Tomohiro Mizoshiri1, Morikatsu Yoshida1, Seitaro Oda2, Shota Tsumagari1, Takeshi Nakaura2, Kazunori Harada3, Osamu Ikeda2.
Abstract
ABSTRACT: We investigated the feasibility of non-contrast three-dimensional modified Dixon (mDixon) magnetic resonance angiography (MRA) to evaluate the carotid artery.We studied 30 normal patients who underwent non-contrast mDixon and conventional time-of-flight (TOF) MRA of the neck with a clinical 3T MR scanner. Carotid artery signal-to-noise ratio (SNR) and contrast-to-noise ratio were compared between mDixon-MRA and TOF-MRA. Two readers independently evaluated vessel sharpness, image contrast, and overall image quality using a 4-point scale.SNR was significantly higher on mDixon-MRA than TOF-MRA (P < .01). There was no significant difference in contrast-to-noise ratio. The visual score for vessel sharpness was significantly higher on mDixon-MRA than TOF-MRA (P < .01), whereas the score for contrast was significantly higher on TOF-MRA (P < .01).Although non-contrast three-dimensional mDixon-MRA showed lower visual contrast than conventional TOF-MRA, it provided images with significantly higher SNR and better vessel sharpness than TOF-MRA.Entities:
Mesh:
Substances:
Year: 2021 PMID: 34941146 PMCID: PMC8702219 DOI: 10.1097/MD.0000000000028351
Source DB: PubMed Journal: Medicine (Baltimore) ISSN: 0025-7974 Impact factor: 1.817
Magnetic resonance imaging sequences and parameters.
| TOF-MRA | mDixon-MRA | |
| Spatial resolution (mm) | 0.5 × 0.79 × 1.1 | 1.2 × 1.19 × 1 |
| TR (ms) | 24 | 13 |
| TE (ms) | 3.5 | 1.43/2.6 |
| Flip angle (°) | 20 | 5 |
| Matrix | 400 × 189 | 168 × 123 |
| Field of view (mm) | 200 × 150 | 200 × 148 |
| SENSE (phase × slice) | 3 × 1 | 1 × 1 |
| Number of slices | 159 | 150 |
| Acquisition time (min) | 2:43 | 1:51 |
mDixon = modified Dixon, MRA = magnetic resonance angiography, TE = echo time, TOF = time-of-flight, TR = repetition time.
Quantitative analysis comparing mDixon-MRA and TOF-MRA.
| Signal-to-noise ratio | |||
| TOF-MRA | mDixon-MRA | ||
| CCA | 58.3 ± 17.9 | 66.3 ± 20.3 | <.01 |
| ICA origin | 57.0 ± 17.4 | 67.0 ± 20.8 | <.01 |
| Mid-portion of ICA | 58.5 ± 18.0 | 68.1 ± 21.9 | <.01 |
Values are mean ± standard deviation. P < .05 was considered significant.
CCA = common carotid artery, ICA = internal carotid artery, mDixon = modified Dixon, MRA = magnetic resonance angiography, TOF = time-of-flight.
Qualitative analysis comparing mDixon-MRA and TOF-MRA.
| TOF-MRA | mDixon-MRA | ||
| Image contrast (both sides) | 3.6 ± 0.4 | 2.9 ± 0.5 | <.01 |
| Left-side | 3.5 ± 0.4 | 3.0 ± 0.4 | <.01 |
| Right-side | 3.6 ± 0.4 | 3.1 ± 0.3 | <.01 |
| Vessel sharpness (both sides) | 3.2 ± 0.6 | 3.8 ± 0.4 | <.01 |
| Left-side | 3.2 ± 0.6 | 3.7 ± 0.4 | <.01 |
| Right-side | 3.4 ± 0.6 | 3.8 ± 0.2 | <.01 |
| Overall image quality (both sides) | 3.5 ± 0.3 | 3.5 ± 0.2 | .40 |
| Left-side | 3.6 ± 0.3 | 3.5 ± 0.3 | .11 |
| Right-side | 3.7 ± 0.3 | 3.5 ± 0.2 | .07 |
Values are mean ± standard deviation. P < .05 was considered significant.
mDixon = modified Dixon, MRA = magnetic resonance angiography, TOF = time-of-flight.
Number of arteries judged as inappropriate image quality (score = 1 or 2).
| TOF-MRA | mDixon-MRA | ||
| Image contrast | 0 | 9 | <.01 |
| Vessel sharpness | 11 | 2 | <.01 |
| Overall image quality | 2 | 2 | .69 |
P < .05 was considered significant.
mDixon = modified Dixon, MRA = magnetic resonance angiography, TOF = time-of-flight.
Figure 1Maximum intensity projections of the carotid bifurcation in a 64-year-old man who underwent MRA of the neck. The visual scores for image contrast, vessel sharpness, and overall image quality were 4, 3, and 4, respectively, for TOF-MRA, and 4, 4, and 4, respectively, for mDixon-MRA. mDixon = modified Dixon, MRA = magnetic resonance angiography, TOF = time-of-flight.
Figure 2Maximum intensity projections of the carotid artery bifurcation in a 66-year-old man who underwent MRA of the neck. The visual scores for image contrast, vessel sharpness, and overall image quality were 4, 3, and 4, respectively, for TOF-MRA, and 3, 4, and 3, respectively, for mDixon-MRA. mDixon = modified Dixon, MRA = magnetic resonance angiography, TOF = time-of-flight.