Traditional liquid phase extraction techniques that use optically responsive ligands provide benefits that enable cost-efficient and rapid measurements. However, these approaches have limitations in their excessive use of organic solvents and multistep procedures. Here, we developed a simple, nanoscale extraction approach by replacing the macroscopic organic phase with hydrophobic polymeric nanoparticles that are dispersed in an aqueous feed. The concentration of analytes in polymeric nanoparticle suspensions is governed by similar partition principles to liquid-liquid phase extraction techniques. By encasing optically responsive metal ligands inside polymeric nanoparticles, we introduce a one-step metal quantification assay based on traditional two-phase extraction methodologies. As an initial proof of concept, we encapsulated bathophenanthroline (BP) inside the particles to extract then quantify Fe2+ with colorimetry in a dissolved supplement tablet and creek water. These Fe2+ nanosensors are sensitive and selective and report out with fluorescence by adding a fluorophore (DiO) into the particle core. To show that this new rapid extraction assay is not exclusive to measuring Fe2+, we replaced BP with either 8-hydroxyquinoline or bathocuproine to measure Al3+ or Cu+, respectively, in water samples. Utilizing this nanoscale extraction approach will allow users to rapidly quantify metals of interest without the drawbacks of larger-scale phase extraction approaches while also allowing for the expansion of phase extraction methodologies into areas of biological research.
Traditional liquid phase extraction techniques that use optically responsive ligands provide benefits that enable cost-efficient and rapid measurements. However, these approaches have limitations in their excessive use of organic solvents and multistep procedures. Here, we developed a simple, nanoscale extraction approach by replacing the macroscopic organic phase with hydrophobic polymeric nanoparticles that are dispersed in an aqueous feed. The concentration of analytes in polymeric nanoparticle suspensions is governed by similar partition principles to liquid-liquid phase extraction techniques. By encasing optically responsive metal ligands inside polymeric nanoparticles, we introduce a one-step metal quantification assay based on traditional two-phase extraction methodologies. As an initial proof of concept, we encapsulated bathophenanthroline (BP) inside the particles to extract then quantify Fe2+ with colorimetry in a dissolved supplement tablet and creek water. These Fe2+ nanosensors are sensitive and selective and report out with fluorescence by adding a fluorophore (DiO) into the particle core. To show that this new rapid extraction assay is not exclusive to measuring Fe2+, we replaced BP with either 8-hydroxyquinoline or bathocuproine to measure Al3+ or Cu+, respectively, in water samples. Utilizing this nanoscale extraction approach will allow users to rapidly quantify metals of interest without the drawbacks of larger-scale phase extraction approaches while also allowing for the expansion of phase extraction methodologies into areas of biological research.
Entities:
Keywords:
extraction; metal determination; nanoparticles; nanosensors; optical probes
Quantifying
metals is an ongoing
scientific problem that has implications in a wide range of disciplines.[1−4] Traditional measurement techniques that use large-scale equipment
(e.g., atomic absorption, X-ray absorbance or fluorescence, and inductively
coupled plasma-mass spectrometry) have benefits such as high sensitivity,
identification of oxidation states, and multiple element determination.
However, there are disadvantages such as complicated operations, expensive
equipment and reagents, and time-intensive sample preparation. For
researchers that prefer a technique with minimal sample preparation
and a simple, sensitive quantification, these techniques are impractical.
Alternatively, there are small-scale analytical techniques that can
be applied to a wide range of applications and are easier to integrate
into routine experiments in the laboratory without the need for expertise.[5,6]A typical, simpler approach for metal quantification uses
liquid–liquid
extraction (LLE) procedures.[7] These assays
function on the basis of metal–ligand complex formation and
its relative solubility in one of two immiscible liquids, typically
an aqueous phase and an organic phase. While the metal salt resides
in the aqueous phase, the organic phase consists of an optically responsive,
hydrophobic ligand that binds to the metal in the hydrophobic phase,
stabilizes it, and changes the optical properties to enable detection.
