Stephan Warnke1, Ahmed Ben Faleh1, Thomas R Rizzo1. 1. Laboratoire de Chimie Physique Moléculaire, École Polytechnique Fédérale de Lausanne, EPFL SB ISIC LCPM, Station 6, CH-1025 Lausanne, Switzerland.
Abstract
Infrared (IR) spectroscopy is a powerful tool used to infer detailed structural information on molecules, often in conjunction with quantum-chemical calculations. When applied to cryogenically cooled ions, IR spectra provide unique fingerprints that can be used for biomolecular identification. This is particularly important in the analysis of isomeric biopolymers, which are difficult to distinguish using mass spectrometry. However, IR spectroscopy typically requires laser systems that need substantial user attention and measurement times of tens of minutes, which limits its analytical utility. We report here the development of a new high-throughput instrument that combines ultrahigh-resolution ion-mobility spectrometry with cryogenic IR spectroscopy and mass spectrometry, and we apply it to the analysis of isomeric glycans. The ion mobility step, which is based on structures for lossless ion manipulations (SLIM), separates glycan isomers, and an IR fingerprint spectrum identifies them. An innovative cryogenic ion trap allows multiplexing the acquisition of analyte IR fingerprints following mobility separation, and using a turn-key IR laser, we can obtain spectra and identify isomeric species in less than a minute. This work demonstrates the potential of IR fingerprinting methods to impact the analysis of isomeric biomolecules and more specifically glycans.
Infrared (IR) spectroscopy is a powerful tool used to infer detailed structural information on molecules, often in conjunction with quantum-chemical calculations. When applied to cryogenically cooled ions, IR spectra provide unique fingerprints that can be used for biomolecular identification. This is particularly important in the analysis of isomeric biopolymers, which are difficult to distinguish using mass spectrometry. However, IR spectroscopy typically requires laser systems that need substantial user attention and measurement times of tens of minutes, which limits its analytical utility. We report here the development of a new high-throughput instrument that combines ultrahigh-resolution ion-mobility spectrometry with cryogenic IR spectroscopy and mass spectrometry, and we apply it to the analysis of isomeric glycans. The ion mobility step, which is based on structures for lossless ion manipulations (SLIM), separates glycan isomers, and an IR fingerprint spectrum identifies them. An innovative cryogenic ion trap allows multiplexing the acquisition of analyte IR fingerprints following mobility separation, and using a turn-key IR laser, we can obtain spectra and identify isomeric species in less than a minute. This work demonstrates the potential of IR fingerprinting methods to impact the analysis of isomeric biomolecules and more specifically glycans.
Glycans are the most
abundant class of biopolymers, being present
either as free species or as glycoconjugates such as glycoproteins,
glycolipids, and other glycosides. It is thus unsurprising that glycobiology
has become a key frontier in biomedical science. For example, understanding
the involvement of glycans in viral pathogenesis[1] is important for analyzing the mechanism by which a virus
shields itself from vaccines.[2] The observation
of disease-specific alteration of protein and lipid glycosylation
on the surface of cells has led to the use of glycans as biomarkers
in various types of cancer,[3−6] inflammatory diseases,[7,8] and liver disease.[9,10] The glycosylation pattern of biotherapeutics, such as monoclonal
antibodies (mAbs), which are used to treat a wide range of pathologies,[11] directly affects their effectiveness, toxicity,
and long-term stability[12] and thus plays
an important role in quality control as well as in the regulatory
approval of biosimilars.[13]Unfortunately,
glycans are significantly more difficult to analyze
than other classes of biopolymers, which is a result of their complex
stereochemistry and possibility to form branched structures. Consequently,
no single analytical tool can currently resolve all forms of glycan
isomerism.[14] While NMR has the potential
to yield glycan sequence as well as stereochemical details,[15] it requires relatively large amounts of the
purified analyte, which is challenging to obtain from biological samples.
