First Transfemoral Implantation of a Novel Transcatheter Valve in an LVAD Patient With Aortic Insufficiency.

Lauren S Ranard1, Ryan Kaple2, Omar K Khalique1, Vratika Agarwal2, Lavanya Bellumkonda2, Pramod Bonde2, Isaac George1, Nir Uriel1, Martin B Leon1, Torsten P Vahl1.   

Abstract

An 80-year-old man with a destination left ventricular assist device (LVAD) presented with decompensated heart failure. Evaluation demonstrated numerous LVAD high power spike events, significant aortic regurgitation, and hemolysis. He underwent successful aortic valve replacement with a novel transcatheter valve and LVAD pump exchange that resulted in an improvement in his clinical status. (Level of Difficulty: Advanced.).
© 2021 The Authors.

Entities:  

Keywords:  AR, aortic regurgitation; CO, cardiac output; CTA, computed tomography angiography; LVAD, left ventricular assist device; LVEDD, left ventricular end diastolic dimension; NYHA, New York Heart Association; PA, pulmonary artery; PCWP, pulmonary capillary wedge pressure; RA, right atrial; RV, right ventricular; TAVR; TAVR, transcatheter aortic valve replacement; TEE, transesophageal echocardiogram; THV, transcatheter heart valve; TTE, transthoracic echocardiogram; aortic regurgitation

Year:  2021        PMID: 34917959      PMCID: PMC8642726          DOI: 10.1016/j.jaccas.2021.08.026

Source DB:  PubMed          Journal:  JACC Case Rep        ISSN: 2666-0849


History of Presentation

An 80-year-old man with ischemic cardiomyopathy after implantation of a HeartMate II (Abbott) left ventricular assist device (LVAD) in 2017 presented with New York Heart Association (NYHA) functional class IV symptoms. On examination, his heart rate was 82 beats/min, his mean arterial pressure was 92 mm Hg, and cardiovascular examination revealed an LVAD hum but was otherwise unremarkable. There was laboratory evidence of hemolysis (Table 1), and his echocardiogram demonstrated pancyclic, moderate to severe aortic regurgitation (AR). The LVAD pump speed was 9,200 rpm, and numerous high-power spike events were appreciated (Figure 1).
Table 1

Laboratory Evaluation

TestsAdmissionDischarge From Hospital6 Months Follow-Up
Hemoglobin, g/dL12.39.413.1
Platelets, 109/L169202236
Lactate dehydrogenase, U/L1,354371198
Plasma free hemoglobin, mg/dL12.1152.0
Haptoglobin, mg/dL<20265223
Albumin, g/L4.02.93.8
Total bilirubin, mg/dL1.50.70.4
Direct bilirubin, mg/dL0.40.3<0.2
Creatinine, mg/dL1.881.71.61
eGFR, mL/min/1.73m2384750

eGFR = estimated glomerular filtration rate.

Figure 1

Left Ventricular Assist Device Device Evaluation

Device evaluation demonstrated numerous high-power alarms (examples indicated by blue arrows).

Learning Objectives

To describe the evaluation of AR in patients with LVADs. To understand the management of AR and demonstrate the use of a novel TAVR system for patients with LVADs. Laboratory Evaluation eGFR = estimated glomerular filtration rate. Left Ventricular Assist Device Device Evaluation Device evaluation demonstrated numerous high-power alarms (examples indicated by blue arrows).

Past Medical History

The patient’s additional medical history included sleep apnea, chronic kidney disease, atrial fibrillation, and a previous transient ischemic attack.

