Tao Hu1, Wengen Li1, Kangkai Xu1, Ke Chen1, Xiao Li1, Hong Yi1, Zhonghua Ni1. 1. School of Mechanical Engineering, and Jiangsu Key Laboratory for Design and Manufacture of Micro-Nano Biomedical Instruments, Southeast University, Nanjing 211189, China.
Abstract
Diabetes is a public health problem characterized by hyperglycemia, high mortality, and morbidity. A simple, rapid, and sensitive glucose detection method for diabetes screening and health self-management of patients with diabetes is of great significance. Therefore, an attractive urine glucose (UG) analyzer with advantages of fastness, sensitivity, and portability was developed. A cadmium telluride quantum dots (CdTe QDs)@glucose oxidase (GOx) aerogel circular array sensor can emit visible red fluorescence when excited by a 365 nm ultraviolet light source inside the analyzer. When urine samples containing glucose were dropped onto the sensor, glucose was oxidized by GOx to produce hydrogen peroxide (H2O2), which quenched the red fluorescence of CdTe QDs. The fluorescence images of the sensor were obtained using a CCD camera, and the linear relationship between the glucose concentration and the gray value of the fluorescence image was established. The analyzer shows good sensitivity (LOD, 0.12 mM) with a wide linear range of 0.12-26 mM. Based on the linear relation, the software of the analyzer was written in the C++ language, which can automatically give the gray value of the image and the corresponding glucose concentration. The UG analyzer was used for the detection of a large number clinical samples and compared with a variety of UG test papers, which all showed good detection performance. The novel analyzer we proposed has an important significance in the screening of diabetes and the self-management of diabetic patients.
Diabetes is a public health problem characterized by hyperglycemia, high mortality, and morbidity. A simple, rapid, and sensitive glucose detection method for diabetes screening and health self-management of patients with diabetes is of great significance. Therefore, an attractive urine glucose (UG) analyzer with advantages of fastness, sensitivity, and portability was developed. A cadmium telluride quantum dots (CdTe QDs)@glucose oxidase (GOx) aerogel circular array sensor can emit visible red fluorescence when excited by a 365 nm ultraviolet light source inside the analyzer. When urine samples containing glucose were dropped onto the sensor, glucose was oxidized by GOx to produce hydrogen peroxide (H2O2), which quenched the red fluorescence of CdTe QDs. The fluorescence images of the sensor were obtained using a CCD camera, and the linear relationship between the glucose concentration and the gray value of the fluorescence image was established. The analyzer shows good sensitivity (LOD, 0.12 mM) with a wide linear range of 0.12-26 mM. Based on the linear relation, the software of the analyzer was written in the C++ language, which can automatically give the gray value of the image and the corresponding glucose concentration. The UG analyzer was used for the detection of a large number clinical samples and compared with a variety of UG test papers, which all showed good detection performance. The novel analyzer we proposed has an important significance in the screening of diabetes and the self-management of diabetic patients.
Diabetes is a global
disease; more than 400 million people have
been tortured by diabetes since 2015, and the number of patients is
still increasing. It is expected that the number will exceed 600 million
by 2045.[1,2] As a result, both the growth of medical
expenditure and the decline in the quality of life are becoming serious
challenges.[3] Early diagnosis is of great
significance for the management of diabetes and preventing or delaying
the occurrence of diabetes complications.[4] However, most patients with type 2 diabetes are asymptomatic, and
more than 30% of patients with type 2 diabetes are not diagnosed until
systemic complications occur.[5] Therefore,
screening for diabetes, especially in high-risk areas, is essential
to reduce the negative effects of diabetes.Until now, traditional
methods of diabetes screening and monitoring
include evaluation of fasting plasma glucose (FPG), hemoglobin A1c
(HbA1c), and FPG in combination with 2 h plasma glucose (2 h-PG) after
a 75 g oral glucose tolerance test.[6−9] However, due to low efficiency, a long time,
and high cost, the application of these methods is limited.[10] More importantly, frequent blood sampling leads
to poor patient compliance, which is not conducive to diabetes screening
and diabetic self-management. Therefore, a simple, user-friendly assistance
approach with excellent sensitivity and selectivity is urgently needed.The measurement of urine glucose (UG) appears to be an attractive
choice. Current research shows that UG detection can reliably reflect
the blood glucose concentration;[11,12] the higher
the UG concentration, the more the patient’s blood glucose
exceeds the normal. Also, previous studies have shown that compared
with blood self-monitoring, quantitative UG monitoring has similar
effects in maintaining blood glucose control.[13,14] The appearance of glucose in urine is an important indicator of
many related diseases.[15] UG detection has
great advantages in early screening of diabetes, which avoids the
pain of blood sampling needles and simplifies the testing process.At present, various methods are used in UG detection, including
electrochemical devices,[16−18] surface-enhanced Raman scattering
(SERS),[19−21] colorimetric sensors,[22−24] and so on. Glucose reacts
with enzymes or enzyme-like substances to generate electrons to cause
changes in current. Electrochemical workstations were used to capture
the changes in current to determine UG concentrations in the electrochemical
methods; in the SERS detection method, glucose causes the detection
probe to agglomerate or change the surface structure, which changes
the Raman intensity. Researchers use this Raman spectrum change to
determine the UG concentration to be measured. In colorimetric detection,
glucose produces H2O2 under the action of glucose
oxidase, and hydrogen peroxide changes the color of the color probe;
commonly used color probes are quantum dots (QDs),[25,26] metal nanoparticles,[27,28] TMB,[29−31] etc. The above
methods have been widely used in UG detection; these methods have
a wider linear range, high sensitivity, and a low detection limit.
