Hongguang Yun1, Marco Carlone2,3, Zheng Liu1.
Abstract
PURPOSE: Medical linear accelerators (linacs) can fail in a multitude of different manners due to complex structures. An unclear identification of failure modes occurring constantly is a major obstacle to maintenance arrangements, thereby may increasing downtime. This study aims to use natural language processing techniques to deal with the unformatted maintenance logs to identify the linac failure modes and trends over time.Entities:
Keywords: failure modes; latent Dirichlet allocation; linac; natural language processing; topic modeling
Mesh:
Year: 2021 PMID: 34842335 PMCID: PMC8803305 DOI: 10.1002/acm2.13477
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Linacs specification and service date
| Treatment room | Manufacturer | Model | Starting service date | End service date |
|---|---|---|---|---|
| A | Elekta | SL75 | April 1998 | October 2008 |
| A | Varian | Clinac iX | July 2009 | September 2021 |
| B | Elekta | SL75 | April 1998 | December 2009 |
| B | Varian | Clinac iX | September 2010 | January 2021 |
| C | Elekta | SL 20 | July 1998 | July 2010 |
| C | Varian | TrueBeam | March 2011 | Present |
| D | Elekta | SL 20 | July 1998 | February 2011 |
| D | Varian | TrueBeam | August 2011 | Present |
| E | Varian | Clinac iX | November 2009 | Present |
Example entries of linac maintenance logs
| TaskKey | Comments | Repair Description | Date |
|---|---|---|---|
| Corrective | Touchguard interlock could not be removed with iView detector in place. | Adjusted touchguard microswitches at the locking pin end for proper contact. Adjusted the alignment nuts on both sides of detector for easier locking pin insertion. | April 1998, 16 |
| Major repairs | CARR/FOIL W29 cable repaired. Error 7F when calibrating carrousel, was also getting error 70 when exiting calibration, pointing to switch S16. | Cleaned all five switches and reseated connectors J82 and J83. Adjusted the Carr pot voltage to 5.05 V from 4.74 V. Replaced and adjusted S16 on the carrousel switch assy, no change. We could reproduce the fault by moving the gantry from Zero degrees to 350°. Lubed Carr chain with TriFlow. Replaced the PWM pcbA4, no change. | January 2020, 29 |
FIGURE 1Graphical representation of latent Dirichlet allocation model
FIGURE 2Metrics used to select the optimal number of topic
FIGURE 3Top‐ranked words within topic 3 and corresponding failure mode
Keywords within topics and identified failure modes of linacs
| Subsystem | Topic (words frequency) | Keywords within topic | Identified failure modes |
|---|---|---|---|
| Topic 2 (6.5%) | Fuse supply power tube generator relay rectifier blown outage black bridge magnetron replaced checked | Fuse blown; power outage; bridge rectifier, magnetron, modulator failure | |
| Electrical | Topic 3 (6.4%) | MLC leaf motor stuck moving emitter reflector replaced pushing initialized cleaned | MLC motor failure |
| Topic 21 (2.2%) | Reflector reference line detector lost locked verified adjusted calibration reset | MLC leaf lost reflector; reflector out of calibration | |
| Control | Topic 4 (5.9%) | Console physics keyswitch sound timing frame board replaced check | Keyswitch replaced; buzzing sound from board |
| Topic 8 (4.4%) | PCB controller carriage program fitting driver change clear tightened | Replaced controller PCB; tightened up connection | |
| Topic 14 (3.4%) | New Varian software version device configuration old installed upgrade completed | Software upgrade; device configuration error | |
| Topic 20 (2.6%) | Prescription external pair receive communication timer database clinical failure close error reboot | Reboot computer and timer; external prescription error; communication error | |
| Topic 25 (1.9%) | Assembly secondary carrousel radial feedback captured replaced calibrate initialize | Recalibrated carrousel; lost secondary feedback | |
| Topic 26 (1.7%) | OBI stand modulator charge tapping charge capacitor box assisted replaced unstuck | OBI modulator error; OBI supervisor/computer | |
| Mechanical | Topic 5 (5.3%) | Gantry connector cable head noise display rotating patch bending monitor installed filled tightened | Gantry position error; bending system failure; tightened connector in head |
| Topic 9 (4.4%) | Couch readout long tolerance rotation jaw chain final hair collimator strip encoder calibrated adjust | Gantry readout calibration; couch long calibration; adjust collimator position | |
| Topic 11 (3.6%) | Drive limit key bearing height thumbwheel clutch movement replaced adjusted lubricated cleaned | Carriage bearing replaced; thumbwheel replaced; couch movement failure | |
| Topic 12 (3.6%) | ODI lamp bulb battery green burnt emergency grid socket mod red defective pointer light replaced used reseated | ODI lamp reseated; bulb burnt; emergency battery replaced | |
| Topic 13 (3.4%) | Arm tray collision switch touchguard sensor actuator cover plate pin angle holder fix removed adjusted attached | Arm position error; touchguard adjustment; sensor replaced/cleaned | |
| Topic 15 (3.3%) | Field laser mirror alignment capital output coincidence wall vertical dim light adjusted check closed | Field lamp replaced; mirror position adjustment; coincidence check | |
| Topic 16 (3.1%) | Pendant tight hand latch facility information door fiber optic ribbon terminal reset replaced collect | Reset latch; pendant terminal strip; optic cable failure | |
| Topic 17 (1.9%) | Pot fine pro mismatch offset shaft minor gear jaw edge replaced calibrated | Jaw calibration; replaced pot | |
| Topic 7 (4.6%) | Energy Thyratron beam URDS tuned interlock pulse voltage dequing tuned troubleshooting replaced | Thyratron failure; de‐quing and thyratron voltage low | |
| Beam | Topic 18 (2.9%) | Gun image breaker circuit HT RF button bolt filament balance tripped pulse adjusted reset checked | Gun driver failure; adjust gun filament voltage; reset HT circuit breaker |
| Topic 23 (2.1%) | MV symmetry profile wedge jaws resolve steering improved adjustment repositioned | Steering adjustment | |
| Support | Topic 10 (4.2%) | Flow fan pressure air cooled loop valve water conditioner chiller wave pumped changed shut tuned | Water contaminated; fan cooled failure; chiller pumped; conditioner fan noisy |
| Topic 14 (3.4%) | Hose pump water leak tank pump calibrated gas pop up | Hose water leak; water tank calibrated; pop‐up gas | |
| Dosimetry topic 24 (2.1%) | Ion chamber channel reading aria responding treatment dose dosimetry replaced check set calibrated | Replaced ion chamber; dose board failure; dosimetry calibrated | |
FIGURE 4Trends of topics with periodic pattern over years
FIGURE 5Trends of topics with increasing trend over years
FIGURE 6Trends of topics with decreasing trend over years
Dataset used in recent published applications using topic modeling
| Number of entries | Average length of each entry | Number of topics | Research field | Reference |
|---|---|---|---|---|
| 55,612 | Unknown | 30 | Patient online review analysis | [5] |
| 19,543 | 28.3 | 10 | Product review analysis (wireless mouse) | [14] |
| 30,268 | 27.6 | 10 | Product review analysis (diffuser) | [14] |
| 25,706 | 73.5 | 17 | Aviation safety report analysis | [3] |