Investigations

Chest computed tomography angiography (CTA) was performed to evaluate the LVAD inflow and outflow cannulas, and no evidence of thrombus or obstruction was seen. Right-sided heart catheterization demonstrated the following: elevated filling pressures; mean right atrial (RA) pressure, 15 mm Hg; right ventricular (RV) pressure, 65/14 mm Hg, mean pulmonary arterial (PA) pressure, 45 mm Hg; pulmonary capillary wedge pressure (PCWP), 28 mm Hg; PA saturation, 49%; cardiac output (CO), 3.2 L/min; and cardiac index, 1.6 L/min/m2. His transesophageal echocardiogram (TEE) confirmed pancyclic AR with a regurgitant volume of 59 mL (Video 1). An echocardiographic ramp study was performed. Baseline transthoracic echocardiogram (TTE) demonstrated an ejection fraction of 10% to 15% and a left ventricular end-diastolic dimension (LVEDD) of 6.0 cm. With higher pump speeds, there was worsening of AR and limited reduction in the LVEDD.

Management

After volume and medical optimization with afterload reduction, repeat right-sided heart catheterization was performed: RA pressure, 8 mm Hg; RV pressure, 34/9 mm Hg; mean PA pressure; 25 mm Hg; PCWP, 9 mm Hg; PA saturation, 63.8%; CO, 4.74 L/min; and cardiac index, 2.37 L/min/m2. Despite the improvement in filling pressures, these conservative measurements failed to resolve the patient’s symptoms and hemolysis. Pump thrombosis and AR both increase shear stress, thus leading to hemolysis and reduced left ventricular unloading. In patients with an LVAD and AR, blood is circulating directly from the outflow cannula back to the pump, and this can result in hemolysis. Thus, it was difficult to differentiate whether pump thrombosis or AR was causing the clinical syndrome because a positive ramp study result can also be seen in both conditions. The working diagnosis was that both processes existed. The patient was a high surgical risk, and therefore the decision was made to first proceed with transcatheter aortic valve replacement (TAVR). A subsequent LVAD pump exchange could be performed using a minimally invasive approach. Cardiac CTA demonstrated no aortic valve calcium to anchor a commercially available TAVR valve. Therefore, a novel TAVR system, the JenaValve Trilogy (JenaValve Technology, Inc.) was considered, given its ability to anchor in pure AR in the absence of calcified leaflets. The locator technology prevents migration of the device into the left ventricle after deployment, a feature that is particularly advantageous in patients with LVADs (Figure 2). The valve is under investigational use for native AR in the United States in the ALIGN-AR (JenaValve Pericardial TAVR Aortic Regurgitation Study; NCT04415047) pivotal trial; however, in May 2021, it received CE mark approval for treatment of both aortic stenosis and AR.
Figure 2

JenaValve Trilogy Transcatheter Heart Valve System

The valve consists of a nitinol frame with porcine pericardial leaflets. The delivery system has a deflector that allows for coaxial alignment and a controller knob that facilitates axial rotation of the valve to align the radiopaque locators in the middle of each coronary cusp.

JenaValve Trilogy Transcatheter Heart Valve System The valve consists of a nitinol frame with porcine pericardial leaflets. The delivery system has a deflector that allows for coaxial alignment and a controller knob that facilitates axial rotation of the valve to align the radiopaque locators in the middle of each coronary cusp. Cardiac CTA demonstrated suitable anatomy, and the valve was implanted under an emergency use protocol with the patient under general anesthesia and TEE guidance (Figure 3). Transfemoral access was obtained. An 18-F dedicated sheath was advanced to the sinotubular junction. An aortogram was performed to determine the deployment view (Video 2). Next, the valve was advanced through the sheath. The locators were centered within each of the aortic cusps (Videos 3 and 4). After confirmation of the correct position, the valve was deployed without changes in LVAD speed (Video 5). The procedure was uncomplicated, requiring only 16 minutes from sheath insertion to sheath removal. TEE assessment demonstrated a well-positioned valve with closed leaflets and no evidence of transvalvular or paravalvular regurgitation (Video 6).
Figure 3

Preprocedural Cardiac Computed Tomography Angiography Analysis

Preprocedural cardiac computed tomography angiography demonstrated an annulus perimeter of 82.9 mm, corresponding to a 27-mm valve. The aortic angulation was 50° with adequate distance to the ascending aorta. LM = left main [coronary artery]; RCA = right coronary artery; VAD = ventricular assist device.