While somewhat easy to operate, these techniques can require large
amounts of organic solvents, which can be toxic and hard to dispose
of. Further advances like solid phase extraction (SPE), cloud-point
extraction, or dispersive liquid–liquid microextraction (among
others) have been developed to quantify analytes while using significantly
smaller volumes of harmful solvents;[8] however,
these techniques can require expensive equipment, extensive procedures,
and a need for additional expertise.Classic examples of LLE
and subsequent metal quantification originate
from assays involving probes related to 2,2′-bipyridine.[9,10] These complexes generally absorb in the visible wavelength range
due to their metal-to-ligand charge transfer band and can be quantified
using UV–vis spectroscopy.[9] Building
on this premise, related species that have selectivity for Fe2+ have gained popularity.[11−13] For example, 4,7-diphenyl-1,10-phenanthroline
(bathophenanthroline, BP) has enabled sensitive detection of Fe2+ down to low micromolar concentrations even in the presence
of Fe3+ and other potential competing analytes.[14,15] While widely accepted and seemingly straightforward, this technique
has required acid digestion for sample pretreatment and Fe(BP)32+ extraction into a bulk organic phase, resulting
in the same LLE issues stated above. Using the same ligand, some have
performed SPE to measure Fe2+,[16] but these methods tend to have poor response times.[17]Polymeric nanosensors (PNS) are an analytical tool
that gives the
user the ability to measure ions or small analytes by encasing optically
responsive sensing components into a nanoscale hydrophobic polymer
matrix. Most of the PNS are plasticized polymer nanoparticles solubilized
by a surfactant coating and dispersed in an aqueous phase. There are
many sensing mechanisms for PNS discussed in fundamental reviews of
the field,[17,18] and more recent mechanistic developments
are discussed elsewhere.[19−22] Conventionally, the target ion is extracted from
the aqueous phase and stabilized in the hydrophobic particle core
by the recognition group. Upon extraction, the luminescent output
of the nanoparticle is altered. This can be a result of altering the
protonation degree of a pH-sensitive dye, displacing a solvatochromic
dye into the aqueous phase, or the quenching of either the fluorescence
or phosphorescence of a particular transducer inside the particle.[19−22] Polymeric nanosensors have been used to measure a wide range of
analytes in both biological (in vitro and in vivo) and chemical systems and have shown advantageous
character over their electrode and bulk membrane counterparts.[17]One example of nanoparticle-based measurement
systems is Bakker
and coworkers’ exploitation of the technology to make more
sensitive and selective complexometric titration reagents to measure
alkali, alkaline earth metals, and the anions NO3– and ClO4–.[23−25] The reagents
had optical readouts by encapsulating solvatochromic dyes or chromoionophores,
which significantly simplified the identification of the endpoint
in their titrations.[24,26] The use of nanoemulsions in this
regime resulted in heterogeneous complexometric reagents that are
more sensitive than the homogeneous-based methods and allow for highly
selective hydrophobic ligands to be utilized in titrimetry. Creating
heterogeneous complexometric reagents further illuminated the potential
for PNS to be used in a wide range of analytical analyses.The
extraction process of classic LLE and the sensing mechanism
for PNS follow similar principles, with partitioning of the analyte
between two immiscible phases and optical changes in a reporter group.
Accordingly, we hypothesized that the encapsulation of optically responsive
ligands into polymer nanoparticles would enable a rapid quantification
of metals of interest without the drawbacks of traditional LLE. While
nanoemulsions have been utilized as complexometric reagents, our nanoparticle
approach has shown that ligands that are typically dissolved in bulk
organic phases can be encapsulated into polymeric nanoparticles while
maintaining their metal-sensing functionality. This approach enables
easier analytical procedures and allows for adaptations of general
LLE approaches to novel biological applications where LLE is not feasible.
By utilizing the components of PNS as an extraction matrix, metals
that are usually quantified through bulk liquid phase extraction methods
can instead be measured with smaller volumes of harmful reagents and
much higher throughput. In this manuscript, we describe the combination
of optically responsive ligands traditionally used for LLE and PNS
as a nanoparticle extraction method for quantifying metals in water
samples (demonstrated with Fe2+, Al3+, and Cu+) toward a general platform for the adaptation of LLE-based
methods.