However, they suffer from high cost, professional operation, complicated
material synthesis, and modification processes. Moreover, most of
these methods can only be performed in the laboratory. A universal,
complete, and independent detection method is still vacant.In this work, a portable UG analyzer was designed and manufactured.
Its size is much smaller compared to medical instruments, which is
very convenient to carry. Its core is the CdTe QDs@GOx aerogel circular
array sensor. Through the fluorescence quenching reaction between
glucose and the QDs@GOx aerogel, the linear relationship between the
glucose concentration and the gray value of the fluorescence image
is established. The linear range is 0–26 mM, and the detection
limit is as low as 0.12 mM. The analysis software is written on the
basis of this linear model; it can automatically read the gray value
of the fluorescence picture and display the UG concentration, which
greatly simplifies the data processing process. The UG analyzer was
used for the detection of large clinical samples and showed good detection
performance.
Results and Discussion
Detection Mechanism and
Process
The detection principle
and process are shown in Scheme . When UG solution is dropped onto the circular sensors,
glucose produces H2O2 under the catalytic oxidation
of GOx. H2O2 destroys the surface structure
of CdTe QDs and quenches the fluorescence of QDs.[34,35] Fluorescence changes in the sensors are captured by high-resolution
industrial cameras and transmitted to the PC. The gray value of the
picture is measured by the analysis software and calculated to show
the glucose concentration. The whole process is very fast and efficient,
and professional operation training is not needed.
Scheme 1
Detection Principle
and Process
Characteristics of the
CdTe QD Aerogel
As described
in the Experimental Section, CdTe QDs were
synthesized by a simple one-step hydrothermal method. Figure a shows the fluorescence spectrum
and ultraviolet (UV) spectrum of the synthesized CdTe QD aerogel.
As shown in Figure a, the UV absorption peak wavelength of the prepared CdTe QDs is
580 nm, and the fluorescence emission peak wavelength is 610 nm, which
is dazzling red under a UV lamp. Figure b is a scanning electron micrograph of the
CdTe QD aerogel prepared by the freeze-drying method in the circular
reaction area. The three-dimensional (3D) porous structure is clearly
shown in Figure b;
this 3D porous structure has a large specific surface area, which
improves the reaction rate. Besides, the stabilizer of QDs is a polypeptide,
which makes the 3D structure with good biocompatibility and provides
a good site for the attachment of glucose oxidase, which improves
the stability of the sensor.
Figure 1
(a) Fluorescence and UV spectra of CdTe QDs;
(b) SEM images of
CdTe-GOx aerogels in the reaction region.
(a) Fluorescence and UV spectra of CdTe QDs;
(b) SEM images of
CdTe-GOx aerogels in the reaction region.
Hardware Structure of the Portable UG Analyzer
The
UG analyzer presents a closed box as a whole, which was made with
3D printing. The physical map can be seen in Figure S2. Since our analysis principle is fluorescence change, the
completely enclosed structure can avoid the interference of ambient
light and improve the stability and accuracy. The schematic diagram
of the structure of the UG analyzer is shown in Figure . There is a sensor placement area on the
base of the instrument. In order to excite the light intensity field
uniformly, two UV light sources are symmetrically arranged on both
sides of the placement area. The UV excitation wavelength is 365 nm.
