Comparative healing of swine skin following incisions with different surgical devices.

Yuting Zhong1,2, Yufan Wei2,3, Ningning Min2,3, Qingyu Guan2,3, Jin Zhao2, Junyong Zhu2, Huayu Hu2,3, Rui Geng1,2, Chenyan Hong2,3, Yashuang Ji2,3, Jie Li2, Yiqiong Zheng2, Yanjun Zhang2, Xiru Li2.   

Abstract

BACKGROUND: Electrosurgical technology is widely used in surgical dissection and hemostasis, but the generated heat creates thermal injury to adjacent tissues and delays wound healing. The plasma blade (PB) applies pulsed radiofrequency (RF) to generate electrical plasma along the edge of a thin, flat, insulated electrode, minimizing collateral tissue damage. This study aimed to evaluate wound healing in swine skin following incision with a new surgical system that applies low-temperature plasma (NTS-100), a foreign PB, conventional electrosurgery (ES), and a scalpel blade.
METHODS: In vitro porcine skin and an in vivo porcine skin model were used in this study. Full-thickness skin incisions 3 cm in length were made on the dorsum of each animal for each of the 5 surgical procedures at 0, 21, 28, 35, and 42 days. The timing of the surgical procedures allowed for wound-healing data points at 1, 2, 3, and 6 weeks accordingly. Local operating temperature and blood loss were quantified. Wounds were harvested at designated time points, tested for wound tensile strength, and examined histologically for scar formation and tissue damage.
RESULTS: Local operating temperature was reduced significantly with NTS-100 (cut mode 83.12±23.55 °C; coagulation mode 90.07±10.6 °C) compared with PB (cut mode 94.46±11.48 °C; coagulation mode 100.23±6.58 °C, P<0.05) and ES (cut mode 208.99±34.33 °C, P<0.01; coagulation mode 233.37±28.69 °C, P<0.01) in vitro. Acute thermal damage from NTS-100 was significantly less than ES incisions (cut mode: 247.345±42.274 versus 495.295±103.525 µm, P<0.01; coagulation mode: 351.419±127.948 versus 584.516±31.708 µm, P<0.05). Bleeding, histological scoring of injury, and wound strength were equivalent for the NTS-100 and PB incisions.
CONCLUSIONS: The local operating temperature of NTS-100 was lower than PB, and NTS-100 had similarly reliable safety and efficacy. 2021 Annals of Translational Medicine. All rights reserved.

Entities:  

Keywords:  Plasma blade (PB); electrosurgery (ES); scar formation; thermal damage; wound healing

Year:  2021        PMID: 34790720      PMCID: PMC8576679          DOI: 10.21037/atm-21-3090

Source DB:  PubMed          Journal:  Ann Transl Med        ISSN: 2305-5839


Introduction

Electrosurgery (ES) is used in over 80% of surgical operations (1), among which high-frequency electrotome (conventional ES) is currently the most widely used electrosurgical device (2). ES applies radiofrequency (RF) alternating current to generate heat at the cutting blade tip and tissue interface, allowing surgeons to effectively cut tissue and achieve rapid hemostasis simultaneously, resulting in a shortened operation time and improved operational efficiency. The operating temperature reaches 200–300 °C, and the generated heat creates thermal injury to adjacent tissues, causing delayed wound healing (3-5). In recent years, the plasma blade (PB) has been a breakthrough in the history of the development of electrosurgical surgery. It uses high-intensity RF pulses to induce the formation of electrical plasma along the edge of the electrode, limiting this effect on the tissue surface. Thus, PB reduces local operating temperature, creating less thermal damage to adjacent tissues (6-8). At present, this technology is widely used in surgical procedures abroad (9,10). However, there has been inconsistent evidence regarding the comparative effectiveness of PB and ES in terms of the length of procedure time, drainage volume (11,12), postoperative seroma (11,12), and postoperative infection rates (9,13,14). Therefore, more improvements should be made for better clinical effectiveness. On this basis, our research team developed a new surgical system that applies low-temperature plasma (NTS-100) (). This system also adopts pulsed RF energy, which improves the high-frequency generator, enhances the frequency of the main engine, and reduces the width of the slit of the surgical electrode to reduce thermal damage while enhancing hemostasis ability. To evaluate the efficacy and safety of NTS-100, we compared the incision healing of NTS-100 with ES and a foreign PB using an in vitro and in vivo porcine skin model. Local operating temperature and blood loss were quantified. Wounds were harvested at designated time points, tested for wound tensile strength, and examined histologically for scar formation and tissue damage. We present the following article in accordance with the ARRIVE reporting checklist (available at https://dx.doi.org/10.21037/atm-21-3090).
Figure 1

New-typed low temperature plasma (NTS-100) and its electrode.