Experimental Section
Materials
Polyvinyl
chloride (PVC), bis ethylhexyl
sebacate (DOS), bathophenanthroline (BP), 8-hydroxyquinoline (8HQ),
bathocuproine (BC), tetrahydrofuran (THF), dichloromethane (DCM),
sodium tetrakis-[3,5-bis(trifluoromethyl)phenyl]borate (NaBARF; Selectophore),
sodium acetate, glacial acetic acid, Dulbecco’s phosphate-buffered
saline (PBS), ammonium iron(II) sulfate hexahydrate, iron(III) chloride,
cobalt(II) sulfate hydrate, copper(II) sulfate, nickel(II) sulfate,
magnesium sulfate, and thioglycolic acid were all purchased from Sigma-Aldrich
(St. Louis, MI, USA). 3,3′-Dioctadecyloxacarbocyanine perchlorate
(DiO; Invitrogen) and 1,1′-dioctadecyl-3,3,3′,3′-tetramethylindodicarbocyanine,
4-chlorobenzenesulfonate salt (DiD; Invitrogen) were purchased from
Thermo Fisher Scientific (Waltham, MA, USA). Microdialysis hollow
fibers (13 kDa cutoff) were purchased from Spectrum Laboratories,
Repligen (Waltham, MA, USA). Ferrous sulfate supplement tablets were
purchased from Signature Care. Creek water was obtained from Clear
Creek in Golden, CO, USA on February 21, 2021.
Fabrication
The
procedure for all preparticle mixtures
started by placing polyvinyl chloride (PVC) into 2 mL glass vials
with Teflon-lined caps. DOS was added to the PVC and set aside. The
ionic additive (NaBARF) was brought into 250 μL of THF and added
to the PVC vial. This vial was vortexed until the PVC was fully dissolved.
The ligand, which was dissolved in DCM, was added to the PVC vial
and vortexed for 45 s. The fluorescent Fe2+ NS required
that DiO (also dissolved in DCM) be added at the same point as the
ligand. All preparticle mixtures were stored in a refrigerator at
4 °C. For the mass of each component, see Table S1.All NS batches were made with the same protocol,
following a similar procedure to Dubach et al.[27] Two milligrams of the PEG-lipid was put into
a 4-dram scintillation vial. While the PEG-lipid is used for the solubilization
of these nanoparticle sensors, other surfactants have been used in
similar nanosensors (such as triblock copolymers).[28,29] Acetate buffer (pH 4.6 unless stated otherwise) was added to the
scintillation vial at a volume of 5 mL and then sonicated with a probe
tip sonicator (Branson Digital Sonifier 450; Branson Ultrasonics Corporation,
Danbury, CT) for 30 s at 20% of its maximum power. A total of 100
μL of the preparticle mixture was added to the acetate solution
and sonicated for 3 min under 20% maximum power. After sonication,
the excess polymer was removed with a 0.8 μm filter (Pall Corporation,
Port Washington, NY, USA), and the filtrate was stored in a 1.5-dram
screw cap glass vial at room temperature in the dark.
Nanosensor
Characterization
The calibration curves
of the NS were obtained by loading 100 μL of analyte standard
dilutions (in acetate buffer) and 100 μL of NS solution into
the wells of a Nunc MicroWell 96-well optical-bottom plate (Nalgene
Nunc International, Roskilde, Denmark). This yielded a final volume
of 200 μL in each well and concentrations ranging from 0.5 to
150 μM Fe2+, 0.5 to 100 μM Cu+,
or 1 μM to 1 mM Al3+ for each respective NS calibration.
The selectivity of the Fe2+ NS was determined with the
separate-solution method against Cu2+, Fe3+,
Co2+, and Ni2+ dilutions. The Cu+ and Al3+ NS selectivities were obtained by two-point,
separate-solution calibrations of each competing analyte at 100 μM.
