| Literature DB >> 34786534 |
Eiji Nakajima1, Hiroko Tsunoda2, Mariko Ookura3, Kanako Ban4, Yuko Kawaguchi4, Mami Inagaki5, Norihiko Ikeda6, Kinya Furukawa1, Takashi Ishikawa7.
Abstract
OBJECTIVE: To compare the performance of two-dimensional synthetic mammography (SM) combined with digital breast tomosynthesis (DBT) (SM/DBT) and full-field digital mammography (FFDM) including women with DBT (FFDM/DBT) undergoing secondary examination for breast cancer.Entities:
Keywords: Breast cancer; and full-field digital mammography; digital breast tomosynthesis; mammography; two-dimensional synthetic mammography
Year: 2021 PMID: 34786534 PMCID: PMC8570192 DOI: 10.5334/jbsr.2457
Source DB: PubMed Journal: J Belg Soc Radiol ISSN: 2514-8281 Impact factor: 1.894
Characteristics of the breasts analyzed.
|
| |||
|---|---|---|---|
| AGE | BREAST COMPOSITION | CANCER (N = 52) | HISTOLOGICALLY BENIGN (N = 28) |
|
| |||
| 31–82 years | A: 1 /B: 22 /C: 57 /D: 13 | DCIS 9 (17.3%) | Fibroadenoma 7 |
|
| |||
186 breasts in 93 nationality women.
A. fatty, B. scattered fibroglandular densities, C. heterogeneously dense, D. extremely dense, DCIS, ductal carcinoma in situ; IDC, invasive ductal carcinoma; Negative, no indication of cancer.
Radiation dose for image acquisition.
|
| |||
|---|---|---|---|
| RADIATION DOSE (MGY) | MEDIAN | ||
|
| |||
| FFDM | MLO | 0.69–6.29 | |
|
| |||
| CC | 0.70–6.18 | 1.47 | |
|
| |||
| DBT (=SM/DBT) | MLO | 0.92–5.43 | |
|
| |||
| CC | 0.93–5.03 | 1.50 | |
|
| |||
| FFDM/DBT | MLO | 1.64–11.72 | |
|
| |||
| CC | 1.63–11.21 | 2.95 | |
|
| |||
FFDM, full-field digital mammography; DBT, digital breast tomosynthesis; SM, two-dimensional synthetic mammography; MLO, mediolateral oblique; CC, craniocaudal.
Inter-rater reliability with Cohen’s Kappa coefficients.
|
| ||
|---|---|---|
| COMPARISON | COHEN’S KAPPA COEFFICIENTS | |
|
| ||
| Readers | I and II reading FFDM/DBT | 0.780 (substantial agreement) |
|
| ||
| III and IV reading SM/DBT | 0.773 (substantial agreement) | |
|
| ||
| Interpretation | FFDM/DBT and SM/DBT | 0.745 (substantial agreement) |
|
| ||
FFDM, full-field digital mammography; DBT, digital breast tomosynthesis; SM, two-dimensional synthetic mammography.
FFDM/DBT compared with SM/DBT for the detection of breast cancer.
|
| |||
|---|---|---|---|
| INTERPRETATION | FISHER’S EXACT TEST | ||
|
| |||
| Sensitivity | FFDM/DBT | 38/52 (73.1%) | |
|
| |||
| SM/DBT | 37/52 (71.2%) | ||
|
| |||
| Specificity | FFDM/DBT | 109/134 (81.3%) | |
|
| |||
| SM/DBT | 111/134 (82.8%) | ||
|
| |||
| Accuracy | FFDM/DBT | 147/186 (79.0%) | |
|
| |||
| SM/DBT | 148/186 (79.6%) | ||
|
| |||
FFDM, full-field digital mammography; DBT, digital breast tomosynthesis; SM, two-dimensional synthetic mammography.