Katsuhiro Masago1, Shiro Fujita2, Yuko Oya3, Yusuke Takahashi3, Hirokazu Matsushita4, Eiichi Sasaki1, Hiroaki Kuroda3.
Abstract
Background andEntities:
Keywords: PCR-based; UV absorbance measurement; next-generation sequencing
Mesh:
Substances:
Year: 2021 PMID: 34946321 PMCID: PMC8709233 DOI: 10.3390/medicina57121375
Source DB: PubMed Journal: Medicina (Kaunas) ISSN: 1010-660X Impact factor: 2.430
Figure 1DNA (A) and RNA (B) quality measured by A260/A280 and grouped according to tissue type.
Figure 2DNA concentration measured by Qubit and NanoDrop and grouped according to tissue type (A): Surgical specimens, Cell block specimens, Biopsy specimens and Scatter plot for the correlation analysis between Qubit and NanoDrop measurements (B). The R coefficients were 0.587 for the surgical specimens, 0.819 for the cell block specimens, and 0.199 for the biopsy specimens.
Figure 3RNA concentration measured by Qubit and NanoDrop and grouped according to tissue type (A): Surgical specimens, cell block specimens, Biopsy specimens) and Scatter plot for the correlation analysis between Qubit and NanoDrop measurements (B). The R coefficients were 0.535 for the surgical specimens, 0.597 for the cell block specimens, and 0.1 for the biopsy specimens.
Characteristics of the tissues used for genomic tests.
| Tissue Type | DNA (%) | RNA (%) |
|---|---|---|
| Surgery specimen | 249 (64.3) | 336 (60.2) |
| Cell block specimen | 31 (8.1) | 51 (9.2) |
| Biopsy specimen | 107 (27.6) | 171 (30.6) |
| Total | 387 (100) | 558 (100) |
Success rate of routine genomic tests using DNA or RNA with concentrations that are outliers when calculated with Qubit.
| DNA Based Genomic Tests | Success Rate (%) |
|---|---|
| CycleaveTM
| 26/27 (96.3) |
| Fragment | 26/27 (96.3) |
| CycleaveTM | 25/27 (92.6) |
| CycleaveTM | 26/27 (96.3) |
| Fragment | 26/27 (96.3) |
| RNA based genomic tests | |
| 86/153 (56.2) | |
| 54/150 (36.0) | |
| 91/150 (60.7) | |
| 60/144 (41.7) | |
| 59/150 (39.3) | |
| 78/148 (52.7) |
Required sample concentrations and *R2 values.
| Sample Type | Required Concentration | Required *R2 Values |
|---|---|---|
| DNA | ≥0.83 ng/µL | ≥0.99 |
| RNA | ≥1.43 ng/µL | ≥0.98 |
*R2 values should be evaluated only if the standard curve includes 3 or more points.
Success rate of NGS panel sequence using DNA or RNA with concentrations that are outliers when calculated with Qubit.
| Methods | Success Rate (%) |
|---|---|
| AmpliSeqTM DNA panel sequence | 14/18 (77.8) |
| AmpliSeqTM RNA panel sequence | 43/47 (91.5) |