Sylvia P Nauta1,2, Pascal Huysmans3, Gabriëlle J M Tuijthof3, Gert B Eijkel1, Martijn Poeze4,5, Tiffany Porta Siegel1, Ron M A Heeren1. 1. Maastricht MultiModal Molecular Imaging (M4i) Institute, Division of Imaging Mass Spectrometry, Maastricht University, Universiteitssingel 50, 6229ER Maastricht, The Netherlands. 2. Department of Orthopedic Surgery and Trauma Surgery, Maastricht University Medical Center, P. Debyelaan 25, 6229 HX Maastricht, The Netherlands. 3. Research Engineering, Maastricht University, Universiteitssingel 50, 6229ER Maastricht, The Netherlands. 4. Department of Surgery, Division of Trauma Surgery, Maastricht University Medical Center, P. Debyelaan 25, 6229 HX Maastricht, The Netherlands. 5. NUTRIM, School for Nutrition and Translational Research in Metabolism, Maastricht University, Universiteitssingel 40, 6229 ER Maastricht, The Netherlands.
Abstract
The analysis of samples with large height variations remains a challenge for mass spectrometry imaging (MSI), despite many technological advantages. Ambient sampling and ionization MS techniques allow for the molecular analysis of sample surfaces with height variations, but most techniques lack MSI capabilities. We developed a 3D MS scanner for the automated sampling and imaging of a 3D surface with laser-assisted rapid evaporative ionization mass spectrometry (LA-REIMS). The sample is moved automatically with a constant distance between the laser probe and sample surface in the 3D MS Scanner. The topography of the surface was scanned with a laser point distance sensor to define the MS measurement points. MS acquisition was performed with LA-REIMS using a surgical CO2 laser coupled to a qTOF instrument. The topographical scan and MS acquisition can be completed within 1 h using the 3D MS scanner for 300 measurement points on uneven samples with a spatial resolution of 2 mm in the top view, corresponding to 22.04 cm2. Comparison between the automated acquisition with the 3D MS scanner and manual acquisition by hand showed that the automation resulted in increased reproducibility between the measurement points. 3D visualizations of molecular distributions related to structural differences were shown for an apple, a marrowbone, and a human femoral head to demonstrate the imaging feasibility of the system. The developed 3D MS scanner allows for the automated sampling of surfaces with uneven topographies with LA-REIMS, which can be used for the 3D visualization of molecular distributions of these surfaces.
The analysis of samples with large height variations remains a challenge for mass spectrometry imaging (MSI), despite many technological advantages. Ambient sampling and ionization MS techniques allow for the molecular analysis of sample surfaces with height variations, but most techniques lack MSI capabilities. We developed a 3D MS scanner for the automated sampling and imaging of a 3D surface with laser-assisted rapid evaporative ionization mass spectrometry (LA-REIMS). The sample is moved automatically with a constant distance between the laser probe and sample surface in the 3D MS Scanner. The topography of the surface was scanned with a laser point distance sensor to define the MS measurement points. MS acquisition was performed with LA-REIMS using a surgical CO2 laser coupled to a qTOF instrument. The topographical scan and MS acquisition can be completed within 1 h using the 3D MS scanner for 300 measurement points on uneven samples with a spatial resolution of 2 mm in the top view, corresponding to 22.04 cm2. Comparison between the automated acquisition with the 3D MS scanner and manual acquisition by hand showed that the automation resulted in increased reproducibility between the measurement points. 3D visualizations of molecular distributions related to structural differences were shown for an apple, a marrowbone, and a human femoral head to demonstrate the imaging feasibility of the system. The developed 3D MS scanner allows for the automated sampling of surfaces with uneven topographies with LA-REIMS, which can be used for the 3D visualization of molecular distributions of these surfaces.
Entities:
Keywords:
3D mass spectrometry imaging; LA-REIMS; automation; surface sampling
Over
the past decades, mass spectrometry imaging (MSI) has been
increasingly used to investigate biological and biomedical samples.
Nevertheless, the surface sampling of large areas with topographical
height variations with 3D MSI is very limited. The analysis of samples
with large and uneven surfaces has a potentially broad range of applications,
including for the analysis of whole fruits or vegetables as well as
clinically for analysis of large surface areas on bones and joints.
The analysis of these types of surfaces can be of added value in reproducible
molecular mapping of samples surfaces to study the molecular changes
on the sample surface. For clinical applications, this can be used
to improve molecular understanding of processes related to structural
changes, which can be applied to tissue classification model building
for identification of diseases. In addition, the number of sampling
spots, and thus, tissue damage can be reduced by determining the optimal
sampling point due to the improved molecular understanding of the
sample surface for further applications.Samples with uneven
surfaces are more challenging to analyze with
MSI, as mass analyzers, like time-of-flight (TOF) analyzers, are frequently
affected by small height variations in the sample resulting in signal
intensity variations or peak shifts.[1] For
example, a linear TOF is more sensitive to height variation than an
orthogonal TOF, as in an orthogonal TOF the ionization source is decoupled
from the mass analysis.[2] Height variations
of even a few micrometers in the sample can cause defocusing of the
laser beam during analysis for MSI technique relying on laser beams,
resulting in decreasing sensitivity.[3] Because
of these challenges, most MSI techniques require a constant distance
between the sample, the ionization source, and the extraction to the
mass analyzer. Samples prepared for 2D MSI techniques, such as matrix-assisted
laser desorption/ionization (MALDI) and desorption electrospray ionization
(DESI), are often sectioned to create planar surfaces for easy analysis.[3−6] Nevertheless, sectioning is not always an option for certain tissue
types or surface analyses.Some techniques have been developed
that can compensate for small
variations in sample height. These compensation techniques are based
on obtaining surface height measurements by, for example, camera image
analysis,[7,8] atomic force microscopy,[9−12] shear force microscopy,[13] scanning near-field optical microscopy,[14,15] laser triangulation,[5,16] or confocal distance sensor.[3] The height variation tolerance remains limited
to only hundreds of μms and still requires the (biological)
samples to be relatively flat for all these compensation techniques.