Liquid chromatography based methods,[16,17] although significantly
more sensitive, often require sample derivatization and analysis times
ranging from tens of minutes to hours for a single separation. Moreover,
certain positional isomers pose a particular problem for LC separation.[18] Ion mobility spectrometry (IMS) has made important
inroads as a fast separation technique that can differentiate between
certain kinds of glycan isomers;[19−23] however, it still has difficulty resolving the most
subtly different isomeric species. Several groups have recently employed
vibrational spectroscopy to distinguish isomeric glycan ions, either
at room temperature[24−31] or when cooled to cryogenic temperatures using collisions with cold
buffer gas[32] or embedded in liquid helium
nanodroplets.[33−36] While room-temperature IRMPD studies have clearly demonstrated the
ability to distinguish small isomeric glycans, they have difficulties
resolving larger species. Cooling glycans to cryogenic temperatures
extends the size limit of species that can be distinguished, although
resolving isomeric mixtures of large species still presents a formidable
challenge. Moreover, cooling ions in liquid-helium droplets is not
particularly well suited for incorporation into routine analytical
instruments.Our group has previously demonstrated that high-resolution
IMS
combined with cryogenic infrared (IR) spectroscopy can unambiguously
distinguish between structurally related glycan isomers,[37−39] and we have recently reported structure-specific IR fingerprints
for glycans consisting of up to 10 monosaccharides.[40] The robustness and accuracy of this technique is based
on the fact that an IR spectrum is an intrinsic property of the analyte
molecule and requires neither continuous calibration of the instrument
nor internal standards. Once recorded and stored in a database, an
IR fingerprint serves as a unique identifier that can be reproduced
across laboratories. The structural specificity of this technique
is demonstrated, for example, by its ability to distinguish between
α and β reducing-end anomers,[41] which differ in the orientation of only a single OH group. Confident
identification based on an IR fingerprint can be achieved as long
as the signal-to-noise ratio of a spectrum is sufficiently high. This
has previously been accomplished by extensive signal averaging, typically
over tens of minutes, making the approach impractical for high-throughput
applications or those where small amounts of sample do not permit
long analysis times.To overcome this drawback, we present here
a new multiplexing approach
that enables acquiring IR fingerprints of multiple mobility-separated
species simultaneously. We describe our instrument in which we combine
ultrahigh-resolution IMS using structures for lossless ion manipulation
(SLIM) with a state-of-the-art, segmented, cryogenic ion trap, where
multiple ensembles of ions can be held spatially separated while being
spectroscopically investigated. The increased acquisition speed and
repeatability of IR fingerprint spectra available using this approach
opens the door for the use of cryogenic IR spectroscopy as a broadly
applicable analytical tool for glycans as well as other classes of
isomeric biomolecules.
Experimental Approach
Materials
Methylated disaccharides and milk oligosaccharides
(LNT and LNnT) were purchased from Carbosynth Ltd. (GBR) and Dextra
(GBR), respectively, and used without further purification. For nanoelectrospray
ionization (nESI), 10 μM solutions of the analytes were prepared
in 50/50 MeOH/H2O, and approximately 1 equiv of sodium
acetate was added to aid in the formation of sodiated species. In-house
prepared borosilicate glass emitters were used to inject samples into
the instrument. All gases were of 99.9999% purity.
Instrument
Overview
The instrument, shown schematically
in Figure A, was designed
for the acquisition of high-resolution IR fingerprint spectra of mobility
separated ions in a high-throughput approach. Ions are generated in
a nESI source and transported toward the ion-mobility region (described
below) through a dual ion funnel assembly (MassTech, USA). An 80 mm
long ring-electrode funnel guides ions onto the SLIM printed circuit
board (PCB) assembly, where trapping and mobility separation is carried
out. Subsequently, a second ring-electrode funnel transfers ions through
a conductance limit and a series of multipole guides, which take them
through three differentially pumped regions (10–2, 10–4, and 10–7 mbar). Ions
then enter the trap region, held at approximately 10–6 mbar, where a short octupole guide and an einzel lens assembly directs
them into the cryogenic, planar multipole trap described below. Depending
on the mode of operation, an ion ensemble can be trapped and spectroscopically
interrogated or simply guided through to be analyzed by a reflectron
time-of-flight mass spectrometer (ToF-MS) (TofWerk, CHE). All DC,
radiofrequency, and traveling-wave electrical potentials applied from
the ion source to the cryogenic ion trap are generated by five MIPS-series
power supplies from GAA Custom Electronics, LLC (USA).