On the top of the instrument, an industrial camera is fixedly arranged,
which is used to clearly capture the fluorescence image of the sensor
and store and transmit it to the PC to perform software analysis and
result readout. All subsequent experiments in this article are done
using this instrument (Figure ).
Figure 3
(a) Linear relationship between the change
of fluorescence gray
value and glucose concentration; (b) sensor fluorescence gradient
image [from left to right: (a) 0 mM, (b) 0.085 mM, (c) 0.17 mM, (d)
0.34 mM, (e) 0.675 mM, (f) 1.35 mM, (g) 4.33 mM, (h) 10.8 mM, and
(i) 26 mM].
Figure 2
Schematic diagram of the structure of the UG analyzer.
Schematic diagram of the structure of the UG analyzer.
Optimization of Analytical Instruments
In order to
get the fastest response time, the optimal detection time is explored.
10 μL of 10.8 mM glucose solution was dropped into the circular
reaction area of the sensor. Fluorescence images are obtained at regular
intervals and analyzed for gray values. The relationship between the
gray value of the fluorescence image and the reaction time is obtained,
as shown in Figure S3. The concentration
of quantum dots was diluted 2.5,5,10 and 15 times, respectively. Glucose
solutions of 4.33, 10.8, and 26 mM were dropped onto the sensors prepared
with different concentrations of QDs. After 5 min of reaction, the
fluorescence image of the reaction area was taken and calculated in
the gray scale. As shown in Figure S4,
with the reaction progressing, the fluorescent gray value of the reaction
zone gradually decreases. When the reaction time reaches 5 min, the
gray value tends to be stable. In order to improve the sensitivity
of the sensor and the need for rapid detection, 5 min was determined
as the best reaction time.Different concentrations of QDs will
affect the QDs fluorescence quenching. The gray value
change rate of different QD concentrations is calculated, as shown
in Table S1. When the QD stock solution
is diluted 10 times, the rate of change of the gray value caused by
4.33 and 10.8 mM glucose is the largest, reaching 54.57 and 72.01%,
respectively, and the rate of change of the gray value caused by 26
mM is not far from other concentrations. This means that when diluted
10 times, the QD probe has the greatest detection sensitivity and
the most sensitive response to changes in glucose concentration. Therefore,
comprehensively, the dilution of 10 is determined as the optimal concentration
of the QD solution.
Establishment of the Glucose Detection Curve
Different
concentrations of UG solutions are dropped onto the circular reaction
area of the sensor. After 5 min of reaction, the fluorescence image
of the reaction area is captured by the camera; the result is shown
in Figure b. It can be observed from the figure that when the
glucose concentration gradually increases, the red fluorescence in
the reaction area gradually decreases. By comparing and analyzing
the relationship between glucose concentration and the gray value,
a linear curve of glucose concentration detection is established,
as shown in Figure a. It can be seen from Figure a that from 0.085 to 26 mM, there is a good linear relationship
between I/I0 and glucose
concentration, and the correlation coefficient is 0.9933 (the gray
value of the blank sample is defined as I0, and the gray value of the reaction area with different glucose
concentrations is defined as I). The linear equation
is (glucose concentration is defined as C)(a) Linear relationship between the change
of fluorescence gray
value and glucose concentration; (b) sensor fluorescence gradient
image [from left to right: (a) 0 mM, (b) 0.085 mM, (c) 0.17 mM, (d)
0.34 mM, (e) 0.675 mM, (f) 1.35 mM, (g) 4.33 mM, (h) 10.8 mM, and
(i) 26 mM].The limit of detection (LOD) is
0.12 mM.
Software Interface and the Operation Process
The software
interface diagram based on the OPENCV library is shown in Figure . The interface is
simple and intuitive. “Read” is used to read the fluorescence
picture of the sensor transmitted by the camera. After the image is
read, it will be displayed in the center of the interface, and the
fluorescence changes of the sensor can be observed visually by the
naked eye. This provides the qualitative judgment basis. When “Get
CCNC” is clicked, the data bar on either side of the interface
quickly displays the gray values for each circular area and the corresponding
glucose concentration, and precise quantitative results are measured.