Therefore, the analysis of biological sample surfaces with large height
variations in terms of millimeters, for example, bones and joints
removed during surgery, is not possible with these developments.It is important to determine the topography of the surface that
needs to be analyzed to achieve MSI of 3D surfaces. This topography
will determine the position of the sample, ionization source, and
MS inlet to keep a constant distance and orientation relative to each
other by moving the sample or the ionization source and MS inlet.
Only a few MSI studies have been performed on large 3D surfaces with
sampling MS techniques that applied the concept of aligning the sampling
probe with the uneven surface to create 3D distributions and, therefore,
are described as 3D MSI. In 2014, Bennett et al. coupled direct analysis
in real time (DART) with robotic probe surface sampling via a needle
and showed the distribution of different dyes on the polystyrene hemisphere.[4] In 2018, Li et al. used a robotic arm with a
needle probe combined with an open port sampling interface (OPSI)
and showed the relative abundance of molecules on a plastic coffee
cup cover, medicine blister pack, football, and six-well spot plate.[6] In 2021, Ogrinc et al. coupled a robotic arm
with the SpiderMass (a water-assisted laser desorption ionization
MS technique) for in vivo surgical applications.[17] In this setup, 3D topographical imaging of the
surface is combined with molecular surface sampling with the SpiderMass.
Topological images and 3D mass distributions were shown ex
vivo for a sponge spotted with lipid standards, an apple
core with seeds, a skin biopsy, and parts of a whole-body mouse model
and in vivo on a human fingertip. Although these
prior methods are innovative, they have not been applied to samples
that are difficult to sample, like bone and cartilage of joints.The development of different ambient sampling and ionization mass
spectrometry techniques, for in vivo applications,
allows for the sampling of sample surfaces without sample preparation
independent of the flatness of the sample.[4,16,18] One of these techniques is rapid evaporative
ionization mass spectrometry (REIMS),[19] combined with either electrocautery or laser ablation sampling.[20,21] REIMS works by analyzing the smoke containing aerosolized molecules
created when electrocautery or laser ablation is applied to biological
tissue.[19,21] This smoke is rich in lipids and metabolites
and is aspirated to a mass spectrometer via tubing connected to a
Venturi pump.[19,21] Schäfer et al. showed
that laser desorption ionization MS with a CO2 laser could
be applied to biological tissue to generate similar spectra as those
created with electrocautery when combined with REIMS.[20] Recently, Genangeli et al. demonstrated the use of a CO2 laser coupled with REIMS for the sampling of bone, cartilage,
and bone marrow.[21] Of all the developed
ambient techniques, the combination of a CO2 laser with
REIMS (LA-REIMS) and the SpiderMass are the only techniques that were
demonstrated to be able to sample hard tissue samples, like bone and
cartilage.[21,22] LA-REIMS was selected, as this
technique is promising for the 3D MSI of samples with large height
variations and can be used on bone and cartilage. Laser-assisted REIMS
(LA-REIMS) has been used for different tissue classification, metabolic
screening, and other applications.[18,21,23−27] LA-REIMS and similar techniques can be used for sampling of 3D surfaces,
but the created mass spectra are often not spatially correlated back
to the three-dimensional sample location to create distribution images.
For LA-REIMS, a high-throughput and automated platform has been developed
for the metabolic screening.[18,23−27]Despite the promising field of applications, LA-REIMS has
not yet
been applied to the analysis of complex 3D sample surfaces, to the
authors’ knowledge. In addition, LA-REIMS currently lacks imaging
capabilities, despite the development of automated screening platforms.
The coordinates in the three-dimensional space of the sample locations
as well as the order of the sampling locations need to be correlated
back to the obtained mass spectra per sampling point to enable imaging
capabilities for LA-REIMS. Automation of 3D surface sampling with
LA-REIMS would allow for (1) increased reproducibility between measurement
points, as the distance and the angle between the sampling device
and the sample surface is constant; (2) increased precision, as all
sampling positions will be defined by a sampling grid with constant
distances between measurement points; (3) faster analysis, as the
sample or the laser probe is moved to the next measurement position
automatically; and (4) the creation of molecular distribution images,
as the spatial location of each measurement point can be coupled to
the mass spectrum of the sample location. Besides, 3D MSI of the surfaces
of, for example, bones and joints, can improve the generation of clinical
classification models, as they will be more robust and small changes
can be related back to the annotation of the sample surface.Herein, we describe the development and demonstrate the effectiveness
of such an automated device—the “3D MS scanner”—for
the sampling of a 3D surface with LA-REIMS. We focus on the comparison
between manual analysis of a 3D surface with LA-REIMS and the automated
analysis. Furthermore, we show for the first time for LA-REIMS the
added value of this automated sampling technique in terms of retaining
spatial information on each sampling location, which allows for 3D
visualization of the acquired area.
Materials and Methods
Materials
and Samples
2-Propanol and leucine-enkephalin
were purchased from Honeywell (Seelze, Germany) and Sigma-Aldrich
(St. Louis, MO), respectively. A 0.5 mM sodium formate solution was
used for calibration of the mass spectrometer.Apples and cow
marrowbone were purchased at the local supermarket for different types
of experiments with the 3D MS scanner. The apple and marrowbone were
used without any further sample preparation, except for analysis of
the bone of the marrowbone for which the remaining soft tissue was
removed with a scalpel.Human femoral heads were collected from
five patients that underwent
total hip replacement surgeries at Maastricht University Medical Center
(MUMC+). The Medical Ethical Committee of the MUMC+ (approval number:
09-2-123) approved the use of “waste” materials from
surgeries for scientific research without an additional written informed
consent of the patient. The human femoral heads were stored in a −80
°C freezer until analysis and were left to thaw at room temperature
before analysis. No further sample preparation was performed.
Surgical
CO2 Laser
An AcuPulse Class IV
CO2 surgical laser (Lumenis GmbH, Germany) with a wavelength
of 10.6 μm was used for all experiments. The laser has a maximum
power of 60 W and can be operated in the modalities continuous wave
(CW), pulsed wave (PW), and super pulsed wave (SPW). The used laser
settings depended on the sample type. An overview of the different
laser settings used for the different samples (apple, cow marrowbone,
and human femoral head) during the different experiments (comparison
between manual and automated sampling, 3D visualization of molecular
distributions, and MS/MS experiments) is provided in Table .
