Kosuke Ino1, Hao-Jen Pai2, Kaoru Hiramoto2, Yoshinobu Utagawa2, Yuji Nashimoto1,3, Hitoshi Shiku1. 1. Graduate School of Engineering, Tohoku University, 6-6-11 Aramaki-aza Aoba, Aoba-ku, Sendai 980-8579, Japan. 2. Graduate School of Environmental Studies, Tohoku University, 6-6-11 Aramaki-aza Aoba, Aoba-ku, Sendai 980-8579, Japan. 3. Frontier Research Institute for Interdisciplinary Sciences, Tohoku University, 6-3 Aramaki-aza Aoba, Aoba-ku, Sendai 980-8578, Japan.
Abstract
It is important to clarify the transport of biomolecules and chemicals to tissues. Herein, we present an electrochemical imaging method for evaluating the endothelial permeability. In this method, the diffusion of electrochemical tracers, [Fe(CN)6]4-, through a monolayer of human umbilical vein endothelial cells (HUVECs) was monitored using a large-scale integration-based device containing 400 electrodes. In conventional tracer-based assays, tracers that diffuse through an HUVEC monolayer into another channel are detected. In contrast, the present method does not employ separated channels. In detail, a HUVEC monolayer is immersed in a solution containing [Fe(CN)6]4- on the device. As [Fe(CN)6]4- is oxidized and consumed at the packed electrodes, [Fe(CN)6]4- begins to diffuse through the monolayer from the bulk solution to the electrodes and the obtained currents depend on the endothelial permeability. As a proof-of-concept, the effects of histamine on the monolayer were monitored. Also, an HUVEC monolayer was cocultured with cancer spheroids, and the endothelial permeability was monitored to evaluate the metastasis of the cancer spheroids. Unlike conventional methods, the device can provide spatial information, allowing the interaction between the monolayer and the spheroids to be monitored. The developed method is a promising tool for organs-on-a-chip and drug screening in vitro.
It is important to clarify the transport of biomolecules and chemicals to tissues. Herein, we present an electrochemical imaging method for evaluating the endothelial permeability. In this method, the diffusion of electrochemical tracers, [Fe(CN)6]4-, through a monolayer of human umbilical vein endothelial cells (HUVECs) was monitored using a large-scale integration-based device containing 400 electrodes. In conventional tracer-based assays, tracers that diffuse through an HUVEC monolayer into another channel are detected. In contrast, the present method does not employ separated channels. In detail, a HUVEC monolayer is immersed in a solution containing [Fe(CN)6]4- on the device. As [Fe(CN)6]4- is oxidized and consumed at the packed electrodes, [Fe(CN)6]4- begins to diffuse through the monolayer from the bulk solution to the electrodes and the obtained currents depend on the endothelial permeability. As a proof-of-concept, the effects of histamine on the monolayer were monitored. Also, an HUVEC monolayer was cocultured with cancer spheroids, and the endothelial permeability was monitored to evaluate the metastasis of the cancer spheroids. Unlike conventional methods, the device can provide spatial information, allowing the interaction between the monolayer and the spheroids to be monitored. The developed method is a promising tool for organs-on-a-chip and drug screening in vitro.
Endothelial
and epithelial cell barriers play critical roles in
regulating the transport of biomolecules and chemicals to tissues.
In particular, vascular endothelial (VE) cell barriers are important
because blood vessels are utilized to transport nutrients and drugs.
As endogenous mediators such as histamine increase the endothelial
permeability, the mechanism by which such molecules affect cell barriers
has been investigated.[1] In addition, interactions
between cell barriers and cancers play a crucial role in metastasis.
