degradation of the tissue, and therefore contained formaldehyde. Formaldehyde causes known shortening of the T1 and T2 values, reducing the differences between the myelin and intracellular/extracellular compartments. The stochastic region contraction search algorithm used in mcDESPOT may not be able to consistently differentiate the signal from the different water pools in this situation. Care was taken to choose appropriate fitting ranges (Table 2), but derived MWF values are highly dependent on the selected parameter ranges, especially when the stochastic region contraction algorithm is used for fitting. , An alternate approach such as the Bayesian priors of Bouhrara et al may improve fitting accuracy. Unfixed tissue is challenging to scan in a suitable time frame after death in humans and may show decomposition over a long scan time; however, if possible, further validation in unfixed tissue would be desirable, as the water pools should exhibit similar separation as in vivo tissue. The IEWF maps also show good definition between tissue types, with high values in lesional tissue. This implies more space in larger cells and between the cells in these areas. This has been postulated previously by Christiansen et al, who found that the apparent diffusion coefficient increased in MS lesions and suggested that this may be related to an increase in the extracellular space due to oedema and demyelination. They hypothesized the myelin removed is replaced by water, resulting in an increase of the extracellular space; thereby, the formerly restricted diffusion becomes less restricted. Despite the issues related to fitting the mcDESPOT model, the MWF and IEWF showed good contrast and anatomic correspondence to the histology images, whereas the single‐component T1 and T2 maps showed comparatively low contrast. However, the aPD maps also showed excellent correspondence with histology, with increased values in lesion areas. It should be noted that because MWF and IEWF are defined as fractions, the aPD is effectively removed from the mcDESPOT measures; hence, the contrast in the aPD maps is orthogonal information to that in the mcDESPOT parameter maps. Our results imply that if, for example, only the MWF is considered, then an important additional source of information (i.e., the total water content in a voxel) will be missed.

CONCLUSIONS

High‐resolution quantitative MR parameter maps offer an alternative modality to histology for ex vivo quantitative assessment of MS spinal cord lesions. The MWF maps showed good contrast between lesional and non‐lesional tissue. Proton density, which reflects total water content, correlated significantly in MS lesions and offers additional information compared with the mcDESPOT parameter maps. The mcDESPOT model may benefit from the use of a Bayesian priors algorithm for more challenging conditions.

CONFLICT OF INTEREST

GJB received honoraria for teaching from GE Healthcare at the time of this study. There are no other known conflicts of interest to disclose. FIGURE S1 Whole spinal cord secured in a glass tube and immersed in perfluoropolyether ready for scanning FIGURE S2 Example hematoxylin and eosin (H&E) stained section (left, A) with four square regions of interest (ROIs) positioned in the red outlined region (top right, B); gray matter is outlined in black. Zoomed section (bottom right, C) shows detail in square ROI. Cellularity and axonal counts were determined by counting the number of cells in four square ROIs (size: 120 × 120 μm2) cast onto lesional and non‐lesional white matter on the H&E and SMI‐31 stained sections FIGURE S3 Expanded from Figure 2, showing only MBP, proton density (PD), myelin water fraction (MWF), intracellular/extracellular water fraction (IEWF), and free water fraction (FWF) to facilitate comparison FIGURE S4 Box and whisker plots of all MR parameters in each tissue type. Red points are control samples, for which there are no lesional data. Differences in all parameters were present between nonlesional white matter (NLWM) and white‐matter lesion (WML). Center line is the mean with whiskers of 1 SD FIGURE S5 Graphs of correlation for MWF separated into individual white‐matter types expanded from Figure 2: WML and MLWM. Correlations with nuclei against MWF in WML and NLWM in multiple sclerosis (MS) cords and control white matter in control cords (A and C, respectively) and axons against MWF in WLM and NLWM in MS cords and control white matter in control cords (B and D, respectively). Control cord white‐matter data are shown by the red data points. The best fit line is shown with the 95% confidence intervals FIGURE S6 Graphs of correlation for proton density separated into individual white‐matter types expanded from Figure 2: WML NLWM. Correlations with nuclei against proton density in WML and NLWM in MS cords and control white matter in control cords (A and C, respectively) and axons against proton density in WML and NLWM in MS cords and control white matter in control cords (B and D, respectively). Control cord white‐matter data are shown by the red data points. The best fit line is shown with the 95% confidence intervals TABLE S1 Correlation coefficients for graphs shown in Figure 2 Click here for additional data file.
  45 in total

1.  Cortical lesions in multiple sclerosis: combined postmortem MR imaging and histopathology.

Authors:  Jeroen J G Geurts; Lars Bö; Petra J W Pouwels; Jonas A Castelijns; Chris H Polman; Frederik Barkhof
Journal:  AJNR Am J Neuroradiol       Date:  2005-03       Impact factor: 3.825

2.  Transverse relaxation time (T2) mapping in the brain with off-resonance correction using phase-cycled steady-state free precession imaging.