Table 1
Overview
of Laser Settings for Different
Samples during the Different Experimentsa
comparison
3D visualization
MS/MS
M
P
D
E
M
P
D
E
M
P
D
E
apple
CW
25
0.1
2.5
CW
25
0.1
2.5
CW
25
0.4
10.0
cow marrowbone
PW
30
0.4
12.0
CW
30
0.1
3.0
CW
30
0.4
12.0
human femoral head
CW
20
0.3
6.0
CW
20
0.1
2.0
CW
20
0.4
8.0
M is the operation
mode, P is the laser power (in W), D is the laser
pulse duration (in s), and E is the corresponding
sampling energy (in J) per measurement point.
M is the operation
mode, P is the laser power (in W), D is the laser
pulse duration (in s), and E is the corresponding
sampling energy (in J) per measurement point.
Safety Considerations
Appropriate safety measures were
put in place, in accordance with local safety regulations when working
with a Class IV laser CO2 surgical laser. The laser was
placed fully shielded in either the purposely designed and constructed
mechanical CO2 laser probe holder or the 3D MS scanner
to prevent holding the laser in the hand while it fires. The experimental
area was shielded with Class IV laser safety curtains, and all people
inside this area wore CO2 laser-grade safety goggles. An
anodized aluminum plate was placed underneath the laser beam to prevent
scattering of the CO2 laser beam outside of the class II
biosafety cabinet. All experiments were performed inside a class II
biosafety cabinet that ventilated the created smoke safely. The CO2 laser has a safety interlock to stop the laser on the laser
body, which could be pressed in case of emergency.
Manual Setup:
Mechanical CO2 Laser Probe Holder
A custom-built,
articulated, mechanical arm was used to secure
the CO2 laser probe (the end of the laser arm through which
the laser beam passes) for safety considerations and to allow for
operation of the laser at a fixed position when the laser was fired.
This mechanical CO2 laser probe arm has been previously
described by Genangeli et al.[21] In short,
the arm is made of aluminum alloy 6082, which was glass bead blasted.
The moving parts of the arm were kept in place with air-powered breaks,
which cause small movements in the position of the arm when releasing
them. The laser probe in the holder could be moved by hand to the
required position, where it remained while the laser was fired. The
laser was fired with a two-button safety box outside of the biosafety
cabinet for safety reasons. The sample is placed underneath the laser
at a plate at the bottom of the biosafety cabinet.
Automated Setup:
3D MS Scanner Development
The 3D MS
scanner was developed to measure the surface of a sample in an automated
way using LA-REIMS. The main components of the 3D MS scanner are the
different axes, the sample holder, distance sensor, clamps for fixation
of the laser probe, and the control box (see Figure ). For the operation of the 3D MS scanner,
a sample is placed in a custom-designed sample holder positioned underneath
the laser probe for safety considerations. The sample holder is positioned
in the middle of the biosafety cabinet. The sample is automatically
moved for each measurement point while maintaining a constant distance
and angle between the laser probe and sample surface. The aim of this
movement is to place the sample underneath the laser probe in such
a way that the laser beam is aligned with the surface normal at the
measurement point as perfect as achievable within the implemented
degrees of freedom. The setup has four degrees of freedom: three translational
axes (x, y, and z axes) and one rotational axis (around the x axis).
Electric slides (models EGSK and EGSC, FESTO, Esslingen am Neckar,
Germany) in combination with stepper motors (model EMMS, FESTO) were
used for the motion along the translational axes. A rotary driver
(model ERMO, FESTO) was used for the rotational motion with a manufacturer
specified repeatability of ±0.05°. A high-performance laser
point distance sensor (model BA_OM70, Baumer Holding AG, Frauenfeld,
Switzerland) was used for the measurement of the topography of the
surface sample. The distance sensor was placed at a fixed distance
from the laser probe and has an optimal measurement range of 40 mm.
Therefore, the maximum z-height variation in the
acquisition areas should not extend this value. The laser probe was
held by two magnetic clamps made of the plastic Stratasys-Verowhiteplus
that would release the laser probe if a collision between the sample
and laser probe occurred to prevent damage to the fragile laser arm.
The base frame and sample holder of the 3D MS scanner were made of
aluminum and an anodized aluminum plate was attached at the bottom.
Figure 1
Images
of the 3D MS scanner. (A) Picture of the 3D MS scanner setup
in the biosafety cabinet with the CO2 laser probe inserted.
Besides the laser probe, the following elements are indicated: the
tubing through which the created smoke is aspirated toward the mass
spectrometer (MS), the distance sensor that is used for the topographical
measurement, and the sample holder. (B) Design image of the 3D MS
scanner with indication of the three translational axes, the rotational
axis, and the control box.
Images
of the 3D MS scanner. (A) Picture of the 3D MS scanner setup
in the biosafety cabinet with the CO2 laser probe inserted.
Besides the laser probe, the following elements are indicated: the
tubing through which the created smoke is aspirated toward the mass
spectrometer (MS), the distance sensor that is used for the topographical
measurement, and the sample holder. (B) Design image of the 3D MS
scanner with indication of the three translational axes, the rotational
axis, and the control box.The 3D MS scanner was operated using home-built control electronics
that were programmed with custom LabVIEW software (version 2020, National
Instruments). The control software was running on a Microsoft Surface
Pro 6 using the custom-build LabVIEW application MS3D version 2.01.
Analog Laser Trigger Registration with Mass Spectral Data
The position (in three dimensions) of a measurement point must
be registered with the correct corresponding mass spectrum for each
sampling event for correct processing of the data and the generation
of images. Registration of an analogue laser trigger into the mass
spectrometer next to the total ion current (TIC) signal is important
for this alignment. An eSAT/IN box (D16EST083M, Waters Corp.) was
used for the registration of an analog signal during the MS acquisition.
An analogue trigger was registered next to the TIC chromatogram in
MassLynx (version 4.1, Waters Corp.) during the MS measurement of
the sample whenever the surgical CO2 laser was fired. See Figure S1 for an example of an overlay image
between the analogue signal and the TIC chromatogram.
Acquisition
Process
The acquisition process consists
of three steps: (1) homing of the motors, (2) performing a topographical
scan, and (3) performing the MS measurement. The three translational
motors are homed, and the center of the rotational motor is determined
to ensure accuracy and reproducibility at the start of each experiment.