Therefore, for anticancer drug screening in vitro, it is important
to prepare a model for endothelial permeability. As a cell barrier
model, the transwell system is widely used. In this model, a cell
monolayer is prepared on a porous membrane. To assess the barrier
function, fluorescent tracers are injected on top of the monolayer,
and tracers that diffuse to the bottom are monitored. As an alternative
to this assay, the electric resistance of a monolayer of cells, called
the transepithelial/endothelial electrical resistance (TEER), can
be monitored.[2−4] Several devices have been reported for TEER measurements,
including a piezoelectric biosensor,[5] a
3D tubular vascular channel,[6] and a combination
of multiple electrode arrays for assessing the cell barrier function
and electrical activity.[7] TEER measurements
have been applied for organs-on-a-chip with integrated electrodes[8] and a blood–brain barrier model consisting
of astrocytes and a microvascular endothelial cell monolayer.[9]In addition to these methods using fluorescent
tracers and TEER
measurements, electrochemical tracers have recently been used for
endothelial permeability assays.[10,11] In this approach
using electrochemical tracers, electrodes are placed near an endothelial
monolayer, and electrochemical tracers are loaded on the opposite
side of the monolayer. The tracers that diffuse through the monolayer
to the electrodes are monitored amperometrically, and the current
is used to evaluate the endothelial permeability. This assay is superior
to that using fluorescent tracers because it eliminates the need for
manual sampling and complex optical instrumentation. Although this
electrochemical approach is very attractive for detecting cell barriers,
its applications are limited because information about different local
areas cannot be obtained owing to the use of a single working electrode.
For example, to evaluate interactions between an endothelial monolayer
and cancer spheroids, it is necessary to visualize the local cell
permeability, which is difficult when using a single sensor. In addition,
it is a little complex to separate bottom and top channels. Furthermore,
replacing the initial solution with the assay solution containing
tracers can lead to chaotic mixing and errors. To solve these problems,
we proposed a novel strategy using an electrode array device.Several types of electrode array devices are available for bioanalysis.[12] Large-scale integration (LSI)-based devices
consisting of electrode arrays have been used for various bioanalysis
applications, such as cellular characterization,[13] label-free imaging of adenosine 5′-triphosphate,[14] proton imaging of dynamics in the living brain,[15] tumor cell counting,[16] bacteria counting,[17] DNA detection,[18] glucose detection,[19] and imaging of metabolites in biofilms.[20,21] We previously developed an LSI-based amperometric/potentiometric
device containing 20 × 20 sensors with a pitch of 250 μm.[22] Taking advantage of the high-throughput measurements
that can be achieved with the LSI device, we have reported the real-time
imaging of respiratory activity[23] and the
cell differentiation level of multiple cell aggregates.[24,25] In addition, this device has been applied for dopamine detection,[26] plant evaluation,[27] immunoassays,[28] cell adhesion evaluation,[29] and electrochemical motion tracking of microorganisms.[30] Furthermore, multiple analytes have been simultaneously
visualized,[31] and the developed imaging
system has been named electrochemicolor imaging.In the present
study, the LSI-based device was applied to the assessment
of endothelial permeability. Briefly, a monolayer of human umbilical
vein endothelial cells (HUVECs) is placed in a solution containing
electrochemical tracers, [Fe(CN)6]4–,
on the device. Under an applied potential, the electrochemical tracers
near the electrode are consumed, which induces diffusion through the
monolayer from the bulk solution to the electrodes. Therefore, the
current at the electrodes depends on the endothelial permeability.
As the strategy does not require the preparation of top and bottom
channels, it is superior to conventional transwell systems. In addition,
owing to the packed electrode arrays, local information can be obtained
in multiple areas. Therefore, this imaging strategy can be used to
visualize local endothelial permeability. As a proof-of-concept of
this novel electrochemical imaging system, the effects of histamine
on an HUVEC monolayer were monitored. Next, as metastatic cancers
invade HUVEC monolayers, the proposed strategy was applied to visualize
the changes in the endothelial permeability of a coculture model consisting
of an HUVEC monolayer and cancer spheroids (HepG2, MCF-7, and MDA-MB-231).