New-typed low temperature plasma (NTS-100) and its electrode.

Methods

Infrared analysis of the instrument operating temperatures

Infrared temperature images of the operation, which were conducted using NTS-100, PB (PEAK Surgical, Inc.), and ES (Valleylab Inc. USA), were captured by an infrared thermal imaging instrument (NS9300, NEC Corporation).

Incisional wound model

In vitro experiment

One piece of isolated pigskin with a size of 30×20×5 cm was used. The experiment comprised six groups: (I) conventional electrosurgery on cut 40 mode (ESC), (II) conventional electrosurgery on coagulation 40 mode (ESCo), (III) the plasma blade on cut 6 mode (PBC), (IV) the plasma blade on coagulation 8 mode (PBCo), (V) new low-temperature plasma surgery system on cut 60 mode (NPC), (VI) new low-temperature plasma surgery system on coagulation 80 mode (NPCo). All 3 power settings had similar widths of thermal damage (data not shown).

Porcine skin model

The study was conducted on 6 healthy Bama miniature pigs (weight: 21–23 kg; age: 10–13 months). Experimental animals were provided by Beijing Shichuangshiji Miniature Pig Breeding Base [SCXK (Beijing) 2013-0008] and were bred in the Animal Experiment Center of the First Medical Center of the Chinese PLA General Hospital. Animal experiments were performed under a project license (No. 20160010703) granted by the Ethics Committee of Chinese PLA General Hospital. All animal experiments complied with the National Institutes of Health Guide for the Care and Use of Laboratory Animals. The animals were acclimated for 1 week, and there was no abnormal situation. The experiment comprised 7 groups: the scalpel group (SC) and the above six groups.

Surgical procedure

Incisions were made on the isolated pigskin with 3 electrosurgical devices (NTS-100, PB, and ES) on cut and coagulation modes, as per the grouping and setting mode power described above. A total of 10 full-thickness skin incisions in each group were made at the same uniform speed. Full-thickness skin incisions 3 cm in length were made on the dorsum of each pig for each of the 5 surgical procedures at 0, 21, 28, 35, and 42 days. The most medial incision was placed 3 cm from the dorsal spinal processes, with each subsequent incision placed 2 cm lateral to the prior incision, all in parallel orientation (). The surgical procedure at each time point consisted of 1 incision made with a scalpel and ES, PB, and NTS-100 cut and coagulation modes (n=6 in each group). All wounds were closed with a 3-0 nylon suture, covered with antibiotic ointment, and removed the sutures on day 14. The timing of the surgical procedures allowed for wound-healing data points at 1, 2, 3, and 6 weeks, and the animals were sacrificed immediately after the surgical procedure on day 42.
Figure 2

Incisional wound model. (A) Design of incisions (full-thickness skin incisions 3 cm in length 2 cm apart were made on the dorsum of each animal for each of the 5 surgical procedures at 0, 21, 28, 35, and 42 days. In order to prevent errors due to different skin thicknesses, the operation sequence of the skin incisions on the dorsum of the 6 animals was performed by 7 surgical modes in turn). (B) The appearance of the in vitro incisions made with ES, PB, and NTS-100, respectively. (C) Acute histology of the skin incisions produced by (a) ES, (b) PB, and (c) NTS-100, respectively. Arrows indicate the zone of coagulation necrosis. Acute thermal injury was determined as the maximum width of the zone of coagulation necrosis using hematoxylin and eosin sections. Scale bar: 0.5 mm. ES, conventional electrosurgical instrument; PB, plasma blade; NTS-100, new surgical system which applies low temperature plasma.

