Juan J Portero-Portaz1, Juan G Córdoba-Soriano1, Arsenio Gallardo-López1, Jesús M Jiménez-Mazuecos1.
Abstract
BACKGROUND: Pulmonary veins stenosis (PVS) after atrial fibrillation radiofrequency ablation is an uncommon complication. When it occurs, percutaneous treatment is the preferred approach. There is a lack of standardized procedures, and when stenting restenosis is relatively common. CASEEntities:
Keywords: Atrial fibrillation; Case report; Drug-coated balloon; IVUS; In-stent restenosis; Pulmonary vein
Year: 2021 PMID: 34870083 PMCID: PMC8637799 DOI: 10.1093/ehjcr/ytab384
Source DB: PubMed Journal: Eur Heart J Case Rep ISSN: 2514-2119
| May 2016 | Atrial fibrillation diagnosed. |
| March 2017 | Pulmonary veins radiofrequency ablation. |
| May 2017 | Haemoptysis and dyspnoea of new appearance. |
| July 2017 | Pulmonary veins stenosis (PVS) diagnosed. |
| August 2017 | Percutaneous treatment of PVS. |
| November– December 2017 | Haemoptysis and dyspnoea of new appearance. |
| May 2018 | In-stent restenosis in pulmonary veins and intravascular ultrasound-guided drug-coated balloon angioplasty. |
| May 2019 | One-year follow-up. No symptoms. |
| Nov 2020 | No Symptoms. Sinusal rhythm. |
Device type, localization, measures, inflation pressure and hemodynamic gradient pre/post.
| Vein | Device | Size (mm) | Inflation pressure (atm) | Pre/post gradient (mmHg) |
|---|---|---|---|---|
| LSPV | Pantera LuxTM (Biotronik, Berlin, Germany) | 4 × 15 | 8 | 13/6 |
| LIPV | RangerTM (Boston Scientific, Massachusetts, USA) | 8 × 30 | 6 | 20/3 |
| RSPV | Pantera LuxTM (Biotronik, Berlin, Germany) | 4.0/4.0 × 15 | 8/8 | |
| RSPV* | MustangTM (Boston Scientific, Massachusetts, USA) | 10 × 40 | 8 | 20/4 |
| RIPV | RangerTM (Boston Scientific, Massachusetts, USA) | 7 × 30 | 6 | 18/3 |
In RSPV two drug coated balloons were inflated in parallel due to the lack of appropriate size coated balloon, finally a peripheral regular balloon angioplasty was realized.