Recommended performance characteristics for displays used to interpret medical images
| Characteristic | Suggested Performance Level for Medical Displays |
|---|---|
| Ambient illuminance | 25-75 lux |
| Ambient luminance | <0.2 cd/m2 |
| Minimum luminance | 0.8 cd/m2 |
| Maximum luminance | 350 cd/m2 |
| Luminance ratio | 350 |
| Luminance response function | Within 10% of DICOM Grayscale Standard Display Function |
| Luminance uniformity | LUDM <15%, no clinically significant nonuniformities |
| Pixel pitch | Approximately 0.200 mm at typical radiology viewing distances |
cd = candela; LUDM = luminance uniformity deviation from the median.