Preprocedural Cardiac Computed Tomography Angiography Analysis Preprocedural cardiac computed tomography angiography demonstrated an annulus perimeter of 82.9 mm, corresponding to a 27-mm valve. The aortic angulation was 50° with adequate distance to the ascending aorta. LM = left main [coronary artery]; RCA = right coronary artery; VAD = ventricular assist device. Given the persistent hemolysis and power spikes indicating LVAD pump dysfunction, the patient later underwent a minimally invasive LVAD pump exchange (HeartMate II [Abbott] explant, HeartMate III implant) through subcostal and left thoracotomy access. The pump rotor bearing area was noted to have thrombus present on explantation.

Discussion

AR is a frequent complication related to long-term LVAD support and progresses with increased support duration. At least moderate AR is estimated to develop in ∼30% of patients who remain on continuous-flow LVAD support for 3 years and by 4% per month of support (1,2). The development of AR is associated with an increase in hospitalizations, reduced survival, and adverse left ventricular remodeling (3,4). The assessment of AR in patients with LVADs is complex and requires a multiparametric approach. Traditional TTE parameters commonly underestimate the severity because regurgitant flow is typically pancyclic, thus making vena contracta and the regurgitant jet appear narrow. An outflow cannula examination, including assessment of the systolic-to-diastolic peak velocity ratio and diastolic acceleration, is often helpful to assess severity. Moderate or greater AR is indicated by a systolic-to-diastolic ratio <5.0 and/or diastolic acceleration >49.0 cm/s2 (5). The optimal management of LVAD-associated AR remains unknown. Medical management focuses on volume optimization and afterload reduction. LVAD device management should target the lowest possible pump speed because the development of a reverse transaortic pressure gradient will result in worsening of AR. Surgical treatment options include aortic valve replacement; however, the operative mortality in these patients is significant. There is limited experience with off-label TAVR for AR in patients with LVADs. Single-center studies have demonstrated variable rates of periprocedural mortality and valve migration requiring implantation of a second TAVR device (6,7). The novel transcatheter heart valve (THV) used in this patient has a unique design compared with most other TAVR systems; the locator technology aids with anchoring and prevents ventricular embolization, perhaps making it a more optimal THV in patients with LVADs. A first generation of this valve has been successfully implanted transapically in this clinical setting (8). An alternative transcatheter approach is aortic valve closure with an Amplatzer Septal Occluder (Abbott); an improvement in hemodynamics has been observed with this approach; however, this makes patients pump dependent and vulnerable if there is pump malfunction or failure (9). Finally, if the patient is a transplant candidate, urgent heart transplantation can be considered.

Follow-Up

The patient demonstrated symptomatic improvement after TAVR and LVAD exchange. At 6-month clinical follow-up, he described NYHA functional class II symptoms, and his laboratory test results demonstrated resolution of hemolysis. On 6-month TTE, the prosthesis was stable without any paravalvular or transvalvular AR.

Conclusions

The use of commercially available TAVR devices is challenging because of issues with valve anchoring. This first case example demonstrates that this novel THV may be a feasible option for TAVR in the patients with LVADs.

Funding Support and Author Disclosures

Dr Khalique is part of a core laboratory contracting with JenaValve but has not received any direct compensation; has received consulting fees from Abbott Structural and Boston Scientific; and has received Speakers Bureau fees from Edwards Lifesciences. Dr George has received consulting fees from Cardiomech, Mitremedical, Atricure, Vdyne, Valcare Medical, DurVena, MITRx, and Johnson & Johnson. Dr Leon has reported institutional clinical research grants from Abbott, Boston Scientific, Edwards Lifesciences, Medtronic, and JenaValve. Dr Vahl has reported institutional funding to Columbia University Irving Medical Center from Boston Scientific, Edwards Lifesciences, JenaValve, Medtronic, and Siemens Healthineers; and has received consulting fees from Abbott Vascular, Boston Scientific, and Siemens Healthineers. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.
  8 in total

1.  Accurate Quantification Methods for Aortic Insufficiency Severity in Patients With LVAD: Role of Diastolic Flow Acceleration and Systolic-to-Diastolic Peak Velocity Ratio of Outflow Cannula.