Thus, in the present study, we proposed the novel electrochemical
imaging system of endothelial permeability using the LSI-based device,
and the proof-of-concept was conducted.
Results and Discussion
Schemes
Figure A,B shows
the schemes of preparation of a HUVEC monolayer
and detection, respectively. They are described in detail in the Materials and Methods section.
Figure 1
Strategy and process
for evaluating endothelial permeability. (A)
Fabrication of a HUVEC monolayer on a porous PET membrane. (B) Electrochemical
imaging using the LSI-based device. The schematic illustrations are
not to scale. (C) Photographs of the detection process: (i,ii) collection
of the HUVEC monolayer on the membrane; (iii) placement of the membrane
on the LSI-based device; (iv) PDMS frame; (v) pushing of the membrane
onto the device surface using the frame; and (vi) electrochemical
detection. The reference and counter electrodes are inserted in the
assay solution.
Strategy and process
for evaluating endothelial permeability. (A)
Fabrication of a HUVEC monolayer on a porous PET membrane. (B) Electrochemical
imaging using the LSI-based device. The schematic illustrations are
not to scale. (C) Photographs of the detection process: (i,ii) collection
of the HUVEC monolayer on the membrane; (iii) placement of the membrane
on the LSI-based device; (iv) PDMS frame; (v) pushing of the membrane
onto the device surface using the frame; and (vi) electrochemical
detection. The reference and counter electrodes are inserted in the
assay solution.
HUVEC Monolayer
HUVECs were attached to a collagen-coated
membrane and formed a monolayer, which remained after culturing for
7 days (Figure A).
In the phase contrast image, the microholes in the membrane could
also be observed. Although green fluorescent protein (GFP)-expressing
HUVECs were used, the cells could be observed using phase-contrast
imaging owning to the high transparency of the membrane. Therefore,
a discussion of cell imaging using a GFP is omitted from this paper.
Figure 2
HUVEC
monolayer cultured for 7 days. (A) Phase contrast image of
the monolayer on the membrane. (B,C) Fluorescence images of VE-cadherin
(green) and nuclei (red) in the monolayer on the culture plate: (B)
control and (C) after incubation for 30 min in the culture medium
containing [Fe(CN)6]4–.
HUVEC
monolayer cultured for 7 days. (A) Phase contrast image of
the monolayer on the membrane. (B,C) Fluorescence images of VE-cadherin
(green) and nuclei (red) in the monolayer on the culture plate: (B)
control and (C) after incubation for 30 min in the culture medium
containing [Fe(CN)6]4–.During electrochemical imaging, the monolayer was exposed
to [Fe(CN)6]4– for several minutes. Therefore,
the
effect of [Fe(CN)6]4– on VE-cadherin
at intercellular junctions was investigated. Immunostaining showed
that VE-cadherin was retained after culturing for 30 min in the medium
containing 5 mM [Fe(CN)6]4– (Figure B,C), indicating
that the toxic influence of [Fe(CN)6]4– is neglectable during the measurements.
Amperometric Measurements
of Endothelial Permeability
Figure shows the
amperograms obtained using the LSI device under various conditions.
First, using phosphate-buffered saline (PBS) containing [Fe(CN)6]4–, steady currents of approximately 22
nA were obtained in the absence of the membrane. Assuming a diffusion
coefficient of 6.5 × 10–10 m2/s
for [Fe(CN)6]4–,[32] the theoretical current for a single electrode[33,34] is approximately 19 nA, which is similar to the experimental value.
In contrast, when the sensing area was covered with a collage-coated
membrane, the currents decreased to approximately 13 nA, indicating
that the diffusion of [Fe(CN)6]4– was
blocked by the membrane. According to results of a current simulation,
the gap between the device surface and the membrane is roughly estimated
to be 10 μm (Figure S1). Thus, the
current values are determined by the diffusion of [Fe(CN)6]4– from the bulk solution through the pores of
the membrane to the electrodes. When an HUVEC monolayer was prepared
on the membrane, the currents decreased again to 6–8 nA because
the diffusion of [Fe(CN)6]4– was further
blocked by the HUVEC monolayer. These results indicate that the change
in current can be used to evaluate the endothelial permeability. Next,
to provide better conditions for the cells, a cell culture medium
[endothelial cell growth medium 2 (ECGM2)] was used instead of PBS.
