John M Boyce1, Patricia A Farrel2, Dana Towle2, Renee Fekieta1, Michael Aniskiewicz1.
Abstract
OBJECTIVE To evaluate ultraviolet C (UV-C) irradiance, UV-C dosage, and antimicrobial effect achieved by a mobile continuous UV-C device. DESIGN Prospective observational study. METHODS We used 6 UV light sensors to determine UV-C irradiance (W/cm2) and UV-C dosage (µWsec/cm2) at various distances from and orientations relative to the UV-C device during 5-minute and 15-minute cycles in an ICU room and a surgical ward room. In both rooms, stainless-steel disks inoculated with methicillin-resistant Staphylococcus aureus (MRSA), vancomycin-resistant Enterococcus (VRE), and Clostridium difficile spores were placed next to sensors, and UV-C dosages and log10 reductions of target organisms achieved during 5-minute and 15-minute cycles were determined. Mean irradiance and dosage readings were compared using ANOVA. RESULTS Mean UV-C irradiance was nearly 1.0E-03 W/cm2 in direct sight at a distance of 1.3 m (4 ft) from the device but was 1.12E-05 W/cm2 on a horizontal surface in a shaded area 3.3 m (10 ft) from the device (P4 to 1-3 for MRSA, >4 to 1-2 for VRE and >4 to 0 log10 for C. difficile spores, depending on the distance from, and orientation relative to, the device with 5-minute and 15-minute cycles. CONCLUSION UV-C irradiance, dosage, and antimicrobial effect received from a mobile UV-C device varied substantially based on location in a room relative to the UV-C device. Infect Control Hosp Epidemiol 2016;37:667-672.Entities:
Mesh:
Year: 2016 PMID: 27004524 PMCID: PMC4890342 DOI: 10.1017/ice.2016.35
Source DB: PubMed Journal: Infect Control Hosp Epidemiol ISSN: 0899-823X Impact factor: 3.254
FIGURE 1Mean ultraviolet C (UV-C) and standard deviation (SD) irradiance levels (W/cm ) measured with 6 UV-C sensors placed at different distances and orientations relative to a mobile UV-C device in an intensive care unit room (left column at each position) and a room on a surgical ward (right columns). Sensor positions included in direct sight at distances of 1.3 m (4 ft) and 3.3 m (10 ft), at a 0° angle relative to the device at distances of 1.3 m and 3.3 m, and in shaded areas 1.3 m and 3.3 m from the device.
Mean UV-C Dosages (µWsec/cm2) Received by Sensors Located at 6 Positions in an ICU Room and in a Room on a Standard Surgical Ward (Non-ICU Room)
| UV-C Sensor Distance from Device | Orientation of UV-C Sensor | 5-Min Cycle, ICU Room, Mean Dosage, µWsec/cm | 5-Min Cycle, Non-ICU Room, Mean Dosage, µWsec/cm | 15-Min Cycle, ICU Room, Mean Dosage, µWsec/cm | 15-Min Cycle, Non-ICU Room, Mean Dosage, µWsec/cm |
|---|---|---|---|---|---|
| 1.3 m (4 ft) | 90° angle, direct sight | 358,667 (5,131) | 280,667 (577) | 750,667 (4,041) | 819,667 (6,110) |
| 1.3 m | 0° angle, direct sight | 55,867 (586) | 60,533 (351) | 152,333 (1,528) | 169,333 (1,527) |
| 1.3 m | 0° angle, indirect light | 7,223 (70) | 8,713 (60) | 21,400 (346) | 27,000 (360) |
| 3.3 m (10 ft) | 90° angle, direct sight | 69,233 (987) | 80,300 (100) | 205,000 (1,000) | 247,667 (1527) |
| 3.3 m | 0° angle, direct light | 10,533 (115) | 9,733 (367) | 28,767 (404) | 35,267 (611) |
| 3.3 m | 0° angle, indirect light | 4,047 (21) | 3,737 (49) | 8,407 (59) | 11,767 (153) |
UV-C, ultraviolet C; ICU, intensive care unit; SD, standard deviation.
Triplicate fixed-position readings were obtained in each room with 5-minute and 15-minute cycles.
FIGURE 2Results obtained with qualitative photochromic dosage indicator labels (red, pink, and white rectangles attached to sensors) and dosages (µWsec/cm ) measured using radiometric ultraviolet C (UV-C) light sensors placed at various distances from and orientations relative to the UV device after a 15-minute cycle. UV-C light normally enters the radiometric sensors through openings that were covered by circular elevated caps at the time photographs were taken.
Range of Log10 Reductions of MRSA, VRE, and Clostridium difficile Achieved with Inoculated Disk Carriers Exposed to UV-C for 5-Minute and 15-Minute Cycles on 3 Occasions at Each Cycle Time
| Distance and Orientation of Disks Relative to UV-C Device | Mean UV-C Dosage Measured Adjacent to Disks for 5-Min Cycles, µWsec/cm | Range of Log10 Reduction with 5-Min Cycles, by Pathogen | Mean UV-C Dosage Measured Adjacent to Disks for 15-Min Cycles, µWsec/cm | Range of Log10 Reduction with 15-Min Cycles, by Pathogen |
|---|---|---|---|---|
| 1.3 m (4 ft), direct | 342,667 | MRSA: >4 log | 842,000 | MRSA: >4 log |
| VRE: 4 to >4 log | VRE: >4 log | |||
|
|
| |||
| 1.3 m, 0° angle | 53,900 | MRSA: 4 to >4 log | 148,667 | MRSA: >4 log |
| VRE: 3 to >4 log | VRE: 3–4 log | |||
|
|
| |||
| 1.3 m, shaded | 8,547 | MRSA: 1–4 log | 24,467 | MRSA: >4 log |
| VRE: 2–3 log | VRE: 2–3 log | |||
|
|
| |||
| 3.3 m (10 ft), direct | 67,567 | MRSA: 4 to >4 log | 202.667 | MRSA: >4 log |
| VRE: 3 to >4 log | VRE: >4 log | |||
|
|
| |||
| 3.3 m, 0° angle | 10,767 | MRSA: 4 to >4 log | 29,000 | MRSA: 4 to >4 log |
| VRE: 2 log | VRE: 3 log | |||
|
|
| |||
| 3.3 m, shaded | 3,395 | MRSA: 1–3 log | 8,880 | MRSA: 3 log |
| VRE: 1–2 log | VRE: 1–2 log | |||
|
|
|
MRSA, methicillin-resistant Staphylococcus aureus; VRE, vancomycin-resistant Enterococcus; UV-C, ultraviolet C.