Authors:  Jonathan Grinstein; Eric Kruse; Gabriel Sayer; Savitri Fedson; Gene H Kim; Ulrich P Jorde; Colleen Juricek; Takeyoshi Ota; Valluvan Jeevanandam; Roberto M Lang; Nir Uriel
Journal:  JACC Cardiovasc Imaging       Date:  2015-12-09

2.  Short-term results with transcatheter aortic valve replacement for treatment of left ventricular assist device patients with symptomatic aortic insufficiency.

Authors:  Amin Yehya; Vivek Rajagopal; Christopher Meduri; James Kauten; Morris Brown; Lynn Dean; Julie Webster; Arun Krishnamoorthy; Tara Hrobowski; David Dean
Journal:  J Heart Lung Transplant       Date:  2019-03-11       Impact factor: 10.247

3.  Percutaneous transcatheter aortic valve closure successfully treats left ventricular assist device-associated aortic insufficiency and improves cardiac hemodynamics.

Authors:  Kishan S Parikh; Amit K Mehrotra; Mark J Russo; Roberto M Lang; Allen Anderson; Valluvan Jeevanandam; Benjamin H Freed; Jonathan D Paul; Janet Karol; Sandeep Nathan; Atman P Shah
Journal:  JACC Cardiovasc Interv       Date:  2013-01       Impact factor: 11.195

4.  Prevalence, significance, and management of aortic insufficiency in continuous flow left ventricular assist device recipients.

Authors:  Ulrich P Jorde; Nir Uriel; Nadav Nahumi; David Bejar; Jose Gonzalez-Costello; Sunu S Thomas; Jason Han; Kerry A Morrison; Sophie Jones; Susheel Kodali; Rebecca T Hahn; Sofia Shames; Melana Yuzefpolskaya; Paolo Colombo; Hiroo Takayama; Yoshifumi Naka
Journal:  Circ Heart Fail       Date:  2014-01-10       Impact factor: 8.790

5.  Concomitant transcatheter aortic valve and left ventricular assist device implantation.

Authors:  Christina Baum; Moritz Seiffert; Hendrik Treede; Hermann Reichenspurner; Tobias Deuse
Journal:  ASAIO J       Date:  2013 Jan-Feb       Impact factor: 2.872

Review 6.  De novo aortic insufficiency during long-term support on a left ventricular assist device: a systematic review and meta-analysis.

Authors:  Salil V Deo; Vikas Sharma; Yang Hyun Cho; Ishan K Shah; Soon J Park
Journal:  ASAIO J       Date:  2014 Mar-Apr       Impact factor: 2.872

7.  Impact of worsening of aortic insufficiency during HeartMate 3 LVAD support.

Authors:  Teruhiko Imamura; Nikhil Narang; Gene Kim; Daisuke Nitta; Takeo Fujino; Ann Nguyen; Jonathan Grinstein; Daniel Rodgers; Takeyoshi Ota; Jayant Raikhelkar; Valluvan Jeevanandam; Gabriel Sayer; Nir Uriel
Journal:  Artif Organs       Date:  2020-10-23       Impact factor: 3.094

8.  Aortic Insufficiency During Contemporary Left Ventricular Assist Device Support: Analysis of the INTERMACS Registry.

Authors:  Lauren K Truby; A Reshad Garan; Raymond C Givens; Brian Wayda; Koji Takeda; Melana Yuzefpolskaya; Paolo C Colombo; Yoshifumi Naka; Hiroo Takayama; Veli K Topkara
Journal:  JACC Heart Fail       Date:  2018-11       Impact factor: 12.035