Similar currents were obtained in both ECGM2 and PBS, indicating that
the effects of the cell culture medium on the electrochemical reaction
of 5 mM [Fe(CN)6]4– are neglectable.
Therefore, ECGM2 containing 5 mM [Fe(CN)6]4– was used in all subsequent experiments.
Figure 3
Chronoamperometry for
determining cell permeability using the LSI
device. Amperograms in PBS (without the membrane, with the membrane
but without the HUVEC monolayer, and with the HUVEC monolayer on the
membrane) and in the medium containing [Fe(CN)6]4– (with the HUVEC monolayer on the membrane).
Chronoamperometry for
determining cell permeability using the LSI
device. Amperograms in PBS (without the membrane, with the membrane
but without the HUVEC monolayer, and with the HUVEC monolayer on the
membrane) and in the medium containing [Fe(CN)6]4– (with the HUVEC monolayer on the membrane).Thus, a concentration gradient formed by the consumption of [Fe(CN)6]4– in an electrochemical reaction can be
used as a driving force for diffusion, resulting in the successful
detection of endothelial permeability. As a further proof-of-concept,
the influences of histamine and cancer spheroids on the endothelial
permeability were monitored.
Electrochemical Imaging of Endothelial Permeability
of HUVEC
Monolayers Stimulated with Histamine
The binding of histamine
to receptors on endothelial cell surfaces opens calcium channels,
causing cytoskeleton constriction and damage to barrier functions. Figure shows the electrochemical
images of the endothelial permeability of an HUVEC monolayer. The
addition of a histamine solution at approximately 528 s caused the
currents to increase rapidly owing to the mixing of the solution on
the membrane (Figure A). However, within 10–20 s, the currents returned to the
initial value because mixing was complete. Once stable, the current
increased gradually over a few minutes, indicating that the endothelial
permeability increased following cell damage by histamine. Han et
al.[5] reported that histamine influences
the permeability of a cell monolayer within a few minutes to 0.5 h,
which is similar to the response time observed in the present study.
In contrast, when PBS solution was added as a control, the current
values did not increase (Figure B). The chronoamperograms at the sensors clearly showed
the effect of the histamine (Figure C). Although biomolecules from the cells due to the
stimulation could be reacted at 0.5 V, the redox current at the 40
μm disk electrode increased by more than 1 nA and persisted
for a long time, indicating that the redox currents were mainly derived
from 5 mM [Fe(CN)6]4– rather than biomolecules
from the cells across the porous membrane. Thus, the change in endothelial
permeability caused by histamine was successfully monitored using
the proposed electrochemical imaging strategy. Although the effects
of varying the histamine concentration have not been explored, these
results provide a proof-of-concept.
Figure 4
Electrochemical imaging of the HUVEC monolayer
on the membrane
after the addition of histamine. (A) Electrochemical images following
histamine addition (normalized by subtraction of the image at 528
s). (B) Control images following PBS addition (normalized by subtraction
of the image at 218.4 s). (C) Chronoamperograms from the sensors marked
with red and yellow “x” in (A,B), respectively.
Electrochemical imaging of the HUVEC monolayer
on the membrane
after the addition of histamine. (A) Electrochemical images following
histamine addition (normalized by subtraction of the image at 528
s). (B) Control images following PBS addition (normalized by subtraction
of the image at 218.4 s). (C) Chronoamperograms from the sensors marked
with red and yellow “x” in (A,B), respectively.