Each calibration was obtained in triplicate using a Synergy H1 microplate
reader with absorbance readings at 533 and 478 nm or a fluorescence
endpoint at 505 nm for Fe2+, Cu+, or Al3+, respectively.The Fe2+ NS fluorescence
readout was obtained by exciting each sample at 450 nm and measuring
fluorescence of DiO at 501 nm. All standards were made new for each
calibration to limit errors due to analyte oxidation. To obtain the
Cu+ solution, 0.2% thioglycolic acid was added to Cu(II)SO4 (1:10 v/v). Further standard dilutions were made from this
stock.All NS calibrations were fit to either a simple linear
regression
or a 4-parameter logistic equation with GraphPad Prism 9.1.2 software
(San Diego, CA, USA). The statistical analyses for all calibrations
were also done in GraphPad Prism 9. The limits of detection were calculated
according to ref (30).The reversibility of the Fe2+ NS was tested using
the
procedure by Ferris et al., with some variation.[22] Briefly, the Fe2+ NS were concentrated
to a 10× stock, put into a hollow fiber microdialysis tube (MWCO,
13 kDa; Spectrum Laboratories), and sealed at each end and then secured
into an untreated 6-well culture plate with epoxy. After the epoxy
dried, acetate buffer was put into the same well and conditioned for
3 h. Then, the initial absorbance (533 nm) and fluorescence (λ
= 450/501) reading was taken on the Synergy H1 microplate reader using
the area scan settings (11 × 5, 1600 × 1600 μm spacing).
The acetate buffer was removed, and the well was rinsed before a dilution
of 150 μM Fe2+ was put into the well and incubated
for 30 min. After incubation, a measurement was taken with the same
settings, and the Fe2+ dilution was removed before another
rinse and addition of acetate-buffered solution. This cycle was repeated
twice in triplicate.The Fe2+ NS functional lifetime
was measured by calibrating
the fluorescence response according to the procedure on days 0, 2,
4, and 7.
Analysis of Real Samples
The reliability of the measurements
was tested with the Fe2+ NS against a ferrous sulfate supplement
tablet. One tablet (205 mg of anhydrous ferrous sulfate) was dissolved
in half a liter of acetate buffer and filtered with a 0.8 μm
PES membrane syringe filter. A series of two-fold dilutions were made
until the theoretical concentration (calculated by the dilution factor
and the 205 mg tablets) was within the NS linear range. These dilutions
were mixed with NS into the wells of a 96-well plate, and an absorbance
value at 533 nm was obtained with the Synergy H1 microplate reader.
With the absorbance values of the supplement dilutions, the respective
concentrations were back calculated with the regression line of the
standard curve. Taking the calculated concentrations and dividing
them by their dilution factor (stock = 1, dilution 1 = 0.5, dilution
2 = 0.25, etc.) allowed us to calculate the concentration of ferrous
sulfate in the stock solution. We also attempted to measure Fe2+ in creek water with the standard addition and external standard
method.[31] The accuracy of the Fe2+ NS measurements was confirmed by a comparison to a Hanna colorimeter—Checker
HC (Woonsocket, Rhode Island, USA)—that is used for selectively
measuring total iron.
Results
Our Fe2+ nanosensors
(Fe2+ NS) incorporate
bathophenanthroline (BP, log P = 3.37[32]) as the recognition agent and reporter for Fe2+. Instead of extracting Fe2+ from the aqueous
to the organic liquid phase, our approach uses BP to extract Fe2+ into a polymer nanoparticle phase with a 3:1 BP:Fe2+ ratio (Scheme ).
Scheme 1
Sensing Mechanism of the Fe2+ Nanosensors
Without Fe2+ in
the system, the sample does not absorb light at 533 nm. BP selectively
extracts Fe2+ into the polymer phase and undergoes a large
increase in absorbance at 533 nm, creating a red sample.
Sensing Mechanism of the Fe2+ Nanosensors
Without Fe2+ in
the system, the sample does not absorb light at 533 nm. BP selectively
extracts Fe2+ into the polymer phase and undergoes a large
increase in absorbance at 533 nm, creating a red sample.In the absence of Fe2+, BP does not absorb
light from
300 to 700 nm. Increasing Fe2+ in the system results in
a BP-mediated Fe2+ extraction and stabilization in the
polymer phase, causing an extensive increase in absorbance with a
peak at 533 nm (Figure ). Using these data, a calibration curve was obtained using a set
of standards ranging from 0 to 150 μM ferrous sulfate. The Fe2+ NS have a linear range from 1 to 30 μM (R2 = 0.994) and a limit of detection (LOD) of 0.8 μM
(Figure S2). This linear range is sufficient
for immediate implementation into a variety of applications like environmental
water testing[33,34] and serum-based diagnostics[35] (potentially requiring additional sample preparation
to release transferrin-bound iron).
Figure 1
Increasing Fe2+ results in
a large absorbance peak at
533 nm from the nanosensors. Linear and semilog calibration curves
from this data are provided in Figure S2.
