Matevž Topolovec1, Ingrid Milošev1,2, Andrej Cör1. 1. Valdoltra Orthopaedic Hospital, Jadranska c. 31, 6280 Ankaran, Slovenia. 2. Department of Physical and Organic Chemistry, Jožef Stefan Institute, Jamova 39, 1000 Ljubljana, Slovenia.
If a fracture of CoC bearing couple in THA is suspected in an asymptomatic patient, a CT scan should be performed as soon as possible and the failed bearing should be replaced with CoC or CoP and not to MoP, because of the risk for metallosis and further revision operations.
Introduction
Alumina ceramic-on-ceramic (CoC) bearing couple has been introduced in the 1970s as an alternative for younger and more active patients, based on the lower wear rates [1, 2]. Unfortunately, the favorable wear properties of CoC have been largely suppressed by the brittleness of the material, which resulted in occasional fractures of the bearing couple [3]. Despite the fact that the fracture rate of the fourth-generation (Biolox delta®) ceramic significantly decreased compared with the third generation ceramic [4], the fear how to cope with the revision operation for ensuring the long-term survival of a new bearing couple remains.
Case Presentation
In December 1999, a 50-year-old man with post-traumatic painful right hip arthrosis underwent a cementless total hip arthroplasty with a third generation alumina CoC bearing couple (Biolox Forte®). The implants used were Cremascoli Ortho with ANCA-FIT acetabular cup and ANCA-FIT modular femoral stem. The patient had good clinical results until March 2007, when he experienced little pain and crepitus noise. After 3 months, the operating surgeon examined him. The hip was pain free with full range of motion; however, a radiograph indicated some small calcinations which could also represent fractured ceramic particles. A hip arthrography excluded aseptic loosening or an infection. However, after 3 years since the patient experienced the first symptoms in 2007, he was in October 2010 presented with more pain, more crepitus noises and limited hip motion, all characteristics of ceramic fracture. A radiographic image taken (Fig. 1) now clearly revealed a ceramic liner fracture (Fig. 2a). The ceramic head was not fractured but covered with severe black metal tracks (Fig. 2b); these originated from the action of sharp ceramic debris against metal parts of the implant. Metal particles thus formed became embedded within the articulating surface of the fractured CoC bearing. The patient underwent revision surgery in November 2010. A direct lateral approach true the primary scar was used. The ceramic fragments were removed, a polyethylene liner was inserted and the ceramic head was replaced with a CoCr femoral head. Therefore, metal-on-polyethylene (MoP) bearing was introduced. An extensive synovectomy was done. Only manual tissue rinsing was done, because of the absence of pulsatile lavage system. Periprosthetic tissue samples were sent for histological analysis, where numerous ceramic particles, some even quite large, were found (Fig. 3). After revision, the patient fully recovered without any pain and discomfort in the hip.
Figure 1
Radiograph showing fractured ceramic liner with marked fragmentation (arrow) taken in 2010 prior revision due to fracture of CoC bearing.
Figure 2
Picture of a retrieved ceramic liner fracture (a) and ceramic head covered with severe black metal tracks (b).
Figure 3
Periprosthetic tissue retrieved at revision operation in 2010 following the fracture of CoC bearing with ceramic particles (a); some of them are quite large. (b) (arrows) (Hematoxylin and eosin; obj mag. ×20).
Radiograph showing fractured ceramic liner with marked fragmentation (arrow) taken in 2010 prior revision due to fracture of CoC bearing.Picture of a retrieved ceramic liner fracture (a) and ceramic head covered with severe black metal tracks (b).Periprosthetic tissue retrieved at revision operation in 2010 following the fracture of CoC bearing with ceramic particles (a); some of them are quite large. (b) (arrows) (Hematoxylin and eosin; obj mag. ×20).At a regularly post-operative check-up about 3 years later (July 2013) a radiograph revealed a worn metal head with a radio-dense shadow by the great trochanter (Fig. 4). The patient underwent a second revision in October 2013. The same operative approach and skin scar were used as in the first two operations. The black-stained pseudocapsule was removed together with the black periprosthetic tissue. There was an obvious infiltration of the muscles with metallosis; however, no clear necrotic sections were seen, so an extensive necrectomy was not necessary. The worn CoCr femoral head (Fig. 5) was replaced with a 4th-generation ceramic head (Biolox delta) with titanium sleeve. Furthermore, we removed the acetabular cup, although well fixed, and replaced it for a cemented Durasul cup with a cross-linked polyethylene liner. The femoral component was also well fixed and the femoral taper was intact so we decided to leave it in. Periprosthetic tissue samples were once again sent for histological investigation with slides now demonstrating significant metallosis with lymphocyte infiltration (Fig. 6). At approximately 10 years after the first and 7 years after the second revision operation the patient is doing fine and fully returned to normal daily activities. A control radiograph (Fig. 7a) taken just after the second revision and a follow-up radiograph (Fig. 7b) taken in July 2020 revealed no osteolytic changes or evidence of complication. Harris hip score [1, 2] for the patients on a last control was 73.