Figure 1
(A) Cut-away schematic
of the newly developed instrument utilized
for this work. (B) Schematic representation of the cryogenic, segmented
trap electrodes, where those used to apply trapping DC potentials
are highlighted in red. (C) Schematic representation of the DC potentials
applied when trapping in five separate segments. The trap is loaded
from the left side of (B), and thus, the rightmost trap is loaded
and unloaded first.
(A) Cut-away schematic
of the newly developed instrument utilized
for this work. (B) Schematic representation of the cryogenic, segmented
trap electrodes, where those used to apply trapping DC potentials
are highlighted in red. (C) Schematic representation of the DC potentials
applied when trapping in five separate segments. The trap is loaded
from the left side of (B), and thus, the rightmost trap is loaded
and unloaded first.
Ion Mobility Module
Mobility separation is performed
using structures for lossless ion manipulation (SLIM), originally
developed by Smith and co-workers.[42−44] Using traveling-wave
electrical potentials applied to electrodes on a PCB “sandwich”
structure, SLIM offers superior IMS resolving power[43] and peak capacity for its size and cost. The ability of
SLIM to guide ions through corners and turns as well as the possibility
of cycling them through the separation region as many times as necessary
in a nearly lossless manner enable extremely long drift lengths, and
hence high resolution, on a relatively small device. The SLIM module
we designed for our instrument features a 2 m accumulation region
that is constantly filled from the ion source, and a separation region
with a 10 m single-cycle path length. As demonstrated by Li et al.,[45] the high-volume accumulation region allows for
an increased ion utilization efficiency compared to previous implementations
using ion funnel traps for ion storage prior to ion-mobility separation.
Although not used for the study presented here, our SLIM module also
features five on-board trapping regions that can be used for intermediate
ion storage, enrichment of low-abundance analyte ions, and CID of
mobility-separated species.[46] Typical operating
parameters for the results presented here were: N2 drift
gas pressure, 2 mbar; traveling wave (TW) height, 30 V; TW speed,
160 m/s; RF amplitude, 120 Vpp; and RF frequency, 800 kHz.We demonstrate the separation power of our new instrument using
the reverse-sequence pair of peptides GRGDS/SDGRG, which differ by
1.5% in collisional cross section in their singly sodiated form.[47] The separation in time and the peak widths of
these ions were used to calculate the resolving power as a function
of the number of separation cycles, shown in Figure . The resolving power in a TW ion mobility
instrument scales with the square-root of the number of TW segments,[48] which is proportional to the drift length or
the number of separation cycles. The solid line in Figure represents a fit according
to the relationship f(x) = a√x, where x corresponds
to the drift length. A single separation cycle yields a resolving
power of approximately 200 and reaches almost 1000 after 200 m or
20 separation cycles.
Figure 2
Ion-mobility resolving power of the 10 m cyclic SLIM IMS
module
as a function of drift length, determined using the singly charged
ion-mobility standard peptides GRGDS and SDGRG.
Ion-mobility resolving power of the 10 m cyclic SLIM IMS
module
as a function of drift length, determined using the singly charged
ion-mobility standard peptides GRGDS and SDGRG.
Cryogenic Multitrap
Our new cryogenic trap, which is
based on our previous design,[49] was conceived
to accommodate a variable number of simultaneous trapping regions.