Incisional wound model. (A) Design of incisions (full-thickness skin incisions 3 cm in length 2 cm apart were made on the dorsum of each animal for each of the 5 surgical procedures at 0, 21, 28, 35, and 42 days. In order to prevent errors due to different skin thicknesses, the operation sequence of the skin incisions on the dorsum of the 6 animals was performed by 7 surgical modes in turn). (B) The appearance of the in vitro incisions made with ES, PB, and NTS-100, respectively. (C) Acute histology of the skin incisions produced by (a) ES, (b) PB, and (c) NTS-100, respectively. Arrows indicate the zone of coagulation necrosis. Acute thermal injury was determined as the maximum width of the zone of coagulation necrosis using hematoxylin and eosin sections. Scale bar: 0.5 mm. ES, conventional electrosurgical instrument; PB, plasma blade; NTS-100, new surgical system which applies low temperature plasma.

Blood loss evaluation

Filter paper used to absorb blood was weighed before and after surgery. Only 1 type of filter paper was used. After the incision operation of each instrument, a piece of sterile filter paper was immediately used to cover the incision for 1 minute until blood was completely absorbed on the filter paper during this period. Blood loss was measured by weighing the filter paper. No suction was used. The total weight during the operation of the blood filter paper was weighed by an electronic analysis balance (JA1003B, Shanghai Yueping Scientific Instrument Co., Ltd.).

Histological examination

A 2×1 cm specimen containing the incision was excised and fixed in 10% buffered formalin for 24 hours and embedded in paraffin. Hematoxylin and eosin (HE) staining and immunohistochemistry were performed to observe the histomorphology. All specimens were coded and evaluated by light microscopy. Acute thermal injury (time point at week 0) was determined as the maximum width of the zone of coagulation necrosis using HE sections (15). All sections were examined for the width of the scar and the number of T lymphocytes and macrophages (time points at 1, 2, 3, 6 weeks). Images for measuring scar width were processed using Image J software (National Institutes of Health). The number of T lymphocytes and macrophages were counted using the Image-Pro Plus software program.

Wound strength

A 2×1 cm full-thickness strip of tissue with the incision at the midpoint was excised using a steel template. The wound burst strength of the tissues was measured using a biological force tester (MTS Bionix 858, MTS company).

Statistical analysis

The experimental data were statistically analyzed and processed by SPSS 17.0 software. The data were reported as mean ± SEM. One-way analysis of variance (ANOVA) was used to compare the means of multiple groups. The value of P<0.05 was considered statistically significant, and the value of P<0.01 was considered highly statistically significant.

Results

Instrument operating temperatures

The color of isolated skin after the procedure changed obviously in the in vitro experiment. The incisions made by ES, PB, and NTS-100 were charred black, reddish-brown, and light yellow, respectively (). The difference in thermal damage was visible to the naked eye. The operating temperature of isolated pigskin with NTS-100 measured by an infrared camera was 83.12±23.55 °C in cut mode and 90.07±10.6 °C in coagulation mode, which were lower than ES (208.99±34.33 °C in cut mode and 233.37±28.69 °C in coagulation mode, P<0.01) and the PB (100.23±6.58 °C in coagulation mode, P<0.05) (n=10 in each group). NTS-100 system cut (0.046±0.034 g) and coagulation (0.041±0.024 g) modes demonstrated reduced bleeding compared with scalpel incisions (0.410±0.291 g, P<0.001), and there was no statistically significant difference compared with PB (cut mode 0.020±0.009 g; coagulation mode 0.018±0.011 g) and EC (cut mode 0.014±0.007 g; coagulation mode 0.012±0.010 g). The results indicated that NTS-100 had the same hemostatic effect as PB and ES.

Zone of coagulation necrosis

Acute thermal damage from NTS-100 (cut mode 247.345±42.274 µm; coagulation mode 351.419±127.948 µm) was significantly less than ES incisions (cut mode 495.295±103.525 µm, P=0.002; coagulation mode 584.516±31.708 µm, P=0.012), and there was no statistically significant difference compared with PB (cut mode 349.120±30.470 µm; coagulation mode 459.359±67.160 µm) ().

Scar formation

Histological scoring of injury with the NTS-100 was significantly less than ES (cut mode: 847.9250±146.943 versus 1,806.617±254.372 µm at 1 week, P<0.01; 593.696±232.071 versus 907.152±180.281 µm at 2 weeks, P<0.05; 861.145±219.715 versus 412.890±160.994 µm at 3 weeks, P<0.05; 669.394±182.444 versus 366.206±100.260 µm at 6 weeks, P<0.05; coagulation mode: 2,285.386±334.232 versus 893.7277±151.942 µm at 1 week, P<0.01; 1,591.094±254.284 versus 830.904±143.469 µm at 2 weeks, P<0.05; 1,075.201±490.276 versus 728.059±127.176 µm at 3 weeks; 722.071±167.581 versus 392.539±81.264 µm at 6 weeks, P<0.05), and there was no statistically significant difference compared with the PB ().
Figure 3

