Mst Noorjahan Begum1, Mohammad Jubair1, Kamrun Nahar1, Sezanur Rahman1, Muhammad Talha1, Md Safiullah Sarker1, A K M Nasir Uddin2, Saifuddin Khaled2, Mohammed Shehab Uddin3, Zhibiao Li4, Tianlin Ke4, Mohammed Ziaur Rahman1, Mustafizur Rahman1.
Abstract
BACKGROUND: Globally, real-time reverse transcription-polymerase chain reaction (rRT-PCR) is the reference detection technique for SARS-CoV-2, which is expensive, time consuming, and requires trained laboratory personnel. Thus, a cost-effective, rapid antigen test is urgently needed. This study evaluated the performance of the rapid antigen tests (RATs) for SARS-CoV-2 compared with rRT-PCR, considering different influencing factors.Entities:
Keywords: COVID-19 diagnosis; Rapid antigen test, RAT; SARS-CoV-2; kit evaluation
Mesh:
Substances:
Year: 2021 PMID: 34942043 PMCID: PMC8842313 DOI: 10.1002/jcla.24203
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Nasopharyngeal and nasal swab: rapid antigen kits vs rRT‐PCR (RdRp)
| rRT‐PCR | InTec Rapid SARS‐CoV−2 Antigen Test | SD Biosensor Rapid SARS‐CoV−2 Antigen Test | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Positive | Negative | Positive | Negative |
Sensitivity (95% CI) | Specificity (95% CI) | Accuracy (95% CI) | Positive | Negative |
Sensitivity (95% CI) | Specificity (95% CI) | Accuracy (95% CI) | ||
| NP | Ct <25 | 86 | NA | 85 | NA |
98.8 (93.7 to 99.9) | NA | NA | 84 | NA | 97.7 (91.9 to 99.7) | NA | NA |
| Ct 25–30 | 10 | NA | 8 | NA | 83.3 (51.6 to 97.9) | NA | NA | 4 | NA | 62.5 (35.4 to 84.8) | NA | NA | |
| Ct >30 | 16 | NA | 8 | NA | 50.0 (24.7 to 75.4) | NA | NA | 7 | NA | 43.8 (19.8 to 70.1) | NA | NA | |
| Overall | 112 | 102 | 101 | 102 | 90.2 (83.1 to 95.0) | 100.0 (96.5 to 100.00) | 94.9 (91.0 to 97.4) | 95 | 99 | 84.8 (76.8 to 90.9) | 97.1 (91.6 to 99.4) | 90.7 (85.9 to 94.2) | |
| NS | Ct <25 | 55 | NA | 55 | NA | 100.0 (93.5 to 100.0) | NA | NA | 54 | NA | 98.2 (90.3 to 99.9) | NA | NA |
| Ct 25–30 | 24 | NA | 19 | NA | 82.8 (64.2 to 94.2) | NA | NA | 18 | NA | 80.0 (61.4 to 92.3) | NA | NA | |
| Ct >30 | 21 | NA | 6 | NA | 28.6 (11.3 to 52.2) | NA | NA | 6 | NA | 28.6 (11.3 to 52.2) | NA | NA | |
| Overall | 100 | 114 | 80 | 111 | 80.0 (70.8 to 87.3) | 97.4 (92.5 to 99.5) | 89.6 (84.3 to 93.1) | 78 | 108 | 78.0 (68.6 to 85.7) | 94.7 (88.9 to 98.0) | 86.9 (81.7 to 91.1) | |
NA: Not applicable since negatives have no Ct values.
FIGURE 1Scatter dot plot for all specimens; X‐axis denotes rRT‐PCR Ct value for RdRp gene using nasopharyngeal swab specimen, and Y‐axis denotes rRT‐PCR Ct value for RdRp gene using nasal swab specimen
FIGURE 2Comparison of the rapid test result. (A) Box plot of rRT‐PCR Ct value (positive only) compared with rapid test–positive and –negative result using a nasopharyngeal swab. (B) Box plot of rRT‐PCR Ct value (positive only) compared with rapid test–positive and –negative result using a nasal swab
Comparison of the rapid test based on the days of symptom onset and viral load using NP and NS specimens
| Ct value | Test | NP | NS | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 0–3 days (n = 29) | 4–7 days (n = 117) | >7 days (n = 68) | 0–3 days (n = 29) | 4–7 days (n = 117) | >7 days (n = 68) | ||||||||
| Positive | Sensitivity | Positive | Sensitivity | Positive | Sensitivity | Positive | Sensitivity | Positive | Sensitivity | Positive | Sensitivity | ||
| <25 | rRT‐PCR | 7 | 46 | 32 | 5 | 28 | 16 | ||||||
| InTec | 7 | 100 | 46 | 100 | 31 | 96.9 | 5 | 100 | 28 | 100 | 16 | 100 | |
| SD Biosensor | 7 | 100 | 46 | 100 | 31 | 96.9 | 5 | 100 | 28 | 100 | 16 | 100 | |
| 25–30 | rRT‐PCR | 0 | 5 | 6 | 1 | 17 | 14 | ||||||
| InTec | 0 | 5 | 100 | 4 | 66.7 | 0 | 0 | 15 | 88.2 | 12 | 85.7 | ||
| SD Biosensor | 0 | 2 | 40.0 | 2 | 33.3 | 0 | 0 | 15 | 88.2 | 10 | 71.4 | ||
| >30 | rRT‐PCR | 0 | 8 | 8 | 1 | 8 | 10 | ||||||
| InTec | 0 | 6 | 75.0 | 2 | 25.0 | 0 | 0 | 2 | 25.0 | 2 | 20.0 | ||
| SD Biosensor | 0 | 5 | 62.5 | 2 | 25.0 | 0 | 0 | 2 | 25.0 | 2 | 20.0 | ||
| Overall | rRT‐PCR | 7 | 59 | 46 | 7 | 53 | 40 | ||||||
| InTec | 7 | 100 | 57 | 96.6 | 37 | 80.4 | 5 | 71.4 | 45 | 84.9 | 30 | 75.0 | |
| SD Biosensor | 7 | 100 | 53 | 89.8 | 35 | 76.1 | 5 | 71.4 | 45 | 84.9 | 28 | 70.0 | |