It is similar to a SLIM device in that the electric fields used for
trapping and guiding ions are created by electrodes on two mirrored
PCBs in a sandwich assembly. The 110 mm long electrode configuration
found on each PCB, displayed in Figure B, consists of four RF electrodes for confinement as
well as 3 rows of 51 individually controlled DC electrodes that define
the electric field gradient that pushes ions toward the exit of the
trap (right side). Individual DC electrodes can also be used to block
and trap mobility-separated ion distributions in predefined segments.
A typical DC potential gradient applied to form five individual trapping
segments is displayed schematically in Figure C, where the five potential spikes represent
the blocking voltage used to separate one trap compartment from the
next. Typical DC gradients across the entire trap are 10–20
V, with blocking potentials approximately 20 V above those of neighboring
electrodes. Solid copper electrodes placed on either side of the trap
fill the 5 mm space between the two PCBs and represent a physical
as well as a potential barrier for the ions. To confine ions in the
horizonal direction, we apply to these “side-barrier”
electrodes a DC potential of 5–10 V higher than that of the
on-track DC electrodes. The trap assembly is attached to a closed-cycle
He cryostat (Sumitomo, JPN) and typically held at a temperature of
45 K. It is worth noting that the number of trapping regions can be
reconfigured simply by changing which DC electrodes are used for blocking,
and this is done external to the machine.
Modes of Operation
(1) In MS mode, ions
are continuously introduced from the source into the IMS region and
transmitted without separation. The cryogenic multitrap region is
held at a constant He pressure of 10–6 mbar to enable
collisional damping,[50] thereby homogenizing
the kinetic energy of the ions before they enter the ToF MS. This
mode is used to acquire mass spectra of an unknown sample rapidly.
(2) In IMS mode (Figure S1A in the Supporting Information), a pulse of ions is released by an
ion gate into the SLIM separation region from the accumulation region,
and this represents the start of an instrument cycle. Ions of different
mobilities are separated with as many cycles as necessary before being
guided toward the ToF MS. During one instrument cycle, mass spectra
are acquired at 10 kHz repetition rate, thus creating nested IMS-MS
data[51] with 100 μs time resolution.
The length of an instrument cycle depends on the IMS separation/cycling
requirements and can range from 50 ms to more than 1 s. (3) In spectroscopy mode (Figure S1B),
after determination of the drift times of mobility-separated ions,
DC blocking voltages in the cryogenic multitrap are switched in an
accurately timed sequence to ensure trapping of each ion ensemble
of the same mobility in a separate compartment. We then perform messenger-tagging
infrared spectroscopy[52] of the trapped
ions, as follows. First, we introduce an intense gas pulse of an 80/20
He/N2 mixture through a solenoid-actuated pulsed valve
(Parker) a few ms before arrival of the first ions to ensure collisional
damping of excess kinetic energy and cooling of internal degrees of
freedom until the ion energies are in equilibrium with the surrounding
buffer gas, which is then pumped out for the remainder of the instrument
cycle. Three-body collisions involving N2 molecules lead
to the formation of ion-N2 clusters, which can be observed
as a +28 Da mass shift for ions in each compartment of the multitrap.
