Poly(acrylamide-co-N,N'-methylenebisacrylamide) Monoliths for High-Peak-Capacity Hydrophilic-Interaction Chromatography-High-Resolution Mass Spectrometry of Intact Proteins at Low Trifluoroacetic Acid Content.

Marta Passamonti1,2, Chiem de Roos1, Peter J Schoenmakers1,2, Andrea F G Gargano1,2.   

Abstract

In this study, we optimized a polymerization mixture to synthesize poly(acrylamide-co-N,N'-methylenebisacrylamide) monolithic stationary phases for hydrophilic-interaction chromatography (HILIC) of intact proteins. Thermal polymerization was performed, and the effects of varying the amount of cross-linker and the porogen composition on the separation performance of the resulting columns were studied. The homogeneity of the structure and the different porosities were examined through scanning electron microscopy (SEM). Further characterization of the monolithic structure revealed a permeable (Kf between 2.5 × 10-15 and 1.40 × 10-13 m2) and polar stationary phase suitable for HILIC. The HILIC separation performance of the different columns was assessed using gradient separation of a sample containing four intact proteins, with the best performing stationary phase exhibiting a peak capacity of 51 in a gradient of 25 min. Polyacrylamide-based materials were compared with a silica-based particulate amide phase (2.7 μm core-shell particles). The monolith has no residual silanol sites and, therefore, fewer sites for ion-exchange interactions with proteins. Thus, it required lower concentrations of ion-pair reagent in HILIC of intact proteins. When using 0.1% of trifluoroacetic acid (TFA), the peak capacities of the two columns were similar (30 and 34 for the monolithic and packed column, respectively). However, when decreasing the concentration of TFA to 0.005%, the monolithic column maintained similar separation performance and selectivity (peak capacity 23), whereas the packed column showed greatly reduced performance (peak capacity 12), lower selectivity, and inability to elute all four reference proteins. Finally, using a mobile phase containing 0.1% formic acid and 0.005% TFA, the HILIC separation on the monolithic column was successfully hyphenated with high-resolution mass spectrometry. Detection sensitivity for protein and glycoproteins was increased and the amount of adducts formed was decreased in comparison with separations performed at 0.1% TFA.

Entities:  

Mesh:

Substances:

Year:  2021        PMID: 34807576      PMCID: PMC8655738          DOI: 10.1021/acs.analchem.1c03473

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


Introduction

Monolithic stationary phases are single porous pieces of stationary phase obtained through thermal- or photo-polymerization of a mixture consisting of monomer, cross-linker, porogens, and initiator. These stationary phases have been extensively investigated because of their inherent high permeability, low resistance to mass transfer, fast and simple preparation in small formats,[1−5] and a plethora of possible chemistries and retention mechanisms.[6−8] In the mid-1990s, the first acrylamide-based monoliths were developed by Hjertén and co-workers.[9,10] In 1997, Xie et al. studied an alternative approach to generate more rigid materials.[11] Monoliths with pore sizes up to 1000 nm were prepared by adjusting the synthesis conditions. No application of these materials for chromatographic purposes was reported. To the best of our knowledge, the first applications of hydrophilic-interaction chromatography (HILIC) to the study of intact proteins involved the separation of histone proteins on a weak-cation-exchange column[12−14] and later the separation of membrane proteins.[15,16] Zhang,[17] Pedrali,[18] Lauber,[19] and Periat[20]et al. discussed the unique selectivity of HILIC in the separation of glycoforms of subunits and intact glycoproteins. These results have drawn attention to HILIC–MS, with recent applications reporting high-resolution separations of neo-glycoproteins,[21] biopharmaceuticals,[22−25] biotechnological products,[26] and serum immunoglobins.[27] For these proteins, HILIC resolved proteoforms that would co-elute in RPLC. The majority of the separations of intact proteins and subunits by HILIC use silica particles with amide selectors and mobile phases based on acetonitrile, water, and acidic ion-pairs, such as trifluoroacetic acid (TFA). Negatively charged ion-pair reagents are thought to neutralize the polar basic groups present in proteins, improving peak shapes and increasing the contribution of neutral polar groups, such as glycans, to retention. However, TFA and other ion-pair reagents greatly reduce MS sensitivity and can lead to gas-phase ion adducts, which complicate the mass spectra.[28] The residual acidity of silica-based materials (free silanols) reputedly is a source of non-glycan-specific ion-exchange interactions and, therefore, increases the need for ion-pair reagents in the mobile phase to enhance the separation efficiency and reduce peak tailing for glycoforms. To reduce the amount of TFA in HILIC separations of proteins, Zhang and co-workers[17] bonded a polyacrylamide layer on non-porous silica particles of 700 nm size packed in a 2.1 μm ID (internal diameter) column. They obtained baseline separation of the five proteoforms of ribonuclease B, lowering the amount of TFA to 0.05% and adding 0.5% formic acid (FA). However, on polymeric materials that are not based on silica, but, for example, exclusively on polyacrylamide, can ion-exchange interaction with silanols be excluded? To date, no fully organic material for HILIC of intact proteins has been reported. In this study, we set out to synthesize and optimize an acrylamide-based monolithic stationary phase for protein separations in HILIC. We aimed to explore variations in the reaction mixture (monomer, cross-linker, and porogens) and to characterize the different capillary columns in terms of permeability, hydrophilicity, morphology, and porosity. Our aim was to obtain efficient columns for HILIC separations of intact proteins and proteoforms, which could be operated with very low concentrations of TFA as ion-pair reagent, so as to greatly enhance the sensitivity of MS detection in comparison with capillary columns packed with commercial silica-based amide stationary phases.