Increasing Fe2+ results in
a large absorbance peak at
533 nm from the nanosensors. Linear and semilog calibration curves
from this data are provided in Figure S2.The sensor mechanism and the efficiency
of BP incorporation into
the nanoparticles were tested by centrifugal filtration and subsequent
testing on the filtrates. No optical response was seen when adding
50 μM Fe2+ to the nanosensor filtrate, and minimal
absorbance (533 nm) was observed in the filtrate of the nanosensors
that already responded to 50 μM Fe2+ (Figure S1).Before the analysis of a real
sample, the Fe2+ NS selectivity
was tested against Fe3+ chloride and Cu2+, Co2+, and Ni2+ sulfates. All resulting response curves
for potentially interfering metals showed negligible changes in absorbance
at 533 nm (Figure ). Phenanthrolines have also been shown to chelate other metals and
absorb light at some capacity at a range of wavelengths.[36] Absorbance spectra for the other metals are
shown in Figure S3 and confirm no significant
increase in absorbance between 300 and 700 nm.
Figure 2
Calibration curves obtained
with other transition metal standards
show a negligible response at 533 nm deeming the sensors highly selective
for Fe2+ over the other analytes. n =
3, and error bars are smaller than data points.
Calibration curves obtained
with other transition metal standards
show a negligible response at 533 nm deeming the sensors highly selective
for Fe2+ over the other analytes. n =
3, and error bars are smaller than data points.Other BP-based assays have shown poor selectivity over Cu species,[12] as phenanthroline
and its derivatives can form a 2:1 cuprous complex, which has an absorption
maximum at ∼478 nm and is extractable into an organic phase.[37] Thus, more confirmation of the selectivity over
Cu2+ and its impact on our Fe2+ calibrations
was needed. According to the U.S. Agency for Toxic Substances and
Disease Registry, the average amount of copper in river water is 0.16
μM.[38] Accordingly, we calibrated
the Fe2+ NS to a new set of standard dilutions that were
spiked with a range of relevant Cu2+ concentrations (0–5
μM). Figure S4 shows low Cu2+ interference on the Fe2+ response. The Fe2+ calibrations with spiked Cu2+ showed no difference compared
to the standards without Cu2+. All calibrations held a
good linear range between 1 and 30 μM.After demonstrating
Fe2+ NS selectivity, Fe2+ in a ferrous sulfate
supplement tablet was measured. Dissolving
the supplement tablet in 0.5 L of acetate buffer yielded a stock concentration
of 2.7 mM. Measuring the two-fold dilutions that fell within the linear
range of the assay resulted in a final calculated mass of 206 ±
8 mg of ferrous sulfate. A high correlation was obtained when comparing
these measurements to those from a commercially available method for
the determination of total iron (Figure ). In addition to measuring Fe2+ in a supplement tablet, the sensors were also used to measure water
collected from Clear Creek in Golden, Colorado with the standard addition
method. Statistical analysis of the regressions from spiked, unknown
samples and standards showed no difference between slopes, revealing
negligible matrix effects in the sample (Figure S5). The method of external standards showed no measurable
Fe2+ in the creek water, which was also confirmed with
the Fe-selective Hanna colorimeter.
Figure 3
Measured
iron concentrations from dilutions of a supplement sample
obtained with the Fe2+ nanosensors (y-axis)
highly correlate with those obtained with a standard commercial iron
selective colorimeter (x-axis). The dashed line indicates
a slope of 1 between the two methods. n = 3, and
error bars are smaller than the data points.
Measured
iron concentrations from dilutions of a supplement sample
obtained with the Fe2+ nanosensors (y-axis)
highly correlate with those obtained with a standard commercial iron
selective colorimeter (x-axis). The dashed line indicates
a slope of 1 between the two methods. n = 3, and
error bars are smaller than the data points.Following prior work with a fluorescence gating mechanism,[21] we incorporated a fluorescent reporter, which
is not sensitive to Fe2+ (DiO) into the particle core to
exhibit the user-tunability of this approach to a broader set of applications.
