| Literature DB >> 34917780 |
Jan Hofmaier1, Franziska Walter1, Indrawati Hadi1, Maya Rottler1, Rieke von Bestenbostel1, George Dedes2, Katia Parodi2, Maximilian Niyazi1, Claus Belka1,3, Florian Kamp1,4.
Abstract
Margin concepts in proton therapy aim to ensure full dose coverage of the clinical target volume (CTV) in presence of setup and range uncertainty. Due to inter-observer variability (IOV), the CTV itself is uncertain. We present a framework to evaluate the combined impact of IOV, setup and range uncertainty in a variance-based sensitivity analysis (SA). For ten patients with skull base meningioma, the mean calculation time to perform the SA including 1.6 × 104 dose recalculations was 59 min. For two patients in this dataset, IOV had a relevant impact on the estimated CTV D95% uncertainty.Entities:
Keywords: Inter-observer variability; Proton therapy; Range uncertainty; Sensitivity analysis; Setup uncertainty
Year: 2021 PMID: 34917780 PMCID: PMC8645917 DOI: 10.1016/j.phro.2021.11.005
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Convergence plots for Sl and STl for one patient (panels A and B) and DVHs for CTVSTAPLE for the four patients with the largest overall D95% uncertainties (panels C to F). The variability of the DVH in presence of setup uncertainty, range uncertainty and IOV is visualized by the shaded areas (68% and 95% CIs). The solid line indicates the mean value over all simulated error scenarios.
Uncertainty and sensitivity analysis results for the D95% for (CTV)STAPLE. For each patient, the mean value and 95% and the 68% CIs have been calculated. The relative contribution to the overall uncertainty is broken down to first order indices Ssetup, Srange and SIOV, higher order indices with involvement of IOV (SIIOV) and higher order indices without involvement of IOV (SIother).
| pat. | mean [Gy] | CI95% [Gy] | CI68% [Gy] | Ssetup | Srange | SIOV | SIIOV | SIother |
|---|---|---|---|---|---|---|---|---|
| 1 | 1.71 | 1.62–1.74 | 1.69–1.73 | 0.21 | 0.29 | 0.11 | 0.10 | 0.29 |
| 2 | 1.56 | 1.16–1.73 | 1.35–1.72 | 0.12 | 0.35 | 0.34 | 0.09 | 0.10 |
| 3 | 1.66 | 1.49–1.73 | 1.59–1.71 | 0.14 | 0.11 | 0.62 | 0.01 | 0.12 |
| 4 | 1.70 | 1.59–1.73 | 1.68–1.72 | 0.19 | 0.14 | 0.10 | 0.12 | 0.45 |
| 5 | 1.73 | 1.71–1.74 | 1.72–1.74 | 0.37 | 0.31 | 0.00 | 0.14 | 0.18 |
| 6 | 1.73 | 1.70–1.74 | 1.72–1.73 | 0.20 | 0.29 | 0.03 | 0.04 | 0.44 |
| 7 | 1.67 | 1.45–1.73 | 1.61–1.73 | 0.25 | 0.53 | 0.00 | 0.02 | 0.20 |
| 8 | 1.69 | 1.58–1.74 | 1.65–1.73 | 0.23 | 0.49 | 0.04 | 0.05 | 0.19 |
| 9 | 1.62 | 1.25–1.73 | 1.48–1.71 | 0.13 | 0.70 | 0.01 | 0.00 | 0.16 |
| 10 | 1.73 | 1.70–1.74 | 1.72–1.74 | 0.26 | 0.24 | 0.02 | 0.06 | 0.42 |