Electrochemical Imaging of Endothelial Permeability
of HepG2
Spheroids
As a control for the coculture model, HepG2 spheroids
on a membrane without an HUVEC monolayer were electrochemically imaged. Figure A,B shows the results
after culturing for 1 day, whereas Figure C,D shows the results after culturing for
4 days. Figure E shows
the current values in areas under the spheroids and in those without
spheroids. In the areas without spheroids, after culturing for 1 day,
the [Fe(CN)6]4– oxidation currents decreased
to approximately 9 nA (from 13 nA on day 0, Figure ) with a further decrease to approximately
6 nA observed after culturing for 4 days. These results indicate that
some of the pores in the membrane were filled with components of the
culture medium even in the areas without cells. Although a single
pixel in Figure A
has a significantly lower current, this might not be derived from
cells but might be noise due to the defective sensor.
Figure 5
Electrochemical imaging
of HepG2 spheroids on the membrane. (A,B)
Day 1 and (C,D) day 4. (A,C) Optical (left) and electrochemical images
obtained at 299.8 s (right). (B,D) Illustrations of the spheroids
on the membranes. The illustrations are not to scale. (E) Currents
from the sensors indicated by red and yellow x in (A,C).
Electrochemical imaging
of HepG2 spheroids on the membrane. (A,B)
Day 1 and (C,D) day 4. (A,C) Optical (left) and electrochemical images
obtained at 299.8 s (right). (B,D) Illustrations of the spheroids
on the membranes. The illustrations are not to scale. (E) Currents
from the sensors indicated by red and yellow x in (A,C).Although the currents under the spheroids decreased slightly
after
culturing for 1 day (Figure A), the currents were similar to those in the areas without
the spheroids (Figure E), indicating that the spheroids were weakly adhered to the membrane
(Figures B and S2A). In contrast, the oxidation currents under
the spheroids decreased dramatically to approximately 3 nA after culturing
for 4 days (Figure C,E), suggesting that the spheroids strongly were adhered to the
membrane and the membrane pores were filled with cells (Figures D and S2B). In addition, the currents decreased in areas where the
cells proliferated and formed a monolayer (Figure C,D). According to these results, cell adhesion
on the membrane also affects the electrochemical signal.
Electrochemical
Imaging of Endothelial Permeability of HUVEC
Monolayers Cocultured with HepG2, MCF-7, or MDA-MB-231 Spheroids
As a model for analyzing the interactions of a tumor with blood
vessel walls, tumor spheroids were cultured on endothelial cell monolayers.[35] Once attached to a blood vessel, tumor cells
can breach or invade the endothelial-based lamina to reach the extracellular
connective tissue space and establish a secondary tumor. The cells
attach and penetrate the junctions during the earliest stage of the
invasion process. As this step is crucial for the development of metastases,
it is important to investigate these interactions and effects using
in vitro models. Tumor metastasis processes depend on several factors
such as cell adhesion molecules, direct cell-to-cell interactions,
reactive oxygen species (ROS), angiogenic growth factors, receptors,
and proteolytic enzymes.[36] Previously,
a coculture model consisting of tumor spheroids and an HUVEC monolayer
was applied to evaluate ROS production.[36] Microfluidic devices have also been applied to coculture spheroids,
mammary epithelial cells, and stroma fibroblasts.[37] In the present study, as a model of metastasis through
an endothelial monolayer, cancer spheroids were seeded on HUVEC monolayers. Figure A shows optical and
electrochemical images of HepG2 spheroids on the HUVEC monolayer.