The examination of scar width. (A) Comparison of scar width at the same time after incision for (a) ESC, (b) ESCo, (c) PBC, (d) PBCo, (e) NPC, and (f) NPCo. All sections were stained with hematoxylin and eosin stain. All specimens were coded and evaluated by light microscopy (100×) by a single pathologist in a blinded manner. Scale bar: 0.5 mm. (B) The width of the scar in the wounds at 1, 2, 3, and 6 weeks after incision (*, compared with NTS-100). SC, scalpel; ES, conventional electrosurgical instrument; ESC, conventional electrosurgical instrument in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PB, plasma blade; PBC, plasma blade in cut mode; PBCo, plasma blade in coagulation mode; NP, new surgical system which applies low temperature plasma; NPC, new surgical system which applies low temperature plasma in cut mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode.

The examination of scar width. (A) Comparison of scar width at the same time after incision for (a) ESC, (b) ESCo, (c) PBC, (d) PBCo, (e) NPC, and (f) NPCo. All sections were stained with hematoxylin and eosin stain. All specimens were coded and evaluated by light microscopy (100×) by a single pathologist in a blinded manner. Scale bar: 0.5 mm. (B) The width of the scar in the wounds at 1, 2, 3, and 6 weeks after incision (*, compared with NTS-100). SC, scalpel; ES, conventional electrosurgical instrument; ESC, conventional electrosurgical instrument in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PB, plasma blade; PBC, plasma blade in cut mode; PBCo, plasma blade in coagulation mode; NP, new surgical system which applies low temperature plasma; NPC, new surgical system which applies low temperature plasma in cut mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode.

Inflammatory markers

The distribution of T lymphocyte and macrophage immunohistochemical staining differed in each equipment group at each time point. Generally, the infiltration of T lymphocytes and macrophages for each device increased until the number reached the maximum at 1 week. After that, the infiltration at weeks 2, 3, and 6 gradually decreased with time. The number of infiltrating T lymphocytes and macrophages in the scalpel group was significantly lower than that in the other groups when the healing time was 1, 2, and 3 weeks (P<0.05). At 6 weeks, there was no statistically significant difference between the scalpel group and the NTS-100 cutting mode group. As for the comparison of the 3 electrosurgical devices, at 1 and 2 weeks, there were no statistically significant differences among them. At 3 and 6 weeks, the incisions created by NTS-100 contained fewer T lymphocytes than ES incisions both in cut and coagulation mode (cut mode: 165.67±65.332 versus 247.83±24.045 at 3 weeks, P<0.01; 75.17±14.497 versus 130.33±17.061 at 6 weeks, P<0.01; coagulation mode: 210.33±65.938 versus 288.00±27.225 at 3 weeks, P<0.01; 101.00±17.967 versus 186.67±21.087 at 6 weeks, P <0.01) ().
Figure 4

The examination of T lymphocytes. (A) Immunohistochemistry staining results of T lymphocytes by (a) scalpel, (b) ES, (c) PB, and (d) NTS-100, respectively. All sections were examined for T lymphocytes (CD3 M7254; Dako, Carpinteria, CA, USA). Sections were examined under high-power magnification (100×) and the number of cells was counted by a blinded observer. (B) Number of T lymphocytes in the wounds at 0, 1, 2, 3, and 6 weeks after surgery (*, compared with NTS-100). SC, scalpel; ES, conventional electrosurgical instrument; ESC, conventional electrosurgical instrument in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PB, plasma blade; PBC, plasma blade in cut mode; PBCo, plasma blade in coagulation mode; NP, new surgical system which applies low temperature plasma; NPC, new surgical system which applies low temperature plasma in cut mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode.