Due to the sufficient spectral overlap between the absorbance of Fe(BP3)2+ and the emission of DiO, the Fe2+ NS show a large turn-off response to increasing Fe2+ (Figure ). These calibrations
resulted in a response midpoint of 2.1 μM (Figure C). In conjunction with the
colorimetric responses shown in Figure , the fluorescence calibrations for the Fe2+ NS against the competing analytes show good selectivity (K based on a 50% maximal potential signal change
shows no response for off-target analytes due to minimal response)
(Figure C). Moreover,
the midpoint of the fluorescence response of these sensors is stable
for at least 4 days (Figure S6). Due to
the prevalence of fluorescence quenching by transition metals, including
Fe2+,[39] controlling for the
possibility that other targets would quench DiO was a necessity. Sensors
without BP (which should not respond to Fe2+) had no DiO
response to increasing analyte concentrations (Figure S7).
Figure 4
(A) Introducing a lipophilic fluorescent indicator (DiO,
λem of 501 nm) into the colorimetric Fe2+ NS results
in a fluorescence assay for Fe2+ determination. (B) Exposure
to Fe2+ results in high absorbance at 533 nm, which sufficiently
gates the DiO emission, causing a decrease in measured fluorescence.
(C) Sensors are selective for Fe2+. n =
3 for panels (B, error bars not shown) and (C).
(A) Introducing a lipophilic fluorescent indicator (DiO,
λem of 501 nm) into the colorimetric Fe2+ NS results
in a fluorescence assay for Fe2+ determination. (B) Exposure
to Fe2+ results in high absorbance at 533 nm, which sufficiently
gates the DiO emission, causing a decrease in measured fluorescence.
(C) Sensors are selective for Fe2+. n =
3 for panels (B, error bars not shown) and (C).At circumneutral pH (pH ∼6–8), Fe2+(aq) is not thermodynamically stable and is rapidly oxidized
to Fe3+ in the presence of oxygen.[40] As BP is easily protonated at acidic pH, it was necessary to determine
the impact of pH on the Fe2+ NS response. The response
at 50 μM Fe2+ was measured at pH 3.6, 4.6, and 5.6
(Figure S8). Indeed, the sensor response
at pH 3.6 was amplified relative to the signal at pH 5.6. Prior analyses
of both colorimetric and fluorescent phenanthroline-based probes show
an irreversible response to Fe2+.[41] As expected, our approach is also irreversible (Figure S9).To test how general our nanoparticle extraction
is, we used the
same approach with optically responsive ligands for both Al3+ and Cu+. For sensing Al3+, we used the responsive
probe 8-hydroxyquinoline (8HQ, log P = 1.72[42]). 8HQ extracts and stabilizes the free Al3+ from the aqueous phase into the sensor core with a 3:1 stoichiometry.[43] Increasing Al3+ in the system enhances
the fluorescence at 505 nm due to the formation of the Al(8HQ)3 complex inside the particle. Using standard buffer solutions,
the Al3+ NS show a linear range from 10 to 100 μM
(Figure A). Al3+ is often leached into streams and other water bodies when
the pH is below circumneutral.[44] The range
of the Al3+ NS allows for application in the analysis of
acid mine drainage or acid rain runoff where Al3+ can easily
exceed 10 μM.[45] While 8HQ can chelate
other metals, our selectivity tests (100 μM analyte) show that
the Al3+ NS are selective (Figure A, inset). The spectra from the NS–analyte
mixtures show no significant fluorescence peak from 300 to 800 nm
(Figure S10).
Figure 5
This approach is not
exclusive to Fe2+ extraction and
quantification. Using an optically responsive ligand (8HQ, λem = 505 nm) for Al3+ results in (A) sensitive and
(A, inset) selective NS for Al3+ (R2 = 0.9985). (B) Using BC, Cu+ NS show an increased
absorbance at 478 nm (R2 = 0.9989) and
(inset) high selectivity for Cu+. Calibrations are in standard
buffer sets with n = 3. Where not visible, the error
bars are smaller than data points.