Experimental Section

Materials

Acrylamide (AA, electrophoresis grade, 99%), N,N′-methylenebisacrylamide (MbA, 99%), 3-(trimethoxysilyl)propyl methacrylate (γ-MAPS, 98%), 2,2′-azobisisobutyronitrile (AIBN, 98%), 1-octanol (OctOH, 99%), dimethyl sulfoxide (DMSO >99.9%), toluene, sodium hydroxide (NaOH), lysozyme from chicken egg white (Lys), carbonic anhydrase from bovine heart (CA, >90%), myoglobine from equine heart (Myo, >90%), cytochrome C (CC), bovine serum albumin (BSA), ribonuclease A from bovine pancreas (RnA), ribonuclease B from bovine pancreas (RnB, ≥80%), transferrin from human serum (Tf), cytosine (Cyt), trifluoroacetic acid (TFA, ≥99%), and formic acid (FA, analytical grade > 98%) were purchased from Sigma Aldrich (St. Louis, Missouri, United States). Methanol (MeOH) and acetonitrile (ACN) were purchased from Biosolve (Valkenswaard, The Netherlands). Hydrochloric acid 37% (HCl) was obtained from Acros (Geel, Belgium). High-purity (HP) water (18.2 MΩ·cm) was produced by a Sartorius (Göttingen, Germany) Arium 611UV Ultrapure-Water System. The capillary (0.20 mm ID, 0.36 mm OD) was purchased from CMScientific (Silsden, UK), and glass-lined tubing (300 mm length × 0.80 mm ID) was from VICI (Houston, TX).

Preparation of Poly(acrylamide-co-N,N′-methylenebisacrylamide) Monolithic and Silica Packed Columns

The surface of the fused-silica capillary was etched using NaOH and HCl and subsequently silanized using a 20% (v/v) γ-MAPS solution in toluene, as described by Courtois et al.[29] Thereafter, the capillary was flushed with toluene and dried with nitrogen. Several acrylamide-based monoliths were synthesized inside the capillaries starting from polymerization conditions similar to those described by Xie et al.11 using the recipes summarized in Table . To solubilize the monomers, the MbA was added to the porogens and sonicated for 1 h at a temperature between 30 and 40 °C. Once the cross-linker was dissolved, we added the monomer and kept sonicating for 45 min at the same temperature. Finally, the AIBN (1% w/w with respect to the monomers) was added and the mixture was vortexed. Thereafter, the silanized capillaries were filled with the polymerization mixture and the ends were closed using rubber pieces. Polymerization took place for 24 h at 60 °C in a thermostated glass tube. Finally, the monolithic capillary columns were thoroughly flushed with MeOH.
Table 1

Composition of the Polymerization Mixtures (Monomers and Porogens) Used to Prepare Different Acrylamide-Based Columns (Abbreviations of Monomers and Porogens Are Listed in the Materials Section)a

columnAA%MbA%OctOH%DMSO%DMF%
A5012.512.521.553.5 
A5513.7511.2521.553.5 
A60151021.553.5 
A6516.258.7521.553.5 
13DMF12.512.521.546.86.69
25DMF12.512.521.540.113.38
37DMF12.512.521.533.4420.06
50DMF12.512.521.526.7526.75

All the percentages are by weight. One percent (by weight; with respect to the monomers) of AIBN was added to each polymerization mixture.