A continuous-wave (cw) fiber-pumped IR laser (CLT series, IPG, USA)
with a tuning range of 3250–4000 cm–1 operating
at a constant power of 1 W and a resolution of ∼1 cm–1 irradiates ions in all trap compartments simultaneously for the
entirety of the trapping time (approximately 50 ms), before the content
of individual trap compartments are successively analyzed in the ToF
over the last 5 ms of an instrument cycle. Mass spectra of each compartment
are therefore recorded separately. The absorption of a single photon
by a messenger-tagged ion leads to loss of the weakly bound N2 tag molecule and is observed in the MS data as a decrease
in intensity of the tagged ions and an increase in the intensity of
the untagged ions. Figure S2 shows an example
of the tagged and untagged ion signals obtained with the laser on
and off an absorption band. By plotting the ratio of tagged vs the
sum of tagged and untagged ion signal for each individual trap compartment
as a function of the laser wavelength, the IR spectra of each separated
species can be reconstituted in real time by the data acquisition
program. Spectra presented here were recorded two to three times and
averaged. With 1 W power and an unfocused laser beam, we avoid multiple-photon
absorption that might distort the spectrum. The use of a cw laser
system also allows one to freely choose the length of an instrument
cycle as required for ion separation in the IMS module. This flexibility
was not previously available with the pulsed OPO laser systems often
used in messenger-tagging spectroscopic studies.[37,53]
Results and Discussion
The performance of our new apparatus
was first tested with a mixture
of the structurally related human milk oligosaccharide isomers lacto-N-traose
(LNT) and lacto-N-neotetraose (LNnT). The difficulty to resolve these
isomers as sodium adduct species[54] represents
a formidable challenge. Both glycans have been reported to exhibit
two distinct gas-phase isomers,[37] which
have not been previously resolved using IMS. An arrival time distribution
(ATD) obtained on our apparatus from a mixture of LNT/LNnT after 100
m separation (10 cycles) is displayed in Figure . The IMS resolving power under these conditions
is approximately 700 (see Figure ), which indicates that the nearly baseline-separated
species at 780 to 790 ms drift times differ by only approximately
0.1% in their CCS. Accurately determining CCSs under these high-resolution
conditions remains extremely challenging,[55,56] and thus identifying such closely related isomers based a CCS value
alone would be tenuous.
Figure 3
IMS arrival time distribution of a singly sodiated
mixture of LNT
and LNnT after 100 m mobility separation (10 cycles).
IMS arrival time distribution of a singly sodiated
mixture of LNT
and LNnT after 100 m mobility separation (10 cycles).To assign the drift peaks in Figure to particular isomers, we recorded cryogenic
IR fingerprint
spectra of all four species simultaneously by making use of the new
segmented ion trap, where mobility-separated species are confined
in their respective compartments while being interrogated by the IR
laser. The spectra from this measurement are shown in Figure (blue), each of which display
a series of absorption bands originating from relatively free OH oscillators
(3580–3680 cm–1) and more strongly hydrogen-bonded
OH and NH oscillators (below 3550 cm–1). Each spectrum
is clearly unique over a broad spectral range, even though their underlying
structures are so closely related. This observation is typical for
cryogenic IR spectra and is what enables their use as an identifying
fingerprint for any compound with structure-sensitive IR active vibrational
transitions, such as glycans. The spectra from the individual trap
compartments were then compared to previously recorded reference spectra
of LNT and LNnT, shown in purple in Figure . A simple visual comparison allows us to
assign the first and third ATD features as conformers of LNT and the
second and fourth features as conformers of LNnT. Based on previous
work,[39] we suggest that each pair of peaks
likely corresponds to the α and β anomers at the C1 carbon
of the reducing end.
Figure 4
Cryogenic IR fingerprint spectra of the four IMS-separated
species
from the ATD (Figure ) of an LNT/LNnT mixture recorded simultaneously using the segmented
cryogenic ion trap (blue spectra). Spectra are matched to reference
spectra (purple) of previously individually recorded IR spectra for
identification of individual IMS drift peaks. The spectra are normalized
between 0 and 1, and the maximum depletion of tagged species is typically
80–90%.
