Uwe Gründler1, Eva Ekesbo2, Martin Löwe1, Adelheid Gauly3.
Abstract
PURPOSE: Innovative hemodialysis systems are designed to ensure user safety and reduce operational time to allow health-care personnel to focus on patient care. The 6008 CareSystem has been developed to simplify the extracorporeal circuit of the system through a disposable cassette, automate operation steps, and facilitate handling in comparison to its predecessor - the 5008 CorDiax. The present investigations were performed with the aim of evaluating usability, safety, and ergonomic aspects of the new therapy system.Entities:
Keywords: ergonomics; hemodialysis equipment; safety; usability
Year: 2021 PMID: 34819757 PMCID: PMC8607128 DOI: 10.2147/MDER.S316610
Source DB: PubMed Journal: Med Devices (Auckl) ISSN: 1179-1470
Figure 1Blood lines for the extracorporeal circuit in the 5008 CorDiax HD system and safeline needed for online hemodiafiltration treatments and priming (filling and reinfusion).
Figure 2Extracorporeal circuit for the 6008 CareSystem as all-in-one cassette (6008 CareSet).
Figure 3Operator steps critical to successful hemodialysis-machine use.
Handling steps and critical contact points per treatment necessary to prepare, operate, and dismantle the two hemodialysis systems
| 5008 CorDiax | 6008 CareSystem | |
|---|---|---|
| Patients with fistula as vascular access | 122 | 92 |
| Patients with catheter as vascular access | 145 | 113 |
| Patients with fistula as vascular access | 31 | 23 |
| Patients with catheter as vascular access | 36 | 28 |
Time needed for handling steps for 5008 CorDiax and 6008 CareSystem
| 5008 CorDiax | 6008 CareSystem | ||||||
|---|---|---|---|---|---|---|---|
| n | Mean (minutes) | SD (minutes) | n | Mean (minutes) | SD (minutes) | Time savings (minutes) | |
| Machine preparation | 5 | 3.70 | 0.75 | 7 | 1.94 | 0.27 | 1.76 |
| Puncturing of a fistula patient using a connection set | 3 | 0.94 | 0.02 | 3 | 0.62 | 0.17 | 0.32 |
| Connection of a patient using a catheter-connection set | 2 | 0.68 | 0.00 | 5 | 0.60 | 0.16 | 0.08 |
| Reinfusion (assumed with simultaneous AV reinfusion) | 4 | 0.61 | 0.16 | 4 | 0.12 | 0.03 | 0.49 |
| Dismantling of 6008 | 4 | 1.23 | 0.21 | 4 | 0.98 | 0.16 | 0.26 |
| Sum fistula | 6.48 | 3.65 | 2.83 | ||||
| Sum catheter | 6.23 | 3.64 | 2.59 | ||||
Abbreviation: AV, arteriovenous.
Hand activity derived from ergonomic video analysis of each simulation study with the two hemodialysis systems
| 5008 CorDiax | 6008 CareSystem | |
|---|---|---|
| Left | 175 | 86 |
| Right | 174 | 88 |
| Left (with strong pressure) | 119 (13) | 46 (0) |
| Right (with strong pressure) | 158 (14) | 66 (0) |
Figure 4Classification of the HD systems on an HAL TLV graph, with threshold -limit value (TLV) and action limit as per Bernard.11 Reproduced from Bernard TE. ACGIH TLV for Hand Activity. American Conference of Governmental Industrial Hygienists (ACGIH), 2002. From ACGIH®, Documentation of the Threshold Limit Values and Biological Exposure Indices, 9th Edition. © 2021. Statement of Position Regarding the TLVs® and BEIs®. Available from: . Policy Statement on the Uses of TLVs® and BEIs®. Available from: .11