The measurement settings and the rectangular acquisition area are
defined. Currently, a control software limitation requires all acquisition
areas to be rectangles. In the next step, the topographical scan is
used to determine the topography of the surface and translate this
into motor positions for each measurement point. The topographical
scan consists of two phases: (1) a quick scan of the sample surface
where a spline is fitted through the points to define the mesh for
the measurement points and (2) adjustment of the motor positions at
each measurement point based on the mesh and optimization of the distance
between the distance sensor and the surface. When the topographical
scan is finished, the MS measurement is performed. The control box
will send a trigger to the XEVO G2-XS to start the MS acquisition,
which also starts the acquisition of the analogue laser trigger signal.
The following steps are repeated for each measurement point: (a) the
surgical CO2 laser is triggered to fire one laser pulse
and an analog trigger is registered at the eSAT/IN box at the same
time, (b) the sample is held still for a short waiting period (optimized
value: 0.1 s) to allow for the aspiration of the ablated smoke to
the mass spectrometer and to prevent disturbance of the smoke by movement,
and (c) the sample is repositioned to the next measurement point.
A file is generated that contains the x, y, and z coordinates of each of the measurement
points at the end of the acquisition. This file is used to assign
the 3D coordinates to the mass spectral data of each measurement point
in postprocessing.
REIMS-qTOF Instrumentation
The smoke
created with the
CO2 laser was analyzed with REIMS by aspiration through
a flexible tube by a Venturi pump to the source. A benchtop XEVO G2-XS
qTOF (Waters Corp., Manchester, UK) equipped with a REIMS source was
used for the ionization and detection of the molecules from this smoke.
The measurements were acquired in negative ion mode, in sensitivity
mode, and with a mass resolution of around 15000 full width at half-maximum
(fwhm) at m/z 600. The acquired m/z range for all samples was 100–1500.
The instrument was calibrated daily with a 0.5 mM solution of sodium
formate. A lock mass solution of leucine-enkephalin (Leu-Enk) in 2-propanol
(final concentration 0.075 ng/μL) was continuously infused at
a flow rate of 150 μL/min in the REIMS source using a syringe
pump. A lock mass solution was constantly infused as it has been shown
that infusion of a solvent matrix during the REIMS experiment can
significantly improve the ion intensity.[28]
Comparison between Manual and Automated Measurements
Manual
and automated measurements of 50 points were acquired to compare
the reproducibility for the manual (see the Manual
Setup section) and automated (see the Automated
Setup section) setups in data acquisition. The manual data
acquisition was always performed subsequent to the automated data
acquisition, as the manual measurements points were placed in between
the automated measurement points to prevent overlap between the sample
location of the measurement points. The data acquisition for the manual
and automated setups were performed within the same rectangular acquisition
area to keep the biological variance in both data sets similar. The
comparisons were performed on an apple, the outside surface of a marrowbone
(bone tissue), and five human femoral heads. Human femoral heads are
molecularly more heterogeneous than apples and bone, and therefore,
areas were selected that were homogeneous by eye.The manual
and automated setups were compared based on the overall intensity,
the intensity of 10 selected m/z values, and the variance between measurement points. These m/z values were defined per sample to be
specific for the sample signal and to cover the measured mass range
(Table S1). The overall signal intensity
was determined by (1) the maximum TIC peak intensity value and (2)
the area of the TIC peak. The maximum TIC peak intensity values corresponding
to each measurement point were determined by taking the maxima in
the chromatogram trace of the measurement. The TIC peak per measurement
point in the chromatogram trace of the measurement was integrated
to calculate the area of the TIC peak. The maximum TIC peak intensity
value represents the maximum total signal intensity in one scan for
each measurement point, while the TIC peak area represents the total
signal intensity for the whole measurement point. Therefore, both
values are compared between manual and automated measurements. The
10 selected m/z values were compared based on the
calculated areas of the extracted ion chromatograms (EICs) for the
measurement points. Results are reported as average ± standard
deviation (SD) and coefficient of variance (CV,%). The variance between
measurement points was compared using principle component analysis
(PCA) of the manual and automated setups per sample. The PCA plots
can be related to the variance, as the closer the different measurement
points (represented by spheres) are together the lower the variance
in between them and vice versa.MassLynx (version 4.1, Waters
Corp.) was used for the analysis
of the acquired MS data. The coefficients of variances (CV) for the
manual and automated data were calculated to determine whether they
were significantly different with a coefficient of variance test.
The determination threshold for significance was set at p-values <0.05. Abstract Model Builder (AMX) (version 1.0.1581.0,
Waters Corp.) was used for the PCA analyses. The mass spectra were
binned with a bin size of m/z 0.2
for the mass range of m/z 100–1500.
The mass spectra of each measurement point were lock mass corrected,
background subtracted and removed, and normalized (TIC normalization)
before building the a 10 component PCA model.
3D MSI Measurements
Three experiments were set up to
show the use of the 3D MS scanner for the 3D visualization of molecular
distributions on different types of samples: (1) 300 measurements
points on an apple of which part of the peel was removed, (2) 200
measurement points on the outside of a marrowbone, and (3) 150 measurement
points on the outside of a human femoral head. The spatial resolution
was set to 2 mm, and the MS acquisition time per measurement point
was 5.0 s.Home-built MATLAB (version 2018b, Mathworks) algorithms
were used for data conversion and 3D visualization of the measurements
acquired with the 3D MS scanner.
Molecular Identifications
Tentative molecular identifications
were performed on the basis of MS/MS and database searches for the m/z values that are displayed in the different
figures in the Results. MS/MS fragmentation
was performed on a REIMS-qTOF in negative ionization mode with varying
collision energies (CEs) between 10 and 40 arbitrary units. Alex123 lipid calculator, METLIN, LIPID MAPS Structure Database
(LMSD), and MassLynx (version 4.1, Waters Corp.) were used for the
identification of the m/z values.
The data sets were lock-mass corrected using mMass (Open Source Mass
Spectrometry Tool, version 5.5.0) when the internal lock mass (Leu-Enk)
showed a ppm error above 10 ppm. The lock mass corrected m/z values were used for the identifications.