All the percentages are by weight. One percent (by weight; with respect to the monomers) of AIBN was added to each polymerization mixture. An attempt at scaling-up the column dimensions was made using glass-lined tubing of 0.8 mm ID (150 mm length). The tubing was silanized, vinylized, and polymerized as described before. The A55 polymerization mixture was used without further optimization. A separation of a protein mixture was performed (see SI, Figure S7 for details). The column was packed using 2.7 μm AdvanceBio Glycan Mapping 120 Å (Agilent, Santa Clara, CA, USA) particles following the procedure described in ref (27).

Instruments and Conditions

SEM Experiments

SEM images of cross sections of the monoliths were recorded on a FEI Verios 460 instrument (Thermo Fisher Scientific, Eindhoven, The Netherlands) equipped with an Everhart–Thornley detector (EDT) using a 2 kV electron beam. The samples were sputter-coated with a 15 nm gold layer on top of a 5 nm layer of palladium.

Permeability Calculations

The permeability and hydrophilicity of the monolithic stationary phases were evaluated using an M-Class Acquity UPLC system (Waters, Milford, MA, USA) equipped with a binary solvent manager, a thermostated autosampler (1 μL sample loop), and a dual-wavelength tunable UV–vis detector with a 100 nL flow cell. The permeability (Kf) was determined by flushing MeOH through the acrylamide-based monolithic columns at various flow rates between 0.5 and 3 μL/min and using Darcy’s law, as follows: where Fm is the flow rate of the solvent, η is the dynamic viscosity of the solvent, L is the length of the monolith, ΔP is the pressure drop across the monolithic column, and r is the radius of the capillary.

Evaluation of Monolith Performance

Separations of intact proteins were performed on an UltiMate RSLCnano system (Thermo Fisher Scientific, Breda, The Netherlands) equipped with an autosampler (5 μL loop); thermostated column compartment with a 10-port, two-position valve; a loading-pump system (NCS-3500RS); and a UV–vis detector (VWD-3400RS). A trap column (5 × 0.3 mm ID; C4 stationary phase, 5 μm particle size, 300 Å pore size; Thermo Fisher Scientific) was used to inject samples from water-based solutions using a setup similar to that described by Gargano et al.30 Each reference protein was diluted in a solution of water/ACN = 95:5 with 0.1% TFA. The sample was loaded on the trap column at 20 μL/min using the loading pump for 5 min with a mobile phase (A) of 2% ACN in water with 0.1% TFA. The valve was switched after 3 min from the beginning of the method. Three washing cycles that consisted of fast linear gradients from 20 to 80% and back to 20% of B (2% of water in ACN) in 2 min were performed at the end of every separation to avoid carryover. Further information about the method can be found in the figure captions. The efficiency of the columns was assessed by calculating the peak capacity (nc) using the following equation: where tG is the gradient time and w0.5 is the average peak width at half-height. The asymmetry of the peaks (AS) was calculated using the Chromeleon software (Thermo Fisher Scientific) according to the following equation: where RW5% and LW5% are the right and left peak width at 5% of the peak height, respectively. The resolution for the separation of the five RnB proteoforms was estimated using the peak–valley ratio (P) equation suggested by Christophe:[31] where v is the signal intensity at the valley between two peaks and h is the intensity of the peak taken into consideration (see explanation in Figure S6).

HILIC–HRMS Conditions

A Q-Exactive Plus Biopharma (Thermo Fisher Scientific, Bremen, Germany) instrument was operated in positive-ion mode using steel-based emitters, a nano-ESI source at a voltage of 2 kV, a transfer capillary temperature of 300 °C, and S-lens of RF 70. The analyzed samples consisted of a 0.5 mg/mL solution of four proteins, viz., CC, CA, BSA, and Tf, and a 0.5 mg/mL solution of RnB. The reference proteins were diluted in 10% ACN in water. The volume injected was 0.5 μL for the protein mixture and 1 μL for the RnB sample. Acquisition parameters for the protein mix were as follows: m/z range, 400–6000; resolution @200 m/z, 17,500; 10 microscans; max IT (injection time), 200 ms; AGC (automatic gain control), 106; in-source CID (collision-induced dissociation), 20 eV; and HMR (high mass range) activated (trapping gas 1.2). Acquisition parameters for RnB were as follows: m/z range, 400–3000; resolution, 14,000; 4 microscans; max IT, 200 ms; AGC, 106; in-source CID, 20 eV; and intact protein mode activated (trapping gas 0.2). Total-ion current (TIC) and extracted-ion current (EIC) chromatograms and deconvoluted RnB masses were obtained using the Freestyle software (Thermo Fisher Scientific), while deconvoluted masses of the other proteins were obtained using UniDec.[32] Charge range, mass range, and sample-mass interval were 1–30, 12,000–13,000 Da, and 5 Da for CC; 1–50, 28,000–29,500 Da, and 1.0 Da for CA; 1–50, 66,000-68,000, and 1.0 Da For BSA; and 1–80, 79,000–83,000 Da, and 1.0 Da for Tf (MassIVE Repository link: ftp://massive.ucsd.edu/MSV000087977/).

