| Literature DB >> 34984207 |
Gyanendra Swaroop Mittal1, Deepak Sundriyal2, Niranjan B Naik3, Amit Sehrawat2.
Abstract
Background Chemoport (totally implantable venous access device) and its catheter system are used to administer long-term chemotherapy in cancer patients. The objective of this study was to analyze the complications associated with chemoport insertion in various cancer patients. Material and Methods A total number of 168 chemoports along with polyurethane catheters were inserted in various cancer patients over a period of 3 years. 9.6 F polyurethane catheters were put by a team of surgical oncologists in operation theater under general or local anesthesia. Analysis of the complications was done until the chemoport was removed due to any reason. Results Out of 168 patients, 30 (17.85%) developed complications. Complications included arterial puncture, malposition of the catheter tip, pneumothorax, hematoma, seroma, deep vein thrombosis, fracture of the catheter, a reversal of port, infections, and thrombosis of the catheter. Only a few required premature port and catheter removal. Conclusion There was a low rate of complications associated with chemoport using a polyurethane type of catheter system. However, infection-related complications were comparatively more common in our series. Chemoport requires expert handling, patient education, strict follow-up, and dedicated teamwork to minimize complications. MedIntel Services Pvt Ltd. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).Entities:
Keywords: TIVAD; chemoport; polyurethane
Year: 2021 PMID: 34984207 PMCID: PMC8719977 DOI: 10.1055/s-0041-1739041
Source DB: PubMed Journal: South Asian J Cancer ISSN: 2278-330X
Various malignancies
| Malignancies | ||
|---|---|---|
|
| Breast | 141 (83.92) |
|
| Lung | 09 (5.35) |
|
| Gastrointestinal | 8 (4.76) |
|
| Leukemia | 6 (3.57) |
|
| Ovary | 4 (2.38) |
| Total | 168 |
Complications
| Complications | ||
|---|---|---|
|
| Malposition of catheter tip | 2 (1.19) |
|
| Pneumothorax | 1 (0.59) |
|
| Hematoma, seroma | 2 (1.19) |
|
| Deep vein thrombosis | 1 (0.59) |
|
| Fracture of catheter | 1 (0.59) |
|
| Reversal of port | 1 (0.59) |
|
| Excess length of catheter | 1 (0.59) |
|
| Infective complications | 16 (9.52) |
|
| Catheter kinking | 1 (0.59) |
|
| Catheter obstruction | 4 (2.38) |
| Total | 30 (17.85) |