The examination of T lymphocytes. (A) Immunohistochemistry staining results of T lymphocytes by (a) scalpel, (b) ES, (c) PB, and (d) NTS-100, respectively. All sections were examined for T lymphocytes (CD3 M7254; Dako, Carpinteria, CA, USA). Sections were examined under high-power magnification (100×) and the number of cells was counted by a blinded observer. (B) Number of T lymphocytes in the wounds at 0, 1, 2, 3, and 6 weeks after surgery (*, compared with NTS-100). SC, scalpel; ES, conventional electrosurgical instrument; ESC, conventional electrosurgical instrument in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PB, plasma blade; PBC, plasma blade in cut mode; PBCo, plasma blade in coagulation mode; NP, new surgical system which applies low temperature plasma; NPC, new surgical system which applies low temperature plasma in cut mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode. Likewise, at 3 and 6 weeks, the incisions created by NTS-100 contained fewer macrophages than ES incisions both in cut and coagulation mode (cut mode: 122.00±13.416 versus 161.83±33.469 at 3 weeks, P<0.01; 44.17±5.636 versus 83.17±8.329 at 6 weeks, P<0.01; coagulation mode: 151.33±24.468 versus 180.17±33.689 at 3 weeks, P<0.01; 79.33±9.331 versus 107.83±9.109 at 6 weeks, P<0.01) (). There was no significant difference between NTS-100 and PB. The results indicated that NTS-100 had less infiltration of inflammatory cells in the late stage of skin healing, and the scar formation was better ().
Figure 5

The examination of macrophages. (A) Immunohistochemistry results of macrophages by (a) scalpel, (b) ES, (c PB, and (d) NTS-100, respectively. All sections were examined for macrophages (CD68 NCL-L-CD68, Novacostra, Newcastle, UK). Sections were examined under high-power magnification (100×) and the number of cells was counted by a blinded observer. (B) Number of macrophages in the wounds at 0, 1, 2, 3, and 6 weeks after surgery (*, compared with NTS-100). SC, scalpel; ES, conventional electrosurgical instrument; ESC, conventional electrosurgical instrument in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PB, plasma blade; PBC, plasma blade in cut mode; PBCo, plasma blade in coagulation mode; NP, new surgical system which applies low temperature plasma; NPC, new surgical system which applies low temperature plasma in cut mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode.

Table 1

Comparison of indexes for wound healing during cutting and coagulation modes of 3 electrosurgical devices and a scalpel () (n=6)

Evaluation indicatorsSCElectrosurgical devices
Cutting modeCoagulation mode
ESCPBCNPCESCoPBCoNPCo
The width of the scar (μm)
   1w336.587±162.890*1,806.617±254.372**1,023.3680±266.121847.9250±146.9432,285.386±334.232**1,259.6817±165.343893.7277±151.942
   2w130.605±23.055*907.152±180.281*830.199±360.154593.696±232.0711,591.094±254.284*1,163.824±211.416830.904±143.469
   3w102.447±18.675*861.145±219.715*904.347±55.709412.890±160.9941,075.201±490.276931.086±92.515728.059±127.176
   6w90.667±31.182*669.394±182.444*464.391±90.935366.206±100.260722.071±167.581*633.405±84.487*392.539±81.264
The number of T lymphocytes
   1w215.00±28.242*366.67±50.262355.50±49.209337.50±44.948401.67±60.328375.50±82.718365.00±52.242
   2w156.50±24.006*299.67±40.272288.83±31.346285.67±19.500325.33±29.070316.00±36.797308.50±31.463
   3w102.17±15.497*247.83±24.045**169.33±38.640165.67±65.332288.00±27.225**216.67±54.581210.33±65.938
   6w58.17±12.287130.33±17.061**88.33±10.65275.17±14.497186.67±21.087**116.33±13.441101.00±17.967
The number of macrophages
   1w132.33±16.597**214.67±43.871208.33±15.488210.67±41.495246.00±33.251241.17±20.634241.00±39.350
   2w101.33±15.756*195.00±40.492177.17±19.395187.83±42.546211.17±29.587211.83±12.057210.17±33.891
   3w77.67±16.452**161.83±33.469**139.33±21.021122.00±13.416180.17±33.689**166.83±16.339151.33±24.468
   6w39.00±12.96183.17±8.329**61.00±17.561*44.17±5.636107.83±9.109**96.50±8.550*79.33±9.331
Wound burst strength (N)
   1w9.71±4.54000000
   2w48.81±37.0810.91±2.727.32±5.1612.25±6.278.52±5.6711.02±4.0417.32±7.43
   3w116.62±60.70*30.39±16.68*29.50±12.1259.52±13.3640.01±27.49*30.55±16.9368.91±22.40
   6w334.08±92.65*146.65±50.45159.77±57.54168.05±38.46124.98±107.85118.88±49.46167.62±47.24

**, P<0.01; *, P<0.05, compared with NTS-100. SC, scalpel; ESC, conventional electrosurgical instrument in cut mode; PBC, plasma blade in cut mode; NPC, new surgical system which applies low temperature plasma in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PBCo, plasma blade in coagulation mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode.