Results
This study consisted of three parts: (1) the characterization of
the 3D MS scanner, (2) comparison between the mechanical CO2 laser probe holder controlled by hand and the automated 3D MS scanner,
and (3) 3D visualization of the data acquired with the 3D MS scanner.
Characterization
of the 3D MS Scanner
The developed
3D MS scanner (see Figure and Figure S2) allows for the
mass spectrometer analysis of sample surfaces with large topographical
variations in 3D. The whole measurement process of the 3D MS scanner
consists of three main phases: homing of the axes for alignment, topographical
scan of the sample surface, and MS data acquisition with LA-REIMS.
These phases take less than 60 s, 5.1 s per measurement point, and
5.0 s per measurement point, respectively, after optimization. Three
hundred measurement points can be acquired within 1 h taking into
account the whole process. The sample positions are defined based
on the topography of the surface such that there is a constant distance
and angle between the sample surface and the laser probe. The spatial
resolution is constant in the top view with a distance of 2 mm between
measurement points, which is difficult to achieve with manual analysis.
The 3D MS scanner allows for the analysis of 3D sample surfaces in
an automated and fast way with LA-REIMS. See Supporting Information S1 for further information about the optimization
and characterization of the 3D MS scanner.
Comparison between Manual
and Automated Measurements
The CVs were calculated as a representation
of the reproducibility
between measurement points. The reproducibilities of automated acquisitions
were compared with the reproducibilities of manual acquisitions for
different sample types. Three of the seven CVs were significantly
lower for the automated setup using the 3D MS scanner in comparison
to the manual setup looking at the TIC peak intensity values (see Table ). In addition, the
CVs for the TIC peak intensity for the manual setup for femoral heads
3 and 5 are lower than the CVs of the automated setup, but the CVs
are not significantly different. The CVs for TIC peak area values
(see Table ) are lower
for the automated setup than the manual setup with significant differences
in the CVs for five samples. The results of the TIC peak intensities
and TIC peak areas show the same trend, with lower CVs for the automated
setup. These lower CVs for the TIC peak intensity and TIC peak area
show the improved signal reproducibility between measurement points
for the automated setup compared to the manual setup. This is further
emphasized by the CVs for the selected m/z values, most of the CVs are lower or similar for the automated
setup in comparison to the manual setup (see Table S1). The variation between the measurement points for the whole
mass range (m/z 100–1500)
for both setups can be assessed using PCA plots (see Figure S3). These plots display a smaller or similar variation
between the measurement points of the automated setup than between
those of the manual setup.
Table 2
Comparison of TIC
Peak Intensities
between Automated and Manual Measurementsa
TIC
peak intensity
automated
manual
avg ± SD
CV (%)
avg ±
SD
CV (%)
P-value
apple
5.97E7 ± 1.08E7
18.13
7.99E7 ± 1.47E7
18.46
0.9026
marrowbone
1.27E8 ± 2.66E7
20.90
6.09E7 ± 2.76E7
45.31
<0.0001
femoral head 1
1.25E8 ± 3.24E7
25.88
7.49E7 ± 2.70E7
36.08
0.0347
femoral head 2
2.10E8 ± 3.25E7
15.48
1.02E8 ± 3.79E7
37.23
<0.0001
femoral head 3
1.46E8 ± 3.51E7
23.98
1.19E8 ± 2.14E7
18.03
0.0577
femoral head
4
2.04E8 ± 3.57E7
17.46
1.57E8 ± 3.35E7
21.41
0.1691
femoral head 5
1.53E8 ± 3.21E7
20.99
1.23E8 ± 2.42E7
19.79
0.6916
Comparison of the absolute maximum
TIC peak values in the chromatogram per measurement point between
automated and manual measurements for the different samples; apple,
marrowbone, and five femoral heads. The average maximum TIC peak intensity
value is given with the standard deviation as well as the coefficient
of variance (CV) as a percentage. In addition, the result of the coefficient
of variance test is provided, indicating if the CVs are significantly
different (p < 0.05).
Table 3
Comparison of TIC Peak Areas between
Automated and Manual Measurementsa
TIC
peak area
automated
manual
avg ± SD
CV (%)
avg ±
SD
CV (%)
P-value
apple
1.39E6 ± 1.42E5
10.17
1.45E6 ± 2.31E5
15.94
0.0024
marrowbone
3.29E6 ± 3.20E5
9.73
1.35E6 ± 5.60E5
41.48
<0.0001
femoral head 1
3.66E6 ± 7.23E5
19.74
1.52E6 ± 5.08E5
33.49
0.0007
femoral head 2
5.83E6 ± 4.37E5
7.49
2.64E6 ± 1.13E6
42.70
<0.0001
femoral head 3
4.29E6 ± 6.55E5
15.25
2.47E6 ± 4.22E5
17.05
0.4464
femoral head
4
4.82E6 ± 4.57E5
9.48
4.39E6 ± 7.48E5
17.03
0.0001
femoral head 5
4.00E6 ± 4.67E5
11.68
2.74E6 ± 6.09E5
22.24
<0.0001
Comparison of the areas under
the TIC peaks in the chromatogram per measurement point between automated
and manual measurements for the different samples; apple, marrowbone,
and five femoral heads. The areas are determined with integration
of the TIC peaks. The average TIC peak area value is given with the
standard deviation as well as the coefficient of variance (CV) as
a percentage. In addition, the result of the coefficient of variance
test is provided, indicating if the CVs are significantly different
(p < 0.05).
Comparison of the absolute maximum
TIC peak values in the chromatogram per measurement point between
automated and manual measurements for the different samples; apple,
marrowbone, and five femoral heads. The average maximum TIC peak intensity
value is given with the standard deviation as well as the coefficient
of variance (CV) as a percentage. In addition, the result of the coefficient
of variance test is provided, indicating if the CVs are significantly
different (p < 0.05).Comparison of the areas under
the TIC peaks in the chromatogram per measurement point between automated
and manual measurements for the different samples; apple, marrowbone,
and five femoral heads. The areas are determined with integration
of the TIC peaks. The average TIC peak area value is given with the
standard deviation as well as the coefficient of variance (CV) as
a percentage. In addition, the result of the coefficient of variance
test is provided, indicating if the CVs are significantly different
(p < 0.05).
