| Literature DB >> 34917959 |
Lauren S Ranard1, Ryan Kaple2, Omar K Khalique1, Vratika Agarwal2, Lavanya Bellumkonda2, Pramod Bonde2, Isaac George1, Nir Uriel1, Martin B Leon1, Torsten P Vahl1.
Abstract
An 80-year-old man with a destination left ventricular assist device (LVAD) presented with decompensated heart failure. Evaluation demonstrated numerous LVAD high power spike events, significant aortic regurgitation, and hemolysis. He underwent successful aortic valve replacement with a novel transcatheter valve and LVAD pump exchange that resulted in an improvement in his clinical status. (Level of Difficulty: Advanced.).Entities:
Keywords: AR, aortic regurgitation; CO, cardiac output; CTA, computed tomography angiography; LVAD, left ventricular assist device; LVEDD, left ventricular end diastolic dimension; NYHA, New York Heart Association; PA, pulmonary artery; PCWP, pulmonary capillary wedge pressure; RA, right atrial; RV, right ventricular; TAVR; TAVR, transcatheter aortic valve replacement; TEE, transesophageal echocardiogram; THV, transcatheter heart valve; TTE, transthoracic echocardiogram; aortic regurgitation
Year: 2021 PMID: 34917959 PMCID: PMC8642726 DOI: 10.1016/j.jaccas.2021.08.026
Source DB: PubMed Journal: JACC Case Rep ISSN: 2666-0849
Laboratory Evaluation
| Tests | Admission | Discharge From Hospital | 6 Months Follow-Up |
|---|---|---|---|
| Hemoglobin, g/dL | 12.3 | 9.4 | 13.1 |
| Platelets, 109/L | 169 | 202 | 236 |
| Lactate dehydrogenase, U/L | 1,354 | 371 | 198 |
| Plasma free hemoglobin, mg/dL | 12.1 | 15 | 2.0 |
| Haptoglobin, mg/dL | <20 | 265 | 223 |
| Albumin, g/L | 4.0 | 2.9 | 3.8 |
| Total bilirubin, mg/dL | 1.5 | 0.7 | 0.4 |
| Direct bilirubin, mg/dL | 0.4 | 0.3 | <0.2 |
| Creatinine, mg/dL | 1.88 | 1.7 | 1.61 |
| eGFR, mL/min/1.73m2 | 38 | 47 | 50 |
eGFR = estimated glomerular filtration rate.
Figure 1Left Ventricular Assist Device Device Evaluation
Device evaluation demonstrated numerous high-power alarms (examples indicated by blue arrows).
Figure 2JenaValve Trilogy Transcatheter Heart Valve System
The valve consists of a nitinol frame with porcine pericardial leaflets. The delivery system has a deflector that allows for coaxial alignment and a controller knob that facilitates axial rotation of the valve to align the radiopaque locators in the middle of each coronary cusp.
Figure 3Preprocedural Cardiac Computed Tomography Angiography Analysis
Preprocedural cardiac computed tomography angiography demonstrated an annulus perimeter of 82.9 mm, corresponding to a 27-mm valve. The aortic angulation was 50° with adequate distance to the ascending aorta. LM = left main [coronary artery]; RCA = right coronary artery; VAD = ventricular assist device.