This approach is not
exclusive to Fe2+ extraction and
quantification. Using an optically responsive ligand (8HQ, λem = 505 nm) for Al3+ results in (A) sensitive and
(A, inset) selective NS for Al3+ (R2 = 0.9985). (B) Using BC, Cu+ NS show an increased
absorbance at 478 nm (R2 = 0.9989) and
(inset) high selectivity for Cu+. Calibrations are in standard
buffer sets with n = 3. Where not visible, the error
bars are smaller than data points.The Cu+ NS response is mediated by a BP derivative,
bathocuproine (BC, log P = 4.47[46]). Inside the particle, the Cu(BC2)+ undergoes metal-to-ligand charge transfer and increases absorbance
with a peak at 478 nm in response to Cu+ (Figure S11). The calibration of these sensors with standard
buffer sets resulted in a linear range between 1 and 20 μM (Figure B). The Cu+ sensors also showed good selectivity over other metals (Figure B, inset).Similar to the Fe2+ NS, the characterization of the
extraction mechanism and confirmation of sufficient ligand partition
were done through centrifugal filtration and testing of the filtrate
(Figure S1). Additionally, as expected,
both the Al3+ and Cu+ NS responses were impacted
by pH (Figure S8).
Discussion
The
quantitation of metals is a challenge that has implications
in medicine, microbiology, water chemistry, and geology.[1−4] Innovations upon demanding analytical procedures that allow for
quick and high-throughput analysis are a necessity. Our nanoparticle
extraction assay provides a platform that enables easy quantification
of metals by adopting bulk phase extraction reagents and principles.
The partition principles that govern the extraction of the analyte
of interest are similar to those of traditional LLE.[47] Namely, the metal ion is poorly soluble inside the hydrophobic
extraction phase without the stabilization by the selective ligand;
with this component, the ion is stabilized in the particle, while
the other analytes remain in the aqueous phase.Others have
used similar materials (plasticized PVC) for improved
extraction methods of Fe2+, Cu+, and many other
analytes into a polymer membrane instead of an organic liquid.[16,48] Alternative materials have also been used for solid phase extraction
(SPE); Martínez et al. investigated the partition
of the phenanthroline–Fe2+ complex in modified acrylamide
hydrogels.[49] They determined the free Fe2+ concentrations in milk, providing a method for spectrophotometric
measurement in opaque samples. Similar to some metal-selective bulk
optode membranes,[50,51] these SPE approaches have resulted
in a long response time (i.e., a t95 of
15–50 min[49,52]) when analyzing real samples
due to their dependence on both diffusion of the ligand inside the
matrix as well as mass transport of the small amount of the analyte
in the aqueous phase.[17] For sensors with
a polymeric foundation, transitioning from a planar surface to a nanoparticle
dispersion has allowed for significantly lower response times due
to a much larger surface area-to-volume ratio.[17] In this work, the response time is too fast to measure
with our equipment and approach. We know that the sensor responds
faster than the time it takes between sample preparation and sample
measurement (∼15 s), but the response time is likely closer
to the ∼ms regime as demonstrated with similar sensors.[53]By adopting the structure from polymeric
nanoparticle sensors (PNS),
our extraction approach not only benefits from a faster response time
but also adds the potential application in systems where LLE and its
variations are not feasible. Nanoparticle sensors with this structure
have been used to characterize complex biological systems such as
clinically grown biofilms of Pseudomonas aeruginosa, eukaryotic intracellular environments, and animal models by probing
for a range of diverse analytes.[53−55] A large factor to consider
when applying analytical techniques to live models is the biocompatibility
of the reagents used as well as the assay’s procedure itself.[56] The reagents used in classic LLE and variations
thereof are often toxic to live samples. For example, applying a large
amount of an organic solvent to a live biological sample while using
traditional LLE will quickly destroy the sample’s integrity.