The examination of macrophages. (A) Immunohistochemistry results of macrophages by (a) scalpel, (b) ES, (c PB, and (d) NTS-100, respectively. All sections were examined for macrophages (CD68 NCL-L-CD68, Novacostra, Newcastle, UK). Sections were examined under high-power magnification (100×) and the number of cells was counted by a blinded observer. (B) Number of macrophages in the wounds at 0, 1, 2, 3, and 6 weeks after surgery (*, compared with NTS-100). SC, scalpel; ES, conventional electrosurgical instrument; ESC, conventional electrosurgical instrument in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PB, plasma blade; PBC, plasma blade in cut mode; PBCo, plasma blade in coagulation mode; NP, new surgical system which applies low temperature plasma; NPC, new surgical system which applies low temperature plasma in cut mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode. **, P<0.01; *, P<0.05, compared with NTS-100. SC, scalpel; ESC, conventional electrosurgical instrument in cut mode; PBC, plasma blade in cut mode; NPC, new surgical system which applies low temperature plasma in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PBCo, plasma blade in coagulation mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode.

Wound burst strength evaluation

Scalpel and NTS-100 incisions exhibited greater tensile strength compared with PB and ES incisions at all time points. By 3 weeks, the healed wound strength of NTS-100 (cut mode 59.52±13.36 N, coagulation mode 68.91±22.40 N) was stronger than ES (cut mode 30.39±16.68 N, coagulation mode 68.91±22.40 N), and the scar formation was better (, ).
Figure 6

Wound burst strength at 1, 2, 3, and 6 weeks after surgery (*, compared with NTS-100). SC, scalpel; ES, conventional electrosurgical instrument; ESC, conventional electrosurgical instrument in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PB, plasma blade; PBC, plasma blade in cut mode; PBCo, plasma blade in coagulation mode; NP, new surgical system which applies low temperature plasma; NPC, new surgical system which applies low temperature plasma in cut mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode.

Wound burst strength at 1, 2, 3, and 6 weeks after surgery (*, compared with NTS-100). SC, scalpel; ES, conventional electrosurgical instrument; ESC, conventional electrosurgical instrument in cut mode; ESCo, conventional electrosurgical instrument in coagulation mode; PB, plasma blade; PBC, plasma blade in cut mode; PBCo, plasma blade in coagulation mode; NP, new surgical system which applies low temperature plasma; NPC, new surgical system which applies low temperature plasma in cut mode; NPCo, new surgical system which applies low temperature plasma in coagulation mode.

Safety evaluation

During the NTS-100 operation, the animals had a stable heart rate and respiration. There was no unexpected morbidity and mortality, and the physiological condition of the experimental animals was normal, including heart rate, respiration, diet, and activities, both in intraoperative and postoperative anesthesia recovery. The NTS-100 surgical system had the same safety profile as other electrosurgical devices.

