Carla Russo1, Desirée Bartolini2, Cristina Corbucci3, Anna Maria Stabile2, Mario Rende2, Antimo Gioiello4, Gabriele Cruciani5, Antonella Mencacci1, Francesco Galli4, Donatella Pietrella1.
Abstract
BACKGROUND: Healthcare-associated infections caused by multi-drug resistant (MDR) pathogens are associated with increased mortality and morbidity among hospitalized patients. Inanimate surfaces, and in particular high-touch surfaces, have often been described as the source for outbreaks of nosocomial infections. The present work aimed to evaluate the efficacy of a last-generation mobile (robotic) irradiation UV-C light device R2S on MDR microorganisms in inanimate surfaces and its translation to hospital disinfection.Entities:
Keywords: MDR microorganisms; UV-C light device; disinfection; high-touch surface; photoreactivation
Mesh:
Substances:
Year: 2021 PMID: 34948629 PMCID: PMC8701089 DOI: 10.3390/ijerph182413019
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 3.390
Figure 1Workflow of the R2S system. R2S robot is equipped with a dedicated app able to read a QR code in which the mission is memorized (all the steps and the time of irradiation of each stop point).
Figure 2R2S mission performed in two different outpatient rooms (Room 1 and 2) of the “Santa Maria della Misericordia” Hospital in Perugia and the schematic representation of the sampling spots. The test was carried out in the two rooms following the indicated steps. Outpatient rooms 1: the distances measured between the UV-C light device R2S irradiation step 1 and the sampling point of surfaces, such as internal door handle (1), floor (2), the right arm of the patient chair (3), and computer keyboard and mouse (5), which were 28, 45, 62, and 50 cm, respectively. The distance between R2S irradiation step 2 and the sampling spot on the bench (4) was 30 cm. Outpatient rooms 2: the distances measured between the UV-C light device R2S irradiation step 1 and the sampling point of surfaces, such as internal door handle (1) and the right armrest of the patient’s chair (2), which were 25 and 50 cm, respectively. The distances between R2S irradiation step 2 and the sampling points on the sink (3) and floor (4) were 38 and 40 cm, respectively. The distance between R2S irradiation step 3 and the sampling point on the bench (5) was 30 cm.
Figure 3Outpatient room disinfection with the UV-C robotic systems R2S compared to standardized manual protocol based on chlorine solution nebulization. Mean ± SD data were calculated from the two determinations described in Materials and Methods. The CFU number of bacteria, yeast, and molds was assessed as described in the Materials and Methods section. Black bars: CFU before sanification; gray bars: after manual disinfection; white bars: after R2S disinfection.
Bacterial inactivation and photoreactivation after in vitro UV-C irradiation.
| Baseline | CFU at Seeding | Inactivation Test * | Photoreactivation Test | |
|---|---|---|---|---|
|
| 2 × 104 | 1.4 × 104 ± 3.8×103 | 0.0 ± 0.0 | 0.0 ± 0.0 |
| 2 × 105 | 1.6 × 105 ± 5.9 × 104 | 0.3 ± 0.6 | 0.3 ± 0.6 | |
| 2 × 106 | 1.9 × 106 ± 5.7 × 105 | 1.3 ± 1.2 | 2.7 ± 4.6 | |
|
| 2 × 104 | 2.2 × 104 ± 2.8 × 102 | 0.0 ± 0.0 | 0.0 ± 0.0 |
| 2 × 105 | 4.9 × 105 ± 9.3 × 104 | 2.0 ± 3.5 | 0.7 ± 0.6 | |
| 2 × 106 | 1.9 × 106 ± 8.4 × 105 | 1.3 ± 0.6 | 0.7 ± 1.2 | |
|
| 2 × 104 | ND | ND | ND |
| 2 × 105 | 3.4 × 105 ± 9.8 × 104 | 0.0 ± 0.0 | 0.0 ± 0.0 | |
| 2 × 106 | 1.9 × 106 ± 4.8 × 105 | 0.0 ± 0.0 | 0.0 ± 0.0 | |
|
| 2 × 104 | 5.2 × 104 ± 2.8 × 103 | 0.0 ± 0.0 | 0.0 ± 0.0 |
| 2 × 105 | 2.8 × 105 ± 2.3 × 104 | 0.0 ± 0.0 | 0.7 ± 1.2 | |
| 2 × 106 | 2.7 × 106 ± 7.8 × 104 | 0.0 ± 0.0 | 2.0 ± 2.0 | |
|
| 2 × 104 | 8.5 × 103 ± 2.2 × 103 | 0.0 ± 0.0 | 0.0 ± 0.0 |
| 2 × 105 | 1.5 × 105 ± 5.2 × 104 | 8.7 ± 9.0 | 1.7 ± 2.1 | |
| 2 × 106 | 1.8 × 106 ± 4.9 × 105 | 7.0 ± 11.2 | 16.7 ± 8.1 | |
|
| 2 × 104 | 6.4 × 103 ± 2.7 × 102 | 0.0 ± 0.0 | 0.0 ± 0.0 |
| 2 × 105 | 2.3×105 ± 6.0 × 104 | 1.3 ± 1.2 | 1.3 ± 1.2 | |
| 2 × 106 | 2.2 × 106 ± 4.2 × 105 | 6.7 ± 11.5 | 14.7 ± 8.3 |
* Inactivation was assessed after a time of exposure to 12.68 mJ/cm2 of UV-C for S. aureus and P. aeruginosa, and 22.97 mJ/cm2 of UV-C for K. pneumonia strains. ND: Not determined.
