| Literature DB >> 34951713 |
Yong-Hong Chai1, Su-Ya Han2, Yu-Xin Zhu3, Jin-Jie Hou4, Xiao-Hui Guan5, Xin-Xin Yin3, Feng-Ying Zhang5, Qin-Zeng Qiao6, Ling-Min Han7, JiangHua Li2.
Abstract
BACKGROUND: We compared the methods of electrocardiogram (ECG) and X-ray localization of the peripherally inserted central catheter (PICC) tip position, in order to find a more convenient, practical, and safe method.Entities:
Keywords: advanced cancer; application study; critically ill bedridden; electrocardiographic localization; peripherally inserted central catheter; tip position
Mesh:
Year: 2021 PMID: 34951713 PMCID: PMC8916555 DOI: 10.1111/anec.12918
Source DB: PubMed Journal: Ann Noninvasive Electrocardiol ISSN: 1082-720X Impact factor: 1.468
Basic PICC information (n = 137)
| Classification | Category | Number of patients | Percentage |
|---|---|---|---|
| Catheter tip type | Opening | 101 | 73.85 |
| Three‐way valve | 36 | 26.15 | |
| Placement in upper extremity | Upper left arm | 36 | 23.85 |
| Right upper arm | 100 | 75.38 | |
| Right cubital fossa | 1 | 0.77 | |
| Inserted vein | Basilic vein | 86 | 64.61 |
| Median cubital vein | 1 | 0.77 | |
| Brachial vein | 47 | 33.85 | |
| Cephalic vein | 3 | 0.77 |
Age comparison of patients with specific P waves and patients without P‐wave change ( ± s years)
| Specific P waves | Number of patients ( | Proportion (%) | Mean age | Standard deviation |
|
|
|---|---|---|---|---|---|---|
| Yes | 130 | 94.89 | 71.58 | 14.84 | −6.704 | <.001 |
| No | 7 | 5.11 | 87.29 | 5.15 |
The consistency between localization with three types of specific P waves and chest X‐ray localization
| Specific P waves | Chest X‐ray (marked by thoracic vertebra) | Accuracy (%) | |||
|---|---|---|---|---|---|
| P/R | Number of patients ( | <T6 ( | T6–T8 ( | >T8 ( | |
| >80% | 65 | 1 | 62 | 2 | 95.38 |
| 50%–80% | 51 | 4 | 43 | 4 | 84.31 |
| <50% | 14 | 1 | 13 | 0 | 92.85 |
| Total | 130 | 6 | 118 | 6 | 90.76 |
Comparison of electrocardiographic localization vs. X‐ray localization (%)
| Specific P waves | Number of patients ( | Ideal position ( | Malposition ( | Accuracy (%) | Sensitivity (%) | Specificity (%) |
|---|---|---|---|---|---|---|
| Yes | 130 | 118 | 12 | 124/137 | 118/119 | 6/18 |
| No | 7 | 1 | 6 | |||
| Total | 137 | 119 | 22 | 90.51% | 99.16% | 33.33% |