Discussion

Traditionally, the scalpel has been used as a surgical cutting tool due to the advantages of precision, tissue integrity preservation, and superior wound healing. However, bleeding limits its widespread use in procedures across surgical specialties, as hemostatic control and dissection capability are fundamental to the practice of surgery (7,16). Consequently, numerous electrosurgical devices have been explored to ensure the basic functions of cutting and coagulation (2,17-19). Although hemostasis can be achieved, electrosurgical devices induce thermal damage to adjacent tissues, inferior wound healing, and a poor cosmetic effect, limiting their applications (20,21). A thorough understanding of these technologies and their applications and the potential risks and the appropriate settings can help improve the effectiveness and safety of surgery. For example, in neurosurgery, the application of ES is limited, which may cause acute and subacute nerve damage. In thyroid surgery, the thermal damage induced by electrosurgical equipment may cause damage to the recurrent laryngeal nerves, resulting in dysphonia or even potentially life-threatening asphyxia (22-24). The PB is a relatively new electrosurgical technology developed to achieve cutting and coagulation effects with a relatively low operating temperature, which reduces thermal damage (6,25). It applies high-frequency RF pulses to generate electrical plasma along the edge of the electrode (26). Several studies have shown that this technology allows for the precise dissection of rabbit, bovine, and human retinas and lens capsules, even when completely submerged in a liquid medium (27-30). The PB system has been approved by the Food and Drug Administration (FDA) for use in the cutting and coagulation of soft tissues during otolaryngology surgical procedures. A prospective double-blinded randomized controlled trial showed that PB tonsillectomy achieved a faster postoperative recovery in terms of the time taken to achieve pain-free swallowing (31). However, no significant differences were found between PB and ES regarding postoperative bleeding (11,32,33), postoperative hematoma (9,11,14), and inflammatory responses (13). Therefore, more improvements should be made for better clinical effectiveness. On this basis, our research team developed a new surgical system that applies low-temperature plasma (NTS-100). This system also adopts pulsed RF energy, which improves the high-frequency generator, enhances the frequency of the main engine, and reduces the width of the slit of the surgical electrode to reduce the thermal damage while enhancing hemostasis ability. This study investigated operating temperature, blood loss, thermal injury depth, inflammation, and scarring in swine skin by comparing the histology of incisions made with NTS-100, PB, ES, and a scalpel. The local operating temperature is related to the thermal injury of tissue and the thermal diffusion. Except for thermal injury, the heated tissues may adhere to the hot electrode. Tissue sticking impairs the accuracy and efficiency of the operation process (34). Therefore, operating temperature control is essential to improve instrument performance, reduce tissue sticking, and prevent thermal injury. Our study showed that the operating temperature of isolated pig skin with NTS-100 was much lower than ES in cut mode and coagulation mode and lower than PB. It indicated that NTS-100 had a better performance in preventing thermal damage and reducing tissue sticking. Moreover, the hemostasis effect was comparable among these three devices, indicating that despite the lower working temperature and less thermal damage to tissues of NTS-100, high operating temperature and deep thermal coagulum associated with ES are not essential for hemostasis. Further studies on this effect should be conducted. Wound healing is one of the most complex processes in the human body, which involves various cell types in the phases of hemostasis, inflammation, growth, re-epithelialization, and remodeling (35). Several studies have shown that inflammatory cells are recruited and activated in the early stage of injury to defend against bacteria and combat self and foreign antigens. As the inflammatory phase ends, angiogenesis occurs, and granulation tissue forms, shifting the wound microenvironment from the inflammatory to the growth state (36-38). However, increased inflammation can alter the outcome of wound repair. T lymphocytes and macrophages are critical to normal wound healing and tissue regeneration (39-42), so to some extent, the degree of infiltration of T lymphocytes and macrophages can reflect the healing of wounds. At each designated time point, scar formation was less on the incisions of NTS-100 than on that of ES. At 3 and 6 weeks, the incisions created by NTS-100 contained fewer T lymphocytes and macrophages than incisions created by ES both in cut and coagulation mode. The infiltration of T lymphocytes and macrophages and the burst strength of the incisions created by the NTS-100 and PB were similar. These results indicate that the NTS-100 created less scar formation and reduced inflammation, which might be associated with lower local working temperature and less tissue thermal damage. The levels of T lymphocytes and macrophages were higher in the NTS-100 and PB groups than in the scalpel group, and the wound burst strength was lower in the NTS-100 and PB groups than in the scalpel group. The greater infiltration by T lymphocytes and macrophages during the course of healing in the ES group is consistent with previous research results showing that abnormal inflammation is associated with reduced wound burst strength (8). Wound healing in adults results in a fibrotic scar resulting from differential cellular responses towards mechanical stress within the healing skin, but excessive scarring shifts the balance towards fibrotic states of hypertrophic scarring and keloid formation (43,44). The appearance of scars was strongly related to thermal damage and wound histological examination (31). Histological examinations showed reduced scar width in the NTS-100 and PB groups compared with the ES group, which indicated that increased inflammation influences wound repair. The incisions by NTS-100 can achieve a better cosmetic effect. Further study of scar formation is necessary, especially in the field of plastic and reconstructive surgery. Furthermore, additional benefits of the NTS-100 were observed during experiments. Heated electrodes are associated with surgical smoke, which is harmful to surgeons and may affect an operation by interfering with the surgeon’s visibility (45,46). The difference in smoke production and blade eschars was visible to the naked eye. NTS-100 produced significantly less surgical smoke than ES, and the blade was free of eschar, resulting in better surgical fluency. The reason may be related to the plasma at the edge of the blade. Compared with ES, the plasma prevents burning tissue at a high temperature, which produces smoke and forms eschars. Therefore, follow-up research will focus on determining the content of particulate matter in surgical smoke, the composition of the smoke, especially the composition and content of harmful substances, and find suitable indicators to objectively evaluate and compare the sharpness of the blade and the fluency of the operation. Consequently, NTS-100 tries to reduce the occupational exposure of medical staff and make the operation process more efficient at the same time (47). Nowadays, NTS-100 has already been used in breast surgery, general surgery, and head and neck surgery, and clinical data have been collected to verify its clinical effectiveness. We believe that as an innovative electrosurgical device with a relatively low operating temperature, our surgical system, which applies low-temperature plasma (NTS-100), has a promising future and will be widely used in surgical procedures. Then, more clinical data will be compared with current electrosurgical devices (10) to improve our system continuously.