Cryogenic IR fingerprint spectra of the four IMS-separated
species
from the ATD (Figure ) of an LNT/LNnT mixture recorded simultaneously using the segmented
cryogenic ion trap (blue spectra). Spectra are matched to reference
spectra (purple) of previously individually recorded IR spectra for
identification of individual IMS drift peaks. The spectra are normalized
between 0 and 1, and the maximum depletion of tagged species is typically
80–90%.Separation of the four LNT/LNnT
species required ultrahigh mobility
resolution, which resulted in an instrument cycle of almost one second
(see drift times in Figure ). Spectra were acquired with 220 points across the entire
wavenumber range with only one acquisition (i.e., one instrument cycle)
per point, resulting in a total acquisition time of 3 min for the
four spectra together.Figure A shows
examples of arrival time profiles of ions ejected from the four adjacent
ion trap compartments during acquisition of the spectra of Figure . Each ion distribution
arrives in the ToF extraction region several tens of microseconds
after their release and is few tens of microseconds wide. The exact
time of their ejection from the trap is arbitrary but needs to occur
in order of their arrival. The fast ToF analysis (10 kHz) allows us
to acquire a mass spectrum of each ion distribution (Figure B) within each instrument cycle.
The relative number of tagged and untagged ions after IR irradiation
is directly determined from these data and plotted as an absorption
yield versus the laser wavenumber as shown in Figure . It is noteworthy that the simultaneous
acquisition of multiple spectra does not increase the total measurement
time compared to interrogation of a single mobility-selected species.
This is because IR irradiation and ToF analysis can occur during the
mobility separation step of the next instrument cycle. The use of
our segmented cryogenic ion trap therefore decreases the total measurement
time by a factor equivalent to the number of trap compartments, which
can be dynamically adapted depending on the analytical requirements.
It is ultimately limited by the total length of the trap and the number
of electrodes used to define the electric fields. For the electrode
geometry chosen here, the minimum segment length will be approximately
1 cm.
Figure 5
(A) Arrival time profile of the ion distributions ejected from
the four different trap compartments within one instrument cycle.
(B) Exemplary ToF MS spectra of the IR irradiated ions extracted from
each trap compartment, indicating different amounts of remaining N2 tagging.
(A) Arrival time profile of the ion distributions ejected from
the four different trap compartments within one instrument cycle.
(B) Exemplary ToF MS spectra of the IR irradiated ions extracted from
each trap compartment, indicating different amounts of remaining N2 tagging.
IR Spectral Comparison
Using Principal Component Analysis (PCA)
The spectral identification
of the LNT and LNnT ions contained
in each peak of the ATD in Figure was based on visual comparison to reference spectra.
As a more rigorous approach for spectral comparison and compound identification
for a routine analysis workflow, we used PCA together with a machine-learning
algorithm to assign the spectra of analyte ions to a number of replicate
reference spectra automatically. PCA is frequently applied in different
areas of analytical chemistry[57] to reduce
the dimensionality of large data sets, that is, identify a smaller
number of variables (the principal components) that still sufficiently
describes the original set. The “scikit-learn” software
machine-learning python library[58] was then
applied for an automatic clustering of corresponding reference and
analyte spectra in principal-component space to identify the analyte.
The result of this method, applied to six reference spectra for each
component of LNT and LNnT, respectively, and the spectra of the analyte
ions from the different trap compartments (blue spectra in Figure ), is displayed in Figure . It is important
to note that this approach requires no user interpretation of the
data; the analyte spectra are automatically assigned to their reference
compounds.
Figure 6
First and second component of a PCA of the reference spectra (colored
dots) and spectra of the analyte ions from the four trap segments
(black triangles).
First and second component of a PCA of the reference spectra (colored
dots) and spectra of the analyte ions from the four trap segments
(black triangles).In addition, PCA allows
us to evaluate the uniqueness of different
wavenumber regions of reference spectra. In practice, this can be
implemented in an algorithm that selects a wavenumber region of reference
spectra where the difference in the first PCA component is maximized.
As a result, the spectra of the analyte ions only need to be recorded
over a reduced wavenumber range (i.e., where spectra exhibit the greatest
differences), thereby saving analysis time. To illustrate this, we
performed the PCA and clustering analysis on the data from the LNT/LNnT
mixture over the reduced wavenumber range of 3600–3700 cm–1, which represents less that one-fourth of the original
scan length. The algorithm is still able to assign the different components
of LNT/LNnT to their respective reference compounds unambiguously,
as shown in Figure S3.