The whole operation interface is very user-friendly, and users who
do not need special training can operate. This enhances the utility
potential of the portable diabetes meter. The logic flow of the software
is shown in Figure below. Since the position of the paper-based sensor in the UG analyzer
is fixed, the detection area was selected by the software directly
based on the fixed coordinates, the copyTo() function was used to
cut and enlarge it and display it in the software interface, and then
the cvtColor() function (using the CV_BGR2GRAY parameter) was used
to convert the color picture into a grayscale picture, which is convenient
to directly obtain the grayscale value of the pixel. The getcct()
function is written to calculate the gray value of the detection area
and calculate the UG concentration.
Figure 4
Software interface of the analyzer.
Figure 5
Logic flow chart of UG analysis software.
Software interface of the analyzer.Logic flow chart of UG analysis software.
Selectivity and Stability
The selectivity of the sensor
was studied. 10.8 mM glucose, sucrose, fructose, maltose, lactose,
and urea as well as 0.45 mM uric acid and 0.6 mM ascorbic acid are
configured. The concentrations of uric acid and ascorbic acid are
the upper limits of urine in healthy humans. 10 μL of various
test solutions is dropped onto the circular reaction area of the sensor,
and the resulting fluorescence gray value is analyzed. The results
are shown in Figure a; as can be seen in Figure a, glucose can cause severe fluorescence quenching of the
sensor, and other substances have basically no effect on fluorescence
quenching. The excellent selectivity of the sensor is verified.
Figure 6
(a) Sensor
selectivity; (b) sensor stability. The detection of
real clinical samples.
(a) Sensor
selectivity; (b) sensor stability. The detection of
real clinical samples.The stability of the
sensor was also studied. The sensor is placed
in a plastic box, and the air is evacuated using a vacuum machine.
Then, the plastic box containing the sensor is placed in a refrigerator
at −20 °C. The glucose solutions of 0 and 10.8 mM are
dropped onto the sensor every 15 days, and the fluorescence gray value
is analyzed and compared. The results are shown in Figure b; as can be seen in Figure b, within 45 days,
the gray value of the sensor fluorescence remained basically unchanged,
which meets the requirements of industrial and clinical use.In order to verify the reliability of the UG analyzer for the detection
of different urine samples, 208 clinical samples were detected in
cooperation with Nanjing Zhongda Hospital. In order to stabilize the
pH value of the samples, all urine samples were diluted 10 times with
phosphate buffered saline (PBS) buffer with a pH of 7.4. The measurement
results are compared with those of medical instruments (Uretest-500B
Automatic Urine Chemistry Analyzer of URIT). Table shows the measurement results and comparison
of the two instruments. In contrast, we have a good recovery rate
and a higher positive detection rate, which mean that we have not
missed any possible suspected positive cases, which is of great significance
for the early screening of diabetes.
Table 1
Comparison
of Test Results of Two
Instruments
UG gradient
Uretest-500B
ours
recovery (%)
–(0–5.6 mM)
52
49
94.23
+(5.6–14 mM)
38
40
105.26
++(14–28 mM)
46
38
82.61
≥+++(≥28 mM)
72
81
112.5
In addition, five commonly used UG
test papers were purchased from
the market for comparison tests with our instruments. The sample results
tested using medical instruments were used as a standard control group.
The comparative test result is shown in Figure ; it can be seen intuitively from the stacking
graph that our test results are closest to the standard values, and
the proportions of each concentration gradient are roughly equal.
Figure 7
UG test
paper comparison test stacking chart.
UG test
paper comparison test stacking chart.In order to quantitatively indicate the accuracy difference between
various UG test papers, the standard deviation of each UG test paper
and the standard control group was calculated. The calculation results
are shown in Table . From the comparison of specific values, we can see that the standard
deviation of the instrument we developed is the smallest. From the
above multi-sample test and comparison, it can be seen that the UG
analyzer developed by us has reliable accuracy and stability and has
the value of clinical practical use.
Table 2
Comparison
of Standard Deviation Values
of UG Test Papers
standard
deviation
ACCU
ANJIAN
GAOERBAO
URIT
AVE
ours
standard
27.50
76.58
57.83
28.98
28.53
14.07
Conclusions
Herein,
based on the CdTe QDs@GOx aerogel circular array sensor,
we designed and manufactured a portable, highly sensitive, and selective
UG analyzer. The linear range is 0–26 mM, and the detection
limit is as low as 0.12 mM. User-friendly and easy-to-use analysis
software has also been developed to circumvent the responsible data
post-processing process and improve detection efficiency and capabilities.