Though liquid phase microextraction (LPME) methods decrease the solvent
volume, biological samples still consist of highly complex matrices
that may interfere with liquid–liquid phase interfacial tension,
which can be detrimental to extraction efficiency.[57] As a result, many liquid extraction methods require pretreatment
of the sample, a hindrance not required with PNS.Another benefit
of the PNS methodology is the ability to substitute
the sensing components inside the particles when the analyte of interest
changes.[17] Thus, the nanoparticles in this
work can be used as an extraction phase for more than just one analyte,
provided that there is a hydrophobic ligand that solubilizes inside
the polymer matrix. We show in Figure that this approach can be tailored to an analyte of
interest depending on the ligand encased in the dispersed polymer
phase while using the same structural materials with no additional
steps.Utilizing an optically responsive ligand resulted in
sensors that
coupled the recognition and transduction moieties inside the particle,
though this is not mandatory. Other nanoparticle sensors have separated
these elements by utilizing an optically inert ligand for extraction
and quantification through ion exchange equilibrium with a pH-sensitive
fluorescent dye[17,18] or by adding an additional optically
responsive readout component into the particle.[20−22,58] Fluorescence gating mechanisms similar to our DiO
approach would likely be infeasible in bulk liquid phase extractions
due to the distance between the optically active groups, demonstrating
an additional benefit to embracing nanoparticles for the extraction
matrix.The sensing approach that we use here is similar to
the optical
complexometric reagents that were introduced by Bakker and associates.[23−26] However, in their work, the recognition and transduction elements
are distinct moieties; in this work, they are combined. While mechanistically
simpler than the NS with distinct recognition and transduction components,
there are drawbacks here such that the sensor response is less straightforward
to adjust.The addition of a ratiometric readout to fluorescent
sensors allows
for the quantitation of an analyte in a highly complex sample due
to the multiwavelength measurement.[59] While
this readout method is preferable, especially with complex matrices,
the incorporation of additional fluorophores into the nanoparticle
also adds potential complexity that may be hard to predict. Our approach
does not avoid this complexity. The addition of DiD as a reference
signal into our Fe2+ NS exhibited a favorable ratiometric
(DiO/DiD) response to Fe2+ but a large, unwanted increase
in the presence of Cu2+ (Figure S12). Future works will focus on this off-target interaction. The ratiometric
Fe2+ NS were still selective over metals other than Cu2+, which may still prove to be useful in some settings. Nevertheless,
the other two readout approaches (501 nm emission and 533 nm absorption)
provided high selectivity over analytes that may compete with BP’s
binding moiety (Figures and 4C).Our NS are not ratiometric;
thus, the sensor response is impacted
by sensor concentration. Specifically, by diluting the Fe2+ NS sensor batch by 0.5× and 0.25× stock solutions, both
the dynamic and linear range decreased in the colorimetric mode (Figure S13). As expected, in the fluorescence
mode, the span of the sensors also decreased proportionally to the
dilution factor with the normalized midpoint response (log EC50) also decreasing (Figure S13),
thus providing a slight increase in sensitivity.This nanoscale
extraction approach is beneficial for a researcher
looking to rapidly quantify a metal of interest. Though this approach
follows similar LLE principles, it is specified for sensing and is
not all-encompassing to LLE applications. An instance in which researchers
need to extract then purify their analyte of interest with chromatography
for further analysis is not compatible with our approach because the
metal–ligand complex stays encased in the nanoparticle. Another
scenario where this method would not work is one in which the researcher
needs to separate the two liquid phases after extraction to get rid
of optically opaque matrices. In our method, the nanoparticle suspensions
are injected directly into the sample, and the measurement is taken
on the mixture. For this reason, colored or turbid matrices could
significantly interfere with measuring analytes in the colorimetric
mode of this approach. While there are analytical procedures to mitigate
matrix effects, the fluorescence readout mode is still recommended
in these cases as it may enable quantification even in some colored
samples. Chemical probe synthesis is a popular area of study;[60] however, there is no probe for every application.
The last setting in which this new method would not work is one in
which the researcher is attempting to extract an analyte that does
not have a ligand that is soluble inside a hydrophobic phase. However,
water-soluble probes could be lipophilized to incorporate them into
the particle core.
Conclusions
Here, we introduce a
general platform for an alternative phase
extraction method for quantifying metal ions. Encapsulating hydrophobic
optically responsive ligands inside polymeric nanoparticles enabled
successful quantitation of Fe2+, Al3+, and Cu+ in aqueous samples. The Fe2+ nanosensor response
highly correlates to that of a commercial colorimeter, deeming this
approach a reliable analytical option. Adopting a polymeric nanoparticle
for the extraction matrix provides benefits like a one-step protocol
and simple formulation procedures, reduced amounts of organic solvents,
higher throughput, fast response times, and the potential for a large
expansion of phase extraction applications. Future works will consist
of identifying the unwanted, ratiometric response of the Fe2+ nanosensors to Cu2+, incorporating optically responsive
metal probes for other analytes into the nanoparticles, and applying
this nanoparticle extraction assay to samples with highly complex
matrices to rapidly measure analytes of interest.