Results and Discussion

Fabrication and Optimization of the Monolithic Stationary Phases

The composition of the polymerization mixture strongly affects the morphology and efficiency of organic monoliths. Based on the study of Xie et al.,11 we selected DMSO and OctOH as the porogen system for our polymerization. While DMSO can dissolve the monomers (good solvent), it alone cannot provide sufficiently large pores. Therefore, OctOH is added as a ″poor solvent″ to promote the development of flow-through pores. In our experiments, we first used the ratio AA/MbA = 70:30 described by Xie et al.11 and we polymerized the reaction mixture using DMSO/OctOH in 200 μm ID capillaries using thermal polymerization. The columns could be flushed with methanol. However, the polymeric material deteriorated when exposed to high amounts of water. Therefore, to obtain more rigid and water-resistant polymers, we used a higher percentage of cross-linker (AA/MbA = 50:50) as described in Table . To guarantee an acceptable back pressure and a less-dense polymer structure, the overall amount of porogens employed was increased (from 70% in the original mixture to 75% w/w). This resulted in polymer monoliths that could withstand pressures in excess of 40 MPa (400 bar; maximum pressure at which the columns were operated) under organic and aqueous conditions. Next, we prepared a series of acrylamide-based monolithic columns to study the influence of the composition of the polymerization mixture, with particular focus on the effects of the AA-to-MbA ratio and the porogen system (using DMF as the ternary solvent). The composition of the polymerization mixtures used is reported in Table .

Characterization of Polyacrylamide Monolithic Materials

Permeability

The columns (triplicates for each polymerization mixture) were tested before and after chromatographic separations, and a change of permeability values, as well as a shift of the retention times, was noticed for some of the columns (A60, A65, 13DMF, and 50DMF). The permeability values of these columns are reported in the SI (Figure S1). It appears that a reduced amount of cross-linker (for A60 and A65) and early phase separation in case of ternary porogen mixtures (13DMF and 50DMF) can result in unstable monolithic structures that degrade rapidly. The calculated permeability for the A50 columns obtained from a 50/50 AA/MbA mixture was 4.4 × 10–14 m2. Increasing the amount of monomer to 55% led to a small increase in permeability (Kf = 6.7 × 10–14 m2). The introduction of a third non-polar porogen (DMF) in the polymerization mixture of the 25DMF column appeared to induce an earlier phase separation compared to the binary porogen mixtures. However, the permeability of the resulting monolith decreased when adding 13 or 25% DMF and increased when increasing the DMF content to 37 or 50%, with the lowest permeability being observed for 25% DMF (Kf = 5.4 × 10–15 m2). The latter is of the same order as that observed for a column packed with 2.7 μm core–shell particles (Kf = 5.2 × 10–15 m2).

Scanning Electron Microscopy

SEM images were used to confirm the attachment of the monolith to the capillary walls, to inspect the monolith homogeneity (Figure S2a,b), and to evaluate the influence of the different porogens and the different concentrations of cross-linker used. We used this approach to analyze columns 25DMF, A50, and A55. All the monolithic stationary phases had a homogeneous microglobular structure and a good attachment to the inner wall of the capillary. Figure shows that the monolithic stationary phase in which 25% of the DMSO was substituted with DMF (column 25DMF, Figure A) features smaller globules than column A50 (Figure B) (approximately 0.5 μm compared to 1 μm).
Figure 1

SEM images of cross sections of three different acrylamide-based monolithic capillary columns. (A) 25DMF, (B) A50, and (C) A55. Scale bars, 10 μm.

SEM images of cross sections of three different acrylamide-based monolithic capillary columns. (A) 25DMF, (B) A50, and (C) A55. Scale bars, 10 μm.