3D Visualization of Molecular Distributions on Sample Surfaces
An advantage of data acquisition automation with LA-REIMS is the
possibility of combining the mass spectra for each measurement point
with the coordinates of that point. This allows the study of the molecular
distributions of sample surfaces using LA-REIMS.Comparison
of the before (Figures A and 3A) and after (Figures B, 3B, and 4A) images shows regularly spaced patterns of the
laser burning spots created during the acquisition. Optical differences
in the laser burning spots are assumed to be related to the structural
differences of the sample surface for all three samples. It was possible
to select m/z values with different
molecular distributions that were related to structural differences
visible by eye for each measurement. In addition, m/z values could be selected that showed molecular
distributions that were related to structural differences not visible
by eye. For the apple, m/z 455.4
has a higher intensity in the peel of the apple, m/z 671.5 has a higher intensity in the spots visible
in the part without the peel and a bit higher intensity in the peel,
and m/z 846.6 has a higher intensity
in the parts without the peel (see Figure C). For the outside of a marrowbone, for
example, m/z 393.3 seems to be related
to the whiter part of the marrowbone with the highest intensities
where the white tissue seems to be the thickest, m/z 419.2 seems to be related to the nonwhite parts
of the surface, and m/z 717.5 seems
to be related to the remaining muscle tissue at the edges of the measurement
area (see Figure C).
For the human femoral head, different structures can be seen, including
a white/soft pink part at the right of Figure A that is the formation of bony tissue due
to sclerosis, yellowish tissue in the middle of the image that is
a small piece of cartilage, and red tissue toward the stem of the
femoral head on the left of the image. For example, m/z 205.9 seems to be more related to the red tissue
on the femoral head, m/z 744.5 seems
to be related to part of the white, and m/z 1080.9 seems to be related to the more yellow tissue and
a bit of the red tissue (see Figure B).
Figure 2
3D visualization of molecular distributions obtained from
an apple
with the 3D MS scanner. (A) Apple before analysis with LA-REIMS using
the 3D MS scanner. Parts of the peel have been removed to show different
molecular patterns can be created. (B) Apple after analysis with LA-REIMS
using the 3D MS scanner. The total number of measurement points acquired
was 300 with a spatial resolution of 2 mm, thus covering an area of
28 × 38 mm (15 × 20 measurement points). (C) 3D visualization
of the molecular distributions (absolute intensities) of the imaging
experiment on an apple. Three m/z values were selected based on their different distributions: 455.4,
671.5, and 846.6. Tentative identifications of the m/z values can be found in Table S2.
Figure 3
3D visualization of molecular distributions
obtained from the outside
of a marrowbone with the 3D MS scanner. (A) Outside of a marrowbone
before analysis with LA-REIMS using the 3D MS scanner. (B) Outside
of the marrowbone after analysis with LA-REIMS using the 3D MS scanner.
The total number of measurement points acquired was 200 with a spatial
resolution of 2 mm, thus covering an area of 18 × 38 mm (10 ×
20 measurement points). (C) 3D visualization of the molecular distributions
(absolute intensities) of the imaging experiment on the outside of
a marrowbone. Six m/z values were
selected based on their different distributions: 128.0, 279.2, 419.3,
585.5, 744.6, and 865.8. Tentative identifications of the m/z values can be found in Table S2.
Figure 4
3D visualization
of molecular distributions obtained from part
of a human femoral head with the 3D MS scanner. (A) Outside of a human
femoral head after analysis with LA-REIMS using the 3D MS scanner.
The total number of measurement points acquired was 150 with a spatial
resolution of 2 mm, thus covering an area of 18 × 28 mm (10 ×
15 measurement points). (C) 3D visualization of the molecular distributions
(absolute intensities) of the imaging experiment on the outside of
a human femoral head. Nine m/z values
were selected based on their different distributions: 205.9, 274.1,
281.2, 642.5, 682.6, 699.5, 744.6, 1008.7, and 1081.0. Tentative identifications
of the m/z values can be found in Table S2.
3D visualization of molecular distributions obtained from
an apple
with the 3D MS scanner. (A) Apple before analysis with LA-REIMS using
the 3D MS scanner. Parts of the peel have been removed to show different
molecular patterns can be created. (B) Apple after analysis with LA-REIMS
using the 3D MS scanner. The total number of measurement points acquired
was 300 with a spatial resolution of 2 mm, thus covering an area of
28 × 38 mm (15 × 20 measurement points). (C) 3D visualization
of the molecular distributions (absolute intensities) of the imaging
experiment on an apple. Three m/z values were selected based on their different distributions: 455.4,
671.5, and 846.6. Tentative identifications of the m/z values can be found in Table S2.3D visualization of molecular distributions
obtained from the outside
of a marrowbone with the 3D MS scanner. (A) Outside of a marrowbone
before analysis with LA-REIMS using the 3D MS scanner. (B) Outside
of the marrowbone after analysis with LA-REIMS using the 3D MS scanner.
The total number of measurement points acquired was 200 with a spatial
resolution of 2 mm, thus covering an area of 18 × 38 mm (10 ×
20 measurement points). (C) 3D visualization of the molecular distributions
(absolute intensities) of the imaging experiment on the outside of
a marrowbone. Six m/z values were
selected based on their different distributions: 128.0, 279.2, 419.3,
585.5, 744.6, and 865.8. Tentative identifications of the m/z values can be found in Table S2.3D visualization
of molecular distributions obtained from part
of a human femoral head with the 3D MS scanner. (A) Outside of a human
femoral head after analysis with LA-REIMS using the 3D MS scanner.
The total number of measurement points acquired was 150 with a spatial
resolution of 2 mm, thus covering an area of 18 × 28 mm (10 ×
15 measurement points). (C) 3D visualization of the molecular distributions
(absolute intensities) of the imaging experiment on the outside of
a human femoral head. Nine m/z values
were selected based on their different distributions: 205.9, 274.1,
281.2, 642.5, 682.6, 699.5, 744.6, 1008.7, and 1081.0. Tentative identifications
of the m/z values can be found in Table S2.