Large currents were observed at the edge of the electrochemical image
because the pushing of the monolayer onto the device was not sufficient
and [Fe(CN)6]4– diffusion also occurred
at the edge of the membrane. Therefore, the currents at the edge were
not considered in the following discussion. Surprisingly, the currents
under the HepG2 spheroids were similar to those in the areas without
the spheroids. The result indicates that HepG2 spheroids might not
destroy the HUVEC monolayer (Figure B). When using MCF-7 spheroids, a slight change in
the currents was observed (Figure C), but the trend was similar to that with HepG2, indicating
that the MCF-7 spheroids might not destroy the monolayer (Figure D). In contrast,
a large increase in current was observed under the MDA-MB-231 spheroids
(Figure E), indicating
that the HUVEC monolayer was destroyed by these spheroids (Figure F,G). As the invasion
ability of MDA-MB-231 cells is higher than that of MCF-7 cells,[38] the HUVEC monolayer was more affected by MDA-MB-231
spheroids in the present study. Nikshoar et al.[39] performed invasion assays based on impedance measurements
using metastatic MDA-MB-231 cells and nonmetastatic MCF-7 cells added
to HUVEC monolayers. They reported that the HUVECs retracted due to
the invasion of the MDA-MB-231 cells, whereas the MCF-7 cells did
not induce any perturbation in the endothelial barrier. These results
are similar to those obtained using the electrochemical imaging strategy
although spheroids were used instead of single cells in the present
study.
Figure 6
Electrochemical imaging of cocultured cancer spheroids and HUVEC
monolayers. (A,B) HepG2 spheroids, (C,D) MCF-7 spheroids, and (E,F)
MDA-MB-231 spheroids. (A,C,E) Optical and electrochemical images obtained
at 599.8 s. (B,D,F) Illustrations of spheroid cross sections. The
illustrations are not to scale. (G) Currents from the sensors indicated
by red and yellow x in (A,C).
Electrochemical imaging of cocultured cancer spheroids and HUVEC
monolayers. (A,B) HepG2 spheroids, (C,D) MCF-7 spheroids, and (E,F)
MDA-MB-231 spheroids. (A,C,E) Optical and electrochemical images obtained
at 599.8 s. (B,D,F) Illustrations of spheroid cross sections. The
illustrations are not to scale. (G) Currents from the sensors indicated
by red and yellow x in (A,C).When using larger MCF-7 spheroids, the currents under the spheroids
increased, indicating that the endothelial permeability of the HUVEC
monolayer increased (Figure S3). This behavior
differed from that observed using the small spheroids, suggesting
that some destruction of the HUVEC monolayer occurred. During culturing,
the spheroids block the diffusion of oxygen nutrients from the bulk
solution into the monolayers, and the large spheroids might have a
larger effect on this process than the small ones.Thus, the
present electrochemical imaging method can be utilized
to evaluate cancer metastasis because information about the local
areas can be obtained. However, in this study, we did not consider
the direct adhesion of the cancer spheroids on the membranes after
the destruction of the HUVEC monolayers because this process is very
complicated. As the direct adhesion of the spheroids will affect the
diffusion of [Fe(CN)6]4–, these effects
should be determined to improve the precision of the analysis.
Conclusions
We developed an electrochemical imaging method to evaluate the
endothelial permeability of HUVEC monolayers. Although the presented
results are preliminary, they provide a proof-of-concept. As diffusion
of the electrochemical tracers is induced by the electrochemical reaction
at the packed electrodes, separated channels are unnecessary, which
is an advantage over conventional methods. Moreover, the use of an
electrode array allowed information about local areas to be obtained,
resulting in the successful imaging and evaluation of the metastasis
of cancer spheroids on HUVEC monolayers. However, in the current system,
the HUVEC monolayers were pushed onto the sensor surface by hand,
resulting in large errors, and the membrane with the HUVEC monolayer
was cut by hand, which is not a sophisticated approach. To solve these
problems, a culture system is needed although it will complicate the
detection system.In the present study, cancer spheroids were
used as primary tumor
models. Cancer cells in primary tumors, such as spheroids, invade
a vascular lumen from the tissues via an endothelial monolayer. Herein,
the cancer spheroids and endothelial monolayer were cultured on the
same side of the membrane because the co-culture model was easy to
prepare although the configuration did not match the in vivo configuration.
Therefore, the configuration should be modified in future studies.