Kinetics of bacterial inactivation and photoreactivation after in vitro UV-C irradiation.
| Irradiation | UV-C Exposure | Photoreactivation | |
|---|---|---|---|
|
| 0 | 1.6 × 105 ± 5.9 × 104 | |
| 4.33 | 269.3 ± 249.8 | 330.0 ± 304.1 | |
| 9.66 | 8.7 ± 15.0 | 0.0 ± 0.0 | |
| 12.68 | 0.3 ± 0.6 | 0.0 ± 0.0 | |
|
| 0 | 4.9 × 105 ± 9.2 × 104 | |
| 4.33 | 2.0 ± 3.5 | 3.0 ± 5.2 | |
| 9.66 | 3.3 ± 5.8 | 0.0 ± 0.0 | |
| 12.68 | 0.3 ± 0.6 | 0.0 ± 0.0 | |
|
| 0 | 3.3 × 105 ± 9.8 × 103 | |
| 4.33 | NT | NT | |
| 9.66 | NT | NT | |
| 12.68 | 0.0 ± 0.0 | 0.0 ± 0.0 | |
|
| 0 | 2.8 × 105 ± 2.2 × 104 | |
| 4.33 | 0.0 ± 0.0 | 41.00 ± 69.3 | |
| 9.66 | 0.0 ± 0.0 | 0.0 ± 0.0 | |
| 12.68 | 0.0 ± 0.0 | 0.0 ± 0.0 | |
|
| 0 | 1.5 × 105 ± 5.2 × 104 | |
| 4.33 | 535.0 ± 379.6 | 592.7 ± 375.0 | |
| 9.66 | 1.3 ± 1.2 | 13.3 ± 15.3 | |
| 22.97 | 8.7 ± 9.0 | 1.7 ± 2.1 | |
|
| 0 | 2.3 × 105 ± 6.0 × 104 | |
| 4.33 | 508.0 ± 426.14 | 567.67 ± 417.9 | |
| 9.66 | 1.3 ± 1.2 | 38.0 ± 32.7 | |
| 22.97 | 1.3 ± 1.2 | 1.3 ± 1.2 | |
Effect of UV-C irradiation on different materials.
| Exposure mJ/cm2 | CFU/ at Seeding | Sample Surface Irradiation | CFU/Plate After Exposure | |||
|---|---|---|---|---|---|---|
| Plastic | Glass | Steel | ||||
|
| 9.66 | 2.1 × 106 ± 2.7 × 105 | Aluminium covered | 2.1 × 106 ± 9.9 × 104 | 2.0 × 106 ± 2.0 × 105 | 2.0 × 106 ± 2.4 × 105 |
| Uncovered | 16.0 ± 7.81 ** | 1.00 ± 1.00 ** | 15.33 ± 6.66 ** | |||
|
| 9.66 | 2.2 × 106 ± 3.9 × 105 | Aluminium covered | 2.1 × 106 ± 2.8 × 105 | 2.2 × 106 ± 4.8 × 104 | 1.9 × 106 ± 2.3 × 105 |
| Uncovered | 24.33 ± 4.73 * | 22.33 ± 12.70 * | 3.00 ± 1.73 * | |||
|
| 9.66 | 1.7 × 106 ± 3.1 × 105 | Aluminium covered | 1.7 × 106 ± 1.1 × 105 | 1.6 × 106 ± 6.4 × 104 | 1.7 × 106 ± 1.1 × 104 |
| Uncovered | 2.67 ± 1.53 * | 0.67 ± 1.15 * | 0.67 ± 0.58 * | |||
|
| 9.66 | 1.9 × 106 ± 3.7 × 105 | Aluminium covered | 1.8 × 106 ± 8.5 × 104 | 1.9 × 106 ± 7.0 × 104 | 1.9 × 106 ± 1.5 × 105 |
| Uncovered | 46.67 ± 48 * | 1.67 ± 0.58 * | 46.00 ± 10.15 * | |||
|
| 9.66 | 2.0 × 106 ± 4.1 × 105 | Aluminium covered | 2.0 × 106 ± 2.1 × 105 | 2.0 × 106 ± 1.6 × 105 | 2.0 × 106 ± 1.4 × 105 |
| Uncovered | 19.67 ± 11.68 * | 5.67 ± 5.03 * | 58.33 ± 43.88 * | |||
|
| 9.66 | 2.4 × 106 ± 4.5 × 105 | Aluminium covered | 2.3 × 106 ± 2.8 × 105 | 2.2 × 106 ± 3.6 × 105 | 2.2 × 106 ± 4.0 × 105 |
| Uncovered | 42.33 ± 34.49* | 0.00 ± 0.00 * | 30.67 ± 18.58 * | |||
Aluminium covered surfaces were sampled by mean of swabs diluted in MHB and spread on MHA plates. * p < 0.05, ** p < 0.001, CFU after exposure vs CFU at seeding.