Discussion
Development and Optimization of the 3D MS Scanner
The
3D MS scanner was developed to analyze sample surfaces in 3D using
LA-REIMS. Each of the three phases in the acquisition was optimized,
and the acquisition rate of the 3D MS scanner is 300 points in approximately
1 h, corresponding to 22.04 cm2 with a spatial resolution
of 2 mm. After homing of the axes, the topographical scan and MS acquisition
take 5.1 and 5.0 s per measurement point, respectively. This is faster
than the surface sampling setups developed by Bennett et al. and Li
et al., which were on the order of minutes per measurement point.[4,6] The acquisition time for the 3D MS scanner is slower than the one
reported by Ogrinc et al., which was approximately 0.6 s per measurement
point with optimized parameters and simultaneous collection of topographical
and MS imaging.[17] One of the reasons for
the slower acquisition with the 3D MS scanner is the consecutive acquisition
of the topographical scan and the MS measurement. Simultaneous collection
is not possible with the 3D MS scanner due to the distance between
the laser probe and laser point distance sensor. A risk in slower
acquisitions could be molecular degradation of the biological samples,
but this is expected to be minimal within 1 h. However, molecular
degradation should be taken into consideration for the acquisition
of larger surfaces, which will have longer acquisition times.The 3D MS scanner was developed to keep the distance and angle between
the laser probe and sample surface consistent. With the implemented
four degrees of freedom (three translational axes and one rotational
axis) curvatures along the y-axis are fully compensated.
Curvatures along the x- and z-axis
cannot be fully compensated with these four degrees of freedom, as
this would require additional degrees of freedom in the x- and z-axes. An angle of 10° along one of
these axes will result in a decrease of only 3% in the laser fluency
per measurement point, which is expected to not have a significant
effect on the signal intensity in this setup due to the use of a CO2 laser. Bennett et al., Li et al., and Ogrinc et al. used
a robotic arm with six degrees of freedom for sampling of the surface.[4,6,17] Six degrees of freedom will allow
for better compensation of the surface topography along all axes.
Our in-house built setup with four degrees of freedom can still compensate
for most topographical height variations, although the possibilities
are more limited than for a robotic arm.Another difference
between the 3D MS scanner and previously published
setups using robotic arms is the movement of the sample underneath
the laser probe for the 3D MS scanner, while in previously published
setups the sample lays still. The needle probe connected to the robotic
arm needed to move from the sampling surface to DART ion source or
the OPSI in the setup of Bennett et al. and Li et al., respectively.[4,6] The SpiderMass probe was connected to the robotic arm together with
the transfer line to the mass spectrometer for the robot-assisted
SpiderMass setup.[17] The laser probe is
moved relative to the sample in the setup of Ogrinc et al., as they
aimed to develop a system that could potentially be used in
vivo during surgeries in the future. Our main consideration
for a setup where the sample moves underneath the laser was safety,
as the system is equipped with a Class IV CO2 laser. Another
reason for this setup of the 3D MS scanner is the focus on improved
understanding of molecular distributions on sample surfaces with large
height variations.
Comparison between Manual and Automated Setup
Comparison
between the manual and automated setups showed that the acquisition
with the automated setup has a higher signal reproducibility between
measurement points than with the manual setup. This was shown by the
lower CVs for the TIC peak maximum intensities and areas for most
of the automated acquisition than for the manual acquisitions (Tables and 3), which is according to expectations. The lower CVs for the
automated setup are caused by the higher reproducibility of the tissue
ablation by the CO2 laser due to the constant distance
between the laser probe and the sample surface with the 3D MS scanner.
In addition, the laser probe is placed as perpendicular to the sample
surface as possible. Both considerations are more difficult to achieve
with manual acquisitions, especially due to small movements of the
laser probe holder caused by the air-powered breaks in the manual
setup.The reported CVs for the manual and automated setups
are slightly higher than previously reported by Genangeli et al. for
the CO2 laser, especially for the manual setup.[21] Direct comparison between the CVs of both studies
is complicated by the differences between the studies, which contribute
to the CVs. In this study, more heterogeneous samples, especially
the femoral heads, more measurement points per experiment, and different
sizes of acquisition areas were used in comparison to the work of
Genangeli et al. The CVs for the automated and manual setups were
based on 50 measurement points, which covers a larger area than the
10 points acquired by Genangeli et al. Therefore, part of the higher
CVs might be explained by small differences in the molecular composition
and tissue ablation across the area. Another factor contributing to
the high CVs can be LA-REIMS, although the whole setup is not changed
during acquisitions. LA-REIMS is based on the analysis of the aspirated
smoke created by laser ablation. The aspiration of the smoke might
be different between measurement points, resulting in variations in
the TIC peak intensities and TIC peak areas contributing to higher
CVs. In addition, the surface sample will affect the aspiration of
the smoke, as the topography of the surface will influence the flow
of the smoke. For example, it has been observed that the aspiration
of the smoke can be affected for samples with large height variations.The TIC peak intensities and TIC peak area values were 8.82–58.94%
lower for the manual setup than for the automated setup, except for
the apple acquisitions. A contributing factor to this might be the
different positions of the sample in the biosafety cabinet in both
setups. Changes in the aspiration of the smoke were observed based
on sample position due to the airflow inside the biosafety cabinet.
It is more difficult to keep the same distance between the sample
and laser probe as well as the orientation of the laser probe compared
to the sample with the manual setup, which can affect the aspiration
of the smoke. Decreasing signal intensities with increasing distance
between the sample surface and laser probe were shown for the automated
setup (see Supporting Information S1).
Therefore, a larger distance between the sample surface and laser
probe with the manual setup can contribute to the lower signal intensities.The manual and automated setup both have their advantages. The
manual setup is preferred if only a few measurements points, for example,
for MS/MS data, need to be acquired, as no homing of the axis and
topographical scan needs to be performed, making the manual acquisition
faster than the automated one. Besides the higher signal reproducibility,
one of the advantages of the automated setup is the larger area of
the sample that is available for analysis. For the manual measurements,
the sample is placed on a flat surface or could be placed in a clamp.