Although cells were only cultured on one side of the membrane in this
study, cells can also be cultured on the other side. Therefore, the
present detection system can be applied to develop further complex
models such as a blood–brain barrier model. Furthermore, as
the detection system can provide spatial information, it can be applied
to visualize interactions between several kinds of cells.
Materials and
Methods
Electrochemical Imaging Strategy for Determining Endothelial
Permeability
First, an HUVEC monolayer is prepared on a porous
polyethylene terephthalate (PET) membrane (Figure A). Next, the monolayer is placed in a solution
containing [Fe(CN)6]4– on the LSI device
and pushed onto the surface of the device (Figure B). Under an applied potential of 0.5 V,
[Fe(CN)6]4– is oxidized to [Fe(CN)6]3–, which decreases the [Fe(CN)6]4– concentration near the device surface to nearly
zero because the packed electrode array operates in the microspace
under the monolayer. Therefore, a [Fe(CN)6]4– concentration gradient is formed between the solutions above and
below the monolayer, which induces [Fe(CN)6]4– diffusion. When the endothelial permeability is high, [Fe(CN)6]4– mainly diffuses to the electrodes through
the monolayer, resulting in large oxidation currents of [Fe(CN)6]4– at the electrodes. Thus, the current
value can be used as an indicator of the endothelial permeability.
Compared with a large single electrode, microelectrodes in an array
offer the advantage of sensitivity. Therefore, an LSI-based device
is used as a packed electrode array. In addition, electrochemical
imaging using such an array of individual working electrodes allows
the local endothelial permeability to be evaluated in multiple areas.
Device Fabrication and Detection System
The device
fabrication process is described in our previous papers.[22,31] Briefly, 20 × 20 Pt working electrodes were prepared on an
LSI chip and then covered with SU-8 microwells (diameter: 40 μm,
depth: 5 μm). Therefore, the individual working electrodes consist
of Pt disk electrodes with a diameter of 40 μm. The pitch of
the sensors was 250 μm, and the sensing area was approximately
5 × 5 mm2. Potential control and data acquisition
were performed using a program written in LabVIEW (National Instruments,
USA). Detection instruments and systems were partially developed by
Japan Aviation Electronics Industry, Limited (Japan) and purchased
from Senschip (Japan). Other details, including electronic circuits,
are described in our previous papers.[22,31]
Cell Cultures
GFP-expressing HUVECs were purchased
from Angio-Proteomie (USA). The cells were cultured in ECGM2 (Promo
Cell, Germany) containing 1% penicillin/streptomycin (PS, Gibco, USA).
HepG2 (hepatocellular carcinoma cell line, ATCC, USA) and MDA-MB-231
(breast cancer cell line, ATCC) were cultured in Dulbecco’s
modified Eagle’s medium (4.5 g/L glucose, Gibco) containing
10% fetal bovine serum (FBS, Gibco) and 1% PS (Gibco). MCF-7 (Tohoku
University) was cultured in RPMI1640 (Gibco) containing 10% FBS and
1% PS. All cells were cultured at 37 °C in a humidified atmosphere
containing 5% CO2. Phase-contrast images of the cells were
captured using a microscope (Olympus IX71, Japan).
Preparation
of HUVEC Monolayers on Porous Membranes
Rat tail collagen
type I (Corning, USA) was diluted to 0.0574 μg/mL
using PBS (Nacalai Tesque, Japan), and 0.25 mL of this solution was
placed in each well of a 12-well cell culture insert (Corning) consisting
of a PET membrane (pore size: 8 μm, pore density: 1 × 105 pores/cm2, thickness: 10.5 μm). The culture
insert was incubated at 37 °C for 30 min. After removing the
solution, the culture insert was washed with PBS. Then, HUVECs (3.0
× 104 cells/cm2) were seeded in the culture
insert and cultured in ECGM2 for 7 days to prepare the HUVEC monolayers
(Figure A). The medium
was changed every 2 days. After culturing, the membranes were cut
into pieces of approximately 6 × 6 mm2 for electrochemical
assays.