Chromatographic Performance of Polyacrylamide Monoliths

Polyacrylamide Monolith for the Separation of Intact Proteins

To find a good compromise between analysis time and efficiency, we tested a wide range of flow rates (from 1 to 5 μL/min) on an A55 capillary column, maintaining the gradient volume constant (70 μL, Table S2). The highest peak capacity (nc = 106) was achieved at 1 μL/min using a 70 min gradient. However, to reduce the analysis time, a flow rate of 2 μL/min was further used in our study. To test the different column materials described in Table , we applied the same gradients (see experimental conditions) at 2 μL/min and 60 °C. The columns were used to separate a mixture of reference proteins, i.e., Myo, Lys, CA, and RNA, with different MWs (13 to 30 kDa) and pIs (6.4 to 9.32) and low occurrence of proteoforms that could lead to peak broadening due to partial resolution. In addition, we used RnB as a sample representative of glycoprotein separations. The results of this study are reported in Figure and Figures S4 and S5.
Figure 2

HILIC separation of intact proteins (A–C) on three capillary columns (A50, A55, and 25DMF; top to bottom of 0.2 mm ID × ca. 150 mm). The analysis was performed using a gradient from 90 to 85% (v/v) B in 1 min and then down to 55% B in 25 min at a flow rate of 2 μL/min. The analysis was carried at 60 °C, and the UV wavelength monitored was 214 nm. The chromatograms obtained with all the columns in Table are shown in Figure S4 of the SI.

HILIC separation of intact proteins (A–C) on three capillary columns (A50, A55, and 25DMF; top to bottom of 0.2 mm ID × ca. 150 mm). The analysis was performed using a gradient from 90 to 85% (v/v) B in 1 min and then down to 55% B in 25 min at a flow rate of 2 μL/min. The analysis was carried at 60 °C, and the UV wavelength monitored was 214 nm. The chromatograms obtained with all the columns in Table are shown in Figure S4 of the SI. The peak capacities for the protein mixture vary between 28 on the 13DMF column and 51 on the A55 column (Table S4). The columns present sufficient separation capacity for the RnB glycoforms. The five peaks are well-recognizable in the chromatograms obtained with most of the columns (Figure S5) and partially resolved (Table S5). Figure shows the separation of four proteins using the monolithic stationary phases that gave the best results, i.e., A50, A55, and 25 DMF. Good peak shapes (with minor tailing; asymmetry between 1.13 and 1.47) and good peak capacities (between 36 and 51) were achieved on these columns. The lower permeability and higher retention times of the 25DMF column in comparison with the A55 columns (Table S3) suggest an increased surface area in the former column. Similar peak–valley ratios were found for the A55 and 25DMF columns (Table S5). All these three polymerization mixtures yielded reasonable batch-to-batch repeatability (variation of retention time < 5%). Even though the A55 and 25DMF columns gave similar results, we continued with column A55 because of its better repeatability and higher permeability and, consequently, the potential for the preparation of longer columns.

Comparison of Protein Separations on Polyacrylamide Monoliths vs Silica Particles at Different Percentages of TFA

Next, we compared the performance of our column A55 and a packed capillary column to assess whether our newly developed material can be used with lower amounts of TFA. The packed column was prepared using 2.7 μm core–shell amide silica particles as described in our previous work.[21,24,27] The two HILIC columns were compared using different concentrations of TFA (0.1 and 0.005% TFA; Figure ). To evaluate the performance of the two columns, we used a protein mixture that included CC, CA, BSA, and Tf and the glycoprotein RnB (Figure S8). These proteins cover a wide range of molecular weights (12 to 80 kDa) and isoelectric points (5.60 to 9.59).
Figure 3

Comparison of HILIC separations of four intact proteins on the particle-packed column (0.2 mm ID × 135 mm L; black) and on the monolithic column (0.2 mm ID × 140 mm L; red). The analyses were performed using a gradient from 94 to 83% (v/v) B in 1 min and then down to 65% B in 26 min with 0.1% (v/v) TFA (A) and with 0.005% (v/v) TFA (B) in the mobile phases. Protein elution order: CC, CA, BSA, and Tf. All the analyses were carried out at 45 °C using a flow rate of 2 μL/min, and the monitored UV wavelength was 214 nm.