Conclusions

The local operating temperature of NTS-100 was lower than PB, and NTS-100 had similarly reliable safety and efficacy. The article’s supplementary files as
  46 in total

Review 1.  Complications and recommended practices for electrosurgery in laparoscopy.

Authors:  M P Wu; C S Ou; S L Chen; E Y Yen; R Rowbotham
Journal:  Am J Surg       Date:  2000-01       Impact factor: 2.565

Review 2.  The hazards of surgical smoke.

Authors:  Brenda C Ulmer
Journal:  AORN J       Date:  2008-04       Impact factor: 0.676

3.  Lateral thermal spread induced by energy devices: a porcine model to evaluate the influence on the recurrent laryngeal nerve.

Authors:  Masaru Hayami; Masayuki Watanabe; Shinji Mine; Yu Imamura; Akihiko Okamura; Masami Yuda; Kotaro Yamashita; Tasuku Toihata; Yoshiaki Shoji; Naoki Ishizuka
Journal:  Surg Endosc       Date:  2019-03-07       Impact factor: 4.584

Review 4.  Scalpel versus electrosurgery for major abdominal incisions.

Authors:  Kittipat Charoenkwan; Zipporah Iheozor-Ejiofor; Kittipan Rerkasem; Elizabeth Matovinovic
Journal:  Cochrane Database Syst Rev       Date:  2017-06-14

Review 5.  Scalpel versus electrosurgery for abdominal incisions.

Authors:  Kittipat Charoenkwan; Narain Chotirosniramit; Kittipan Rerkasem
Journal:  Cochrane Database Syst Rev       Date:  2012-06-13

6.  Safety of Ligasure exact dissector in thyroidectomy with continuous neuromonitoring: a porcine model.

Authors:  Tzu-Yen Huang; Yi-Chu Lin; Hsin-Yi Tseng; Gianlorenzo Dionigi; Hoon-Yub Kim; I-Cheng Lu; Pi-Ying Chang; Feng-Yu Chiang; Che-Wei Wu
Journal:  Gland Surg       Date:  2020-06

7.  Wound healing of skin incisions produced by ultrasonically vibrating knife, scalpel, electrosurgery, and carbon dioxide laser.

Authors:  R Hambley; P A Hebda; E Abell; B A Cohen; B V Jegasothy
Journal:  J Dermatol Surg Oncol       Date:  1988-11

8.  An electrical plasma dissection tool for surgical treatment of chronic ulcers: Results of a prospective randomised trial.

Authors:  Giovanni F Marangi; Tiziano Pallara; Daniela Lamberti; Eleonora Perrella; Raffaele Serra; Francesco Stilo; Giovanni De Caridi; Andrea Onetti Muda; Paolo Persichetti
Journal:  Int Wound J       Date:  2018-03-30       Impact factor: 3.315

9.  The invention of microwave surgical scissors for seamless coagulation and cutting.

Authors:  Tohru Tani; Shigeyuki Naka; Soichiro Tani; Hisanori Shiomi; Koichiro Murakami; Atsushi Yamada; Dang Tran Khiem
Journal:  Surg Today       Date:  2018-05-10       Impact factor: 2.549

10.  Preventive Effect on Seroma of Use of PEAK PlasmaBlade after Latissimus Dorsi Breast Reconstruction.

Authors:  Yoshihiro Sowa; Naoki Inafuku; Takuya Kodama; Daiki Morita; Toshiaki Numajiri
Journal:  Plast Reconstr Surg Glob Open       Date:  2018-12-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.