Decreasing
Acquisition Time
Previous instrumentation,
such as that developed in our laboratory,[37,49,59] allowed for the acquisition of cryogenic
IR spectra over a spectral range of 450 cm–1 (i.e.,
the same as that shown in Figure ) on a time scale of tens of minutes. Such time scales
are not practical for a routine analytical workflow, especially when
sample analysis involves LC-MS, where analyte molecules typically
elute over a period of a few seconds to tens of seconds. A range of
instrument improvements affecting the overall signal stability as
well as the use of a continuous-wave laser, which by definition does
not exhibit inherent shot-to-shot intensity variations like pulsed
OPO systems, has helped to increase considerably the signal-to-noise
(S/N) ratio during spectral acquisition. This eases the need for extensive
signal averaging and hence shortens the data acquisition time. To
demonstrate this, we measured IR spectra of a previously studied[41] disaccharide GalNAc-α(1-3)-Gal-OMe under
a range of averaging conditions (Figure ). In a first experiment, the 450 cm–1 range was oversampled in a 7 min scan, allowing us
to average 18 successive data points without losing spectral resolution.
The averaged signal is displayed in red in Figure (top), and the individual data points are
displayed in gray. A second scan of roughly 1 min (Figure , middle) exhibits virtually
identical spectral features, albeit with slightly decreased S/N. Finally,
two replicates of a 15 s scan are displayed at the bottom of Figure . It is striking
to observe that all characteristic absorption bands can be reproduced
in a measurement in a fraction of the time that was previously necessary,
and that a second replicate measurement shows little deviation from
the first, indicating excellent repeatability of the measurements
(compare the red and black traces in Figure , bottom). It is important to note that in
comparison to the measurements shown in Figure , a shorter instrument cycle of 0.2 s was
applied here, which is a result of the lower ion mobility resolution
necessary for these experiments. Nonetheless, at these single separation-cycle
conditions the instrument exhibits an IMS resolving power of approximately
200.
Figure 7
Comparison of cryogenic IR spectra of O-methylated
GalNAc-α(1-3)-Gal acquired at different scan speeds, ranging
from 7 min (top) to 15 s (bottom). The data points for the top two
spectra are displayed in gray, while the red curves represent averaged
data points (adjacent averaging) of the oversampled spectral features.
No data averaging was applied for the 15 s scans, where the two curves
represent two replicate measurements.
Comparison of cryogenic IR spectra of O-methylated
GalNAc-α(1-3)-Gal acquired at different scan speeds, ranging
from 7 min (top) to 15 s (bottom). The data points for the top two
spectra are displayed in gray, while the red curves represent averaged
data points (adjacent averaging) of the oversampled spectral features.
No data averaging was applied for the 15 s scans, where the two curves
represent two replicate measurements.Furthermore, most diagnostic features in glycan IR spectra from
our laboratory[37,40,41,46,60] are observed
in a wavenumber range from 3580 to 3680 cm–1, which
is also supported by the success of the PCA and clustering method
applied to a reduced wavenumber range as described above. An identifying
fingerprint of the GalNAc-α(1-3)-Gal-OMe ions can thus be recorded
in as little as 5 s without losing its most unique identifying features.What are the implications of these results for the use of cryogenic
IR spectroscopy in an analytical environment? The unprecedentedly
short acquisition times alone represent a milestone toward a broader
application of IR fingerprints for the identification of compounds
in mass spectrometry. At a typical nanospray flow rate and sub-micromolar
concentrations that are compatible with the sensitivity of our ion
detection, these acquisition times imply consumption of femtomole
quantities of the analyte in a 1 min identification procedure. In
the future, analysis time will be further reduced when only a few
discrete isomer-specific wavenumbers will be probed spectroscopically
instead of a continuous wavenumber range. The frequencies and number
of points necessary for identification will then be determined by
an algorithm based on the database spectra of candidate compounds.