In the detection of a large number of clinical samples and comparison
with a variety of UG test paper tests, the analyzer has demonstrated
good detection performance. Our work provides an important approach
for diabetes screening, early treatment, and patient self-health management.
Experimental
Section
Materials
All chemicals were of analytical grade. 2.5
cadmium chloride hydrate (CdCl2·2.5H2O)
(Alfa Aesar), sodium borohydride (NaBH4) (Alfa Aesar), l-glutathione (L-GSH) (Alfa Aesar), sodium tellurate (Na2TeO3) (Alfa Aesar), sodium hydroxide (NaOH) (Chengdu
KESHI Company), PBS buffer (Biosharp, 10 mM), isopropanol (C3H8O) (Chengdu KESHI Company), polyimide (Beijing Yinuokai
Technology Co., Ltd.), ethanol (C2H6O) (Sinopharm
Reagent), and polydimethylsiloxane (Dow Corning Corp., USA) were the
chemicals purchased. Ultrapure water (≥18 MΩ·cm2) was prepared using a Millipore water purification system.
Synthesis of Cadmium Telluride QDs (CdTe QDs)
The red-emission
CdTe QDs were synthesized based on the reported method.[32,33] A 100 mL beaker was dried with nitrogen, and 55 mL of ultrapure
water was added to it. 0.1028 g of CdCl2·2.5H2O and 0.1844 g of L-GSH were weighed and added to the beaker;
a Teflon A25 magnetic stirrer was added to the beaker, and NaOH solution
with a concentration of 0.5 mM was prepared. The pH value of the mixed
solution in the beaker was detected using a Mettler pH meter. The
solution was continuously stirred with a magnetic stirrer, and NaOH
solution was added in drops to adjust the pH value to 10.5. Then,
0.0222 g of Na2TeO3 and 0.0038 g of NaBH4 were added to the mixed solution successively and stirred
for 30 min; after the completion of the mixing, the solution was placed
in a flask and was heated in an oil bath at 110 °C for 6 h. After
heating, 10 mL of ethylene glycol was added to the solution and the
CdTe QD solution was obtained by centrifugation at 5000 rpm for 6
min.
Preparation of the CdTe QDs@GOx Aerogel Circular Array Sensor
The PDMS prepolymer and cross-linker were mixed in a weighing boat
at a mass ratio of 10:1; the mixture was stirred for 5 min and then
was put in a vacuum storage system to evacuate until the bubbles were
removed. The polyimide film was flattened on the glass slide, and
the excess was cut off; then, the glass slide was placed on the spin
coating platform of the spin coater, and 4 g of the PDMS mixed reagent
was poured on it, spin coating at 900 rpm for 55 s. After the spin
coating was completed, the glass slide was placed in an oven at 80 °C
for 1 h; a laser processing machine was used to process the cured
composite film into the desired shape. The basic dimensions of the
PDMS film are 50 cm long and 22 cm wide (details can be seen in Figure S1). After the processing is completed,
the composite film is placed in ethanol for ultrasonic cleaning for
5 min and then placed on a dust-free paper to dry naturally. The dried
composite film PDMS is put with the face up with the glass slide into
the plasma cleaner for bonding. After the bonding is completed, it
is placed in a 120 °C oven for 3 h to enhance the bonding strength.
The dried chips were placed with the face up in a plasma cleaner for
2 min to make the chip reaction section hydrophilic; then, 10 μL
of CdTe-GOx solution (the GOX concentration is 25 mM) was added dropwise
to each round reaction section. The chips were freezed at −20
°C for 12 h. After the freezing is completed, it is placed in
a freeze dryer and dried for 3 h. After taking out, a circular array
sensor is obtained.
Establishment of the UG Linear Curve
Glucose was added
to 2 mL of urine from healthy volunteers to get the final concentrations
of 0, 0.085, 0.17, 0.34, 0.675, 1.35, 4.33, 10.8, and 26 mM. 10 μL
of glucose solution of the abovementioned concentrations was dropped
onto the circular array sensor. 5 min later, the change in the fluorescence
color of the chip was captured by a high-resolution industrial camera
under the 365 nm UV-LED of the analyzer and a linear curve for glucose
detection was established.
Authors: N H Cho; J E Shaw; S Karuranga; Y Huang; J D da Rocha Fernandes; A W Ohlrogge; B Malanda Journal: Diabetes Res Clin Pract Date: 2018-02-26 Impact factor: 5.602