Comparison of HILIC separations of four intact proteins on the particle-packed column (0.2 mm ID × 135 mm L; black) and on the monolithic column (0.2 mm ID × 140 mm L; red). The analyses were performed using a gradient from 94 to 83% (v/v) B in 1 min and then down to 65% B in 26 min with 0.1% (v/v) TFA (A) and with 0.005% (v/v) TFA (B) in the mobile phases. Protein elution order: CC, CA, BSA, and Tf. All the analyses were carried out at 45 °C using a flow rate of 2 μL/min, and the monitored UV wavelength was 214 nm. We applied the same gradient (83–65% of B in 26 min) with a constant concentration of 0.1% (v/v) TFA for the separation of the protein mix on both columns. Interestingly, the peak capacity of the two columns was similar (34 vs 30), but the selectivity was different. On the (packed) silica-based column, the protein mix was not fully separated and CA and BSA co-eluted, while on the polyacrylamide (monolithic) column, the four peaks were baseline separated (Figure A and Table S6). The separation of RnB under gradient conditions (76–62% of B in 26 min) led to the partial separation of the five glycoforms on both columns with similar results in terms of peak–valley ratios (Table S7). Comparing the retention of the proteins between the two columns, the proteins eluted at a higher percentage of B (ca. 2% more ACN) from the monolithic column, possibly due to the lower surface area of this support (Tables S6 and S7). The same gradient used with 0.1% TFA was used for separations of the protein mixture with 0.005% TFA (Figure B). In the case of the polyacrylamide column, the retention decreased when moving from a high (0.1%) to a low concentration (0.005%) of TFA, with a minor reduction in the peak capacity (30 vs 23). A possible explanation for the change in retention is the increased hydrophilic character of the water phase due to the 20-fold decrease in the concentration of the fluorinated ion-pairing reagent in the mobile phase. In the case of the packed silica-based column, one of the proteins (Tf) did not elute from the column (not even when increasing the water content up to 90%) and the separation significantly worsened (peak capacity of 12), with only two significantly tailing peaks present in the chromatogram. We speculate that an increase in ion-exchange interactions due to the decreased concentration of ion-pairing reagent in the mobile phase may be responsible for this phenomenon. An in-depth analysis of the effects of different ion-pair agents and their concentrations is currently ongoing. For the separation of RnB (Figure S8), we broadened the gradient in the case of the polyacrylamide column (83–63% B in 26 min), while a shallower gradient (75–68% B in 26 min) was used on the silica-based column. At a low concentration of ion-pairing reagent (0.005% TFA), RnB eluted at 75 and 73% B from the polyacrylamide and particle-packed columns, respectively. Lower percentages (71 and 69%, respectively) were required when the concentration of TFA was 0.1% (v/v). Under optimal conditions for both columns, the monolithic column outperformed the packed silica-based column at 0.005% (v/v) TFA, as evidenced by the peak-valley ratio listed in Table S8.

Polyacrylamide Columns for HILIC–HRMS at Low TFA Content

High-resolution mass spectrometry (HRMS) is an effective tool to characterize proteoforms of proteins, as these can be clearly identified based on mass shifts, with further confirmation by fragmentation experiments.[8,33] HILIC mobile phases are rich in ACN, and this typically favors desolvation. However, in case of HILIC–MS of intact glycoproteins, the presence of TFA (0.1% v/v, 13 mM), which is typically needed to ensure good chromatographic resolution of the glycoforms, causes signal suppression and protein–TFA adducts. The latter makes the mass spectra more complicated and more difficult to interpret. To reduce adduct formation, high activation energies have to be used.[24] However, under these conditions, some proteins may undergo fragmentation. For this reason, we have tested post-column solutions to decrease adduct formations during HILIC–MS in other research.[30,34] Reducing the content of TFA in the mobile phase should reduce adduct formation in HILIC–MS. To investigate this, we tested our polyacrylamide columns using mobile phases containing 0.1% TFA, 0.005%TFA, and a combination of 0.1% FA and 0.005% TFA. We applied the same gradient (87–70% of B in 35 min) for the separations at lower concentration of TFA (0.005%), with and without FA. Figure shows TIC and EIC chromatograms recorded using the methods described in the captions at relatively low in-source CID (20 eV). The EIC chromatograms of the separation performed at higher TFA concentrations showed lower signal intensities and smaller peak areas (Table S8) in comparison with separations carried out with lower amounts of TFA. At high concentrations of TFA, the efficiency of the column was indeed higher than at a lower percentage of TFA (peak capacity, 42 vs 15). The latter chromatogram showed broader peaks that result in a partial loss of resolution.
Figure 4

TIC (a, c) and EIC (b, d) chromatograms obtained using a HILIC–MS method with (a, b) 0.1% (v/v) TFA and (c, d) 0.005% (v/v) TFA + 0.1% (v/v) FA. (a) Gradient from 94 to 83% (v/v) B in 2 min and then down to 67% in 35 min. (b–d) Gradient from 94 to 87% (v/v) B in 2 min and then down to 70% in 35 min. All the analyses were carried at 45 °C using a flow rate of 1.5 μL/min. The EIC chromatograms were obtained by summing the intensities of different charge states for CC (883.8 , 951.7, 1030.9, 1124.4, 1236.8, 1374.1, 1545.7, 1766.2, and 2060.5 m/z), CA (937.3, 968.4, 1001.9, 1037.6, 1076.6, 1117.3, 1161.9, 1210.4, 1262.8, 1320.3, 1383.0, 1452.3, 1528.5, 1613.4, and 1708.2 m/z), BSA (1278.5, 1303.5, 1329.6, 1356.6, 1384.9, 1414.4, 1445.1, 1477.2, 1510.7, 1545.8, 1582.0, 1621.2, 1664.5, and 1704.2 m/z), and Tf (2040.8, 2094.5, 2151.1, 2210.8, 2273.9, 2340.8, 2411.7, 2487.0, 2567.2, 2652.7, and 2744.2 m/z). All the EICs are ±0.2 m/z