Immunostaining of VE-Cadherin in HUVEC Monolayers
To
investigate the effects of [Fe(CN)6]4– on HUVEC monolayers, VE-cadherin in the HUVEC monolayers was stained.
First, HUVECs were cultured in a 24-well culture plate (Corning) for
7 days to prepare HUVEC monolayers. The monolayers were washed with
PBS three times and then immersed in PBS containing 4% paraformaldehyde
(Fujifilm Wako Pure Chemical Co., Japan) for 15 min. Next, after washing
with PBS three times, the monolayers were immersed in 0.1% Triton
X-100 (Sigma-Aldrich Japan, Japan) for 15 min. After removing Triton
X-100, the monolayers were immersed in PBS containing 1% bovine serum
albumin (Sigma-Aldrich Japan) and rabbit anti VE-cadherin antibody
(1:500 dilution, Cell Signaling Technology, USA). After incubating
overnight at 4 °C, the solution was replaced with PBS containing
a secondary antibody, Alexa Fluor 647 conjugated goat anti-rabbit
IgG (1:1000 dilution, Thermo Fisher Scientific, USA), and 45 μM
propidium iodide (PI, double staining kit, Dojindo, Japan). The monolayers
were incubated overnight at 4 °C and then observed under a fluorescence
microscope (Leica DMi8, Leica, Germany).
Cocultures of Cancer Spheroids
and HUVEC Monolayers
A 96 U well plate (Sumitomo Bakelite
Co., Japan) was used to prepare
the cancer spheroids. First, 200 μL of a HepG2, MCF-7, or MDA-MB-231
suspension was introduced into each well (2000 cells/well) and then
cultured for 4 days to form spheroids. The spheroids were carefully
transferred onto the HUVEC monolayer that had been cultured for 7
days on the membrane. The coculture model was further cultured for
7 days. After culturing, the membrane was cut into pieces of approximately
6 × 6 mm2 for electrochemical assays.
Electrochemical
Imaging of Endothelial Permeability
Figure C shows photographs
of the detection process. After preparing the HUVEC monolayers on
the membranes, a piece of the membrane was immersed in ECGM2 containing
5 mM [Fe(CN)6]4– on the device. The membrane
was pushed onto the device surface using a polydimethylsiloxane frame
(Dow Corning Toray, Japan). Then, Ag/AgCl (sat. KCl) reference and
Pt counter electrodes were inserted into the solution. A voltage of
0.5 V was applied to the 400 working electrodes and the current from
each electrode was obtained every 200 ms. The current values were
converted into colors to construct electrochemical images consisting
of 20 × 20 pixels.For the assays using histamine, 45 μL
of 0.333 mM histamine (Merck, Germany) was added into ECGM2 containing
5 mM [Fe(CN)6]4– (final histamine concentration:
10 μM). The change in the current values was monitored in real
time..
Authors: Olivier Y F Henry; Remi Villenave; Michael J Cronce; William D Leineweber; Maximilian A Benz; Donald E Ingber Journal: Lab Chip Date: 2017-06-27 Impact factor: 6.799
Authors: Roman Paduch; Adam Walter-Croneck; Barbara Zdzisińska; Agnieszka Szuster-Ciesielska; Martyna Kandefer-Szerszeń Journal: Cell Biol Int Date: 2005-07 Impact factor: 3.612
Authors: H Eslami Amirabadi; M Tuerlings; A Hollestelle; S SahebAli; R Luttge; C C van Donkelaar; J W M Martens; J M J den Toonder Journal: Biomed Microdevices Date: 2019-11-23 Impact factor: 2.838
Authors: Daniel L Bellin; Hassan Sakhtah; Jacob K Rosenstein; Peter M Levine; Jordan Thimot; Kevin Emmett; Lars E P Dietrich; Kenneth L Shepard Journal: Nat Commun Date: 2014 Impact factor: 14.919