Figure 4
Radiograph taken in 2013 prior second revision of MoP bearing showing radio-dense shadow by the great trochanter representing severe metallosis.
Figure 5
Picture of a retrieved worn CoCr femoral head (left) next to an unused femoral head of the same size (right). The femoral head was retrieved at the second revision of MoP bearing in 2013.
Figure 6
Periprosthetic tissue retrieved at revision operation in 2013 due to severe wear of metal femoral head of MoP bearing representing significant metallosis (a) (thin arrows) and with extensive perivascular lymphocyte infiltration (b) (thick arrow) (Hematoxylin end eosin; obj.mag ×10).
Figure 7
(a) Radiograph taken after the second revision operation (year 2013), (b) Radiograph taken on last follow-up (year 2020).
Radiograph taken in 2013 prior second revision of MoP bearing showing radio-dense shadow by the great trochanter representing severe metallosis.Picture of a retrieved worn CoCr femoral head (left) next to an unused femoral head of the same size (right). The femoral head was retrieved at the second revision of MoP bearing in 2013.Periprosthetic tissue retrieved at revision operation in 2013 due to severe wear of metal femoral head of MoP bearing representing significant metallosis (a) (thin arrows) and with extensive perivascular lymphocyte infiltration (b) (thick arrow) (Hematoxylin end eosin; obj.mag ×10).(a) Radiograph taken after the second revision operation (year 2013), (b) Radiograph taken on last follow-up (year 2020).
Discussion
Fracture of a ceramic component in CoC bearing couple can represent a major problem for long-term outcomes after revision of THA [3, 4, 5]. The first problem is the prompt diagnostics of a ceramic fracture component. Namely, we know that a non-displaced ceramic linear fracture can be asymptomatic [6]. In our case, it took more than 3 years (first symptoms March 2007, confirmed ceramic fracture November 2010) to reach an obvious clinical and radiological presentation. In the meantime, consequences of a long delay can lead to material deterioration and extensive destruction of bone and tissues [4] and can cause pelvic discontinuity [7]. On this basis, if suspicion of a ceramic bearing is suggested, computed tomography rather than plain radiographs should be used [8], because small fragments can be mistaken for a heterotopic ossification [9]. In our case, no CT diagnostic or component alignment radiographic studies were done. This would surly helped in the rapid diagnosis of our asymptomatic patient (3 years before the severe symptoms accrued) and the revision operation would be much easier, especially in the regard struggling of removing all ceramic particles and debris. CT imaging was also not done before the second revision operation because of fear of stale image due to huge metallosis.The second problem represents the dilemma for choosing the appropriate bearing couple following the ceramic fracture. Although the debate for the most appropriate revision bearing couple is still going on [10, 11], more and more studies [5, 8, 11] suggest that the revision bearing couple should be CoC or ceramic-on-polyethylene (CoP), specifically because of the risk of third-body wear. Since revision of fractured CoC bearing for a MoP bearing may pose a potential high risk for metallosis and further cobalt toxicity it should be avoided. Despite the huge metallosis observed on X-ray image (Fig. 3) and the presence of metallic particles in histological periprosthetic tissue sections (Fig. 5), our patient did not experience any symptoms of systemic cobalt toxicity [12] leading to heavy consequences like heart failure. On the other hand, laboratory tests to determine the chromium and cobalt levels are not part of a routine clinical practice but this case indicates that their introduction should be taken into consideration to allow earlier alert. However, regular radiological and clinical follows-ups are now mandatory for this patient, because of a higher risk for aseptic loosening due to huge particle loads from two former bearing couples with two distinctive failure mechanisms – ceramic fracture and heavy metallosis.
Conclusion
Given that the decision of choice of a revision bearing couple after fracture of ceramic component is still at the discretion of the surgeon, this case-report emphasizes the need for establishment of diagnostic routes and recommended protocols, which would allow easier recognition of potential fracture and diminish its consequences.Prompt diagnostic and quick revision operation of a non-displaced CoC bearing couple, together with the right choice of the revision bearing can deprive the patient of much suffering.
Authors: Francesco Traina; Enrico Tassinari; Marcello De Fine; Barbara Bordini; Aldo Toni Journal: J Bone Joint Surg Am Date: 2011-12-21 Impact factor: 5.284
Authors: Rohit Rambani; David M Kepecs; Tatu J Mäkinen; Oleg A Safir; Allan E Gross; Paul R Kuzyk Journal: J Arthroplasty Date: 2017-01-04 Impact factor: 4.757
Authors: Jérôme Allain; Françoise Roudot-Thoraval; Joel Delecrin; Philippe Anract; Henri Migaud; Daniel Goutallier Journal: J Bone Joint Surg Am Date: 2003-05 Impact factor: 5.284