TIC (a, c) and EIC (b, d) chromatograms obtained using a HILIC–MS method with (a, b) 0.1% (v/v) TFA and (c, d) 0.005% (v/v) TFA + 0.1% (v/v) FA. (a) Gradient from 94 to 83% (v/v) B in 2 min and then down to 67% in 35 min. (b–d) Gradient from 94 to 87% (v/v) B in 2 min and then down to 70% in 35 min. All the analyses were carried at 45 °C using a flow rate of 1.5 μL/min. The EIC chromatograms were obtained by summing the intensities of different charge states for CC (883.8 , 951.7, 1030.9, 1124.4, 1236.8, 1374.1, 1545.7, 1766.2, and 2060.5 m/z), CA (937.3, 968.4, 1001.9, 1037.6, 1076.6, 1117.3, 1161.9, 1210.4, 1262.8, 1320.3, 1383.0, 1452.3, 1528.5, 1613.4, and 1708.2 m/z), BSA (1278.5, 1303.5, 1329.6, 1356.6, 1384.9, 1414.4, 1445.1, 1477.2, 1510.7, 1545.8, 1582.0, 1621.2, 1664.5, and 1704.2 m/z), and Tf (2040.8, 2094.5, 2151.1, 2210.8, 2273.9, 2340.8, 2411.7, 2487.0, 2567.2, 2652.7, and 2744.2 m/z). All the EICs are ±0.2 m/z When adding FA to the mobile phase with low TFA content, the separation performances increased (peak capacity 40). Sharper peak shapes were obtained, and the selectivity was restored. Importantly, FA did not influence the MS spectra and the signal intensity since FA did not create adducts. Figure shows averaged spectra of the four proteins obtained from the HILIC–MS separations using different concentrations of TFA. The black traces indicate the protein spectra obtained using a mobile phases with 0.1% v/v TFA, while the red traces represent separations at a low concentration of TFA (0.005% (v/v) with 0.1% (v/v) FA added. The intensity of the red traces is higher than that of the black traces. This increase in sensitivity is most dramatic in the spectra of proteins with higher molecular weights (especially in Figure d). In fact, the signal intensity of the red trace for Tf is almost 1 order of magnitude higher than that of the black trace. The number of TFA–protein adducts is larger for the high-MW proteins than for the smaller proteins that have fewer interaction sites. Due to the formation of a lower number of adducts, a shift to higher charge states (lower m/z) is noticed when using 0.005% v/v TFA in the mobile phase. A lower amount of TFA reduces the charge distribution and shifts the most intense charge state. The corresponding deconvoluted spectra reflect the trend in terms of adduct observed and are shown in Figure S10.
Figure 5

Average spectrum from the chromatographic peak of (a) CC, (b) CA, (c) BSA, and (d) Tf obtained using the HILIC–MS method presented in Figure . The 0.1% (v/v) TFA (black trace) and 0.005% (v/v) TFA + 0.1% (v/v) FA (red trace). The numbers on the mass spectra indicate the charge-state interval.

Average spectrum from the chromatographic peak of (a) CC, (b) CA, (c) BSA, and (d) Tf obtained using the HILIC–MS method presented in Figure . The 0.1% (v/v) TFA (black trace) and 0.005% (v/v) TFA + 0.1% (v/v) FA (red trace). The numbers on the mass spectra indicate the charge-state interval.