The fixed position of the sample in the manual setup limits the area
available for data acquisition, which is not the case for the automated
setup. The sample is fixated in the sample holder by tightening the
screws for the automated measurements. The sample can be orientated
and moved in such a way that the area of interest can be placed underneath
the laser probe in the automated setup. The only limitation is that
the screws should not be tightened at a position that might be of
interest in later acquisitions. Another advantage of the automated
acquisition is the automated movement to the next measurement point
while maintaining a constant distance between the different measurement
points, which prevents any overlap between the laser burning spots
of different measurements points. Some overlap between the measurement
points for the manual setup occurs sometimes due to the small movements
of the holder due to the air-powered breaks. Therefore, the accuracy
of the sampling position is higher for the automated setup due to
the predefined sampling grit in comparison to the manual setup.
Comparison of 3D Visualization between the 3D MS Scanner and
Similar Technologies
3D visualization of molecular distributions
in Figures –4 demonstrates that different m/z values could be selected that represented molecular distributions
related to visible structural differences but also distributions that
were not related to visible structural differences. The spaces between
the laser burning spots were equal for the three different analyzed
sample types when looking from the top view, but differences between
laser burning spots were seen based on structural differences in the
surface. Recently, Ogrinc et al. have shown that their robot-assisted
SpiderMass setup allows for the visualization of 3D molecular distributions
on different samples.[17] The robot-assisted
SpiderMass setup has not been applied to human bone and joints, like
the human femoral heads used in this study. The SpiderMass is mini-invasive,[17] while the CO2 laser leaves clear
burning spots, due to the different equipped laser types. This is
a result of differences in interaction between the biological tissue
and the laser wavelength as well as different laser energies.All the acquisitions with the 3D MS scanner were acquired with a
spatial resolution of 2 mm. This is larger than the spatial resolution
reported by Ogrinc et al., which was 500 μm without oversampling.[17] One of the factors that will limit the spatial
resolution is the diameter of the laser focus point. The diameter
of the focus point of a CO2 laser is larger than the Nd:YAG
laser used in the SpiderMass setup. Therefore, a spatial resolution
of 500 μm cannot be reached with a CO2 laser without
major overlap between laser burning spots. It is noteworthy that the
spatial resolution of the 3D MS scanner could be reduced to the size
of the laser focus point, which is approximately 1 mm in this setup.
Future Applications
The 3D visualizations created from
the data acquired with the 3D MS scanner allow for observing and correlating
changes in molecular intensity to locations on the sample surface
and, therefore, structural differences. The 3D correlative images
can, for example, be used to study the different molecules present
in the different tissue types of a femoral head and the transitions
areas between them (see Figure ). This can increase the molecular understanding of the processes
underlying the structural changes in, for example, osteoarthritis
of the femoral head. Large areas consisting of different tissue types
should be acquired of a large cohort of femoral heads to achieve this.
Furthermore, this molecular information on a large cohort of samples
can be used to improve models build for clinical applications. Another
potential application of the 3D molecular visualizations is to determine
the optimal sampling spot for tissue classification for in
vivo applications of LA-REIMS. On the basis of the molecular
distributions, the most informative region for tissue classification
can be determined, which can be specifically targeted during in vivo applications. This will reduce the amount of tissue
damage created by the CO2 laser as fewer sampling points
will be required.In addition to this specific potential clinical
application for femoral heads, the 3D MS scanner has the potential
to be used in a wide range of biomedical applications either as is
or with small adjustments, for example, to the sample holder. Any
type of sample could potentially be placed in the 3D MS scanner if
it can be ablated with a CO2 laser, although caution is
advised to prevent contamination of the mass spectrometer. Adjustments
could be made to include other types of ambient sampling probes that
could be triggered in a similar way to also enable analysis of samples
that cannot be ablated with a CO2 laser. Two of the biggest
advantages of this 3D MSI setup are the lack of sample preparation
and the ambient conditions. Sectioning of the sample is commonly needed
for 2D MSI techniques, which is not possible for all sample types
or can be difficult to achieve. Furthermore, some samples are too
large to fit on a microscope slide, which would not be an issue with
the 3D MS scanner combined with LA-REIMS. In addition, the ambient
conditions can allow for the analysis of samples that cannot withstand
the vacuum used in, for example, MALDI-MSI. The 3D MS scanner could
have a broad range of 3D MSI applications, for example, in the food
industry as well as biomedical and preclinical applications, as shown
in Figures –4.
Conclusion
This study describes
the development of the 3D MS scanner, which
allows for the analysis of large and uneven 3D sample surfaces with
LA-REIMS under ambient conditions. The 3D MS scanner automates the
positioning of the sample underneath the laser probe and keeps the
distance between them constant. This automated acquisition results
in increased reproducibility of the obtained molecular information
on the sample surface when compared to manual acquisitions. In addition,
the data acquired with the automated setup can be used for the visualization
of molecular distributions on the sample surface in 3D. This application
was shown on an apple, a marrowbone, and a human femoral head, which
showed distinct molecular distributions that could be related to structural
differences. Therefore, the developed 3D MS scanner can be used to
improve molecular understanding based on 3D molecular distributions
and might have a potential application in determining optimal in vivo sampling points to reduce biological tissue damage.
Authors: Olga S Ovchinnikova; Maxim P Nikiforov; James A Bradshaw; Stephen Jesse; Gary J Van Berkel Journal: ACS Nano Date: 2011-06-06 Impact factor: 15.881
Authors: Olga S Ovchinnikova; Tamin Tai; Vera Bocharova; Mahmut Baris Okatan; Alex Belianinov; Vilmos Kertesz; Stephen Jesse; Gary J Van Berkel Journal: ACS Nano Date: 2015-03-18 Impact factor: 15.881
Authors: Glen-Oliver F Gowers; Simon J S Cameron; Alvaro Perdones-Montero; David Bell; Soo Mei Chee; Marcelo Kern; David Tew; Tom Ellis; Zoltan Takáts Journal: ACS Synth Biol Date: 2019-10-30 Impact factor: 5.110
Authors: Michiel Vandenbosch; Sylvia P Nauta; Anastasiya Svirkova; Martijn Poeze; Ron M A Heeren; Tiffany Porta Siegel; Eva Cuypers; Martina Marchetti-Deschmann Journal: Anal Bioanal Chem Date: 2020-09-15 Impact factor: 4.142