In cases where isomers cannot be completely resolved prior spectroscopic
fingerprinting, the spectra can be deconvoluted using isomer-specific
database entries to identify the components of the mixture. When unknown
analytes are encountered in a sample, on-board IMS2 (or
IMSn) methods can be applied, where IR fingerprints of
CID-generated fragments can be used for identification and subsequent
reconstruction of the unknown compound.[46] Further development of such methods will greatly benefit from the
fast multiplexing approach described here, which represents an important
step toward the widespread use of IR fingerprinting as an analytical
tool in high-throughput applications.
Conclusions
While
a cryogenic IR fingerprint serves as an unmistakable molecular
identifier able to distinguish between minute structural differences
in glycans or other isomeric species, the use of IR spectroscopy for
analysis has up to now been hindered by relatively low throughput
and long measurement times. To compensate the drawbacks of adding
a spectroscopic dimension, we have demonstrated a new multiplexing
approach, which along with general improvements in the cryogenic IR
fingerprinting technology, increases both the spectral acquisition
speed and its robustness. In combination with a fast isomer-separation
technique such as ultrahigh-resolution ion mobility spectrometry using
SLIM, our approach allows one to acquire IR fingerprints of multiple
isomeric species simultaneously. Depending on the time requirements
of the IMS separation step, we can record IR fingerprints over a range
of 450 cm–1 within a few seconds to minutes. However,
recording a spectrum at a few structurally characteristic wavenumbers
may already suffice for a confident identification. This will be implemented
in the future with the help of an algorithm to determine the reduced
wavenumber range from the candidate fingerprint database and could
be done without the need for user input. The resulting shortened analysis
time will ultimately allow for direct combination of the cryogenic
IR fingerprinting technique with, for example, established LC workflows
at existing flow rates. Eluting analytes could then be identified
in a high-throughput manner with the benefit of the exceptional confidence
that cryogenic IR fingerprinting provides. Once incorporated with
user-friendly software, this approach not only will have the potential
to profoundly impact the field of glycomics but can in principle help
resolve any analytical challenge where isomeric, IR-absorbing molecules
need to be analyzed.
Authors: Oscar Hernandez; Samantha Isenberg; Vincent Steinmetz; Gary L Glish; Philippe Maitre Journal: J Phys Chem A Date: 2015-04-15 Impact factor: 2.781
Authors: Antoine Masson; Michael Z Kamrath; Marta A S Perez; Matthew S Glover; U Rothlisberger; David E Clemmer; Thomas R Rizzo Journal: J Am Soc Mass Spectrom Date: 2015-06-20 Impact factor: 3.109
Authors: Ailin Li; Gabe Nagy; Christopher R Conant; Randolph V Norheim; Joon Yong Lee; Cameron Giberson; Adam L Hollerbach; Venkateshkumar Prabhakaran; Isaac K Attah; Christopher D Chouinard; Aneesh Prabhakaran; Richard D Smith; Yehia M Ibrahim; Sandilya V B Garimella Journal: Anal Chem Date: 2020-10-26 Impact factor: 6.986
Authors: Maike Lettow; Kim Greis; Márkó Grabarics; Jan Horlebein; Rebecca L Miller; Gerard Meijer; Gert von Helden; Kevin Pagel Journal: J Phys Chem A Date: 2021-05-12 Impact factor: 2.781
Authors: Robert P Pellegrinelli; Lei Yue; Eduardo Carrascosa; Ahmed Ben Faleh; Stephan Warnke; Priyanka Bansal; Thomas R Rizzo Journal: J Am Soc Mass Spectrom Date: 2022-04-19 Impact factor: 3.262
Authors: Ahmed Ben Faleh; Stephan Warnke; Priyanka Bansal; Robert P Pellegrinelli; Irina Dyukova; Thomas R Rizzo Journal: Anal Chem Date: 2022-07-07 Impact factor: 8.008