Optimization of dual energy computed tomography post-processing to reduce lower limb artifacts in gout.

Baptiste Dubief1, Julien Avril2, Tristan Pascart3,4, Marie Schmitt1, Romaric Loffroy2, Jean-Francis Maillefert1,5, Paul Ornetti1,5,6, André Ramon1.   

Abstract

BACKGROUND: In gout, several types of dual-energy computed tomography (DECT) artifacts have been described (nail bed, skin, beam hardening, submillimeter and vascular artifacts), which can lead to overdiagnosis. The objective of this study was to determine the optimal DECT settings for post processing in order to reduce the frequency of some common artifacts in patients with suspected gout.
METHODS: Seventy-seven patients hospitalized for suspected gout (feet/ankles and/or knees) who received a DECT imaging were included (final diagnosis of 43 gout and 34 other rheumatic disorders). Different post-processing settings were evaluated using Syngovia software: nine settings (R1 to R9) were evaluated with a combination of different ratio (1.28, 1.36 and 1.55) and attenuation coefficient (120, 150, 170 HU).
RESULTS: Among the nine settings tested, the R2 setting (170 HU, ratio =1.28) significantly reduced the presence of knee and foot/ankle artifacts compared to the standard R1 setting (85% and 94% decrease in beam hardening and clumpy artifacts in the ankle and foot, respectively (P<0.001); a decrease of 71%, 60% and 88% respectively of meniscal beam hardening, beam hardening and submillimeter artifacts in the knee (P<0.001). Compared to standard settings, the use of R2 settings decreased sensitivity [0.79 (95% CI: 0.65, 0.88) versus 0.90 (95% CI: 0.78, 0.96)] and increased specificity [0.86 (95% CI: 0.71, 0.93) versus 0.63 (95% CI: 0.47, 0.77)] (P<0.001). Settings using an attenuation coefficient to 120 HU and/or a ratio to 1.55 were all associated with a significant increasing of artifacts, especially clumpy and beam hardening artifacts.
CONCLUSIONS: Applying a ratio of 1.28 and a minimum attenuation of 170 HU in DECT post-processing eliminates the majority of artifacts located in the lower limbs, particularly clumpy artifacts and beam hardening. 2022 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Entities:  

Keywords:  Gout; artifacts; dual energy computed tomography; post-processing

Year:  2022        PMID: 34993099      PMCID: PMC8666734          DOI: 10.21037/qims-21-321

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  29 in total

1.  Dual energy CT in gout: a prospective validation study.

Authors:  Hyon K Choi; Lindsay C Burns; Kamran Shojania; Nicole Koenig; Graham Reid; Mohammed Abufayyah; Genevieve Law; Alison S Kydd; Hugue Ouellette; Savvas Nicolaou
Journal:  Ann Rheum Dis       Date:  2012-03-02       Impact factor: 19.103

2.  Dual-energy CT as a potential new diagnostic tool in the management of gout in the acute setting.

Authors:  Savvakis Nicolaou; Charlotte Jane Yong-Hing; Sandro Galea-Soler; Daniel J Hou; Luck Louis; Peter Munk
Journal:  AJR Am J Roentgenol       Date:  2010-04       Impact factor: 3.959

Review 3.  Getting the Most From Your Dual-Energy Scanner: Recognizing, Reducing, and Eliminating Artifacts.

Authors:  Tyler M Coupal; Paul I Mallinson; Sharon L Gershony; Patrick D McLaughlin; Peter L Munk; Savvas Nicolaou; Hugue A Ouellette
Journal:  AJR Am J Roentgenol       Date:  2016-01       Impact factor: 3.959

Review 4.  Classifications in Brief: Kellgren-Lawrence Classification of Osteoarthritis.

Authors:  Mark D Kohn; Adam A Sassoon; Navin D Fernando
Journal:  Clin Orthop Relat Res       Date:  2016-02-12       Impact factor: 4.176

5.  A survey of the management of gout in primary care.

Authors:  D Owens; B Whelan; G McCarthy
Journal:  Ir Med J       Date:  2008-05

6.  Combining non-contrast and dual-energy CT improves diagnosis of early gout.

Authors:  Seul Ki Lee; Joon-Yong Jung; Won-Hee Jee; Jennifer Jooha Lee; Sung-Hwan Park
Journal:  Eur Radiol       Date:  2018-09-17       Impact factor: 5.315

7.  Artifacts in dual-energy CT gout protocol: a review of 50 suspected cases with an artifact identification guide.

Authors:  Paul Ian Mallinson; Tyler Coupal; Clemens Reisinger; Hong Chou; Peter Loren Munk; Savvas Nicolaou; Hugue Ouellette
Journal:  AJR Am J Roentgenol       Date:  2014-07       Impact factor: 3.959

8.  The effect of tube voltage combination on image artefact and radiation dose in dual-source dual-energy CT: comparison between conventional 80/140 kV and 80/150 kV plus tin filter for gout protocol.

Authors:  Ji Young Jeon; Sheen-Woo Lee; Yu Mi Jeong; Han Joo Baek
Journal:  Eur Radiol       Date:  2018-07-09       Impact factor: 5.315

9.  Prevalence of Gout in the Adult Population of France.

Authors:  Thomas Bardin; Stéphane Bouée; Pierre Clerson; Gérard Chalès; René-Marc Flipo; Frédéric Lioté; Vincent Perez; Thierry Poiraud; Thierry Schaeverbeke; Pascal Richette
Journal:  Arthritis Care Res (Hoboken)       Date:  2016-02       Impact factor: 4.794

10.  Dual-energy computed tomography compared with ultrasound in the diagnosis of gout.

Authors:  Michael Gruber; Gerd Bodner; Eva Rath; Gabriela Supp; Michael Weber; Claudia Schueller-Weidekamm
Journal:  Rheumatology (Oxford)       Date:  2013-10-17       Impact factor: 7.580

View more
  2 in total

1.  Prevalence of Monosodium Urate (MSU) Deposits in Cadavers Detected by Dual-Energy Computed Tomography (DECT).

Authors:  Andrea S Klauser; Sylvia Strobl; Christoph Schwabl; Werner Klotz; Gudrun Feuchtner; Bernhard Moriggl; Julia Held; Mihra Taljanovic; Jennifer S Weaver; Monique Reijnierse; Elke R Gizewski; Hannes Stofferin
Journal:  Diagnostics (Basel)       Date:  2022-05-16

2.  The Clinical Utility of Dual-Energy Computed Tomography in the Diagnosis of Gout-A Cross-Sectional Study.

Authors:  Maria Sotniczuk; Anna Nowakowska-Płaza; Jakub Wroński; Małgorzata Wisłowska; Iwona Sudoł-Szopińska
Journal:  J Clin Med       Date:  2022-09-05       Impact factor: 4.964

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.