Authors:  Sean C L Deoni
Journal:  J Magn Reson Imaging       Date:  2009-08       Impact factor: 4.813

3.  Spinal cord lesions: A modest contributor to diagnosis in clinically isolated syndromes but a relevant prognostic factor.

Authors:  Georgina Arrambide; Alex Rovira; Jaume Sastre-Garriga; Carmen Tur; Joaquín Castilló; Jordi Río; Angela Vidal-Jordana; Ingrid Galán; Breogán Rodríguez-Acevedo; Luciana Midaglia; Carlos Nos; Patricia Mulero; Maria Jesús Arévalo; Manuel Comabella; Elena Huerga; Cristina Auger; Xavier Montalban; Mar Tintore
Journal:  Mult Scler       Date:  2017-03-16       Impact factor: 6.312

Review 4.  Is multiple sclerosis a length-dependent central axonopathy? The case for therapeutic lag and the asynchronous progressive MS hypotheses.

Authors:  Gavin Giovannoni; Gary Cutter; Maria Pia Sormani; Shibeshih Belachew; Robert Hyde; Harold Koendgen; Volker Knappertz; Davorka Tomic; David Leppert; Robert Herndon; Claudia A M Wheeler-Kingshott; Olga Ciccarelli; David Selwood; Elisabetta Verdun di Cantogno; Ali-Frederic Ben-Amor; Paul Matthews; Daniele Carassiti; David Baker; Klaus Schmierer
Journal:  Mult Scler Relat Disord       Date:  2017-01-17       Impact factor: 4.339

5.  Problems and Pitfalls of Identifying Remyelination in Multiple Sclerosis.

Authors:  Bjoern Neumann; Sarah Foerster; Chao Zhao; Benedetta Bodini; Daniel S Reich; Dwight E Bergles; Ragnhildur Thóra Káradóttir; Catherine Lubetzki; Luke L Lairson; Bernard Zalc; Bruno Stankoff; Robin J M Franklin
Journal:  Cell Stem Cell       Date:  2020-05-07       Impact factor: 24.633

Review 6.  Classification, diagnosis, and differential diagnosis of multiple sclerosis.

Authors:  Ilana Katz Sand
Journal:  Curr Opin Neurol       Date:  2015-06       Impact factor: 5.710

7.  Linear least-squares method for unbiased estimation of T1 from SPGR signals.

Authors:  Lin-Ching Chang; Cheng Guan Koay; Peter J Basser; Carlo Pierpaoli
Journal:  Magn Reson Med       Date:  2008-08       Impact factor: 4.668

8.  Improved determination of the myelin water fraction in human brain using magnetic resonance imaging through Bayesian analysis of mcDESPOT.

Authors:  Mustapha Bouhrara; Richard G Spencer
Journal:  Neuroimage       Date:  2015-10-22       Impact factor: 6.556

9.  Axonal loss in the multiple sclerosis spinal cord revisited.

Authors:  Natalia Petrova; Daniele Carassiti; Daniel R Altmann; David Baker; Klaus Schmierer
Journal:  Brain Pathol       Date:  2017-05-07       Impact factor: 6.508

10.  Breastfeeding and early white matter development: A cross-sectional study.

Authors:  Sean C L Deoni; Douglas C Dean; Irene Piryatinsky; Jonathan O'Muircheartaigh; Nicole Waskiewicz; Katie Lehman; Michelle Han; Holly Dirks
Journal:  Neuroimage       Date:  2013-05-28       Impact factor: 6.556

View more
  1 in total

1.  High-resolution quantitative MRI of multiple sclerosis spinal cord lesions.

Authors:  Amy R McDowell; Natalia Petrova; Daniele Carassiti; Marc E Miquel; David L Thomas; Gareth J Barker; Klaus Schmierer; Tobias C Wood
Journal:  Magn Reson Med       Date:  2022-01-11       Impact factor: 3.737

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.