Conclusions

Acrylamide-based monolithic stationary phases were successfully created within 200 μm ID capillaries using thermal polymerization. From SEM pictures, the microglobular structure was seen to be homogeneous and well bounded to the capillary wall. These highly hydrophilic monoliths were optimized for efficient HILIC separations of intact proteins. We selected the monolith with the polymerization mixture A55 as the best column due to its high permeability and high efficiency (peak capacity 51). At high concentrations of TFA (0.1% v/v), the A55 monolithic column showed greater selectivity but a similar peak capacity as a packed (amide-modified silica) column. However, at lower concentrations of TFA (0.005%), the monolithic column showed no loss in selectivity and the peak capacity remained fairly high (31 to 22), whereas the packed column showed broader and more-asymmetrical peaks and a dramatic change in efficiency (peak capacity 32 to 12). In coupling the separations with HRMS, it was found that a reduction of the concentration of TFA in the mobile phase resulted in enhanced sensitivity by 1 order of magnitude for the higher-MW proteins (e.g., Tf). The loss in chromatographic efficiency, incurred by reducing the TFA content, can be compensated by the addition of 0.1% (v/v) FA to the mobile phase on the monolithic column, which was found to raise the peak capacity from 15 back to 50. Moreover, from the (deconvoluted) mass spectra, it was clear that our polyacrylamide monolithic column showed a major advantage, as for high-MW proteins a lower amount of TFA resulted in a lower number of TFA–protein adducts.
  31 in total

1.  Hydrophilic interaction chromatography of intact, soluble proteins.

Authors:  Tim Tetaz; Simon Detzner; Arno Friedlein; Birgit Molitor; Jean-Luc Mary
Journal:  J Chromatogr A       Date:  2010-09-17       Impact factor: 4.759

2.  Orthogonal Middle-up Approaches for Characterization of the Glycan Heterogeneity of Etanercept by Hydrophilic Interaction Chromatography Coupled to High-Resolution Mass Spectrometry.

Authors:  Valentina D'Atri; Lucie Nováková; Szabolcs Fekete; Dwight Stoll; Matthew Lauber; Alain Beck; Davy Guillarme
Journal:  Anal Chem       Date:  2018-12-13       Impact factor: 6.986

3.  Potential of hydrophilic interaction chromatography for the analytical characterization of protein biopharmaceuticals.

Authors:  Aurélie Periat; Szabolcs Fekete; Alessandra Cusumano; Jean-Luc Veuthey; Alain Beck; Matthew Lauber; Davy Guillarme
Journal:  J Chromatogr A       Date:  2016-04-22       Impact factor: 4.759

Review 4.  Challenges and Advances in the Fabrication of Monolithic Bioseparation Materials and their Applications in Proteomics Research.

Authors:  Shujuan Ma; Ya Li; Chen Ma; Yan Wang; Junjie Ou; Mingliang Ye
Journal:  Adv Mater       Date:  2019-09-10       Impact factor: 30.849

5.  A cyclic-olefin-copolymer microfluidic immobilized-enzyme reactor for rapid digestion of proteins from dried blood spots.

Authors:  Bert Wouters; Irena Dapic; Thalassa S E Valkenburg; Sam Wouters; Leon Niezen; Sebastiaan Eeltink; Garry L Corthals; Peter J Schoenmakers
Journal:  J Chromatogr A       Date:  2017-01-28       Impact factor: 4.759

6.  Injection molded microfluidic chips featuring integrated interconnects.

Authors:  Dieudonne A Mair; Emil Geiger; Albert P Pisano; Jean M J Fréchet; Frantisek Svec
Journal:  Lab Chip       Date:  2006-07-31       Impact factor: 6.799

7.  Reassembling protein complexes after controlled disassembly by top-down mass spectrometry in native mode.

Authors:  Luis F Schachner; Denise P Tran; Alexander Lee; John P McGee; Kevin Jooss; Kenneth Durbin; Henrique Dos Santos Seckler; Lauren Adams; Erika Cline; Rafael Melani; Ashley N Ives; Benjamin Des Soye; Neil L Kelleher; Steven M Patrie
Journal:  Int J Mass Spectrom       Date:  2021-03-27       Impact factor: 1.934

8.  Characterization of Adeno-Associated Virus Capsid Proteins Using Hydrophilic Interaction Chromatography Coupled with Mass Spectrometry.

Authors:  Anita P Liu; Shailin K Patel; Tao Xing; Yuetian Yan; Shunhai Wang; Ning Li
Journal:  J Pharm Biomed Anal       Date:  2020-07-21       Impact factor: 3.935

9.  Definition of the mitochondrial proteome by measurement of molecular masses of membrane proteins.

Authors:  Joe Carroll; Ian M Fearnley; John E Walker
Journal:  Proc Natl Acad Sci U S A       Date:  2006-10-23       Impact factor: 11.205

10.  Confinement of Monolithic Stationary Phases in Targeted Regions of 3D-Printed Titanium Devices Using Thermal Polymerization.

Authors:  Marta Passamonti; Ischa L Bremer; Suhas H Nawada; Sinéad A Currivan; Andrea F G Gargano; Peter J Schoenmakers
Journal:  Anal Chem       Date:  2020-01-13       Impact factor: 6.986

View more