| Literature DB >> 34903069 |
Gaston F Coutsiers Morell1, Yuli Berlin-Broner1, Carlos Flores-Mir1, Giseon Heo1.
Abstract
OBJECTIVE: To quantify tooth volume differences from extracted teeth when using three different three-dimensional (3D) computed tomography (CT)-based imaging modalities.Entities:
Keywords: cone-beam computed tomography; ex vivo; micro-CT; orthodontically induced root resorption; orthodontics; root resorption; tooth root; tooth volume
Mesh:
Year: 2021 PMID: 34903069 PMCID: PMC9160778 DOI: 10.1177/14653125211066106
Source DB: PubMed Journal: J Orthod ISSN: 1465-3125
Figure 1.Three-dimensional volumetric representation of one of the teeth.
Volume (in mm3) of each tooth scanned by micro-CT, CBCT at 0.25-mm and 0.30-mm voxel sizes and their comparisons in volume difference and percentage increase.
| Tooth | Micro-CT | CBCT 0.25 | CBCT 0.30 | Difference between CBCT 0.25 and micro-CT | Increase between CBCT 0.25 and micro-CT (%) | Difference between CBCT 0.30 and micro-CT | Increase between CBCT 0.30 and micro-CT (%) | Difference between CBCT 0.30 and CBCT 0.25 | Increase between CBCT 0.30 and CBCT 0.25 (%) |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 519.5 | 555.6 | 634.1 | 36.1 | 6.9 | 114.6 | 22.1 | 78.5 | 14.1 |
| 2 | 579.1 | 652.0 | 716.1 | 72.9 | 12.6 | 137.0 | 23.7 | 64.1 | 9.8 |
| 3 | 378.3 | 427.6 | 479.7 | 49.3 | 13.0 | 101.4 | 26.8 | 52.1 | 12.2 |
| 4 | 466.0 | 537.0 | 594.4 | 71.0 | 15.2 | 128.4 | 27.6 | 57.4 | 10.7 |
| 5 | 467.3 | 538.2 | 593.5 | 70.9 | 15.2 | 126.2 | 27.0 | 55.3 | 10.3 |
| 6 | 463.2 | 648.5 | 728.6 | 185.3 | 40.0 | 265.4 | 57.3 | 80.1 | 12.4 |
| 7 | 527.7 | 605.6 | 664.5 | 77.9 | 14.8 | 136.8 | 25.9 | 58.9 | 9.7 |
| 8 | 505.8 | 618.6 | 694.5 | 112.8 | 22.3 | 188.7 | 37.3 | 75.9 | 12.3 |
| 9 | 674.7 | 751.1 | 832.9 | 76.4 | 11.3 | 158.2 | 23.4 | 81.8 | 10.9 |
| 10 | 850.4 | 963.3 | 1067.0 | 112.9 | 13.3 | 216.6 | 25.5 | 103.7 | 10.8 |
| 11 | 554.1 | 594.7 | 657.6 | 40.6 | 7.3 | 103.5 | 18.7 | 62.9 | 10.6 |
| 12 | 721.0 | 793.7 | 877.9 | 72.7 | 10.1 | 156.9 | 21.8 | 84.2 | 10.6 |
| Mean | 558.9 | 640.5 | 711.7 | 81.6 | 15.2 | 152.8 | 28.1 | 71.2 | 11.2 |
CBCT, cone-beam computed tomography; CT, computed tomography.
Mean differences, SE and 95% CIs for all pairwise comparisons.
| Comparisons of scans | Mean Difference (SE) | 95% CI |
|---|---|---|
| CBCT 0.25 vs. micro-CT | 81.6 ± 11.7 | (55.9–107.2) |
| CBCT 0.30 vs. micro-CT | 152.8 ± 14.1 | (121.7–183.9) |
| CBCT 0.30 vs CBCT 0.25 | 71.2 ± 4.4 | (61.5–80.9) |
CBCT, cone-beam computed tomography; CI, confidence interval; CT, computed tomography; SE, standard error.