| Literature DB >> 34994101 |
Eun Young Han1, He Wang1, Tina Marie Briere1, Debra Nana Yeboa2, Themistoklis Boursianis3, Georgios Kalaitzakis3, Evangelos Pappas4, Pamela Castillo1, Jinzhong Yang1.
Abstract
Online magnetic resonance (MR)-guided radiotherapy is expected to benefit brain stereotactic radiosurgery (SRS) due to superior soft tissue contrast and capability of daily adaptive planning. The purpose of this study was to investigate daily adaptive plan quality with setup variations and to perform an end-to-end test for brain SRS with multiple metastases treated with a 1.5-Tesla MR-Linac (MRL). The RTsafe PseudoPatient Prime brain phantom was used with a delineation insert that includes two predefined structures mimicking gadolinium contrast-enhanced brain lesions. Daily adaptive plans were generated using six preset and six random setup variations. Two adaptive plans per daily MR image were generated using the adapt-to-position (ATP) and adapt-to-shape (ATS) workflows. An adaptive patient plan was generated on a diagnostic MR image with simulated translational and rotational daily setup variation and was compared with the reference plan. All adaptive plans were compared with the reference plan using the target coverage, Paddick conformity index, gradient index (GI), Brain V12 or V20, optimization time and total monitor units. Target doses were measured as an end-to-end test with two ionization chambers inserted into the phantom. With preset translational variations, V12 from the ATS plan was 17% lower than that of the ATP plan. With a larger daily setup variation, GI and V12 of the ATS plan were 10% and 16% lower than those of the ATP plan, respectively. Compared to the ATP plans, the plan quality index of the ATS plans was more consistent with the reference plan, and within 5% in both phantom and patient plans. The differences between the measured and planned target doses were within 1% for both treatment workflows. Treating brain SRS using an MRL is feasible and could achieve satisfactory dosimetric goals. Setup uncertainties could be accounted for using online plan adaptation. The ATS workflow achieved better dosimetric results than the ATP workflow at the cost of longer optimization time.Entities:
Keywords: MR Linac; adaptive planning; brain metastases; end-to-end test
Mesh:
Year: 2022 PMID: 34994101 PMCID: PMC8906207 DOI: 10.1002/acm2.13518
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
FIGURE 1(a) Sagittal plane of T1w magnetic resonance (MR) images of the Prime head phantom with a delineation insert containing two brain lesions. (b) Sagittal plan of CT images of the same phantom with the chamber insert (one chamber shown at this cross‐section with chamber head contoured). (c) Measurement setup with two chambers
FIGURE 2(a) Isodose lines (in cGy) in the reference plan, and (b) target volumes (green and orange contours) from the reference plan projected onto the 3° pitch‐rotated magnetic resonance imaging (MRI) by adapt‐to‐position (ATP) workflow
FIGURE 3Plan quality index from (a) adapt‐to‐position (ATP) and (b) adapt‐to‐shape (ATS) workflows to daily setup variations of 3‐mm shift in lateral (X_tra), superior–inferior (Y_tra), anterior–posterior (Z_tra), and 3° rotation in pitch (X_rot), roll (Y_rot), and yaw (Z_rot), separately. In (a) the dashed lines represent the nominal ATP plans, and the solid lines represent the delivered ATP plans. GI and V12 are normalized to the values in reference plan
Target coverages (V95%) and GTV mininum doses (Dmin) of ATP plans on head phantom for the setup variations of 3 mm translational shifts in lateral (X_tra), superior–inferior (Y_tra), anterior–posterior (Z_tra) directions, and 3° rotations in pitch (X_rot), roll (Y_rot), and yaw (Z_rot) directions
|
|
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|---|---|
| V95% (%) | Sup PTV (delivered) | 99.7 | 100.0 | 99.8 | 99.9 | 99.6 | 99.5 | 99.7 |
| (100.0) | (99.9) | (99.9) | (96.1) | (99.4) | (99.5) | |||
| Inf PTV (delivered) | 99.4 | 99.9 | 99.8 | 100.0 | 99.3 | 99.3 | 99.4 | |
| (100.0) | (99.9) | (100.0) | (99.2) | (99.6) | (99.3) | |||
|
| Sup GTV (delivered) | 2016.6 | 2004 | 2026.4 | 2025.7 | 2032.8 | 2002.4 | 2017.9 |
| (2010.4) | (2003.5) | (2016.7) |
|
| (2010.3) | |||
| Inf GTV (delivered) | 2000.0 | 2000.0 | 2000.0 | 2000.0 | 2000.0 | 2000.0 | 2000.0 | |
| (2033.1) | (1993.4) | (2043.7) |
|
|
|
Note: Numbers in the parenthesis represent the results for delivered adapt‐to‐position (ATP) plans calculated on daily image.
Shifts and rotations of the head phantom relative to reference magnetic resonance imaging (MRI) obtained from image registrations of six random daily MRI setups
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| 1 | 1.3 | 10.9 | 0.7 | 2.3 | 0.0 | −0.1 | 11.0 |
| 2 | −3.9 | 11.8 | 0.7 | 1.6 | 0.1 | 3.5 | 12.4 |
| 3 | 1.9 | 11.7 | 7.1 | −0.6 | 0.7 | −0.6 | 13.8 |
| 4 | 2.0 | 11.9 | 8.9 | 2.1 | 1.3 | −0.3 | 15.0 |
| 5 | −4.4 | 19.6 | 4.2 | 2.2 | 3.3 | −0.2 | 20.5 |
| 6 | 3.8 | 24.0 | 8.5 | 2.2 | 1.0 | 0.4 | 25.8 |
FIGURE 4Plan quality index from (a) adapt‐to‐position (ATP) and (b) adapt‐to‐shape (ATS) workflows for six random daily setup variations
FIGURE 5A patient magnetic resonance imaging (MRI) and Monaco reference plan. Red colorwash: GTV; blue colorwash: PTV. Isodose lines are in cGy
Adaptive patient plan: Comparison of adapt‐to‐position (ATP) versus adapt‐to‐shape (ATS) workflows to reference and clinical plan
|
|
|
|
| ||
|---|---|---|---|---|---|
| Reference plan | 3.63 | 0.84 | 24.3 | 162.8 | |
| Clinical plan | 2.95 | 0.91 | 30.5 | 133.6 | |
| Adaptive test 1 | ATP | 3.52 | 0.83 | 25.1 | 160.9 |
| ATS | 3.45 | 0.89 | 26.0 | 148.3 | |
| Adaptive test 2 | ATP | 3.6 | 0.83 | 25.0 | 163.6 |
| ATS | 3.48 | 0.89 | 25.6 | 147.8 | |
Clinical plan was generated in RayStation with VMAT technique.
Chamber measurements for daily adapt‐to‐shape (ATS) and adapt‐to‐position (ATP) plans on head phantom
|
|
| |||
|---|---|---|---|---|
|
|
|
|
| |
| Adaptive plan (cGy) | 2201.0 | 2263.0 | 2201.6 | 2205.6 |
| Measurement (cGy) | 2186.5 | 2255.0 | 2195.5 | 2194.0 |
| % Difference |
|
|
|
|